← Product Code [IZL](/productcode/IZL) · K221803

# PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress. (K221803)

_Sedecal., Sa. · IZL · Jul 18, 2022 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221803

## Device Facts

- **Applicant:** Sedecal., Sa.
- **Product Code:** [IZL](/productcode/IZL.md)
- **Decision Date:** Jul 18, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1720
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

This is a digital mobile diagnostic x-ray system intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

## Device Story

Digital mobile X-ray system; used by physicians or technicians in clinical settings. Inputs: X-ray radiation captured by digital flat-panel detectors (AeroDR series). Operation: System generates X-rays; detectors convert radiation to digital signals; software (CS-7 or Ultra) processes, displays, and stores images. Modification enables Dynamic Digital Radiography (DDR) mode, allowing rapid acquisition of up to 15 frames/sec (300 frames total) using AeroDR P-65/P-75 panels. Output: Diagnostic radiographic images viewed by clinicians to inform clinical decision-making. Benefits: Enables mobile diagnostic imaging for patients unable to be transported to stationary X-ray suites; DDR mode provides serial radiographic data.

## Clinical Evidence

No clinical data. Substantial equivalence established via bench testing and verification of system integration with previously cleared digital X-ray panels and software.

## Technological Characteristics

Mobile X-ray system; 20-50 kW generator; 40-150 kV range. Detectors: AeroDR P-51, P-52, P-61, P-71, P-81, P-65, P-75. Connectivity: Ethernet/Wi-Fi. Software: CS-7 or Ultra. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, 60601-2-28, 60601-1-6, IEC 62304.

## Regulatory Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- PHOENIX ([K212291](/device/K212291.md))

## Reference Devices

- AeroDR SYSTEM 2 ([K151465](/device/K151465.md))
- SKR3000 ([K172793](/device/K172793.md))
- SKR3000 ([K210619](/device/K210619.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

July 18, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Sedecal SA % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114

Re: K221803

Trade/Device Name: PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress. Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL, MQB Dated: June 16, 2022 Received: June 22, 2022

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221803

Device Name

PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress.

#### Indications for Use (Describe)

This is a digital mobile diagnostic x-ray system intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

Type of Use (Select one or both, as applicable)

| <span style="font-size:12px"></span> <span style="font-size:12px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:12px"></span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                                      |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary: 510(k) Number K221803

Image /page/3/Picture/1 description: The image is a logo for SEDECAL. The logo is a teal rounded rectangle with a white border. Inside the rectangle is a white icon of a person sitting at a desk, followed by the word "SEDECAL" in white, bold letters.

SEDECAL SA C/ Pelaya, 9 - 13, Pol. Ind. Río de Janeiro 28110 Algete, Madrid España (Spain) Tel .- +34 91 6280544 Fax.- +34 91 6280574 Date Prepared: July 7, 2022 Contact: Mª Luisa Gómez de Agüero, Quality and Regulatory Manager

- 1) Identification of the Device: Trade/Device Name: PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress. Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 2) Equivalent legally marketed device: K212291 Trade/Device Name: PHOENIX Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Codes: IZL, MQB. Common/Usual Name: Digital Mobile Diagnostic X-Ray System
- 3) Reference devices: We employ these cleared devices without modification: Trade/Device Name: (K151465, K172793, K210619) Konica-Minolta Digital X-ray Panels AeroDR SYSTEM 2 includes P-51 & P-52 and software CS-7 (K151465) SKR3000 includes panels P-71 & P-81 and software CS-7 (K172793) SKR3000 includes panels P-65 & P-75 with software CS-7 (K210619) (The Ultra software was cleared in our predicate K212291) Regulation Number: 21 CFR 892. 1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB
- 4) Indications for Use: This is a digital mobile diagnostic x-ray system intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. Not for mammography.

