← Product Code [FAJ](/productcode/FAJ) · K221683

# CYSTO-NEPHRO VIDEOSCOPE (OLYMPUS CYF-V2), CYSTO-NEPHRO VIDEOSCOPE (OLYMPUS CYF-V2R), CYSTO-NEPHRO VIDEOSCOPE (OLYMPUS CYF-VH) and CYSTO-NEPHRO VIDEOSCOPE (OLYMPUS CYF-VHR) (K221683)

_Olympus Medical Systems Corporation · FAJ · Jan 20, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221683

## Device Facts

- **Applicant:** Olympus Medical Systems Corporation
- **Product Code:** [FAJ](/productcode/FAJ.md)
- **Decision Date:** Jan 20, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

VISERA CYSTO-NEPHRO VIDEOSCOPEs OLYMPUS CYF TYPE V2, CYF TYPE V2R have been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, and kidney. CYSTO-NEPHRO VIDEOSCOPEs OLYMPUS CYF VH, and CYF VHR have been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, and kidney.

## Device Story

Flexible cysto-nephro videoscope for endoscopic diagnosis and treatment in bladder, urethra, and kidney. Inputs: optical images captured via distal color CCD sensor. Operation: device connects to Olympus video system center, light source, and monitor; physician navigates scope into anatomy; images processed by system for real-time visualization (WLI, NBI modes). Output: live endoscopic video feed for clinician viewing. Used in hospital/healthcare facility settings by physicians. Benefits: enables minimally invasive visualization and therapeutic access to urinary tract. Device includes control section, insertion tube, and connector. Reusable; requires reprocessing (AER/manual cleaning, ETO/STERRAD sterilization).

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including reprocessing validation, biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1/2-18), EMC (IEC 60601-1-2), and performance testing (thermal safety, durability, color performance, photobiological safety, image quality, resolution, NBI effectiveness).

## Technological Characteristics

Reusable flexible endoscope. Materials: fluorine resin (insertion tube), fluoro rubber (bending section), polysulfone/polyphenylsulfone (distal end), glass (lenses), stainless steel (channels), PTFE (channel liner). Sensing: 1-chip color CCD. Illumination: light guide. Connectivity: proprietary Olympus video system interface. Reprocessing: ETO, STERRAD NX. Standards: ANSI/AAMI ES 60601-1, IEC 60601-2-18, IEC 60601-1-2, ISO 10993-1, ISO 14971.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2, CYF TYPE VA2, CYF TYPE V2R ([K133538](/device/K133538.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 20, 2023

Olympus Medical Systems Corporation % Teffany Hutto Regulatory Affairs Consultant Olympus Corporation of the Americas 3500 Corporate Parkway P.O. Box 610 Center Valley, PA 18034

Re: K221683

> Trade/Device Name: VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2, CYF TYPE V2R CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VH, CYF-VHR Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FAJ Dated: December 22, 2022 Received: December 22, 2022

Dear Teffany Hutto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Mark J. Antonino -S

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K221683

Device Name

VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2, CYF TYPE V2R

Indications for Use (Describe)

VISERA CYSTO-NEPHRO VIDEOSCOPEs OLYMPUS CYF TYPE V2, CYF TYPE V2R have been designed to be used with an Olympus video system center, light source, documentation equipment,

monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, and kidney.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K221683

## Device Name

CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VH, CYF-VHR

Indications for Use (Describe)

CYSTO-NEPHRO VIDEOSCOPEs OLYMPUS CYF VH, and CYF VHR have been designed to be used with an Olympus video system center, light source, documentation equipment,

monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, and kidney.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are all capitalized and evenly spaced. A registered trademark symbol is located to the right of the letter "S".

