← Product Code FRO · K221720

# Extra Silver Gelling Fiber Dressing (K221720)

_Winner Medical Co., Ltd. · FRO · Dec 21, 2022 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221720

## Device Facts

- **Applicant:** Winner Medical Co., Ltd.
- **Product Code:** FRO
- **Decision Date:** Dec 21, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

Prescription Use: Under the supervision of a healthcare professional. Extra Silver Gelling Fiber Dressing may be used for the management of moderate to heavily exuding chronic and acute wounds as follow: - Partial thickness (second degree) burns; - Pressure ulcers (partial and full thickness); - Leg ulcers (venous stasis ulcers, arterial ulcers and leg ulcers of mixed etiology); - Diabetic foot ulcers; - Surgical wounds that heal by primary intent such as dermatological and surgical incisions; - Surgical wounds left to heal by secondary intention such as dehisced surgical incisions and donor sites; - Traumatic wounds OTC Use: Extra Silver Gelling Fiber Dressing may be used for the management of: - Minor Abrasions - Minor Lacerations - Minor cuts - Minor scalds and burns

## Device Story

Extra Silver Gelling Fiber Dressing is a sterile, non-woven pad or ribbon dressing composed of sodium carboxymethylcellulose (CMC) fibers, strengthening polyester fibers, and 1.2% ionic silver. Used in clinical or home settings for wound management; applied by healthcare professionals or patients. Dressing absorbs wound exudate, transforming into a soft, conformable gel that maintains a moist healing environment. Ionic silver provides antimicrobial activity within the dressing for up to 7 days, reducing bacterial colonization. Strengthening fibers enhance tensile strength compared to predicate. Output is a physical barrier and antimicrobial environment; aids healing by managing exudate and reducing bioburden.

## Clinical Evidence

No clinical data included. Bench testing only. Performance verified via ISO 10993 (cytotoxicity, irritation, sensitization, systemic toxicity), ASTM F88/F88M-15 (seal strength), ASTM F1929-15 (leak detection), and AATCC 100-2012 (antimicrobial efficacy). In vitro testing demonstrated >4 log reduction of 8 bacterial strains (Gram-positive and Gram-negative) over 7 days.

## Technological Characteristics

Non-woven pad/ribbon composed of sodium carboxymethylcellulose, polyester strengthening fibers, and 1.2% ionic silver. Sterilized via gamma radiation (ISO 11137-2:2013). Biocompatibility per ISO 10993-1. Mechanical properties verified via ASTM F88/F88M-15 and ASTM F1929-15. No software or electronic components.

## Predicate Devices

- Silver Gelling Fiber Dressing ([K210466](/device/K210466.md))

## Reference Devices

- AQUACEL™ Ag EXTRATM Hydrofiber™ Dressing with Silver and Strengthening Fiber ([K121275](/device/K121275.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

December 21, 2022

Winner Medical Co., Ltd. Mingni Liu Regulator Affairs Specialist Winner Industrial Park, No. 660 Bulong Road, Longhua District Shenzhen, Guangdong 518109 China

Re: K221720

Trade/Device Name: Extra Silver Gelling Fiber Dressing Regulatory Class: Unclassified Product Code: FRO Dated: November 25, 2022 Received: November 25, 2022

Dear Mingni Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Julie A. Morabito -S

Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K221720

#### Device Name

Extra Silver Gelling Fiber Dressing

#### Indications for Use (Describe)

Prescription Use:

Under the supervision of a healthcare professional. Extra Silver Gelling Fiber Dressing may be used for the management of moderate to heavily exuding chronic and acute wounds as follow:

- · Partial thickness (second degree) burns;
- · Pressure ulcers (partial and full thickness);
- · Leg ulcers (venous stasis ulcers, arterial ulcers and leg ulcers of mixed etiology);
- · Diabetic foot ulcers;
- · Surgical wounds that heal by primary intent such as dermatological and surgical incisions;
- · Surgical wounds left to heal by secondary intention such as dehisced surgical incisions and donor sites;
- · Traumatic wounds

#### OTC Use:

Extra Silver Gelling Fiber Dressing may be used for the management of:

- · Minor Abrasions
- · Minor Lacerations
- Minor cuts
- Minor scalds and burns

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

> Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

The assigned 510(k) Number: K221720

- 1. Date of Submission: December 14, 2022
#### 2. Submitter Identification

#### Winner Medical Co., Ltd.

Winner Industrial Park, No. 660 Bulong Road, Longhua District, Shenzhen City, Guangdong Province, 518109, China

Contact Person: Mingni Liu

Position: Regulatory Affairs Specialist Tel: +86-755 28138888 Fax: +86-755 28134588 Email: 2346@winnermedical.com

### 3. Identification of Proposed Device

Trade/Proprietary Name: Extra Silver Gelling Fiber Dressing Common name: Wound or Burn Dressing

## Regulatory Information

Classification Name: Dressing, Wound, Drug; Classification: Unclassified Product Code: FRO; Review Panel: General & Plastic Surgery;

## 4. Identification of Predicate Device

Predicate Device: 510(k) Number: K210466 Product Name: Silver Gelling Fiber Dressing Manufacturer: Winner Medical Co., Ltd. Reference Device: 510(k) Number: K121275 Product Name: AQUACEL™ Ag EXTRATM Hydrofiber™ Dressing with Silver and Strengthening Fiber Manufacturer: ConvaTec Inc.

