K221795 · Neos Surgery S.L · JDQ · Oct 19, 2022 · Orthopedic
Device Facts
Record ID
K221795
Device Name
STERN FIX Sternal Stabilization System
Applicant
Neos Surgery S.L
Product Code
JDQ · Orthopedic
Decision Date
Oct 19, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The STERN FIX Sternal Stabilization System is intended for closure and stabilization of the sternum following sternotomy, through the intercostal spaces, in order to promote its fusion.
Device Story
Long-term implantable sternal stabilization system; consists of two-part (male/female) telescopic clamping device; curved arms catch sternum halves laterally; ratchet mechanism secures approximation; tightened via application tool; used in OR by surgeons; allows for emergent/long-term re-entry via cutting; provides mechanical stabilization to promote sternal fusion; modified version utilizes carbon-fiber reinforced (CFR) PEEK-OPTIMA™; includes depth gauge, forceps, cutter, and retractor accessories.
Clinical Evidence
No clinical data. Bench testing only, including functional testing (breaking strength, fatigue), biomechanical testing (breaking strength, lateral distraction), and cadaver testing to verify implantation performance.
Technological Characteristics
Implant material: Carbon-fiber reinforced (CFR) PEEK-OPTIMA™ (ASTM F3333-20). Mechanical fixation: Telescopic ratchet mechanism. Dimensions: Five sizes for sternum thicknesses 9.5-17 mm and widths 17-38 mm. Accessories: Depth gauge, forceps, retractor, and cutters. Sterilization: Radiation (ISO 11137).
Indications for Use
Indicated for closure and stabilization of the sternum following sternotomy in patients requiring sternal fusion.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
{0}------------------------------------------------
October 19, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Neos Surgery SL % Cherita James Regulatory Consultant M Squared Associates, Inc. 127 West 30th Street. 9th Floor New York, New York 10001
Re: K221795
Trade/Device Name: STERN FIX Sternal Stabilization System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ Dated: July 21, 2022 Received: July 22, 2022
Dear Cherita James:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221795
Device Name STERN FIX Sternal Stabilization System
Indications for Use (Describe)
The STERN FIX Sternal Stabilization System is intended for closure and stabilization of the sternum following sternotomy, through the intercostal spaces, in order to promote its fusion.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
The following information is provided as required by 21 CFR § 807.87 for the STERN FIX Sternal Stabilization System Traditional 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
- Sponsor: Neos Surgery S.L. Parc Tecnològic del Vallès C/Ceramistes, 2 08290 Cerdanyola del Vallès Barcelona, Spain
- Cherita James Contact: M Squared Associates, Inc. 127 West 30th Street, 9th Floor New York, New York 10001 Ph: 347-954-0624 Fax: 703-562-9797 Email: Cjames@MSquaredAssociates.com
Date of Submission: June 16, 2022 Proprietary Name: STERN FIX Sternal Stabilization System Common Name: Cerclage, fixation Regulatory Class: II Regulation: 888.3010 Bone fixation cerclage Product Codes: JDQ Primary Predicate Device(s): STERN FIX Sternal Stabilization System - K211613
### Device Description
The STERN FIX is a long-term implantable postoperative sternal stabilization system that closes and stabilizes the sternum after a sternotomy. The STERN FIX is a clamping device consisting of two parts, male and female, which match telescopically with one another to form the sternal stabilization system. Both male and female parts have a curved arm that catches one of the two
{4}------------------------------------------------
halves of the sternum laterally and approximates them until the sternum is securely fixed. Five sizes of the STERN FIX are available for use with different sternum thicknesses. The STERN FIX can be cut and removed for emergent, and long-term, re-entry through the sternum. The predicate device was originally made of PEEK-OPTIMA™. This 510(k) introduces a new variant of the product made of carbon-fiber reinforced (CFR) PEEK-OPTIMA™. This new variant also presents some minor geometry changes with respect to the predicate device, to adapt the product to the new material.
## Indications for Use
The STERN FIX Sternal Stabilization System is intended for closure and stabilization of the sternum following sternotomy, through the intercostal spaces, in order to promote its fusion.