Date Prepared: 01/19/2023

# 510(k) Summary

### 1. GENERAL INFORMATION

- 510(k) Submitter: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, Japan
- Contact Person: Teffany Hutto Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 512-508-6550 Fax: 484-896-7128 Email: teffany.hutto@olympus.com
- . Manufacturing site: Shirakawa Olympus Co., Ltd. 3-1 Okamiyama, Odakura, Nishigo-mura, Nishishirakawa-gun, Fukushima 961-8061, Japan

Aizu Olympus Co., Ltd., 3-1-1 Niiderakita, Aizuwakamatsu-shi, Fukushima 965-8520, Japan

### 2. DEVICE IDENTIFICATION

| ■ Trade/Device Name:    | VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF TYPE V2,<br>CYF TYPE V2R<br>CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VH, CYF-VHR |
|-------------------------|------------------------------------------------------------------------------------------------------------------------|
| ■ Model Name:           | OLYMPUS CYF-V2, CYF-V2R, CYF-VH, CYF-VHR                                                                               |
| ■ Common Name:          | CYSTO-NEPHRO VIDEOSCOPE                                                                                                |
| ■ Regulation Number:    | 876.1500                                                                                                               |
| ■ Regulation Name:      | Endoscope and accessories                                                                                              |
| ■ Regulatory Class:     | Class II                                                                                                               |
| ■ Product Code:         | FAJ (Cystoscope and Accessories, Flexible/Rigid)                                                                       |
| ■ Classification Panel: | Gastroenterology/Urology                                                                                               |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are spaced closely together, and the word is slightly angled upwards from left to right. A registered trademark symbol is located at the top right corner of the word.

### 3. PREDICATE DEVICE

#### Predicate device

| Device name                                        | 510(k) Submitter                 | 510(k) No. |
|----------------------------------------------------|----------------------------------|------------|
| VISERA CYSTO-NEPHRO<br>VIDEOSCOPE                  | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K133538    |
| OLYMPUS CYF TYPE V2, CYF TYPE<br>VA2, CYF TYPE V2R |                                  |            |

#### 4. DEVICE DESCRIPTION

#### ■ General Description of the subject device

The subject device has been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, and kidney.

#### Principle of Operation

The subject devices consist of three parts: the control section section, and the connector section. The basic principle including user interface and operation for the subject device are identical to that of the predicate device.

#### ■ List of device components

The VISERA CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-V2/V2R will be packed and offered together with the devices listed in the table below. They can be used with commercially available Olympus devices as described within the Instruction Manual.

| Model No. | Device Name                           |
|-----------|---------------------------------------|
| BW-411B   | Single Use Combination Cleaning Brush |
| MB-156    | ETO Cap                               |

The CYSTO-NEPHRO VIDEOSCOPE OLYMPUS CYF-VH and CYF-VHR will be packed and offered together with the devices listed in the table below. They can be used with commercially available Olympus devices as described within the Instruction Manual.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. A registered trademark symbol is located at the top right corner of the word. 

| Model No. | Device Name                           |
|-----------|---------------------------------------|
| BW-411B   | Single Use Combination Cleaning Brush |
| MAJ-1538  | Sterilization Cap                     |

# 5. INDICATIONS FOR USE

VISERA CYSTO-NEPHRO VIDEOSCOPEs OLYMPUS CYF TYPE V2, CYSTO-NEPHRO VIDEOSCOPEs OLYMPUS CYF-VH and CYF-VHR have been designed to be used with an Olympus video system center, light source, documentation equipment, monitor, EndoTherapy accessories, and other ancillary equipment for endoscopic diagnosis and treatment within the bladder, urethra, and kidney.

# 6. COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject device has the same technological characteristics and design as the predicate device except for the following features:

- Optical performance
- Shape of the distal end

Validation from non-clinical testing demonstrated that these technological features do not raise any new issues of safety or effectiveness of the subject device.