{4}------------------------------------------------

#### 5. Device Description

Extra Silver Gelling Fiber Dressing is a soft, sterile, non-woven pad or ribbon dressing composed of sodium carboxymethylcellulose (CMC) fibers, strengthening fibers and 1.2% ionic silver. This dressing absorbs wound fluid and creates a soft gel that conforms to the wound surface, maintains a moist environment. A moist wound environment supports the body's healing process. The silver antimicrobial may help reduce bacterial colonization within the dressing for up to 7 days.

The dressings are supplied sterile in a range of sizes, ranging in area from 25cm2 to 600cm². All dressings have the exactly the same material, chemical, and physical properties and are different only in size.

All dressings are sterilized and sold after sterilization by gamma radiation using conditions validated following ISO 11137-2:2013.

#### 6. Indications for use

#### OTC Use:

Extra Silver Gelling Fiber Dressing may be used for the management of:

- Minor Abrasions
- Minor Lacerations
- Minor cuts
- Minor scalds and burns

Prescription Use:

Under the supervision of a healthcare professional. Extra Silver Gelling Fiber Dressing may be used for the management of wounds as follow:

- · Partial thickness (second degree) burns;
- · Pressure ulcers (partial and full thickness);
- · Leg ulcers (venous stasis ulcers, arterial ulcers and leg ulcers of mixed etiology);
- Diabetic foot ulcers;
- · Surgical wounds that heal by primary intent such as dermatological and surgical incisions;

• Surgical wounds left to heal by secondary intention such as dehisced surgical incisions and donor sites;

{5}------------------------------------------------

• Traumatic wounds

The indications for use are identical to the primary predicate device (K210466).

#### 7. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications.

The test results demonstrated that the proposed device complies with the following standards:

- ISO 10993-5:2009 Biological Evaluation of Medical Devices- Part 5: Tests For In Vitro Cytotoxicity.
- ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization.
- ISO 10993-11:2017 Biological Evaluation Of Medical Devices- Part 11: Tests For Systemic Toxicity.
- ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials.
- ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical

Packaging by Dye Penetration

USP <85> Bacterial Endotoxins Test AATCC 100-2012 Antibacterial Finishes on Textile Materials: Assessment of

## 8. Clinical Test Conclusion

No clinical study is included in this submission.

## 9. Technological Characteristics Comparison

Extra Silver Gelling Fiber Dressing has the same intended use and indications for use as the predicate device but has an additional tensile strength performance than the predicate device. This is achieved by addition strengthening fibers. The differences in the composition and technology, as detailed below, is explained by the addition of strengthening fibers in the proposed device as compared to the predicate device. There are no differences in operation and directions for use between Extra Silver Gelling Fiber Dressing and the predicate device.

Table 1 Comparison of Intended use, Design and Technological Characteristics

| Item | Proposed Device | Predicate Device (K210466) | Comparison |
|------|-----------------|----------------------------|------------|
|------|-----------------|----------------------------|------------|