{4}------------------------------------------------

- 5) Description of the Device: This is a modified version of our previous predicate mobile PHOENIX. The predicate PHOENIX mobile is interfaced with Konica – Minolta Digital X-ray panels and CS-7 or Ultra software image acquisition. PHOENIX mobile systems will be marketed in the USA by KONICA MINOLTA. Models with the CS-7 Software will be marketed as AeroDR TX m01. Models with the Ultra software will be marketed as mKDR Xpress. The modification adds two new models of compatible Konica-Minolta digital panels, the AeroDR P-65 and AeroDR P-75, cleared in K210619. These newly compatible models are capable of a mode called DDR, Dynamic Digital Radiography wherein a series of radiographic exposures can be rapidly acquired, up to 15 frames per seconds maximum (300 frames).
The CS-7 software was cleared under K151465/K172793, while the Ultra software was cleared within our predicate K212291. The CS-7 is a DIRECT DIGITIZER used with an image diagnosis device, medical imaging device and image storage device connected via the network. This device digitally processes patient images collected by the medical imaging device to provide image and patient information. By contrast the Ultra software is designed as an exam-based modality image acquisition tool. Ultra software and its accompanying Universal Acquisition Interface (UAI) were developed to be acquisition device independent. Basic Features of the software include Modality Worklist Management (MWM) / Modality Worklist (MWL) support, DICOM Send, CD Burn, DICOM Print, and Exam Procedure Mapping. Ultra Software is designed to increase patient throughput while minimizing data input errors. Ultra is made up of multiple components that increase efficiency while minimizing errors. The main components of Ultra are the Worklist, Acquisition Interface and Configuration Utility. These components combine to create a Stable, Powerful, and Customizable Image capture system. The intuitive graphical user interface is designed to improve Radiology, Technologist accuracy, and image quality. Worklist and Exam screens were developed to allow site specific customizations to seamlessly integrate into existing practice workflows. PHOENIX/AeroDR TX m01 employs the "CS-7" software. PHOENIX/mKDR Xpress employs the "Ultra" software. We decided to continue to offer the CS-7 package as a lower cost option for those users who do not need the enhanced features of the Ultra software. Both software packages are of Moderate level of concern and are based on the predicate software. The ONLY DIFFERENCE between the software offered and the predicate is the fact that the DDR function has been enabled because the two newly supported panels (P65, P75) can support that function. As noted in section 3, above, reference devices: All digital flat panel detectors and software have received FDA clearance.