| Item                                 | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                  |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                      | OLYMPUS CYF-V2, CYF-V2R                                                                                                                                                                                                                                                                                                                                                                                        | OLYMPUS CYF<br>TYPE V2, VA2, V2R<br>(K133538)                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for Use                  | VISERA CYSTO-NEPHRO<br>VIDEOSCOPEs OLYMPUS CYF TYPE<br>V2 and CYF TYPE V2R have been<br>designed to be used with an Olympus<br>video system center, light source,<br>documentation equipment, monitor,<br>EndoTherapy accessories, and other<br>ancillary equipment for endoscopic<br>diagnosis and treatment within the<br>bladder, urethra, and kidney.                                                      | This instrument has been designed to<br>be used with an Olympus video<br>system center, light source,<br>documentation equipment, monitor,<br>EndoTherapy accessories, and other<br>ancillary equipment for endoscopic<br>diagnosis and treatment within the<br>bladder, urethra, ureter, and kidney.                                                                                                             |
| Regulation Number                    | 21CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                 | 21CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                    |
| Regulation Name                      | Endoscope and accessories                                                                                                                                                                                                                                                                                                                                                                                      | Endoscope and accessories                                                                                                                                                                                                                                                                                                                                                                                         |
| Regulatory Class                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                       | Class II                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                  |
| Item                                 | OLYMPUS CYF-V2, CYF-V2R                                                                                                                                                                                                                                                                                                                                                                                        | OLYMPUS CYF<br>TYPE V2, VA2, V2R<br>(K133538)                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code                         | FAJ                                                                                                                                                                                                                                                                                                                                                                                                            | FAJ                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification Panel                 | Gastroenterology/Urology                                                                                                                                                                                                                                                                                                                                                                                       | Gastroenterology/Urology                                                                                                                                                                                                                                                                                                                                                                                          |
| Use Environment                      | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                                   | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                                      |
| Reprocessing                         | [Cleaning]<br>AER and manual<br>[Disinfection]<br>AER and manual<br>[Sterilization]<br>- Ethylene oxide gas<br>- STERRAD NX                                                                                                                                                                                                                                                                                    | [Cleaning]<br>AER and manual<br>[Disinfection]<br>AER and manual<br>[Sterilization]<br>- Ethylene oxide gas<br>- STERRAD NX                                                                                                                                                                                                                                                                                       |
| Single-Use/Reuse                     | Reusable                                                                                                                                                                                                                                                                                                                                                                                                       | Reusable                                                                                                                                                                                                                                                                                                                                                                                                          |
| Duration and type of<br>contact      | Surface-contacting device in contact with<br>mucosal membranes. The contact duration<br>is limited exposure (<24 hours).                                                                                                                                                                                                                                                                                       | Surface-contacting device in contact<br>with mucosal membranes. The<br>contact duration is limited exposure<br>(<24 hours).                                                                                                                                                                                                                                                                                       |
| Depth of Field                       | 3-50mm                                                                                                                                                                                                                                                                                                                                                                                                         | 3-50mm                                                                                                                                                                                                                                                                                                                                                                                                            |
| Direction of View                    | 0°(Forward viewing)                                                                                                                                                                                                                                                                                                                                                                                            | 0°(Forward viewing)                                                                                                                                                                                                                                                                                                                                                                                               |
| Field of View                        | 120°                                                                                                                                                                                                                                                                                                                                                                                                           | 120°                                                                                                                                                                                                                                                                                                                                                                                                              |
| Type of CCD                          | Color CCD                                                                                                                                                                                                                                                                                                                                                                                                      | Color CCD                                                                                                                                                                                                                                                                                                                                                                                                         |
| Number of CCD Chip                   | 1                                                                                                                                                                                                                                                                                                                                                                                                              | 1                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Outer Diameter of<br>Distal End      | ø 4.8mm (bullet shape)                                                                                                                                                                                                                                                                                                                                                                                         | ø 4.8mm (bullet shape)                                                                                                                                                                                                                                                                                                                                                                                            |
| Diameter of Insertion<br>Tube        | ø 5.4mm                                                                                                                                                                                                                                                                                                                                                                                                        | ø 5.4mm                                                                                                                                                                                                                                                                                                                                                                                                           |
| Bending Section<br>Angulation        | Up 210° / Down 120°                                                                                                                                                                                                                                                                                                                                                                                            | Up 210° / Down 120°                                                                                                                                                                                                                                                                                                                                                                                               |
