23andMe Personal Genome Service (PGS) Pharmacogenetic Reports
K221885 · 23AndMe, Inc. · QDJ · Oct 26, 2022 · Medical Genetics
Device Facts
Record ID
K221885
Device Name
23andMe Personal Genome Service (PGS) Pharmacogenetic Reports
Applicant
23AndMe, Inc.
Product Code
QDJ · Medical Genetics
Decision Date
Oct 26, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3364
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K221885 · Oct 26, 2022
23andMe Personal Genome Service (PGS) Pharmacogenetic Reports
23AndMe, Inc.
Biobank samples; 23andMe customer database
Retrospective biobank samples were used to evaluate the accuracy of the genotyping test against a reference method. Additionally, the sponsor used their existing customer database to report allele frequencies for the SLCO1B1 variant across various ancestry groups.
Biobank Accuracy Study; Retrospective accuracy study using biobank samples
Samples selected from a biobank; Sample Size: 256 samples (176 TT, 73 CT, 7 CC)
Bidirectional Sanger sequencing
Percent agreement (PPA/NPA) between PGS test and Sanger sequencing
23andMe Database Allele Frequency Analysis; Retrospective analysis of existing customer database
23andMe customers
Not applicable for this study
Allele frequencies for c.521T>C (rs4149056:C) by ancestry
Indications for Use
The 23andMe Personal Genome Service (PGS) is a qualitative genotyping assessment system applied to genomic DNA isolated from human saliva collected using the Oragene Dx OGD-500.001 to simultaneously detect, report, and interpret genetic variants in a broad multigene test. The assessment system is intended to enable users to access information about their genetics that could aid discussions with a healthcare professional. The 23andMe Personal Genome Service Pharmacogenetic Reports are indicated for the reporting of the following variants: Gene: CYP2C19 Variant(s): *2, *3, *17 Gene: CYP2C9 Variant(s): *2, *3, *5, *6, rs7089580 Gene: CYP3A5 Variant(s): *3 Gene: UGT1A1 Variant(s): *6, *28 Gene: DPYD Variant(s):*2A, rs67376798 Gene: TPMT Variant(s): *2, *3C Gene: SLCO1B1 Variant(s): c.521T>C (rs4149056) Gene: CYP2D6 Variant(s): *2, *3, *4, *5, *6, *7, *8, *9, *10, *11, *15, *17, *20, *29, *35, *40, *41
Device Story
Direct-to-consumer genetic testing service; utilizes user-collected saliva samples (Oragene-Dx kit). DNA isolated at CLIA-certified lab; multiplex assay performed on customized Illumina genotyping beadchip. Coregen software determines genotypes for >600,000 SNPs. System generates personalized reports for consumers; provides information on metabolizer/transporter profiles and interpretive drug information for specific medications (citalopram, clopidogrel, simvastatin). Users access reports via online platform; intended to facilitate informed discussions with healthcare providers. Device does not diagnose conditions or predict drug response; results are for informational purposes only. No changes to treatment should be made based on reports.
Clinical Evidence
Bench testing only. Method comparison study against bidirectional Sanger sequencing (n=samples with known variant status) demonstrated 100% PPA and NPA for SLCO1B1 c.521T>C (rs4149056). Precision study (intra-assay, operator, instrument, reagent lot) showed 100% correct genotype calls across multiple sites. Minimum DNA input study confirmed performance at 5 ng/μL. No clinical prospective/retrospective trials conducted.
Technological Characteristics
Qualitative genotyping microarray; Illumina Infinium HTS assay on customized Global Screening Array beadchip. Silicon wafers with silica beads and oligonucleotide capture probes. Energy source: instrumentation for extraction, processing, and scanning. Connectivity: data transmitted to manufacturer for cloud-based analysis. Software: proprietary algorithms for genotype calling and report generation. Sterilization: N/A (collection kit is sterile).
