Image Processing Unit, Model: OPTO-CAM104K, Camera Head, Model: OPTO-CHD104KH
K221781 · Guangdong Optomedic Technologies, Inc. · GCJ · Feb 2, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K221781
Device Name
Image Processing Unit, Model: OPTO-CAM104K, Camera Head, Model: OPTO-CHD104KH
Applicant
Guangdong Optomedic Technologies, Inc.
Product Code
GCJ · Gastroenterology, Urology
Decision Date
Feb 2, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
Image Processing Unit is intended to be used with compatible laparoscopes, camera heads, light sources, monitors, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures. Endoscopic Camera Head is intended to be used with compatible laparoscopes, image processing unit, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.
Device Story
Endoscopic camera system consisting of Image Processing Unit (IPU) and Camera Head; used in laparoscopic procedures. Camera head captures optical images via laparoscope; converts to electronic signals using internal image sensor. IPU acts as control center; processes signals; outputs real-time video to 4K monitors via DVI, HDMI, or SDI cables. System used in surgical settings by physicians; provides real-time visualization to assist surgical decision-making. Benefits include high-resolution imaging for improved anatomical visualization during minimally invasive surgery. Device is reusable; requires high-level disinfection between uses.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (ANSI/AAMI ES 60601-1), electromagnetic compatibility (IEC 60601-1-2), endoscopic equipment safety (IEC 60601-2-18), shipping container performance (ASTM D4169-16), and reprocessing validation (AAMI TIR 30:2016, AAMI TIR 12:2020).
Technological Characteristics
System comprises IPU and Camera Head. Connectivity via LEMO interface; video output via DVI, HDMI, 12G-SDI, 3G-SDI. Power: 110-240V AC, 50/60Hz. Reusable, non-sterile; requires high-level disinfection. Software validated as Moderate Level of Concern. Standards: ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-18, ASTM D4169-16, AAMI TIR 30, AAMI TIR 12.
Indications for Use
Indicated for visualization, image recording, and documentation during laparoscopic procedures in patients requiring endoscopic surgery.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Guangdong OptoMedic Technologies, Inc. Jane Guo Regulatory Affairs Manager Suite 503, Building A. Golden Valley Intellicreation Community. No. 2 Yonganbei Street Foshan, Guangdong 528200 China
February 2, 2023
Re: K221781
Trade/Device Name: Image Processing Unit, Model: OPTO-CAM104K; Camera Head, Model: OPTO-CHD104KH Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: January 4, 2023 Received: January 5, 2023
Dear Jane Guo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Colin K. Chen -S
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221781
Device Name Image Processing Unit: OPTO-CAM104K; Camera Head: OPTO-CHD104KH
#### Indications for Use (Describe)
Image Processing Unit is intended to be used with compatible laparoscopes, camera heads, light sources, monitors, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.
Endoscopic Camera Head is intended to be used with compatible laparoscopes, image processing unit, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow swirl on the left, followed by the word "OptoMedic" in black, bold font. The logo is simple and modern, and it is likely used to represent a company that specializes in eye care.
K221781
# 510(k) Summary
# Prepared in accordance with the requirements of 21 CFR Part 807.92
Date Prepared: July 13, 2022
## I. General Information
| 510(k) Submitter/Owner: | Guangdong OptoMedic Technologies, Inc. |
|------------------------------------|-----------------------------------------------------------------|
| | Suite 503, Building A, Golden Valley Intellicreation Community, |
| | No. 2 Yonganbei Street, Daxu, Guicheng, Nanhai, Foshan, |
| | Guangdong, 528200, P.R. China |
| Establishment Registration Number: | Not yet registered |
| Primary Contact Person: | Jane Guo |
| | Regulatory Affairs Engineer |
| | Tel: +86 (757) 8670 2920 |
| | Email: guoweijuan@optomedic.com |
| Secondary Contact Person: | Alice Lau |
| | Regulatory Affairs Engineer |
| | Tel: +86 (757) 8670 2920 |
| | Email: liuli@optomedic.com |
## II. Device Identification
| Device Trade Name: | Image Processing Unit; Camera Head |
|-----------------------|-------------------------------------------------------------------|
| Common or Usual Name: | Endoscopic Camera System |
| Model: | Image Processing Unit: OPTO-CAM104K<br>Camera Head: OPTO-CHD104KH |
| Regulation Name: | Endoscopes and accessories |
| Regulation Number: | 21 CFR 876.1500 |
| Regulatory Class: | Class II |
| Product Code: | GCJ |
# III. Predicate Device
510(k) Number: K131953
006_6.0_510(k)_Summary
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green and yellow circular graphic on the left, followed by the text "OptoMedic" in a bold, sans-serif font. The "Opto" portion of the text is in a lighter shade of gray, while the "Medic" portion is in a darker, almost black shade of gray.
Product Name: Imagel SPIES
The predicate device has not been subject to a design-related recall.
### IV. Device Description
The Endoscopic Camera System comprises the Image Processing Unit, the Camera Head (including the objective lens), power supply cord and video cables. The Camera Head with integrated cable can be connected to the Image Processing Unit through the LEMO interface.
