The BIOMONITOR III/BIOMONITOR IIIm is indicated to detect the following cardiac arrhythmias: - atrial fibrillation - bradycardia - sudden rate drop - high ventricular rate (HVR) - asystole The BIOMONITOR III/BIOMONITOR IIIm is indicated for use in: - Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias - Patients who experience transient symptoms that may suggest a cardiac arrhythmia The device has not been tested for and it is not intended for pediatric use.
Device Story
Implantable cardiac monitor; records subcutaneous ECG (sECG) and physiological parameters. Automatically detects atrial fibrillation, bradycardia, asystole, sudden rate drop, and high ventricular rate. Patient-activated recording via Remote Assistant III during symptomatic episodes. Data transmitted wirelessly to BIOTRONIK Home Monitoring Service Center for physician review. Used in clinical settings; aids in arrhythmia diagnosis and management. Hardware changes implemented to optimize manufacturing and automation; no changes to fundamental technology or intended use.
Clinical Evidence
No clinical data submitted or relied upon. Bench testing and verification/validation testing performed to support hardware changes.
Technological Characteristics
Subcutaneous insertable cardiac monitor; 47.5 x 8.3 x 4.3 mm (can); 1.9 cc volume; 4.0 g weight. Senses sECG via two integrated electrodes. Wireless connectivity to Home Monitoring Service Center. MR Conditional. Hardware adapted from predicate with minor component changes.
Indications for Use
Indicated for patients at increased risk of cardiac arrhythmias or experiencing transient symptoms suggesting arrhythmia. Not for pediatric use.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 27, 2022
BIOTRONIK, Inc. Jon Brumbaugh Vice President, Regulatory Affairs and New Product Development 6024 Jean Road Lake Oswego, Oregon 97035
## Re: K221856
Trade/Device Name: BIOMONITOR III, BIOMONITOR IIIm Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: June 24, 2022 Received: June 27, 2022
# Dear Jon Brumbaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Hetal Odobasic Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221856
Device Name BIOMONITOR III BIOMONITOR IIIm
Indications for Use (Describe)
The BIOMONITOR III/BIOMONITOR IIIm is indicated to detect the following cardiac arrhythmias:
- atrial fibrillation
- · bradycardia
- · sudden rate drop
- · high ventricular rate (HVR)
- · asystole
The BIOMONITOR III/BIOMONITOR IIIm is indicated for use in:
- · Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- · Patients who experience transient symptoms that may suggest a cardiac arrhythmia
The device has not been tested for and it is not intended for pediatric use.
Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BIOTRONIK. The logo consists of the letters "BIO" inside of a circle that is divided into quadrants. To the right of the circle is the word "BIOTRONIK" in large, bold, dark blue letters. Below the word "BIOTRONIK" is the phrase "excellence for life" in a smaller, lighter font.
# 510(k) Summary
# BIOMONITOR III and BIOMONITOR IIIm, Implantable Cardiac Monitors
Special 510(k) Premarket Notification
- Submission Information 1.
July 26, 2022 Date prepared
- Contact Jon Brumbaugh VP, Requlatory Affairs and New Product Development BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035 Phone (888) 345-0374 jon.brumbaugh@biotronik.com
- BIOTRONIK SE & Co. KG Manufacturer Woermannkehre 1, 12359 Berlin, Germany Registration number 9610139
#### 2. Subject Devices
| Trade Name | BIOMONITOR III and BIOMONITOR IIIm |
|-------------|------------------------------------|
| Common Name | Implantable Cardiac Monitor |
- Classification Name Recorder, Event, Implantable Cardiac (With Arrhythmia Detection)
- Classification Class II (21 CFR 870.1025)
- Product Code MXD
#### 3. Predicate Device
BIOTRONIK BIOMONITOR III and BIOMONITOR IIIm, K201865, cleared December 08, 2020
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for BIOTRONIK. The logo consists of the letters "BIO" inside of a circle that is divided into four quadrants. To the right of the circle is the word "BIOTRONIK" in large, bold, dark blue letters. Below the word "BIOTRONIK" is the phrase "excellence for life" in smaller, light gray letters.
#### Device Descriptions 4.
BIOMONITOR III and BIOMONITOR IIIm are programmable, subcutaneous insertable cardiac monitors able to record subcutaneous ECGs (sECGs) and other physiological parameters.
