K221827 · Saudi Mais Co. For Medical Products · FOZ · Jun 23, 2023 · General Hospital
Device Facts
Record ID
K221827
Device Name
Mais Central Venous Catheter
Applicant
Saudi Mais Co. For Medical Products
Product Code
FOZ · General Hospital
Decision Date
Jun 23, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Mais central venous catheters are indicated to provide short-term access (<30 days) to the central venous system. They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal and central venous pressure monitoring.
Device Story
Mais Central Venous Catheter is a radio-opaque, multi-lumen (double, triple, or quadruple) polyurethane catheter. It is surgically inserted via the Seldinger technique into the superior vena cava (via internal jugular, subclavian, or femoral veins) for short-term (<30 days) central venous access. Used in anesthesia, critical care, and emergency medicine by trained clinicians to administer fluids, blood products, drugs, and parenteral nutrition, or to monitor central venous pressure. The device features centimeter markings, extension tubes with colored clamps, and a soft, pliable tip. It provides pre-operative access for anesthesia and emergency access for conditions like shock or cardiac standstill. Correct tip positioning (superior vena cava, ~2cm above right atrium) is critical for therapeutic success. The device is provided sterile and single-use.
Clinical Evidence
Bench testing only. No clinical data. Device demonstrated conformance to ISO 10555-1 and ISO 10555-3 standards, including leak testing, flow rate by gravity, breaking load (tensile force ≥15N), anti-obstructive performance, radiopacity, and luer connector testing (ISO 80369-7). Biocompatibility testing performed per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, subchronic toxicity, hemolysis, implantation). Particulate matter testing met USP <788> criteria.
Technological Characteristics
Materials: Medical-grade polyurethane (shaft, luer connector, extension legs, junction). Design: Multi-lumen (double, triple, quadruple) radio-opaque tube with soft tip. Dimensions: Variable lengths (5-30cm) and French sizes. Connectivity: None. Sterilization: Ethylene Oxide (ISO 11135:2014). Standards: ISO 10555-1, ISO 10555-3, ISO 80369-7 (luer), ISO 10993-1 (biocompatibility), ISO 11607 (packaging).
Indications for Use
Indicated for patients requiring short-term (<30 days) central venous access for I.V. fluid/drug administration, blood product infusion, parenteral nutrition, blood withdrawal, or central venous pressure monitoring.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
K193158 — ACENT Central Venous Catheter · Global Medikit Limited · Nov 24, 2020
K182252 — Cook Unimpregnated Central Venous Catheter · Cook Incorporated · May 17, 2019
K070451 — ABLE CENTRAL VENOUS CATHETER · Foshan Nanhai Bai HE Medical Technology Co., Ltd. · Jul 31, 2007
K021130 — SPECATH CENTRAL VENOUS CATHETER KITS · Special Medical Co. , Ltd. · Jan 7, 2003
K190855 — BD Acute Central Line · Bard Access Systems, Inc. (Bard Has Joined Bd) · Nov 1, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
June 23, 2023
Saudi Mais Co. for Medical Products % Vaibhav Rajal Official Correspondent for Saudi Mais Co. for Medical Products mdi Consultants Inc 55 Northern Blvd. Suite 200 Great Neck, New York 11021
Re: K221827
Trade/Device Name: Mais Central Venous Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOZ Dated: Mav 24, 2023 Received: May 24, 2023
Dear Vaibhav Rajal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Porsche Bennett
For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221827
Device Name Mais Central Venous Catheter
#### Indications for Use (Describe)
Mais central venous catheters are indicated to provide short-term access (<30 days) to the central venous system.
They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal and central venous pressure monitoring.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Saudi
mais Co.
Image /page/3/Picture/1 description: The image shows a stylized logo in black and white. The logo appears to be in Arabic script, with rounded, interconnected lines forming the letters. Below the main text, there are two small circles, positioned side by side.