| Working Length                       | 380mm                                                                                                                                                                                                                                                                                                                                                                                                          | 380mm                                                                                                                                                                                                                                                                                                                                                                                                             |
| Instrument Channel<br>inner diameter | ø 2.2 mm                                                                                                                                                                                                                                                                                                                                                                                                       | ø 2.2 mm                                                                                                                                                                                                                                                                                                                                                                                                          |
| Observation mode                     | WLI, NBI                                                                                                                                                                                                                                                                                                                                                                                                       | WLI, NBI                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                  |
| Item                                 | OLYMPUS CYF-V2, CYF-V2R                                                                                                                                                                                                                                                                                                                                                                                        | OLYMPUS CYF<br>TYPE V2, VA2, V2R<br>(K133538)                                                                                                                                                                                                                                                                                                                                                                     |
| Patient contacting<br>material       | Insertion tube: Forced dry-hard type of<br>fluorine resin coat                                                                                                                                                                                                                                                                                                                                                 | Insertion tube: Forced dry-hard type<br>of fluorine resin coat                                                                                                                                                                                                                                                                                                                                                    |
|                                      | Bending section: Fluoro Rubber                                                                                                                                                                                                                                                                                                                                                                                 | Bending section: Fluoro Rubber                                                                                                                                                                                                                                                                                                                                                                                    |
|                                      | Glue: Epoxy glue                                                                                                                                                                                                                                                                                                                                                                                               | Glue: Epoxy glue                                                                                                                                                                                                                                                                                                                                                                                                  |
|                                      | Distal End: Polyphenylsulfone                                                                                                                                                                                                                                                                                                                                                                                  | Distal End: Polyphenylsulfone                                                                                                                                                                                                                                                                                                                                                                                     |
|                                      | Objective lens/ Light guide lens: Glass                                                                                                                                                                                                                                                                                                                                                                        | Objective lens/ Light guide lens:<br>Glass                                                                                                                                                                                                                                                                                                                                                                        |
|                                      | Solder: AuSn Solder                                                                                                                                                                                                                                                                                                                                                                                            | Solder: AuSn Solder                                                                                                                                                                                                                                                                                                                                                                                               |
|                                      | Instrument channel pipe/ Instrument<br>channel joint/ Junction/ Instrument<br>channel port: Stainless Steel                                                                                                                                                                                                                                                                                                    | Instrument channel pipe/ Instrument<br>channel joint/ Junction/ Instrument<br>channel port: Stainless Steel                                                                                                                                                                                                                                                                                                       |
|                                      | Instrument Channel:<br>Polytetrafluoroethylene                                                                                                                                                                                                                                                                                                                                                                 | Instrument Channel:<br>Polytetrafluoroethylene                                                                                                                                                                                                                                                                                                                                                                    |
|                                      | Rubber seal: Silicone                                                                                                                                                                                                                                                                                                                                                                                          | Rubber seal: Silicone                                                                                                                                                                                                                                                                                                                                                                                             |
|                                      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                  |
| Item                                 | OLYMPUS CYF-VH, CYF-VHR                                                                                                                                                                                                                                                                                                                                                                                        | OLYMPUS CYF<br>TYPE V2, VA2, V2R<br>(K133538)                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for Use                  | CYSTO-NEPHRO VIDEOSCOPEs<br>OLYMPUS CYF-VH and CYF-<br>VHRhave been designed to be used<br>with an Olympus video system center,<br>light source, documentation<br>equipment, monitor, EndoTherapy<br>accessories, and other ancillary<br>equipment for endoscopic diagnosis<br>and treatment within the bladder,<br>urethra and kidney.                                                                        | This instrument has been designed to<br>be used with an Olympus video system<br>center, light source, documentation<br>equipment, monitor, EndoTherapy<br>accessories, and other ancillary<br>equipment for endoscopic diagnosis<br>and treatment within the bladder,<br>urethra, ureter, and kidney.                                                                                                             |
| Regulation Number                    | 21CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                 | 21CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                    |
| Regulation Name                      | Endoscope and accessories                                                                                                                                                                                                                                                                                                                                                                                      | Endoscope and accessories                                                                                                                                                                                                                                                                                                                                                                                         |
| Regulatory Class                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                       | Class II                                                                                                                                                                                                                                                                                                                                                                                                          |
| Product Code                         | FAJ                                                                                                                                                                                                                                                                                                                                                                                                            | FAJ                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification Panel                 | Gastroenterology/Urology                                                                                                                                                                                                                                                                                                                                                                                       | Gastroenterology/Urology                                                                                                                                                                                                                                                                                                                                                                                          |