{6}------------------------------------------------

| Product<br>Code              | FRO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | FRO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Equivalent |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Class                        | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Unclassified                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Equivalent |
| Indication<br>for Use        | Over-the-Counter Use:<br>Silver Gelling Fiber Dressing may<br>be used for the management of:<br>• Minor abrasions<br>• Minor lacerations<br>• Minor cuts<br>• Minor scalds and burns<br>Prescription Use:<br>Under the supervision of a<br>healthcare professional:<br>Silver Gelling Fiber Dressing may<br>be used for the management of<br>wounds as follow:<br>• Partial thickness (second degree)<br>burns;<br>• Pressure ulcers (partial and full<br>thickness);<br>• Leg ulcers (venous stasis ulcers,<br>arterial ulcers and leg ulcers of<br>mixed etiology);<br>• Diabetic foot ulcers;<br>• Surgical wounds that heal by<br>primary intent such as<br>dermatological and surgical<br>incisions;<br>• Surgical wounds left to heal by<br>secondary intention such as<br>dehisced surgical incisions and<br>donor sites;<br>• Traumatic wounds | Over-the-Counter Use:<br>Silver Gelling Fiber Dressing<br>may be used for the<br>management of:<br>• Minor abrasions<br>• Minor lacerations<br>• Minor cuts<br>• Minor scalds and burns<br>Prescription Use:<br>Under the supervision of a<br>healthcare professional:<br>Silver Gelling Fiber Dressing<br>may be used for the<br>management of wounds as<br>follow:<br>• Partial thickness (second<br>degree) burns;<br>• Pressure ulcers (partial and<br>full thickness);<br>• Leg ulcers (venous stasis<br>ulcers, arterial ulcers and leg<br>ulcers of mixed etiology);<br>• Diabetic foot ulcers;<br>• Surgical wounds that heal<br>by primary intent such as<br>dermatological and surgical<br>incisions;<br>• Surgical wounds left to heal<br>by secondary intention such as<br>dehisced surgical incisions and<br>donor sites;<br>• Traumatic wounds | Equivalent |
| Principle<br>of<br>operation | The dressing absorbs wound fluid<br>and creates a soft, conformable gel,<br>which maintains a moist wound<br>environment to support the healing<br>process. Silver ionic present in the<br>fiber for reducing bacteria<br>colonization in the dressing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | The dressing absorbs wound<br>fluid and creates a soft,<br>conformable gel, which<br>maintains a moist wound<br>environment to support the<br>healing process. Silver ionic<br>present in the fiber for<br>reducing bacteria colonization<br>in the dressing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Equivalent |

{7}------------------------------------------------

| Material<br>componen<br>ts    | Composed of sodium<br>carboxymethylcellulose,<br>strengthening fibers (polyester<br>thread) and ionic silver                                                                                          | Composed of sodium<br>carboxymethylcellulose and<br>ionic silver                                                                                                                                               | Similar    |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Technolog<br>y                | Silver impregnated sodium<br>carboxymethylcellulose fibers<br>needling together and stitch<br>bonded with polyester thread to<br>non-woven pad or ribbon;                                             | Silver impregnated sodium<br>carboxymethylcellulose<br>fibers needling together to<br>non-woven pad or ribbon                                                                                                  | Similar    |
| Characteri<br>stics           | • Sterile<br>• Absorbs exudate (including<br>bacteria)<br>• Forms a soft conformable gel<br>• The silver in the dressing kills<br>bacteria held in the dressing<br>• May require a secondary dressing | • Sterile<br>• Absorbs exudate (including<br>bacteria)<br>• Forms a soft conformable<br>gel<br>• The silver in the dressing<br>kills bacteria held in the<br>dressing<br>• May require a secondary<br>dressing | Equivalent |
| Antibacter<br>ial<br>Duration | 7 days                                                                                                                                                                                                | 7 days                                                                                                                                                                                                         | Equivalent |
| Single Use                    | Yes                                                                                                                                                                                                   | Yes                                                                                                                                                                                                            | Equivalent |
| Sterilizati<br>on             | Gamma Radiation                                                                                                                                                                                       | Gamma Radiation                                                                                                                                                                                                | Equivalent |
| Biocompa<br>tibility          | Biocompatibility in accordance to<br>10993-1(breached or compromised<br>surfaces with prolonged<br>contact(>24h to 30d))                                                                              | Biocompatibility in<br>accordance to 10993-<br>1(breached or compromised<br>surfaces with prolonged<br>contact(>24h to 30d))                                                                                   | Equivalent |

The proposed device has same intended use, principle of operation and characteristics to the predicate device.

## 10. Non-Clinical Testing - Performance

Mechanical testing was conducted for this submission. Absorbency performance is equivalent to the predicate device; tensile strength performance is superior to the predicate device due to the addition strengthening fibers design modification.

Extra Silver Gelling Fiber Dressing in vitro testing has demonstrated > 4 logreduction of four gram positive bacteria (Vancomycin- resistant enterococcus (VRE), Bacillus subtilis, Staphylococcus aureus and Methicillin-resistant Staphylococcus aureus (MRSA)), four 

{8}------------------------------------------------

gram negative bacteria (Serratia marcescens, Pseudomonas aeruginosa, Escherichia coli, and Klebsiella pneumonia) challenge organisms within the dressing for up to 7 days.

## 11. Non-Clinical Testing – Biocompatibility

Biocompatibility testing was conducted for this submission in accordance with the US Food and Drug Administration's guidance entitled Use of International Standard ISO - 10993: 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing' and test results meet the requirements.

## 12. Conclusions

The proposed device and the predicate device underwent evaluation that demonstrates substantial equivalence in the intended use of each device, biocompatibility, safety, efficacy, environment of use, and the principles of operation. Therefore, the proposed device demonstrates substantial equivalence to the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K221720](https://fda.innolitics.com/device/K221720)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