C.R. 1010112398 A Limited Liability Capital S.R. (34,500,000) Fully Paid VAT No .: 3000 5661 72000 03
Image /page/3/Picture/3 description: The image shows a logo with the letters "SMMP" in bold, black font inside a red oval. Below the logo, there is Arabic text that includes the number 55358. The logo appears to be for an organization or company, with the Arabic text possibly indicating a membership or registration number.
ست ۱۰۱۰۱۱۲۳۹۸ شركة ذات مسؤولية محدودة
رأس المال المدفوع ( ٣٤.٥٠٠.٠٠٠) ريال مدفوع بالكامل
الرقم الضريبي : ۰۳ ۳۰۰۰۵۶۶۱۷۲۰۰۰
# K221827- 510(k) SUMMARY
Date Summary Prepared: June 23, 2023
## 1. Submitter's Identification:
Saudi Mais Co. for Medical Products Alkharj Road, 3rd Industrial Area, Street No.256, P.O. Box: 3900 - Riyadh 14335-7599, Saudi Arabia Tel: + 966 11 2650184 Fax: + 966 11 2650139 Email: medical@mais.com.sa
### 2. Contact Person:
Mr. Salman Rashid Saudi Mais Co. for Medical Products Alkharj Road, 3rd Industrial Area, Street No.256, P.O. Box: 3900 - Riyadh 14335-7599, Saudi Arabia Tel: + 966 11 2650184 Email: qa@mais.com.sa
Official Correspondent: Mr. Vaibhav Arvind Rajal, Official Correspondent for Saudi Mais Co. for Medical Products, mdi Consultants, Inc., 55 Northern Blvd., Suite 200, Great Neck, 11021 Cell: (201) 887-3180 or (516) 482-9001 email: vaibhav@mdiconsultants.com
# 3. Subject Device:
Trade/Device Name: Mais Central Venous Catheter Common Name: Short-Term Less Than 30 Days Therapeutic Intravascular Catheter Regulation Description: Intravascular catheter Regulation Number: 21 CFR 880.5200 Regulatory Classification: Class II Product Code: FOZ
{4}------------------------------------------------
## 4. Predicate Device
510(k) Number: K190855 Trade/Device Name: BD Acute Central Line Regulation Number: 21 CFR 880.5200 Requlation Name: Intravascular catheter Regulatory Classification: Class II Product Code: FOZ
## 5. Indications for Use Statement:
Mais central venous catheters are indicated to provide short-term access (<30 days) to the central venous system.
They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal and central venous pressure monitoring.
## 6. Device Description:
Mais Central Venous Catheter is a radio opaque catheter, which is placed mainly in the subclavian, or jugular veins down to the right auricle as a short-term channeling access to the central venous system for central administration of drugs and central infusion of fluids and transfusion of blood.
The device provides pre-operative access to infuse anesthesia drugs and quick central access in emergencies as cardiac standstill, shock, etc.
Measuring the central venous pressure can be achieved by connecting the central lumen to the measuring device tube.
Central venous (CV) catheters can be used in anesthesia, critical care, and emergency medicine.
The correct positioning of the central venous catheter is of key significance for therapeutic success.
For most applications, placement in the superior vena cava approximately 2cm before the opening into the right atrium is viewed as the optimal position for the catheter tip.
The CV Catheter is needle punctured and surgically inserted into the superior vena cava (a large vein) of a patient. It can be planted inside for one-month maximum.
#### Positions for the insertion:
For Mais Central Venous Catheter, there are three optional puncture points depended on the clinical requirement with Seldinger Technique. They respectively are:
- Internal jugular vein; or
- . Subclavian vein; or
- Femoral vein
The Seldinger Technique itself has the specific requirements and procedures to perform this treatment. Therefore, the device must be strictly used by trained, qualified doctors or nurses only.