Indications for Use
Indicated for over-the-counter use by adults >18 years to provide genetic information informing discussions with healthcare professionals regarding therapeutic metabolism. Reports for CYP2C9, CYP3A5, UGT1A1, DPYD, TPMT, and CYP2D6 identify variants associated with metabolism but do not predict drug response or specific therapeutic associations. CYP2C19 report provides interpretive information for citalopram and clopidogrel. SLCO1B1 report provides interpretive information for simvastatin. Not a substitute for professional medical advice; results should not be used to start, stop, or change treatment.
Regulatory Classification
Identification
A pharmacogenetic assessment system is a qualitative in vitro molecular diagnostic system intended to detect nucleic acid variants isolated from human specimens for the purpose of assessing the presence of genetic variants that impact the metabolism, exposure, response, risk of adverse events, dosing, or mechanisms of prescription or over-the-counter medications. The intended use of the device must not include an indication for use in supporting or sustaining human life, being of substantial importance in preventing impairment of human health, or presenting a potential, unreasonable risk of illness or injury.
Special Controls
A pharmacogenetic assessment system must comply with the following special controls:
*Classification.* Class II (special controls). A pharmacogenetic assessment system must comply with the following special controls:(1) Design verification and validation must include:
(i) Data appropriate, as determined by FDA, to demonstrate the analytical accuracy and reliability of the device in intended use specimens, including precision, reproducibility, accuracy, limits of detection, and interferences. This information must include:
(A) Data demonstrating appropriate, as determined by FDA, reproducibility for each genotype using each claimed sample type. Reproducibility data must be evaluated using specimens collected and processed in a manner consistent with the device's instructions for use, or, as determined by FDA, an appropriate alternative sample panel.
(B) Analytical data demonstrating the limits of detection, including the minimum amount of input deoxy-ribonucleic acid (DNA) that will consistently produce accurate results.
(C) Data demonstrating no clinically significant effects from endogenous and exogenous interferents relevant to each intended use specimen type. Interference data must also include an assessment of potentially interfering genetic sequences (
*e.g.,* variants proximal to the variant of interest, pseudogenes).(D) Validation data appropriate, as determined by FDA, to support specimen collection and handling claims.
(E) Clinical data generated in intended use patient populations demonstrating the pharmacogenetic association between the genetic variant tested and any clinical claims or therapy-related recommendations associated with that genotype.
(ii) Results from an appropriate, as determined by FDA, user comprehension study that demonstrate the intended user can use the device safely. The user comprehension study must be designed to include the following:
(A) Study participants from a statistically sufficient sample size and a demographically diverse (
*e.g.,* age, education level) population that is representative of the intended use population and naive to use of the device, and(B) An evaluation of all result comprehension concepts that are critical for safe use of the device.
(2) The labeling required under § 809.10 of this chapter must include:
(i) Clear information, written in language appropriate for the intended user, that describes instructions for how test results should be interpreted. These instructions must be supported by valid scientific evidence and include:
(A) Appropriate explanation of the claimed pharmacogenetic associations for all variants included in the test, any relevant variants not included in the test (
*e.g.,* that may contribute to false negative results), and specific considerations by ethnicity, and(B) Appropriate explanation of non-genetic and non-tested genetic factors that may impact interpretation of the test results;
(ii) Detailed descriptions of analytical performance including, as applicable, precision, reproducibility, accuracy, limits of detection, and interferences as specified in paragraph (b)(1)(i) of this section, in language appropriate for the intended user;
(iii) A warning statement that the patient should not use the test results to stop or change any medication, and that medications should always be taken as prescribed by a healthcare professional;
(iv) A limiting statement explaining that this test is not intended to inform the patient about their current state of health, including whether the patient should or should not take a medication, or how much of a medication the patient should take, as appropriate;
(v) A warning statement that the test does not diagnose any health conditions and is not a substitute for visits to a doctor or other healthcare professional; and
(vi) A prominent and conspicuous limiting statement that the test provides only a preliminary test result that needs to be confirmed using an independent pharmacogenetic test without such a limitation prior to making any medical decisions. Alternatively, appropriate design verification and validation activities, including the generation of robust analytical data demonstrating appropriate analytical accuracy and reliability of test results for each genetic variant included in the test report, must be performed that demonstrate that the test can be used to make well-informed clinical decisions.