| Use Environment                      | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                                   | Healthcare facility/hospital                                                                                                                                                                                                                                                                                                                                                                                      |
| Reprocessing                         | [Cleaning]<br>AER and manual<br>[Disinfection]<br>AER and manual<br>[Sterilization]<br>- Ethylene oxide gas<br>- STERRAD NX                                                                                                                                                                                                                                                                                    | [Cleaning]<br>AER and manual<br>[Disinfection]<br>AER and manual<br>[Sterilization]<br>- Ethylene oxide gas<br>- STERRAD NX                                                                                                                                                                                                                                                                                       |
| Single-Use/Reuse                     | Reusable                                                                                                                                                                                                                                                                                                                                                                                                       | Reusable                                                                                                                                                                                                                                                                                                                                                                                                          |
| Duration and type of<br>contact      | Surface-contacting device in contact<br>with mucosal membranes. The contact<br>duration is limited exposure (<24<br>hours).                                                                                                                                                                                                                                                                                    | Surface-contacting device in contact<br>with mucosal membranes. The contact<br>duration is limited exposure (<24<br>hours).                                                                                                                                                                                                                                                                                       |
| Depth of Field                       | 3-50mm                                                                                                                                                                                                                                                                                                                                                                                                         | 3-50mm                                                                                                                                                                                                                                                                                                                                                                                                            |
| Direction of View                    | 0°(Forward viewing)                                                                                                                                                                                                                                                                                                                                                                                            | 0°(Forward viewing)                                                                                                                                                                                                                                                                                                                                                                                               |
| Field of View                        | 120°                                                                                                                                                                                                                                                                                                                                                                                                           | 120°                                                                                                                                                                                                                                                                                                                                                                                                              |
| Type of CCD                          | Color CCD                                                                                                                                                                                                                                                                                                                                                                                                      | Color CCD                                                                                                                                                                                                                                                                                                                                                                                                         |
| Number of CCD Chip                   | 1                                                                                                                                                                                                                                                                                                                                                                                                              | 1                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Outer Diameter of<br>Distal End      | ø 2.7mm (bullet shape)                                                                                                                                                                                                                                                                                                                                                                                         | ø 4.8mm (bullet shape)                                                                                                                                                                                                                                                                                                                                                                                            |
| Diameter of Insertion<br>Tube        | ø 5.5mm                                                                                                                                                                                                                                                                                                                                                                                                        | ø 5.4mm                                                                                                                                                                                                                                                                                                                                                                                                           |
| Bending Section<br>Angulation        | Up 220° / Down 130°                                                                                                                                                                                                                                                                                                                                                                                            | Up 210° / Down 120°                                                                                                                                                                                                                                                                                                                                                                                               |
| Working Length                       | 380mm                                                                                                                                                                                                                                                                                                                                                                                                          | 380mm                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                      | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                  |
| Item                                 | OLYMPUS CYF-VH, CYF-VHR                                                                                                                                                                                                                                                                                                                                                                                        | OLYMPUS CYF<br>TYPE V2, VA2, V2R<br>(K133538)                                                                                                                                                                                                                                                                                                                                                                     |
| Instrument Channel<br>inner diameter | ø 2.2 mm                                                                                                                                                                                                                                                                                                                                                                                                       | ø 2.2 mm                                                                                                                                                                                                                                                                                                                                                                                                          |
| Observation mode                     | WLI, NBI                                                                                                                                                                                                                                                                                                                                                                                                       | WLI, NBI                                                                                                                                                                                                                                                                                                                                                                                                          |
| Patient contacting<br>material       | Insertion tube: Forced dry-hard type of<br>fluorine resin coat<br>Bending section: Fluoro Rubber<br>Glue: Epoxy glue<br>Distal End/ Light guide lens:<br>Polysulfone<br>Objective lens: Glass<br>Solder: AuSn Solder<br>Instrument channel pipe/ Instrument<br>channel joint/ Junction/ Instrument<br>channel port: Stainless Steel<br>Instrument Channel:<br>Polytetrafluoroethylene<br>Rubber seal: Silicone | Insertion tube: Forced dry-hard type of<br>fluorine resin coat<br>Bending section: Fluoro Rubber<br>Glue: Epoxy glue<br>Distal End: Polyphenylsulfone<br>Objective lens/ Light guide lens: Glass<br>Solder: AuSn Solder<br>Instrument channel pipe/ Instrument<br>channel joint/ Junction/ Instrument<br>channel port: Stainless Steel<br>Instrument Channel:<br>Polytetrafluoroethylene<br>Rubber seal: Silicone |