| Device | STERN FIX<br>(Current product, cleared in K211613) | STERN FIX<br>(Modified) |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company | NEOS Surgery S.L. | NEOS Surgery S.L. |
| 510(k) No. | K211613 | Present document |
| Product Code | JDQ- Cerclage, Fixation | Same |
| Indications for Use | The STERN FIX Sternal Stabilization System<br>is intended for closure and stabilization of the<br>sternum following sternotomy, through the<br>intercostal spaces, in order to promote its<br>fusion. | Same |
| Method of Fixation to Sternum | Peristernally, through the intercostal spaces.<br>Generally, 5 STERN FIX devices are<br>recommended per median sternotomy<br>closure, but they can be used in combination<br>with other sternal closure systems such as<br>wires or plates. | Same |
| Device | STERN FIX<br>(Current product, cleared in K211613) | STERN FIX<br>(Modified) |
| Device Design | The STERN FIX device consists of two parts, male and female, which match telescopically with one another to form the fixation system. Both male and female parts have a curved arm that catches laterally one of the two halves of the sternum and approximates them until the sternum is securely fixed. Both parts present a set of teeth that, together, form a ratchet mechanism that allows their movement in the closing direction and, at the same time, impedes their backward movement. The tightening of both implant parts (male and female) is carried out using an application tool.<br><br>The STERN FIX can be cut and removed for emergent, and long-term, re-entry through the sternum. | Same |
| Range of sternum thicknesses and widths covered | <b>Thickness:</b> From 9.5 to 17 mm<br><b>Width:</b> From 17 to 38 mm | Same |
| Accessories | Four instruments are used to implant the device:<br>• A depth gauge (SCI0000) to select the appropriate STERN FIX size.<br>• Forceps (SCI0501) to tighten and adjust the STERN FIX during its implantation.<br>• A cutter (SCI0200) to cut the excess segment of the male part or cut the device safely in case of need of device removal.<br>• A retractor (SCI0300) to ease the introduction of the STERN FIX in the intercostal spaces. | Besides the four instruments already included in the cleared version of the product (SCI0000, SCI0501, SCI0200, SCI0300), a new alternate cutter (SCI0700) is added. This instrument may be used as an alternative to SCI0200, if preferred by the user, to cut the excess segment of the male part. |
| Material Composition | PEEK-OPTIMA™ LT3, according to ASTM F2026 | PEEK-OPTIMA™ LT1CA30, according to ASTM F3333-20 |
| Sizes | Five sizes that allow implantation in sternal bones of all thicknesses considered (from 9.5 mm to 17 mm). Its length can be adjusted to fit each sternal bone width by cutting the excess segment of the male part. | Same |
| Implant life | Long-term implant | Same |
### Technological characteristics, comparison to predicate device
{5}------------------------------------------------
# Technological Characteristics and Substantial Equivalence
The changes proposed for the modified STERN FIX do not change the indications for use of the original product, nor its contraindications, potential side effects or adverse events. The group of users and the use environment also remain unchanged. The material change does not imply any
{6}------------------------------------------------
new or increased biocompatibility concerns, based on the risk assessment performed. The changes on the geometry do not modify the critical dimensions of the product, nor do they imply a modification of the design requirements or the final product specifications. Similarly, the sternum thickness and width range for which the device can be used, as well as the implantation method, remain the same. Finally, the addition of a new accessory (SCI0700) does not affect the directions of use nor the implantation method of the device.
In summary, the modified STERN FIX is substantially equivalent to its predicate device with regards to indications for use, technological characteristics, principles of operation and performance.
### Performance testing
Performance testing was conducted to characterize the capacity of the modified STERN FIX, as a sternal stabilization system, to withstand the forces that act upon the sternum and the whole fixation system once the products have been implanted. The results of the tests confirm the adequate performance of the device, and that it is able to achieve its intended use.
Performance testing has demonstrated that the technological characteristics of the modified STERN FIX do not raise any new safety or effectiveness issues and that it will perform as intended in the clinical setting. A summary of the tests performed follows:
| Test | Conclusions |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Functional testing | |
| Breaking strength | All tested samples meet the specifications. Functionality of the<br>devices is demonstrated. |
| Fatigue test | All tested samples meet the specifications. Functionality of the<br>devices is demonstrated. |
| Biomechanical testing | |
| Breaking strength | All tested samples meet the specifications. The STERN FIX<br>system has an adequate biomechanical behavior when subject to<br>relevant efforts. |
| Lateral distraction test | All tested samples meet the specifications. The STERN FIX<br>system has an adequate biomechanical behavior when subject to<br>relevant efforts. |
{7}------------------------------------------------
| Cadaver testing | Correct devices implantation is verified in a simulated real-life situation. The devices show adequate performance and safety. |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------|
|-----------------|--------------------------------------------------------------------------------------------------------------------------------|
## Conformance to Standards
The STERN FIX Sternal Stabilization System is in conformance to the following recognized consensus standards:
| Standard | Recognition<br>Number |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| ISO 14971 Third edition 2019-012: Medical devices - Application of risk<br>management to medical devices | 5-125 |
| ISO 14630 Fourth edition 2012-12-01: Non-active surgical implants --<br>General requirements | 11-254 |
| ISO 11607-1 Second edition 2019-02: Packaging for terminally sterilized<br>medical devices - Part 1: Requirements for materials, sterile barrier<br>systems and packaging systems | 14-530 |
| ISO 11607-2 Second edition 2019-02: Packaging for terminally sterilized<br>medical devices - Part 2: Validation requirements for forming, sealing<br>and assembly processes | 14-531 |
| IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION:<br>Medical devices - Part 1: Application of usability engineering to medical<br>devices [Including CORRIGENDUM 1 (2016)] | 5-129 |
| ISO 11137-1 First edition 2006-04-15: Sterilization of health care<br>products - Radiation - Part 1: Requirements for development, validation<br>and routine control of a sterilization process for medical devices<br>[Including: Amendment 1 (2013) and Amendment 2 (2018)] | 14-528 |
| ISO 11137-2 Third edition 2013-06-01: Sterilization of health care<br>products - Radiation - Part 2: Establishing the sterilization dose | 14-409 |
| ISO 11737-1 Third edition 2018-01: Sterilization of health care products<br>- Microbiological methods - Part 1: Determination of a population of<br>microorganisms on product | 14-514 |
| ISO 11737-2 Third edition 2019-12: Sterilization of medical devices - | 14-540 |
| Standard | Recognition<br>Number |
| Microbiological methods - Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process | |
| ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process | 2-258 |
| ISO 10993-5 Third edition 2009-06-01: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity | 2-245 |
| ISO 15223-1 Fourth edition 2021-07: Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements | 5-134 |
| ISO 17665-1 First edition 2006-08-15: Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices | 14-333 |
| ISO 17664 Second edition 2017-10: Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices | 14-515 |
| ASTM F2026-17: Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications | 8-475 |
| ASTM F2503-20: Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment | 8-528 |
{8}------------------------------------------------
## Discussion of clinical testing
No clinical testing was performed to support this submission.
### Conclusion
Based on the comparison of the intended use (based on the indications for use), product technical characteristics, principles of operation and performance, the modified STERN FIX has demonstrated substantial equivalence to the previously cleared STERN FIX product. The differences do not raise new issues of safety or effectiveness, based on the risk assessment
{9}------------------------------------------------
performed and on the results of the verification / validation activities that have been conducted.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.