Hakim Programmable Valves, Hakim Precision Fixed Pressure Valves
K221840 · Integra Lifesciences Production Corporation · JXG · Oct 14, 2022 · Neurology
Device Facts
Record ID
K221840
Device Name
Hakim Programmable Valves, Hakim Precision Fixed Pressure Valves
Applicant
Integra Lifesciences Production Corporation
Product Code
JXG · Neurology
Decision Date
Oct 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5550
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Codman Hakim Precision Fixed Pressure Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus. The Codman Hakim Programmable Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.
Device Story
Implantable, sterile, single-use pressure-regulating valves; manage hydrocephalus by diverting excess cerebrospinal fluid (CSF) from brain ventricles to another body cavity for absorption. Hakim Precision Valves offer fixed opening pressures (5 ranges); Hakim Programmable Valves allow non-invasive adjustment (18 settings). Devices maintain constant intraventricular pressure. Used by neurosurgeons in clinical settings. Output is mechanical regulation of CSF flow; clinical decision-making relies on valve pressure settings to manage intracranial pressure. Benefits include relief of intracranial pressure and management of hydrocephalus symptoms.
Clinical Evidence
No clinical studies performed. Substantial equivalence supported by bench testing and biocompatibility assessments. Bench testing included functional testing (ISO 7197, ASTM F647), shelf-life (ASTM F1980), MRI spatial gradient (ASTM F2052), and X-ray (ASTM F640). Biocompatibility testing followed ISO 10993 series (cytotoxicity, sensitization, irritation, systemic toxicity, implantation, genotoxicity, hemolysis) and USP standards. All tests passed.
Indicated for management of hydrocephalus in patients requiring constant intraventricular pressure regulation and CSF drainage.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
Predicate Devices
Hakim Precision Valves and Hakim Programmable Valves (K172022)
Reference Devices
Codman Certas Plus Programmable Valve (K143111, K182265)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
October 14, 2022
Integra LifeSciences Production Corporation Amanda Erwin Manager, Regulatory Affairs 11 Cabot Boulevard Mansfield, Massachusetts 02048
Re: K221840
Trade/Device Name: Hakim Programmable Valves, Hakim Precision Fixed Pressure Valves Regulation Number: 21 CFR 882.5550 Regulation Name: Central nervous system fluid shunt and components Regulatory Class: Class II Product Code: JXG Dated: July 18, 2022 Received: July 18, 2022
Dear Amanda Erwin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221840
Device Name Hakim Precision Fixed Pressure Valves Hakim Programmable Valves
#### Indications for Use (Describe)
The Codman Hakim Precision Fixed Pressure Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.
The Codman Hakim Programmable Valves Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Hakim® Precision Fixed Pressure Valves and Hakim® Programmable Valves
#### 510(k) Summary
A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| 807.92(a) (1) Submitter Information | |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Integra LifeSciences Production Corporation |
| Address | 11 Cabot Boulevard<br>Mansfield, MA 02048 |
| Telephone number | (609) 627-9053 |
| Primary Contact | Amanda Erwin |
| Date Summary Prepared | June 23, 2022 |
| 807.92(a) (2) Name of Device | |
| Trade or Proprietary Name | Hakim Precision Valves<br>Hakim Programmable Valves |
| Common Name | Central Nervous System Fluid Shunt and Components |
| Classification Name | Central Nervous System Fluid Shunt and Components<br>(21 CFR 882.5550) |
| Device Class | II |
| Product Code | JXG |
| 807.92(a) (3) Predicate Information | |
| Predicate Device | Hakim Precision Valves and Hakim Programmable Valves:<br>K172022<br>The following reference device is used in this submission:<br>Codman Certas Plus Programmable Valve: K143111,<br>K182265 |
| 807.92(a) (4) Device Description | |
The Codman Hakim® Precision Fixed Pressure and Programmable Valves are implantable, sterile, single use devices that provide constant intraventricular pressure and drainage of cerebrospinal fluid (CSF) for the management of hydrocephalus. Hydrocephalus is a condition caused by excessive accumulation of CSF in the ventricles of the brain due to a disturbance of
{4}------------------------------------------------
Hakim® Precision Fixed Pressure Valves and Hakim® Programmable Valves
CSF secretion, flow, or absorption, which causes a rise in intracranial pressure (ICP). To relieve ICP, CSF can be diverted through a shunting device, such as a Hakim Precision Valve or Hakim Programmable Valve, to another body cavity where it is a subsequently absorbed. Both the Codman Hakim Precision Fixed Pressure and Programmable Valves are pressure regulating valves which maintain intraventricular pressure at a constant level. The Hakim Precision valves are fixed pressure valves and are available in 5 different opening pressure ranges. The Codman Hakim Programmable Valves, not having fixed pressures, permit non-invasive adjustment of the valve opening pressure. The Codman Hakim Programmable Valves can be adjusted to 18 different opening pressure settings.
## 807.92(a) (5) Indications for Use
The Codman Hakim Precision Fixed Pressure Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.
The Codman Hakim Programmable Valve Systems are implantable devices that provide constant intraventricular pressure and drainage of cerebral spinal fluid (CSF) for the management of hydrocephalus.