| Characteristic                                       | Predicate: K212291 PHOENIX                                                                                                                                                                                                                                                                                                                                                                                                              | PHOENIX/AeroDR TX m01 and<br>PHOENIX/mKDR Xpress.                                                                                                                                                                                                                                                                                                                                                                                               |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use:                              | This is a digital mobile diagnostic x-ray system<br>intended for use by a qualified/trained doctor<br>or technician on both adult and pediatric<br>subjects for taking diagnostic radiographic<br>exposures of the skull, spinal column, chest,<br>abdomen, extremities, and other body parts.<br>Applications can be performed with the patient<br>sitting, standing, or lying in the prone or supine<br>position. Not for mammography | This is a digital mobile diagnostic x-ray system<br>intended for use by a qualified/trained doctor<br>or technician on both adult and pediatric<br>subjects for taking diagnostic radiographic<br>exposures of the skull, spinal column, chest,<br>abdomen, extremities, and other body parts.<br>Applications can be performed with the patient<br>sitting, standing, or lying in the prone or supine<br>position. Not for mammography. (Same) |
| Configuration                                        | Mobile System with digital x-ray panel and<br>image acquisition computer                                                                                                                                                                                                                                                                                                                                                                | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| X-ray<br>Generator(s)                                | kW rating: 20 kW, 32 kW, 40 kW and 50 kW.<br>kV range: from 40 kV to 150 kV in 1 kV steps.                                                                                                                                                                                                                                                                                                                                              | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Characteristic                                       | Predicate: K212291 PHOENIX                                                                                                                                                                                                                                                                                                                                                                                                              | PHOENIX/AeroDR TX m01 and<br>PHOENIX/mKDR Xpress.                                                                                                                                                                                                                                                                                                                                                                                               |
| mA range: from 10 mA to 630 mA / 640 mA /<br>650 mA. |                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Collimator                                           | Ralco R108F                                                                                                                                                                                                                                                                                                                                                                                                                             | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Photos                                               | Image: K212291 Phoenix                                                                                                                                                                                                                                                                                                                                                                                                                  | Image: PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress                                                                                                                                                                                                                                                                                                                                                                                            |
| Meets US<br>Performance<br>Standard                  | YES 21 CFR 1020.30                                                                                                                                                                                                                                                                                                                                                                                                                      | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Power Source                                         | Universal power supply, from 100 V~ to 240 V~.<br>1 phase, 1.2 kVA                                                                                                                                                                                                                                                                                                                                                                      | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Software                                             | Konica-Minolta control software CS-7 (K151465<br>or K172793) OR Konica-Minolta control<br>software Ultra.                                                                                                                                                                                                                                                                                                                               | CS-7 and Ultra software programs have been<br>modified to accommodate DDR mode.<br>PHOENIX/AeroDR TX m01 employs the "CS-7"<br>software.<br>PHOENIX/mKDR Xpress employs the "Ultra"<br>software.                                                                                                                                                                                                                                                |
| Panel Interface                                      | Ethernet or Wi-Fi wireless                                                                                                                                                                                                                                                                                                                                                                                                              | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                      | DETECTOR TECHNOLOGICAL COMPARISONS                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Image area<br>sizes                                  | AeroDR P-51 14" x 17"<br>AeroDR P-52 14" x 17"<br>AeroDR P-61 14" x 17"<br>AeroDR P-71 17" x 17"<br>AeroDR P-81 10" x 12"                                                                                                                                                                                                                                                                                                               | AeroDR P-51 14" x 17"<br>AeroDR P-52 14" x 17"<br>AeroDR P-61 14" x 17"<br>AeroDR P-71 17" x 17"<br>AeroDR P-81 10" x 12"<br>Similar range of sizes (All previously cleared)<br>Added:<br>AeroDR P-65 14" x 17"                                                                                                                                                                                                                                 |
| Characteristic                                       | Predicate: K212291 PHOENIX                                                                                                                                                                                                                                                                                                                                                                                                              | PHOENIX/AeroDR TX m01 and<br>PHOENIX/mKDR Xpress.                                                                                                                                                                                                                                                                                                                                                                                               |
| Pixel sizes                                          | AeroDR P-51 175 μm<br>AeroDR P-52 175 μm<br>AeroDR P-61 100/200 μm<br>AeroDR P-71 100/200 μm<br>AeroDR P-81 100/200 μm                                                                                                                                                                                                                                                                                                                  | AeroDR P-51 175 μm<br>AeroDR P-52 175 μm<br>AeroDR P-61 100/200 μm<br>AeroDR P-71 100/200 μm<br>AeroDR P-81 100/200 μm<br>Added:<br>AeroDR P-65 100 μm / 200 μm / 400 μm<br>AeroDR P-75 100 μm / 200 μm / 400 μm<br>Similar pixel sizes (All previously cleared)                                                                                                                                                                                |
| Resolutions                                          | AeroDR P-51 1994 x 2430 pixels<br>AeroDR P-52 1994 x 2430 pixels<br>AeroDR P-61 3488 × 4256 pixels<br>AeroDR P-71 4248 × 4248 pixels<br>AeroDR P-81 2456 × 2968 pixels                                                                                                                                                                                                                                                                  | AeroDR P-51 1994 x 2430 pixels<br>AeroDR P-52 1994 x 2430 pixels<br>AeroDR P-61 3488 × 4256 pixels<br>AeroDR P-71 4248 × 4248 pixels<br>AeroDR P-81 2456 × 2968 pixels<br>Added:<br>AeroDR P-65 3,488 × 4,256 pixels<br>AeroDR P-75 4,248 × 4,248 pixels                                                                                                                                                                                        |
| MTF                                                  | AeroDR P-51 0.30 @ 2cy/mm<br>AeroDR P-52 0.30 @ 2cy/mm<br>AeroDR P-61 0.30 @ 2cy/mm<br>AeroDR P-71 0.30 @ 2cy/mm<br>AeroDR P-81 0.30 @ 2cy/mm                                                                                                                                                                                                                                                                                           | AeroDR P-51 0.30 @ 2cy/mm<br>AeroDR P-52 0.30 @ 2cy/mm<br>AeroDR P-61 0.30 @ 2cy/mm<br>AeroDR P-71 0.30 @ 2cy/mm<br>AeroDR P-81 0.30 @ 2cy/mm<br>Added:<br>AeroDR P-65 (Non-binning) 0.62<br>(2x2 binning) 0.58<br>AeroDR P-75 (Non-binning) 0.62<br>(2x2 binning) 0.58                                                                                                                                                                         |
| DQE                                                  | AeroDR P-51 0.62 @ 0 lp/mm<br>AeroDR P-52 0.62 @ 0 lp/mm<br>AeroDR P-61 0.56 @ 1 lp/mm<br>AeroDR P-71 0.56 @ 1 lp/mm<br>AeroDR P-81 0.56 @ 1 lp/mm                                                                                                                                                                                                                                                                                      | AeroDR P-51 0.62 @ 0 lp/mm<br>AeroDR P-52 0.62 @ 0 lp/mm<br>AeroDR P-61 0.56 @ 1 lp/mm<br>AeroDR P-71 0.56 @ 1 lp/mm<br>AeroDR P-81 0.56 @ 1 lp/mm<br>Added:<br>AeroDR P-65 0.56 @ 1 lp/mm<br>AeroDR P-75 0.56 @ 1 lp/mm                                                                                                                                                                                                                        |
| Software                                             | Konica-Minolta control software CS-7 (K151465<br>or K172793) OR Konica-Minolta control<br>software Ultra.                                                                                                                                                                                                                                                                                                                               | Konica-Minolta control software CS-7 (K151465<br>or K172793) OR Konica-Minolta control<br>software Ultra (K212291). Both support DDR.                                                                                                                                                                                                                                                                                                           |
| Panel Interface                                      | Ethernet or Wi-Fi wireless                                                                                                                                                                                                                                                                                                                                                                                                              | SAME                                                                                                                                                                                                                                                                                                                                                                                                                                            |