A side-by-side comparison of the subject device and the predicate device is provided below.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are closely spaced together, creating a solid block of text. The font is sans-serif and appears to be a custom design. There is a registered trademark symbol to the right of the word.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are spaced closely together, and the word is oriented horizontally. There is a registered trademark symbol to the right of the letter "S".

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are closely spaced together, creating a solid block of text. A registered trademark symbol is located in the upper right corner of the image.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The letters are closely spaced together, creating a solid block of text. A registered trademark symbol is located to the right of the letter "S".

# 7. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

#### 1) Reprocessing validation testing

Reprocessing instruction and reprocessing method validation testing were conducted and documentations were provided as recommended by Guidance for Industry and Food and Drug Administration Staff, "Reprocessing Medical Devices in Health Care Setting: Validation Methods and Labeling".

### 2) Biocompatibility testing

Biocompatibility testing were conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:

- Cytotoxicity Study Using the ISO Elution Method
- ISO Intracutaneous Study in Rabbits
- ISO Guinea Pig Maximization Sensitization Test
- ISO Intracutaneous Study in Rabbits
- USP Rabbit Pyrogen Study
- ISO Acute Systemic Toxicity Study in Mice

#### 3) Software verification and validation testing

Software verification and validation testing were conducted, and documentations were provided as

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. There is a registered trademark symbol in the upper right corner of the image. The background is white.

recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".

#### 4) Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted in accordance with the ANSI/AAMI ES 60601-1:2005/(R)2012 and A1:2012 and IEC 60601-2-18:2009 standards for safety and the IEC 60601-1-2:2014 standards for EMC.

#### 5) Performance testing - Bench

Bench testing as listed below were conducted to ensure that the subject device performs as intended and meet design specifications.

- Thermal Safety
- Composite Durability
- Color Performance
- Photobiological Safety
- Noise and Dynamic Range
- Image Intensity Uniformity
- Field of View and Direction of View
- Resolution
- Effectiveness NBI

#### 6) Performance testing - Animal

No animal study was performed to demonstrate substantial equivalence.

#### 7) Performance testing - Clinical

No clinical study was performed to demonstrate substantial equivalence.

#### 8) Risk management

Risk management was performed in accordance with ISO 14971:2007. The design verification tests and their acceptance criteria were performed and identified as a result of this risk management.

### 8. CONCLUSIONS

Based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate device, VISERA CYSTO-NEPHRO VIDEOSCOPEs OLYMPUS CYF TYPE V2, CYF TYPE V2R, CYSTO-NEPHRO VIDEOSCOPEs OLYMPUS CYF-VH and CYF-VHR raise no new issue of safety and effectiveness and are substantially equivalent to the predicate device in terms of safety, effectiveness and performance.

---

**Source:** [https://fda.innolitics.com/device/K221683](https://fda.innolitics.com/device/K221683)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