{5}------------------------------------------------
### Catheter's Technical Description:
Mais Catheters are made from soft polyurethane material and have double, triple or quadruple lumens, and the length is variable. The Mais catheter consists of a white radio opaque multi lumen tube which is tipped with a soft polyurethane material and has many side holes relevant to each lumen. There are markings along the effective length to indicate each centimeter. There is a connector which connects the catheter lumen(s) to a lumen of a transparent polyurethane extension tube so each catheter lumen is connected to a specific extension line. The connector is polyurethane and nearly triangular shaped. Each extension tube is marked to indicate which catheter lumen is attached to it. Distal lumen is used for the CVP monitoring. A flat clamp is put over each extension tube. The flat clamp is colored differently for a given multi lumen catheter.
| # | Product Code | Catheter Description | Flow Rate (ml/min) | | Priming Volume (ml/min) | |
|----|--------------|------------------------|--------------------|--------------|-------------------------|--------|
| | | | Distal Lumen | Medial Lumen | Distal | Medial |
| 1 | 623-0405 | Double-lumen4F-5cm | 19 | 22 | 0.16 | 0.17 |
| 2 | 623-0460 | Double-lumen4F-6cm | 15 | 18 | 0.16 | 0.18 |
| 3 | 623-0408 | Double-lumen4F-8cm | 16 | 16 | 0.16 | 0.172 |
| 4 | 623-0413 | Double-lumen4F-13cm | 13 | 13 | 0.192 | 0.2 |
| 5 | 623-0560 | Double-lumen5F-6cm | 34 | 32 | 0.2 | 0.16 |
| 6 | 623-0508 | Double-lumen5F-8cm | 29 | 32 | 0.2 | 0.17 |
| 7 | 623-0515 | Double-lumen5F-15cm | 23 | 22 | 0.21 | 0.23 |
| 8 | 623-5508 | Double-lumen5.5F-8cm | 25 | 23 | 0.3 | 0.33 |
| 9 | 623-5513 | Double lumen 5.5F-13cm | 26 | 24 | 0.32 | 0.331 |
| 10 | 623-0715 | Double-lumen7F-15cm | 93 | 40 | 0.45 | 0.43 |
| 11 | 623-0720 | Double-lumen7F-20cm | 77 | 35 | 0.46 | 0.42 |
| 12 | 623-0730 | Double-lumen7F-30cm | 68 | 28 | 0.62 | 0.53 |
The Mais Central Venous Catheter is available in different sizes as shown below:
| # | Product Code | Catheter Description | Flow Rate (ml/min) | | | Priming Volume (mL/min | | |
|----|--------------|------------------------|--------------------|--------|-------|------------------------|--------|-------|
| | | | Distal | Medial | Prox. | Distal | Medial | Prox. |
| 13 | 625-0515 | Triple lumen 5F-15cm | 32 | 20 | 22 | 0.17 | 0.39 | 0.35 |
| 14 | 625-5505 | Triple-lumen5.5F-5cm | 39 | 22 | 23 | 0.18 | 0.145 | 0.18 |
| 15 | 625-5506 | Triple lumen 5.5F-6cm | 37 | 21 | 22 | 0.19 | 0.149 | 0.182 |
| 16 | 625-5508 | Triple-lumen5.5F-8cm | 35 | 18 | 19 | 0.201 | 0.156 | 0.18 |
| 17 | 625-5513 | Triple-lumen5.5F-13cm | 29 | 12 | 12 | 0.23 | 0.23 | 0.2 |
| 18 | 625-0715 | Triple-lumen7F-15cm | 63 | 25 | 26 | 0.372 | 0.3 | 0.314 |
| 19 | 625-0720 | Triple-lumen7F-20cm | 54 | 20 | 20 | 0.437 | 0.332 | 0.345 |
| 20 | 625-0730 | Triple-lumen7F-30cm | 46 | 15 | 16 | 0.52 | 0.37 | 0.402 |
| 21 | 625-7520 | Triple lumen 7.5F-20cm | 55 | 22 | 21 | 0.65 | 0.44 | 0.36 |
| 22 | 625-0820 | Triple lumen 8F-20cm | 57 | 24 | 23 | 0.71 | 0.48 | 0.39 |
| # | Product Code | Catheter Description | Flow Rate (ml/min) | | | | Priming Volume (ml/min) | | | |
|----|--------------|----------------------|--------------------|--------|--------|-------|-------------------------|--------|--------|-------|
| | | | Distal | Med. 1 | Med. 2 | Prox. | Distal | Med. 1 | Med. 2 | Prox. |
| 23 | 627-7520 | Quad-lumen7.5F-20cm | 45 | 52 | 35 | 34 | 0.49 | 0.41 | 0.31 | 0.34 |
There are variants available in central venous catheter; double lumen and quad lumen catheters which have various sizes and lengths.