### 6) Substantial Equivalence Chart

{5}------------------------------------------------

{6}------------------------------------------------

7) The technological characteristics, including design, materials, composition, and energy source, are substantially the same, so there are no issues impacting safety and effectiveness.

Safety and Effectiveness, comparison to predicate device. The results of bench testing indicate that the new devices are as safe and effective as the predicate devices. Proper system operation is fully verified upon installation. We verified that the modified combination of components worked properly and

{7}------------------------------------------------

produced diagnostic quality images as good as our predicate generator/panel combination. Here is a summary of the changes we made to the device: We added two more compatible flat panel detectors: AeroDR P-65 and the AeroDR P-75. These two panels are capable of acquiring up to 20 seconds of serial radiography.

- 8) Summary of non-clinical testing: Systems covering all generator/panel combinations were assembled and tested and found to be operating properly. Firmware was unchanged. Because the system uses Wi-Fi and Ethernet, we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff. The CS-7 digital panel software employed was already reviewed by FDA in K172793 and subsequently in our predicate K212291. The Ultra digital panel software was cleared in our predicate and has not been modified.
The PHOENIX Battery Mobile X-Ray Units have been tested to be in compliance with the following International Standards:

- a) IEC 60601-1:2005+A1:2012 (Edition 3.1)
- b) IEC 60601-1-2:2014 (Edition 4.0)
- c) IEC 60601-1-3:2008+A1:2013 (Edition 2.1)
- d) IEC 60601-2-54:2009+A1:2015 (Edition 1.1)
- e) IEC 60601-2-28:2010 (Edition 3.0)
- IEC 60601-1-6:2010 + A1:2013 (Edition 3.1) f)
- g) IEC 62304:2006 + A1:2016 (Edition 1.1)
- 9) Summary of clinical testing: Clinical testing was not required to establish substantial equivalence because all digital x-ray receptor panels have had previous FDA clearance.
- 10) Conclusion: After analyzing bench and non-clinical tests, it is the conclusion of Sedecal SA. that the modified PHOENIX/AeroDR TX m01 and PHOENIX/mKDR Xpress Digital Diagnostic Mobile X-Ray Systems are as safe and effective as the predicate device, have few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K221803](https://fda.innolitics.com/device/K221803)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
