AccurSound Electronic Stethoscope AS-101

K221805 · Heroic Faith International , Ltd. · DQD · Mar 23, 2023 · Cardiovascular

Device Facts

Record IDK221805
Device NameAccurSound Electronic Stethoscope AS-101
ApplicantHeroic Faith International , Ltd.
Product CodeDQD · Cardiovascular
Decision DateMar 23, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1875
Device ClassClass 2
AttributesPediatric

Indications for Use

The AS-101 is an electronic stethoscope intended for the detection of sounds associated with the heart, lungs, arteries, veins, and other internal organs. It can be used on any person undergoing a physical assessment. The device is intended to be operated only by healthcare professionals for diagnostic decision support in clinical settings.

Device Story

Electronic stethoscope system for clinical auscultation; utilizes microphone-equipped disposable sensor patches to detect body sounds; signals processed/filtered by microcontroller-based hub unit; anti-noise function reduces ambient interference; output delivered to earpiece. Multi-channel design supports single-channel handheld or four-channel stationary/continuous monitoring. Operated by healthcare professionals in clinical settings for diagnostic decision support. Provides amplified, filtered audio of internal organs to assist physical assessment.

Clinical Evidence

No clinical or animal performance testing included. Evidence consists of bench testing: electrical safety (ANSI/AAMI ES60601-1), EMC (ANSI/AAMI/IEC 60601-1-2), biocompatibility (ISO 10993-1), risk management (ISO 14971), human factors (IEC 62366-1), and specific acoustic/hardware performance metrics (frequency response, THD, noise, amplification ratio, multi-channel functionality).

Technological Characteristics

Electronic stethoscope with disposable polyethylene foam sensor patches. Microcontroller-based digital signal processing. Frequency range 20-2000 Hz. Powered by external AC-DC adapter (100-240V). User interface: 4-button keypad, 5 LED indicators. Not sterile. Connectivity: wired sensors to hub, wired earpiece output. No Bluetooth.

Indications for Use

Indicated for detection and amplification of heart, lung, arterial, venous, and other internal organ sounds in patients of all ages undergoing physical assessment by healthcare professionals.

Regulatory Classification

Identification

Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.