(3) The test report must include an appropriate description of how the test results should be used by healthcare providers who may receive the test results from their patients, as applicable.
(4) Publicly available pre-purchase labeling with unrestricted access that contains the following information must be provided:
(i) A clear description of the test and its technology, the genotypes detected, and relevant clinical claims associated with each genotype;
(ii) A clear description of what information the test will provide. This includes variant information, the limitations associated with the test, and any precautionary information about the test the user should be aware of before purchase; and
(iii) A discussion of answers to frequently asked questions that is sufficient to provide intended users with an appropriate understanding of information specific to each pharmacogenetic association that is claimed.
(5) The genetic test must use a sample collection device that is FDA-cleared or -approved, or classified as 510(k) exempt, with an indication for in vitro diagnostic use in DNA testing.
(6) The intended use of the device must not include an indication for use in supporting or sustaining human life, being of substantial importance in preventing impairment of human health, or presenting a potential, unreasonable risk of illness or injury.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K221885
B Applicant
23andMe, Inc.
C Proprietary and Established Names
23andMe Personal Genome Service (PGS) Pharmacogenetic Reports
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QDJ | Class II | 21 CFR 862.3364 - Pharmacogenetic Assessment System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
Modification of the SLCO1B1 report
B Measurand:
Genotype of select alleles in Cytochrome P450 2C19 (CYP2C19), 2C9 (CYP2C9), 2D6 (CYP2D6), 3A5 (CYP3A5), UDP glucuronosyltransferase family member A1 (UGT1A1), dihydropyrimidine dehydrogenase (DPYD), thiopurine methyltransferase (TPMT), solute carrier organic anion transporter family member 1B1 (SLCO1B1)
C Type of Test:
Qualitative genotyping microarray
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{2}
K221885 - Page 2 of 11
# III Intended Use/Indications for Use:
## A Intended Use(s):
See Indications for Use below.
## B Indication(s) for Use:
The 23andMe Personal Genome Service (PGS) is a qualitative genotyping assessment system applied to genomic DNA isolated from human saliva collected using the Oragene Dx OGD-500.001 to simultaneously detect, report, and interpret genetic variants in a broad multigene test. The assessment system is intended to enable users to access information about their genetics that could aid discussions with a healthcare professional. The 23andMe Personal Genome Service Pharmacogenetic Reports are indicated for reporting of the following variants:
| Gene | Variant(s) |
| --- | --- |
| CYP2C19 | *2, *3, *17 |
| CYP2C9 | *2, *3, *5, *6, rs7089580 |
| CYP3A5 | *3 |
| UGT1A1 | *6, *28 |
| DPYD | *2A, rs67376798 |
| TPMT | *2, *3C |
| SLCO1B1 | c.521T>C (rs4149056) |
| CYP2D6 | *2, *3, *4, *5, *6, *7, *8, *9, *10, *11, *15, *17, *20, *29, *35, *40, *41 |
This report is for over-the-counter use by adults over the age of 18 and provides genetic information to inform discussions with a healthcare professional about metabolism of therapeutics.
The 23andMe Personal Genome Service pharmacogenetic reports for CYP2C9, CYP3A5, UGT1A1, DPYD, TPMT and CYP2D6 describe if a person has variants associated with metabolism of some therapeutics but does not describe if a person will or will not respond to a particular therapeutic and does not describe the association between detected variants and any specific therapeutic.
23andMe Personal Genome Service pharmacogenetics report for CYP2C19 describes if a person has variants associated with metabolism of some therapeutics and provides interpretive drug information regarding the potential effect of the identified metabolizer phenotype on citalopram and clopidogrel therapy.
23andMe Personal Genome Service pharmacogenetics report for SLCO1B1 describes if a person has variants associated with the processing of some therapeutics and provides interpretive drug
{3}
information regarding the potential effect of the identified transport function phenotype on simvastatin therapy.
The PGS Pharmacogenetic Reports are not a substitute for visits to a healthcare professional. The information provided by this report should not be used to start, stop, or change any course of treatment.
## C Special Conditions for Use Statement(s):
OTC - Over The Counter
Warnings, precautions, and limitations for CYP2C9, CYP3A5, UGT1A1, DPYD, TPMT, CYP2D6:
- Do not use your results to start, stop or change any course of treatment.