During surgical procedures, the Endoscopic Camera System is used to provide real-time visible imaging similar to that provided by conventional imaging system used in surgical endoscopy. The area of interest is illuminated with visible light from the light source and the resulting reflected light is collected by a variety of laparoscopes that are attached to the camera head via equipment coupler. The camera head receives optical image from the connected laparoscope and converts optical image into electronic signal using the incorporated image sensor. The Image Processing Unit serves as the control center of the system and should be connected to a compatible light source in order to provide visualization. It can be connected to a 4K video monitor via one of the video cables provided to display image in real time: DVI cable, HDMI cable, 12G-SDI cable, and 3G-SDI cable.
The proposed system is a reusable device and provided non-sterile. The Camera Head should be cleaned and high-level disinfected prior to the first use and after every subsequent use. The recommended compatible Light Source is OPTO-LED104K (510k exempted, product code: NTN).
#### V. Indications for Use
Image Processing Unit (hereafter referred to as "IPU") is intended to be used with compatible laparoscopes, camera heads, light sources, monitors, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.
Endoscopic Camera Head is intended to be used with compatible laparoscopes, image processing unit, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.
| Description | Subject Device | Predicate Device (K131953) |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number | 21 CER 876.1500 | 21 CER 876.1500 |
| Product Code | GCJ | FET |
| Device class | Class II | Class II |
| Indication for use | Image Processing Unit<br>(hereafter referred to as “IPU”)<br>is intended to be used with<br>compatible laparoscopes,<br>camera heads, light sources,<br>monitors, and other ancillary<br>equipment for visualization, | The Imagel SPIES is a camera control<br>unit (CCU) for use with camera heads or<br>video endoscopes for visualization,<br>image recording and documentation<br>during general endoscopic and<br>microscopic procedures. |
| | | |
| | image recording and<br>documentation during<br>laparoscopic procedures.<br>Endoscopic Camera Head is<br>intended to be used with<br>compatible laparoscopes,<br>image processing unit, and<br>other ancillary equipment for<br>visualization, image recording<br>and documentation during<br>laparoscopic procedures. | |
| Prescription/<br>Over-the-counter<br>use | Prescription | Prescription |
| Testing | ANSI/AAMI ES 60601-1:<br>2005+A2 (R2012) +A1<br>IEC 60601-1-2:2014<br>IEC 60601-2-18:2009<br>ASTM D4169-16<br>AAMI TIR 30:2016<br>AAMI TIR 12:2020 | IEC 60601-1<br>IEC 60601-1-2 |
| Dimensions | IPU:<br>422mm(L)×370mm(W)×149<br>mm(H)<br>Camera Head:<br>92mm(L)×39mm(W)×50mm(<br>H) | Image1 S Connect TC200 :<br>30.5cm×5.4cm×32.0cm (W×H×D)<br>TH 100&TH 101:<br>100mm×48mm×38mm(L×H×W)<br>TH 103:<br>88mm×47mm×35mm(L×H×W)<br>TH 104:<br>100mm×49mm×39mm(L×H×W) |
| Weight | IPU: 9Kg±15%<br>Camera Head: 400(g)±20 | Image1 S Connect TC200: 2.1kg<br>Image1 S H3-Link TC300: 1.86kg<br>Image1 S X-Link TC301: 1.86kg<br>TH100: 270g<br>TH101: 270g<br>TH103: 227g<br>TH104: 299g |
| Power supply | 110-240V AC, 50/60Hz | 100-120VAC, 50/60Hz<br>200-240VAC, 50/60Hz |
| Video Output | 3G-SDI×4<br>HDMI ×1<br>DVI ×1 | 2×DVI-D<br>1×3G-SDI |
| | | |
| | | 12G-SDI ×1 |
| Signal<br>format | output | 4096×2160p, 3840×2160p, 1920×1080p, 50/60Hz<br>1920×1080p,50Hz/60Hz |
| Sensor | | 4K CMOS sensor assembly<br>3×1/3" |
# VI. Comparison of Technological Characteristics with The Predicate Device
Table 1 General Comparison
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "OptoMedic" in a bold, sans-serif font. The "O" in "Opto" is stylized with a green and yellow gradient circle that is open on the right side. The rest of the letters are black.
.
1
..............................................................................................................................................................................
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for OptoMedic. The logo consists of a green circular design on the left, followed by the text "OptoMedic" in black. The circular design has a gradient of green, with the top being a lighter shade and the bottom being a darker shade. The text is in a bold, sans-serif font.
The differences technological characteristics do not raise different questions of safety and effectiveness.
### VII. Performance data
Non clinical tests were conducted to verify that the proposed device met all design specifications as is Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ANSI/AAMI ES 60601-1: 2005+A2 (R2012) +A1 Medical Electrical Equipment- Part 1: . General requirements for basic safety and essential performance.
- IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic . safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests.
- IEC 60601-2-18:2009 Medical electrical equipment Part 2-18: Particular requirements for . the basic safety and essential performance of endoscopic equipment
- ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and . Systems.
- AAMI TIR 30:2016 A compendium of processes, materials, test methods, and acceptance . criteria for cleaning reusable medical devices
- AAMI TIR 12:2020 Designing, testing, and labeling reusable medical devices for reprocessing . in health care facilities: A guide for medical device manufacturers
The software of the proposed device was validated as Moderate level of concern (LoC) in accordance with the following guidance documents: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff
Performance testing were also conducted on the subject device and demonstrate that the proposed system performs according to specifications and functions as intended.
#### VIII. Conclusions
The performance testing summarized above supports a substantial equivalence determination. The performance testing demonstrate that the subject device is as safe and as effective as the legally marketed predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.