807.92(a) (6) Technological Characteristics Compared to Predicate
The proposed Hakim Precision Valves and Hakim Programmable Valves have the same intended use, sterility, design principles and fundamental operation as the predicate valves. The proposed minor change in technological characteristics for the valves consists of a material change and minor dimensional and tolerance changes for various components. The changes do not raise any new questions of safety and/or effectiveness.
| Component Affected | Proposed Modification | Rationale |
|----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Various Internal<br>Components | • Replacement of<br>Polyethersulfone<br>(PES) with<br>Polysulfone (PSU) in<br>Hakim Precision<br>Fixed Pressure<br>Valves and Hakim<br>Programmable Valves | Polyethersulfone will no<br>longer be available for<br>Integra to manufacture the<br>impacted internal<br>components with. Nylon will<br>no longer be available to<br>manufacture the impacted<br>Hakim Programmable |
| Labeling | for various internal<br>components. Minor<br>dimensional and<br>tolerance changes to<br>internal components<br>will also be made to<br>the internal<br>components to<br>accommodate the<br>material change.<br>Replacement of<br>Nylon with<br>Polysulfone (PSU) in<br>Hakim programmable<br>Valve internal<br>component. Minor<br>dimensional changes<br>will be made to<br>accommodate the<br>material change. Updates made to<br>reflect the proposed<br>material change, as<br>well as administrative<br>updates and<br>corrections. Inclusion of Patient<br>Implant Cards and<br>Patient Information<br>Leaflet The labelling for<br>Hakim Programmable<br>Valves has also been<br>updated to comply<br>with Regulation (EU)<br>2017/745. | Valves Needle Guard<br>internal component with.<br>The minor dimensional and<br>tolerance changes to the<br>internal components have<br>been made to accommodate<br>the material change or<br>standardize component<br>design across Integra's<br>hydrocephalus valve<br>portfolio.<br>Update labelling with<br>current information. Patient Implant Cards<br>and Patient<br>Information Leaflets<br>are included as the<br>devices are MR<br>conditional implants. To comply with<br>Regulation (EU)<br>2017/745. |
| 807.92(b) 1-2: Summary of Nonclinical and Clinical Testing Performed | | |
{5}------------------------------------------------
Hakim® Precision Fixed Pressure Valves and Hakim® Programmable Valves
{6}------------------------------------------------
Hakim® Precision Fixed Pressure Valves and Hakim® Programmable Valves
The following performance testing has been conducted in support of the substantial equivalence determination. The testing utilized well-established methods, including those from FDA consensus standards. All testing was performed on production equivalent devices.
| Performance Bench Test Results | |
|---------------------------------------------|------------|
| Test | Conclusion |
| Product Shelf Life Testing per ASTM F1980 | Pass |
| Functional Testing per ISO 7197 | Pass |
| Functional Testing per ASTM F647 | Pass |
| MRI Spatial Gradient Testing per ASTM F2052 | Pass |
| X-Ray Testing per ASTM F640 | Pass |
| Biocompatibility Testing Results | |
|-------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Test | Conclusion |
| MTT and MTS Cytotoxicity Studies per ISO<br>10993-5 | Pass |
| Guinea Pig Maximization Sensitization Studies per<br>ISO 10993-10 | Pass |
| Intracutaneous Irritation Studies in Rabbits per ISO<br>10993-10 | Pass |
| Acute Systemic Toxicity Study in Mice per ISO<br>10993-11 | Pass |
| Rabbit Pyrogen Studies per United States<br>Pharmacopeia (USP 42 – NF 37) | Pass |
| Subcutaneous Implantation Studies in Rabbits, 1<br>Week and 4 Weeks per ISO 10993-6 | Pass |
| Systemic Toxicity and Local Effects Study in<br>Rabbits Following Subcutaneous Implantation, 13<br>Weeks per ISO 10993-6 and ISO 10993-11 | Pass |
| Bacterial Reverse Mutation Studies per ISO 10993-<br>3 and ISO/TR 10993-33 | Pass |
| In Vitro Mouse Lymphoma Studies per ISO 10993-<br>3 and ISO/TR 10993-33 | Pass |
| Hemolysis on Extract Studies per ISO 10993-4 and<br>ASTM F756 | Pass |
There are no changes in sterility method as a result of the proposed material change and minor dimensional and tolerance changes; a sterilization equivalency assessment was performed comparing the proposed device to the predicate device and deemed acceptable.
{7}------------------------------------------------
Hakim® Precision Fixed Pressure Valves and Hakim® Programmable Valves
No clinical studies were required as appropriate verification of the subject device was achieved based on the comparison to the predicate device and from the results of testing.
#### 807.92(b) (3) Conclusion
Based upon the intended use, design, comparison to the predicate device, and testing performed, Integra LifeSciences believes that the proposed modifications to the Hakim Precision Valves and Hakim Programmable Valves do not raise any new questions of safety and effectiveness, and is therefore, substantially equivalent to the predicate Hakim Precision Valves and Hakim Programmable Valves.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.