- Results from this test should not be used to make medical decisions. Results should be confirmed by an independent genetic test prescribed by your own healthcare provider before taking any medical action.
- This test does not provide information on associations between specific DNA variants and any specific therapeutic.
- This test does not diagnose any health conditions, predict drug response, provide medical advice, or determine whether a medication is indicated for the user.
- This test does not determine if a person will or will not respond to a particular therapeutic.
- This test does not detect all genetic variants related to drug metabolism. The absence of a variant tested does not rule out the presence of other genetic variants that may be related to drug metabolism.
- This test is not a substitute for visits to a healthcare professional. You should consult with a healthcare professional if you have any questions or concerns about your results.
- This test may not be able to determine a result for all variants analyzed.
- Different companies offering genetic testing may be measuring different genetic variants for drug metabolism, so you may get different results from a different test.
- As with every test the possibility for an incorrect result exists. Speak to your personal healthcare professional or a genetic counselor if your results are unexpected.
Warning, precautions, and limitations specific for CYP2C19 and SLCO1B1:
- Do not use your results to start, stop or change any course of treatment.
- This test does not diagnose any health conditions, provide medical advice, or determine whether a medication is indicated for the user.
- This test provides interpretive drug information on citalopram and clopidogrel (CYP2C19) and simvastatin (SLCO1B1).
- This test does not determine if a person will or will not respond to a particular therapeutic.
- This test does not detect all genetic variants related to drug metabolism. The absence of a variant tested does not rule out the presence of other genetic variants that may be related to drug metabolism.
- This test is not a substitute for visits to a healthcare professional. You should consult with a healthcare professional if you have any questions or concerns about your results.
- This test may not be able to determine a result for all variants analyzed.
- For CYP2C19 only: This test does not provide interpretive drug information for the CYP2C19 *3/*17 genotype or other CYP2C19 genotype combinations where the predicted metabolizer profile cannot be interpreted. In addition, results for these genotypes should be confirmed by an
K221885 - Page 3 of 11
{4}
independent genetic test that is prescribed by your own healthcare provider before taking any medical action.
- Different companies offering genetic testing may be measuring different genetic variants for drug metabolism, so you may get different results from a different test.
- As with every test the possibility for an incorrect result exists. Speak to your personal healthcare professional or a genetic counselor if your results are unexpected.
## D Special Instrument Requirements:
Same as described in DEN140044.
## IV Device/System Characteristics:
## A Device Description:
The 23andMe PGS is a non-invasive DNA testing service that uses qualitative genotyping. It is a direct-to-consumer, over-the-counter, DNA genetic test. A user's saliva is self-collected using the Oragene-Dx device manufactured by DNA Genotek, Inc. (previously cleared under K141410), which consists of a sealable collection tube containing a stabilizing buffer solution. Once the sample is collected, it is shipped to one of two Clinical Laboratory Improvement Amendments (CLIA)-certified laboratories for testing.
DNA is isolated from the saliva and tested in a multiplex assay using a customized genotyping beadchip, reagents, and instrumentation. The multiplex assay simultaneously tests for more than 600,000 variants, including those for the previously authorized PGS test indications.
The raw data is generated by the scanning instrument's software, and then sent to 23andMe (the Manufacturer). The data are analyzed using the Manufacturer's proprietary software, and a genotype is determined for each tested variant. The results for certain of these variants, as noted in the indications for use, are used to generate personalized reports for users that provide information about the predicted metabolic or transporter function of the tested variants. Personalized reports are generated for each user to provide results of the testing performed. These reports tell the user which variant(s) has/have been detected in their sample and provide information on the predicted metabolizer or transporter profile associated with the variant(s). If no variant is detected, that information is also provided. The personalized reports are intended to present scientific concepts to users in an easy-to-understand format. The reports provide information about the association between the detected variant and the predicted metabolic or transport function that has been associated with the metabolism or transport of some drugs, further described below.
