← Product Code [GEI](/productcode/GEI) · K221815

# Electrosurgical Pencil (OBS-Db, OBS-Dr, OBS-Dc, OBS-Df, OBS-Di, OBS-De); Electrosurgical Pencil (OBS-Dt, OBS-Dd, OBS-Ds, OBS-Dl, OBS-Da) (K221815)

_Baisheng Medical Co., Ltd. · GEI · Nov 23, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221815

## Device Facts

- **Applicant:** Baisheng Medical Co., Ltd.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Nov 23, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Electrosurgical Pencil (OBS-Db, OBS-Dr, OBS-Dr, OBS-Di, OBS-De) is used for cutting and coagulation to remove tissue and control bleeding by using high frequency current during electrosurgical surgery with an electrosurgical generator. The device is disposable and supplied sterile with an electrode tip. The Electrosurgical Pencil (OBS-Dt, OBS-Ds, OBS-Ds, OBS-Da) is used for cutting and coagulation to remove tissue and control bleeding by using high frequency current during electrosurgical surgery with an electrosurgical generator, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system.

## Device Story

Electrosurgical pencil; handheld device used in surgical settings by clinicians. Inputs: high-frequency electrical current from an electrosurgical generator. Operation: delivers current to tissue via electrode tip for cutting/coagulation; specific models include integrated smoke evacuation channel for connection to external suction systems. Output: thermal tissue effect (cutting/coagulation) and smoke removal. Disposable, sterile, single-use. Benefits: facilitates precise tissue management and hemorrhage control; smoke evacuation models reduce surgical plume exposure.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Handheld electrosurgical pencil; disposable; sterile. Features electrode tip for high-frequency current delivery. Some models include integrated smoke evacuation tubing. Operates via connection to standard electrosurgical generators. No software or electronic components.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 23, 2022

Baisheng Medical Co., Ltd. Sylvia Dong Regulatory Affairs Specialist No.11, Fusheng Road, Xinhui District, Jiangmen, Guangdong, Jiangmen, 529100 China

Re: K221815

Trade/Device Name: Electrosurgical Pencil (OBS-Db, OBS-Dr, OBS-Dc, OBS-Df, OBS-De); Electrosurgical Pencil (OBS-Dt, OBS-Dd, OBS-Ds, OBS-D1, OBS-Da) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: September 29, 2022 Received: September 30, 2022

Dear Sylvia Dong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Carr -S

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221815

#### Device Name

Electrosurgical Pencil (OBS-Db, OBS-Dr, OBS-Dc, OBS-Df, OBS-Di, OBS-De); Electrosurgical Pencil (OBS-Dt, OBS-Dd, OBS-Ds, OBS-Dl, OBS-Da)

#### Indications for Use (Describe)

The Electrosurgical Pencil (OBS-Db, OBS-Dr, OBS-Dr, OBS-Di, OBS-De) is used for cutting and coagulation to remove tissue and control bleeding by using high frequency current during electrosurgical surgery with an electrosurgical generator. The device is disposable and supplied sterile with an electrode tip.

The Electrosurgical Pencil (OBS-Dt, OBS-Ds, OBS-Ds, OBS-Da) is used for cutting and coagulation to remove tissue and control bleeding by using high frequency current during electrosurgical surgery with an electrosurgical generator, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/device/K221815](https://fda.innolitics.com/device/K221815)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).