Special Controls

*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. March 23, 2023 Heroic Faith International Ltd. Yuan-Ren Cheng VP & CIO 18F., No. 460, Section 4, Xinyi Road, Xinyi District Taipei City, 110 Taiwan Re: K221805 Trade/Device Name: AccurSound Electronic Stethoscope AS-101 Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD Dated: February 16, 2023 Received: February 21, 2023 Dear Yuan-Ren Cheng: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {2}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Robert T. Kazmierski -S for LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use 510(k) Number (if known) K221805 Device Name AccurSound Electronic Stethoscope AS-101 #### Indications for Use (Describe) The AS-101 is an electronic stethoscope intended for the detection of sounds associated with the heart, lungs, arteries, veins, and other internal organs. It can be used on any person undergoing a physical assessment. The device is intended to be operated only by healthcare professionals for diagnostic decision support in clinical settings. | Type of Use (Select one or both, as applicable) | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff [PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov) *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Heroic Faith. The logo consists of the words "HEROIC" and "FAITH" stacked on top of each other, with a red cross in the middle. To the right of the words is a red hexagon with a white design inside. Below the logo is the tagline "Saving Life Through Innovation." ## 510(k) Summary ### 1. Submission Details Heroic Faith International Ltd. 18F., No. 460, Sec. 4, Xinyi Rd., Xinyi Dist., Taipei City, 110 Taiwan (R.O.C.) Phone: +886-22659-0291 Contract: Yuan-Ren Cheng, VP & CIO #### 2. Date Prepared March 22, 2023 #### 3. Device | Trade name: | AccurSound Electronic Stethoscope AS-101 | |-----------------------|------------------------------------------| | Common name: | Electronic Stethoscope | | Classification name: | Electronic Stethoscope (21 CFR 870.1875) | | Device Class: | Class II | | Product Code: | DQD | | Classification Panel: | Cardiovascular | #### 4. Predicate Device 3M Littmann Electronic Stethoscope, Model 3200 (K083903) #### 5. Device Description The AccurSound Electronic Stethoscope AS-101 ("AS-101") is a device designed for healthcare professionals used in clinical settings. The AS-101 can detect and amplify the sounds of the heart, lungs, arteries, veins, and other internal organs. The microphone-equipped disposable sensor detects and amplifies the sounds from the patient's body. The auscultation sound is digitally processed and filtered, electronically amplified in the hub unit. The anti-noise function reduces ambient noise and echoes, then transferred to the earpiece. The multi-channel design allows healthcare professionals to attach disposable sensors onto patient's body, by switching modes from handheld single-channel recording to four-channel stationery and continuously auscultation based on different requirements of clinical applications or physical assessments. {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows a logo with the words "HEROIC FAITH" in black, with the "T" in "HEROIC" and the "I" in "FAITH" in red. There is also a red horizontal line above the word "FAITH". To the right of the words is a red geometric shape. Below the words is the phrase "Saving Life Through Innovation" in a smaller font. The associated accessories include: - Disposable sensor(s) - · Patient cable(s) - Earphone - Power adapter with power cord #### 6. Intended Use/ Indications for Use The AS-101 is an electronic stethoscope intended for the detection and amplification of sounds associated with the heart, lungs, arteries, veins, and other internal organs. It can be used on any person undergoing a physical assessment. The device is intended to be operated only by healthcare professionals for diagnostic decision support in clinical settings. ### 7. Comparison of technological characteristics with the predicate device The AS-101 is substantially equivalent to the predicate device 3M Littmann Electronic Stethoscope Model 3200 in terms of indications for use, technological characteristics, and safety and effectiveness. The comparison between the subject and predicate devices is provided in the table below. | | Candidate device<br>AccurSound® Electronic<br>Stethoscope AS-101 | Predicate device K083903<br>3M™ Littmann® Electronic<br>Stethoscope Model 3200 | Identical/<br>Similar/<br>Different | |----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------| | | Product classification | | | | Regulatory<br>number | 870.1875 | 870.1875 | Identical | | Product code | DQD | DQD | Identical | | Classification | Class II | Class II | Identical | | Power source | | | | | Power Source | AC-to-DC supply | AA alkaline, lithium, or NiMH battery | Different | | | Intended use | | | | Intended use | The AS-101 is an electronic<br>stethoscope intended for pickup<br>and amplification of sounds<br>associated with the heart, lungs,<br>arteries, veins, and other internal<br>organs. It can be used on any<br>patient undergoing physical<br>examinations. It is intended to be<br>operated by medically qualified<br>personnel for diagnostic purposes<br>in clinics or hospitals. | 3M™ LITTMANN® ELECTRONIC<br>STETHOSCOPE MODEL 3200 is<br>intended for medical diagnostic<br>purposes only. It may be used for the<br>detection and amplification of<br>sounds from the heart, lungs,<br>arteries, veins, and other internal<br>organs with the use of selective<br>frequency ranges. It can be used on<br>any patients undergoing a physical<br>assessment. | Identical | {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows a logo with the words "HEROIC FAITH" in black, with the "T" in "HEROIC" and the "I" in "FAITH" in red. There is a red geometric shape to the right of the words. Below the words, it says "Saving Life Through Innovation" in a smaller font. ## AccurSound Electronic Stethoscope AS-101 510(k) Premarket Notification | Device characteristics | | | | |-------------------------------------|---------------------------------------------------------|--------------------------------------------------------------------------|-----------| | Sterility | Not intended to be sterilized | Not intended to be sterilized | Identical | | | Control Hub:<br>99.0 x 71.0 x 16.6 mm | Total length:<br>690 mm | | | Dimension | Patient Cable: 1450 mm | N/A | Different | | | Sensor (chest-piece) diameter:<br>19.0 x 34.0 x 8.0 mm | Diaphragm diameter:<br>51 mm | | | Patient<br>-contacting<br>materials | Sensor: polyethylene foam single<br>coated medical tape | Chest-piece: metal | Different | | Single-use | Foam sensor is single-use | Reusable chest-piece and the<br>diaphragm can be removed for<br>cleaning | Different | | Technological characteristics | | | | |-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------| | Principle of<br>operation | The Accursound® Electronic<br>Stethoscope AS-101 pick up<br>sounds, such as heart, lung, and<br>other internal organs sound from<br>a patient's body. After<br>amplification and filtering, the<br>sounds are sent to the user<br>through a set of earpiece.