The BIOMONITOR III and BIOMONITOR IIIm are designed to automatically record the occurrence of arrhythmias in a patient. Arrhythmia may be classified as atrial fibrillation (AF), bradvarrhythmia, asystole, sudden rate drop, or high ventricular rate. In addition, the BIOMONITOR III and BIOMONITOR IIIm can be activated by the patient using the Remote Assistant III to record cardiac rhythm during symptomatic episodes. BIOMONITOR III and BIOMONITOR IIIm may be used with the current legally marketed BIOTRONIK Home Monitoring® technology, which is an automatic, wireless, remote monitoring system for management of patients with implantable cardiac monitors.
#### 5. Indications for Use
The indications for use for the BIOMONITOR III and BIOMONITOR IIIm are identical to the predicate device.
The BIOMONITOR III/BIOMONITOR IIIm are indicated to detect the following cardiac arrhythmias:
- . atrial fibrillation
- . bradycardia
- . sudden rate drop
- . high ventricular rate (HVR)
- . asystole
The BIOMONITOR III/BIOMONITOR IIIm is indicated for:
- Patients with clinical syndromes or situations at increased risk of cardiac arrhythmias
- Patients who experience transient symptoms that may suggest a cardiac arrhythmia ●
The device has not been tested for and it is not intended for pediatric use.
#### 6. Technological Characteristics
The substantial equivalence claim between the subject and the predicate device is supported by the information included in the premarket notification. This includes the following information :
- Description of the subject and predicate devices .
- . Intended use of the subject and predicate devices
- Performance of the subject and predicate devices
- . Technological characteristics of the subject and predicate devices
- . Validation testing
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BIOTRONIK. The logo consists of a circular symbol with the letters "BIO" inside, followed by the word "BIOTRONIK" in bold, dark blue letters. Below the word "BIOTRONIK" is the phrase "excellence for life" in a smaller, lighter font.
| Technical Data | Predicate BIOMONITOR<br>III, BIOMONITOR IIIm | Updated<br>BIOMONITOR III and<br>BIOMONITOR IIIm |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| FDA Clearance | K201865, Predicate | Subjects of this 510(k) |
| Indications | Patients with clinical syndromes or situations at increased<br>risk of cardiac arrhythmias<br>Patients who experience transient symptoms that may<br>suggest a cardiac arrhythmia<br>The device has not been tested for and it is not intended<br>for pediatric use | |
| Principle of Operation | The BIOMONITOR III/BIOMONITOR IIIm senses<br>subcutaneous electrocardiograms (SECG) using two<br>integrated electrodes and has the capability of detecting a<br>number of arrhythmias. Like the predicate device,<br>BIOMONITOR III/BIOMONITOR IIIm sends recorded SECG<br>and statistics to the Home Monitoring Service Center. | |
| Dimensions (mm)<br>Length x Width x Height | 47.5 x 8.3 x 4.3 (can)<br>77.5 x 8.6 x 4.6 w/ lead | |
| Volume | 1.9 cc | |
| Weight | 4.0 g | |
| AT/AF | 40 s/episode<br>30 s prior auto activation<br>10 s post auto activation | |
| MR Conditional | Yes | |
## Table 1. Comparison of Updated BIOMONITOR III, BIOMONITOR IIIm and Predicate, BIOMONITOR III
Several device hardware changes were implemented with this Special 510(k). These changes are being made to optimize manufacturing and prepare for further automation of the manufacturing process. The effect on determination of substantial equivalence was minimal as there was no associated change to intended use, indications for use, or fundamental technology and no additional questions regarding safety or effectiveness.
#### 7. Non-Clinical Performance Data
The following performance data are provided in support of the substantial equivalence determination :
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for BIOTRONIK. The logo consists of the letters "BIO" inside of a circle with a plus sign in the middle. To the right of the circle is the word "BIOTRONIK" in large, bold, dark blue letters. Underneath the word "BIOTRONIK" is the phrase "excellence for life" in a smaller, lighter font.
#### 7.1 Validation and Verification Testing:
The BIOMONITOR III and BIOMONITOR IIIm have undergone thorough validation and verification testing to ensure final device functionality. The following categories of tests were performed and passed:
• Verification Testing for BIOMONITOR III and BIOMONITOR IIIm hardware changes
#### 8. Clinical Performance Data
No clinical performance data was submitted or relied upon in support of the substantial equivalence determination.
#### 9. Conclusion
BIOTRONIK concludes that the BIOMONITOR III and BIOMONITOR IIIm with hardware changes are substantially equivalent to BIOTRONIK's predicate BIOMONITOR III and BIOMONITOR IIIm. The subject devices and predicate devices have the same principle of operation and physical device characteristics as well as software features and functionality and there are no new issues of safety or effectiveness. The hardware design is adapted from the predicate with minor component changes. These aspects of equivalence are confirmed by testing provided within the application.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.