The 23andMe PGS Pharmacogenetic Reports detect 33 variants in 8 genes: CYP2C19, CYP2C9, CYP3A5, UGT1A1, DPYD, TPMT, SLCO1B1, and CYP2D6. The 23andMe PGS Pharmacogenetic Reports provide information on the associated enzyme or protein function and the predicted metabolizer or transporter phenotype for variants in drug metabolizing enzymes: CYP2C19, CYP2C9, CYP3A5, UGT1A1, DPYD, TPMT, SLCO1B1, and CYP2D6. The metabolizer or transporter phenotypes and corresponding identification is described in DEN180028.
K221885 - Page 4 of 11
{5}
In this 510(k) submission, the sponsor submitted additional analytical validation data demonstrating sufficient analytical accuracy and reliability such that the limitation statement regarding independent confirmation of the result required per the Special Control (2)(iv) of 21 CFR 862.3364 was no longer required in the labeling for the SLCO1B1 report.
The 23andMe PGS Pharmacogenetic Report for SLCO1B1 includes a drug specific test report for simvastatin. The test report may describe the predicted SLCO1B1 transporter protein function (i.e., normal, decreased, poor) based on the user's genotype inferred transporter phenotype and the predicted response to simvastatin (i.e., likely typical or increased chance of experiencing side effects). Other possible reported outcomes are that the result cannot be determined or the result cannot be interpreted.
## B Principle of Operation:
The PGS is indicated to be performed using a genotyping BeadChip assay, which covers more than 600,000 genetic markers. The BeadChip consists of silicon wafers etched to form wells loaded with silica beads, on which oligonucleotide capture probes are immobilized. DNA from saliva is fragmented and captured on a bead array by hybridization to immobilized SNP-specific primers, followed by extension with hapten-labeled nucleotides. The primers hybridize adjacent to the SNPs and are extended with single nucleotide(s) corresponding to the variant allele. The incorporated hapten-modified nucleotides are detected by adding fluorescently labeled antibodies in several steps to amplify the signals. Instrumentation is used for extraction and processing of the DNA, and for scanning the BeadChip to quantify the results. The genotype content is separated, analyzed, and then integrated into pre-defined report templates specific for each condition associated with each genotype. Genotypes are determined using two software packages. For 23andMe PGS Pharmacogenetics Report, the variants detected are:
| Gene | Variants |
| --- | --- |
| CYP2C19 | *2, *3, *17 |
| CYP2C9 | *2,*3,*5,*6, rs7089580 |
| CYP3A5 | *3 |
| UGT1A | *6, *28 |
| DPYD | *2A, rs67376798 |
| TPMT | *2, *3C |
| SLCO1B1 | c.521T>C (rs4149056) |
| CYP2D6 | *2, *3, *4, *5, *6, *7, *8, *9, *10, *11, *15, *17, *20, *29, *35, *40, *41 |
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
23andMe Personal Genome Service (PGS) Pharmacogenetic Reports
### B Predicate 510(k) Number(s):
K193492
K221885 - Page 5 of 11
{6}
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K221885 | K193492 |
| --- | --- | --- |
| Device Trade Name | 23andMe Personal Genome Service (PGS) Pharmacogenetic Reports | 23andMe Personal Genome Service (PGS) Pharmacogenetic Reports |
| General Device Characteristic Similarities | | |
| Intended Use | Provides genetic information to inform discussion with a healthcare professional about metabolism of therapeutics | Same |
| Collection Kit | Oragene·Dx® saliva collection device (OGD-500.001) | Same |
| Reagents | Illumina Infinium HTS Extra Assay Reagents | Same |
| BeadChip | Illumina Global Screening Array customized for the PGS. The chip is designed to detect specific single nucleotide polymorphisms (SNPs) as well as other genetic variants; all markers refer to specific positions in the National Center for Biotechnology Information (NCBI) reference human genome. | Same |
| Sample Matrix | Saliva | Same |
| General Device Characteristic Differences | | |
| SLCO1B1 Report | Provides interpretive drug information regarding the potential effect of the identified transport function | Does not provide interpretive drug information regarding the potential effect of the identified transport |
K221885 - Page 6 of 11
{7}
K221885 - Page 7 of 11
| | phenotype on simvastatin therapy. | function phenotype on simvastatin therapy. |
| --- | --- | --- |
| | Does not require confirmatory testing. | Requires confirmatory testing. |
VI Standards/Guidance Documents Referenced:
- Special Controls 21 CFR 862.3364, established in DEN180028
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
Precision studies were performed to assess the repeatability and reproducibility of the 23andMe Personal Genome Service (PGS) test for the variant included in the Pharmacogenetics (PGt) report for SLCO1B1 Drug Transport. The studies were designed to determine the imprecision due to assay run, lot, instrument, operator, day, and site. Genotypes of the DNA samples were confirmed using bidirectional Sanger sequencing.