{6}------------------------------------------------
# 7. Comparison to the 510(k) Cleared Device (Predicate Device):
| Attributes | Proposed Subject Device | Predicate Device: Bard Incorporated K190855 | Discussion | | | | | | | | | | | |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|--|--|--|--|--|--|--|--|--|--|--|
| Indications for Use | Mais central venous catheters are indicated to provide short-term access (<30 days) to the central venous system.<br>They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal and central venous pressure monitoring. | Acute central venous catheters are indicated to provide short-term access (<30 days) to the central venous system.<br>They are designed for administering I.V. fluids, blood products, drugs and parenteral nutrition solutions, as well as blood withdrawal, central venous pressure monitoring, and power injection of contrast media. | Different<br>Please see<br>comment# 1 | | | | | | | | | | | |
| | | Catheter Length Lumen Power Injection Flow Rate Maximum Power Injection Pressure Setting 16 cm and 20 cm Distal<br>Medial/Proximal 10mL/sec<br>9ml/sec 325 psi 30 cm Distal<br>Medial/Proximal 9mL/sec<br>7mL/sec | | | | | | | | | | | | |
| Intended Use | Mais Central Venous Catheter System is intended for short-term access to the central venous system for intravenous therapy and blood sampling. | BD Acute Central Lines are intended for short-term access to the central venous system for intravenous therapy and blood sampling. | Same | | | | | | | | | | | |
| Length | 5cm, 6 cm, 8 cm, 10cm, 13 cm, 15 cm, 20 cm, 30 cm | 16 cm, 20 cm and 30 cm | Different<br>Please see<br>comment# 2 | | | | | | | | | | | |
| Device Description | Mais central venous catheters constructed of medical grade polyurethane, is designed for insertion into the central venous line with Radiaopaque, and have a soft tip that is more pliable than the catheter body. Each catheter is provided in a sterile package. | A family of central venous catheters constructed of medical grade polyurethane and is designed for insertion into the central venous catheter with radiopaque, and have a soft tip that is more pliable than the catheter body. Each catheter is provided in a sterile package. | Same | | | | | | | | | | | |
| Catheter Dimension | Double-lumen4F-5cm<br>Double-lumen4F-6cm<br>Double-lumen4F-8cm<br>Double-lumen4F-13cm<br>Double-lumen5F-6cm<br>Double-lumen5F-8cm<br>Double-lumen5F-15cm<br>Double-lumen5.5F-8cm<br>Double lumen 5.5F-13cm<br>Double-lumen7F-15cm<br>Double-lumen7F-20cm<br>Double-lumen7F-30cm<br>Triple lumen 5F-15cm<br>Triple-lumen5.5F-5cm<br>Triple lumen 5.5F-6cm<br>Triple-lumen5.5F-8cm<br>Triple-lumen5.5F-13cm<br>Triple-lumen7F-15cm<br>Triple-lumen7F-20cm<br>Triple-lumen7F-30cm<br>Triple lumen 7.5F-20cm<br>Triple lumen 8F-20cm<br>Quad-lumen7.5F-20cm | 7 Fr Triple Lumen x 16 cm<br>7 Fr Triple Lumen x 20 cm<br>7 Fr Triple Lumen x 30 cm | Different<br>Please see<br>comment# 3 | | | | | | | | | | | |
| Primary Device Materials | Catheter Base Materials<br>Shaft Tubing- Polyurethane<br><br>Luer Connector-<br>Polyurethane | Catheter Base Materials<br>Shaft Tubing- Polyurethane<br><br>Luer Connector- Polyurethane<br><br>Extension Legs- Polyurethane | Same | | | | | | | | | | | |
{7}------------------------------------------------
| | Extension Legs-<br>Polyurethane | Junction- Polyurethane | |
|------------------------------------------------|-------------------------------------------------------------------------------|---------------------------------------|--------------------------------------|
| | Junction- Polyurethane | | |
| Duration of Use | Short term (<30 days) | Short term (<30 days) | Same |
| Means of<br>insertion | Percutaneous | Percutaneous | Same |
| Insertion Site<br>(Anatomical site<br>- route) | Jugular, subclavian, or<br>femoral veins | Jugular, subclavian, or femoral veins | Same |