<br>The disposable sensor patch is<br>designed for use with adult,<br>pediatric, and infant patients.<br>The user interface for the<br>stethoscope includes a 4-button<br>keypad and five LED indicators.<br>Five LED indicators include 4<br>channel indicators and a power<br>indicator.<br>Sound Processing is carried out<br>with a microcontroller and<br>delivered to audio output.<br>Power is supplied by an external<br>power source through a medical-<br>grade AC-DC power adapter,<br>which has an input requirement of<br>100-240V. The power indicator<br>turns solid green when the power<br>is on. | The 3M™ Littman® Electronic<br>Stethoscope Model 3200 picks up<br>sounds, such as heart and lung<br>sounds, from a patient's body. After<br>amplification and filtering, the<br>sounds are sent to the user through a<br>binaural headset. The stethoscope<br>chestpiece is designed for use with<br>adult, pediatric, and infant<br>patients. The user interface for the<br>stethoscope includes a 5-button<br>keypad and an LCD display. Sound<br>processing is carried out with the aid<br>of a digital signal processor.<br>Stethoscope power is provided by a<br>single AA battery in the chestpiece. A<br>power management system is<br>included to prolong battery life.<br>Using its Bluetooth wireless link, the<br>stethoscope can exchange audio data<br>with an external device such as a<br>personal computer (PC). | Similar | {7}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the words "HEROIC FAITH" in black font, with the "T" in "HEROIC" and the line above the "AI" in "FAITH" in red. To the right of the words is a red hexagon with three white lines inside. Below the words is the phrase "Saving Life Through Innovation" in black font. ## AccurSound Electronic Stethoscope AS-101 510(k) Premarket Notification | Functions and performance | | | | |-----------------------------|-----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|-----------| | Audio | headphones | Binaural headset | Different | | Chest-piece | Foam sensors | Metallic chest-piece | Different | | Sound<br>processing | Microcontroller as the digital<br>signal processor | Digital signal processor | Identical | | Display and<br>button panel | A 4-button keypad and five LED<br>indicators. Five LED indicators<br>include 4 channel indicators and a<br>power indicator. | LCD<br>5-button keypad | Different | | Bluetooth | No | Yes | Different | | Sound<br>frequency<br>range | 20-2000 Hz | Bell mode: 20-200 Hz<br>Diaphragm mode: 100-500 Hz<br>Extended range: 50-500 Hz | Similar | | Volume control | Yes | Yes | Identical | | Volume control<br>level | 1-9 level | 1-9 levels | Identical | #### 8. Performance Testing #### Summary discussion of clinical data: This submission does NOT include animal or clinical performance testing. Summary discussion of non-clinical data: The following non-clinical tests were conducted to evaluate the safety and performance of the AS-101 and provide evidence to support the verification and validation of AS-101: - Electrical Safety and EMC tests are in compliance with ANSI/AAMI ES60601-1:2005/(R)2012/A1:2012, C1:2009/(R)2012/A2:2010/(R)2012 and ANSI/AAMI/IEC 60601-1-2:2014 - Biocompatibility evaluation in compliance with ISO 10993-1 was performed. - Software level of concern is moderate. Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" - Risk management according to ISO 14971:2019. - Human factor engineering is in compliance with IEC 62366-1: 2015. {8}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Heroic Faith. The logo consists of the words "HEROIC FAITH" in black font, with a red cross in the middle of the word "HEROIC" and the letter "T" in the word "FAITH". To the right of the words is a red geometric shape. Below the logo is the phrase "Saving Life Through Innovation". 510(k) Premarket Notification - Other bench test: - . LED Indicator Behavior Test Report - Button Selector Test Report - Hardware Sound Band Pass Filter Test Report - Total Harmonic Distortion and Noise Test Report - . Acoustic Frequency Response Test Report - Sample Rate Test Report - Overall Sound Amplification Ratio Test Report - . Multiple Channels Test Report - I2C Communication Test Report - . Audio Comparison of the Foam Sensors Test Report - Data Transmission Test Report - . Test for Degrees of Protection Provided by Enclosures Test Report #### 9. Substantial Equivalence The subject device has same indications for use, technology, and operation principle as well as similar technical characteristics with the predicate device. Verification activities were performed on subject device and all tests were verified to meet the required acceptance criteria. The verification tests demonstrate that the differences in the device do not affect the intended use of the device or raise any unsolved issues. There are no significant differences between subject device and the predicate device(s). We conclude that subject device is substantially equivalent to predicate devices. #### 10. Conclusion Based on the information provided in this premarket notification, Heroic Faith International Ltd. believes the proposed Accursound Electronic Stethoscope AS-101 and its predicate device, the 3M Littmann Model 3200 Electronic Stethoscope are substantially equivalent in their intended use, both devices share similar design and technology characteristics.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...