The studies were performed at two sites across three days using three operator teams at each site. Six saliva samples were used in this study. Samples were genotyped in a blinded fashion using three lots of reagents and three instrument sets. All replicates that failed quality control (FQC) were rerun once. If the sample failed the rerun, it was given a result of FQC (failed QC). If a sample replicate passed overall laboratory QC, but did not produce a result for a specific variant, the result for that variant is designated as a "no call".
Results obtained are summarized below stratified by genotype and site.
{8}
| Genotype | Number of samples (81 replicates per sample) | Number of total replicates (including FQCs* and no calls) | Number of Correct Calls | Number of Incorrect Calls | Number of FQCs* (First run) | Number of No Calls |
| --- | --- | --- | --- | --- | --- | --- |
| Site 1 | | | | | | |
| TT (Homozygous Common) | 2 | 162 | 161 | 0 | 1 | 0 |
| CT (Heterozygous) | 3 | 243 | 236 | 0 | 7 | 0 |
| CC (Homozygous Rare) | 1 | 81 | 81 | 0 | 0 | 0 |
| Site 2 | | | | | | |
| TT (Homozygous Common) | 2 | 162 | 160 | 0 | 2 | 0 |
| CT (Heterozygous) | 3 | 243 | 227 | 0 | 16 | 0 |
| CC (Homozygous Rare) | 1 | 81 | 55 | 0 | 25 | 0 |
*FQC=Failed quality controls
After rerun of the first-run FQCs, there were 162 correct calls for the TT genotype and 243 correct calls for CT at Site 1. At Site 2, after rerun there were 162 correct calls for the TT genotype, 243 correct calls for CT, and 80 correct calls for CC. One CC sample at Site 2 received an FQC result after rerunning. No incorrect calls were observed in this study.
2. Linearity:
Not applicable.
3. Analytical Specificity/Interference:
**Endogenous and Exogenous Substances**
A series of studies were conducted to assess the effects of endogenous substances, exogenous substances, microbial substances, and smoking on the 23andMe PGS Test. The results of the endogenous and exogenous interference studies can be found in the Decision Summary for DEN140044.
**Interfering Mutations**
In the labeling, the sponsor identified the following rare mutations that may interfere with the performance of the test: rs74541382, rs141467543, rs200331427, and rs4149057.
K221885 - Page 8 of 11
{9}
4. Assay Reportable Range:
Not applicable.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
The traceability, stability and expected values for the device were previously reviewed in DEN140044.
6. Detection Limit:
The Limit of Detection (LoD) study was performed to confirm the lowest concentration of DNA, starting from intended use samples, that is necessary for successful assignment of correct genotypes. Study samples were saliva samples and included homozygous common (2 samples), heterozygous common (3 samples), and homozygous rare genotypes (1 sample).
DNA was extracted from the samples according to laboratory SOPs. Three replicates of each sample were diluted to three different DNA concentrations (5, 15, and 50 ng/μl) and genotyped by the PGS test in a blinded fashion using 3 lots of reagents. To confirm the genotype call, each sample was sequenced by bidirectional Sanger sequencing. Genotype calls from the PGS test were compared with genotypes from Sanger sequencing to determine the rates of correct genotype calls at each DNA concentration.
This study yielded 100% correct calls per genotype for all samples across all reagent lots, at all sample concentrations tested. Therefore, the LoD is confirmed at the lowest concentration tested (5 ng/μL). The performance requirement for the PGS test, specified in the laboratory standard operating procedures (SOPs), is set at a minimum of 15 ng/μL DNA and maximum of 50 ng/μL DNA.