| | | | Different<br>Please see comment<br>4 |
| No. of Lumens | Double Lumen, Triple<br>Lumen, Four Lumen | Triple Lumen | Different<br>Please see comment<br>4 |
| Catheter<br>Proximal<br>Configuration | Side-hole skive | Side-hole skive | Same |
| Catheter Distal<br>Configuration | Formed Tip | Formed Tip | Same |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Same |
| Sterility<br>Assurance Limit<br>(SAL) | The SAL is 10-6 Utilizing<br>the existing validated and<br>revalidation cycle | 10-6 | Same |
| EO residuals | 0.006ppm | Unknown | Unknown, Please<br>see comment #5 |
| Catheter Body<br>material | Polyurethane (Medical<br>Grade) | Polyurethane (Medical Grade) | Same |
| Package | Sterile package. | Sterile package. | Same |
Discussions of differences in technological characteristics:
Comment# 1. The basic indications for use statement of the subject device and the predicate device are identical except that the predicate device has an additional feature of power injection of contrast media which is not a part of the subject device. The absence of this feature does not raise new concerns of safety or effectiveness in the subject device presentation.
Comment # 2. The additional catheter lengths are provided for the different age groups of patients and the site of application of the catheter. Based on performance testing conducted such as mechanical testing to ISO 10555-1, 10555-3. Flow Rate by Gravity Testing, Breaking load testing, and Anti obstructive ml/min test, it was established that the additional catheter lengths of 6, 8, 10, 13, 15 and 25 cm do not raise different questions of safety or effectiveness.
Comment # 3. Based on performance testing conducted such as mechanical testing to ISO 10555-1, 10555-3, Flow Rate by Gravity Testing, Breaking load testing, and Anti obstructive ml/min test, it was established that the additional catheter dimensions for the subject device do not raise questions of safety or effectiveness.
Comment # 4. The subject device Mais central venous catheter is available as a double lumen, triple lumen, and quadruple lumen whereas the predicate device is available as triple lumen. Based on performance testing conducted such as mechanical testing to ISO 10555-1, 10555-3, Flow Rate by Gravity Testing, Breaking load testing, and Anti obstructive ml/min test to ISO 10555, ISO 11135 standards demonstrate adequate performance with double, triple, and quadruple lumens and therefore do not raise different questions of safety and effectiveness.
Comment # 5. The EO residuals for the predicate device are unknown; however, the subject device has been tested to ISO 11135:2014 and the residuals are acceptable and therefore, the potential differences do not raise new or different questions to safety and effectiveness.
{8}------------------------------------------------
## 8. Performance Testing
- A. The Mais Central Venous Catheter described in this summary was tested and demonstrated to be in conformance with the following FDA recognized standards:
- ISO 10555-1 2013, Intravascular Catheters Sterile and Single-Use Intravascular ● Catheters - Part 1: General Requirements [Including AMENDMENT 1 (2017)]
- ISO 10555-3 2013 Intravascular Catheters – Sterile and Single-Use Intravascular Catheters- Part3: Central Venous Catheters
- . Leak Test - Test to confirm that the catheter assembly will not leak when the distal end of the catheter is occluded.
- The identity of the length/Dimension Test -Test to measure useful length for ● catheters to ensure compliance with dimensional specification.
- The identification of extension tube/ Extension leq length test - Test to measure and confirm extension leg length compliance with dimensional specification.
- . Flow rate by gravity - Test to measure the gravity flow performance of a full-length catheter.