7. Assay Cut-Off:
Not applicable.
B Comparison Studies:
1. Method Comparison:
Accuracy was evaluated through calculation of agreement of the genetic variant determinations between the 23andMe PGS test results and Sanger bidirectional sequencing (comparator) results. Genotyping results of saliva samples were compared between the PGS test and bidirectional sequencing to calculate positive and negative percent agreement (PPA and NPA) with the sequencing results used as the reference. The overall 95% confidence interval for the correct calls for the c.521T>C variant was 97.1% to 100% (for all c.521T>C genotype combinations). The comparison study results are shown below per sample genotype.
K221885 - Page 9 of 11
{10}
| Genotype | Correct Calls^{1} | Incorrect Calls | No Calls | Failed Quality Controls | % PPA | % NPA | 95% Confidence Interval^{2} |
| --- | --- | --- | --- | --- | --- | --- | --- |
| TT (Homozygous Common) | 45 | 0 | 0 | 0 | 100 | 100 | 93.6-100% |
| CT (Heterozygous) | 30 | 0 | 0 | 0 | 100 | 100 | 90.5-100% |
| CC (Homozygous Rare) | 26 | 0 | 0 | 0 | 100 | 100 | 89.1-100% |
1 Relative to bidirectional Sanger sequencing
2 mid-p method
In a separate study, saliva samples were selected blindly from a biobank to mitigate the risk of false-negative and false-positive results potentially influenced by selecting samples with previously-determined genotypes. Genotyping results were compared between the PGS test and bidirectional sequencing to calculate percent agreement with the sequencing results used as the reference. The comparison study results are shown below per sample genotype. The overall 95% confidence interval for the correct calls for the c.521T>C (rs4149056) variant was 98.8% to 100% (for all genotype combinations).
| Genotype | Correct Calls^{1} | Incorrect Calls | No Calls | Failed Quality Controls | % PPA | % NPA | 95% Confidence Interval^{2} |
| --- | --- | --- | --- | --- | --- | --- | --- |
| TT (Homozygous Common) | 176 | 0 | 0 | 0 | 100 | 100 | 98.3-100% |
| CT (Heterozygous) | 73 | 0 | 0 | 0 | 100 | 100 | 96.0-100% |
| CC (Homozygous Rare) | 7 | 0 | 0 | 0 | 100 | 100 | 65.2-100% |
1 Relative to bidirectional Sanger sequencing
2 mid-p method
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
i. Potential effect of genetic result on simvastatin therapy: The predicted SLCO1B1 transporter phenotypes and the resulting likelihood of side effects from simvastatin were
K221885 - Page 10 of 11
{11}
identified from data in the literature.
ii. User Comprehension Study: The user comprehension study previously described for DEN180028 was leveraged to support the test reports in K221885. Information regarding the user comprehension testing can be found in the Decision Summary for DEN180028.
In addition to supportive user comprehension testing, prior to opening a drug specific test report, the user must actively agree to the following statements:
"I understand that I should always take my medications as directed by my healthcare provider.
I will NOT use my results to start, stop, or change any medication on my own."
## D Clinical Cut-Off:
Not applicable.
## E Expected Values/Reference Range:
The package insert and user test reports include allele frequencies from 23andMe customers. The package insert indicated that this pharmacogenetics report tests for one (1) variant in the SLCO1B1 gene: c.521T>C (rs4149056:C). This variant is found in many ethnicities, at varying allele frequencies. The allele frequencies in the following table are from the 23andMe database and may not be representative of the actual allele frequencies in these populations.
| Ancestry Group | c.521T>C
(rs4149056:C) |
| --- | --- |
| European | 15.99% |
| African American | 5.21% |
| Ashkenazi Jewish | 18.37% |
| East Asian | 12.59% |
| Hispanic/Latino | 13.61% |
| South Asian | 5.00% |
| Middle Eastern | 17.75% |
| Other | 14.49% |
## VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K221885 - Page 11 of 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.