- Breaking Load (Tensile Force) Test to demonstrate the peak tensile force of each ● test piece exceeds the minimum peak tensile force. Result: All the below mentioned parts of the device are within the acceptable range i.e. Force required ≥ 15N as per ISO 10555-1:2013
- o Tube body
- Tube and hub O
- Branch tube and hub O
- Branch tube body O
- o Tip
- Anti-obstructive ml/min Test to estimate anti obstructive ml/min at the proximal and ● distal part of the catheter
- Coordination of slipping buckle (Clamp Engagement) Test to confirm that the ● catheter assembly will not leak when the clamp is engaged.
- Radiopacity Test to demonstrate catheter radio-detectability .
- Luer Testing Testing to ensure that luer connectors meet requirements for ●
- o Leak
- Leak Decay O
- Stress Cracking O
- Resistance to Separation from Axial Load o
- Resistance to Separation from Unscrewing O
- Resistance to Overriding O
- O Gauging as per ISO 80369-7:2017
- Particulate Testing/ Particle pollution Testing to ensure that particulate matter on the ● catheter post-manufacture is not exceeded for prescribed particle sizes. Reference Standard: USP: Sizing and Counting Particulate Matter
The subject device was tested as per the requirements of 10555-3 test standard. The subject device is in compliance with the requirements of 10555-3 test standard and does not raise issues of safety or effectiveness.
- B. Biocompatibility
In accordance with ISO 10993-1 the Central Venous Catheter is classified as: Externally Communicating Device, Blood Path (Circulating blood), Prolonged Contact (>24hr to 30days). The following testing was conducted:
- Cytotoxicity
{9}------------------------------------------------
- Sensitization ●
- Irritation/Intracutaneous ●
- Acute Systemic Toxicity
- Subchronic Toxicity
- Bacterial Reverse mutation assay
- . Hemolysis
- Implantation ●
Particulate matter testing was conducted in accordance with USP <788> Particulate Matter in Injections and met the USP acceptance criteria.
#### 9. Sterilization and Stability Information
The subject device complies with sterilization requirements of ISO 11135:2014, Sterilization of Health Care Products -Ethylene Oxide - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices.
- . Package integrity: We evaluated the device for simulated shipping by testing for all its applicable performance and package testing in compliance with ISO 11607, ASTM F1886, ASTM F1929 and ASTM F88 standards after passing through different transportation medium and found to be free from any kind of change in physical, chemical, microbiological and the package status of the device.
- ං ISO 11607-1:2006 +A1:2014 Packaging For Terminally Sterilized Medical Devices - Part 1: Requirements For Materials, Sterile Barrier Systems And Packaging Systems [Including: Amendment 1 (2014)]
- The subject device, Mais Central Venous Catheter is provided in a sterile package. The ● proposed subject device. Mais Central Venous Catheter is in compliance with the following standards:
- Visual Inspection ASTM F1886 o
- Dye Migration Test ASTM F1929 o
- Seal Peel Test ASTM F88/EN868-5 O
- 11135 Second Edition 2014-07-15 Sterilization Of Health-Care Products - O Ethylene Oxide - Requirements For The Development, Validation And Routine Control Of A Sterilization Process For Medical Devices [Including: Amendment 1 (2018)]
- 11737-1 Third Edition 2018-01 Sterilization Of Health Care Products о Microbiological Methods - Part 1: Determination Of A Population Of Microorganisms On Product
- 11737-2 Second Edition 2009-11-15 Sterilization Of Medical Devices o Microbiological Methods - Part 2: Tests Of Sterility Performed In The Definition, Validation And Maintenance Of A Sterilization Process
- The Shelf Life of 4 years based on the validation documentation of real aging from Study start date of 09-14-2014 to Study Completion date of 07-30-2018.
### Conclusions:
The differences between the Mais Central Venous Catheter and the predicate device do not raise any new or different questions of safety or effectiveness. The subject device, Mais Central Venous Catheter is substantially equivalent to the predicate device, BD Acute Central Line, cleared under K190855.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.