K221606 · Chendu Medart Medical Scientific Co., Ltd. · GXR · Nov 21, 2023 · Neurology
Device Facts
Record ID
K221606
Device Name
Absorbable Cranial Flap Fixation System
Applicant
Chendu Medart Medical Scientific Co., Ltd.
Product Code
GXR · Neurology
Decision Date
Nov 21, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Absorbable Cranial Flap Fixation System is intended to be used for fixation of cranial bone flaps following craniotomy.
Device Story
Absorbable Cranial Flap Fixation System provides rigid fixation and coplanar alignment of cranial bone flaps post-craniotomy. Device consists of lower disc, upper disc, connecting rod, and rotating lock; upper disc moves axially along rod via rotation to grip bone flap. Manufactured from poly-L-lactic acid; supplied sterile for single use. Includes battery-powered Heat/Contouring Pen accessory for trimming excess rod. Used in clinical/surgical settings by neurosurgeons. Output is mechanical stabilization of bone flap; facilitates healing by maintaining alignment. Benefits include rigid fixation using resorbable materials, eliminating need for permanent metal implants.
Clinical Evidence
No clinical data. Evidence is based on bench testing and biocompatibility evaluation. Bench testing compared tripping force, compression force, and torque force against the predicate, showing no statistical difference. Biocompatibility testing included cytotoxicity (ISO 10993-5), sensitization, intracutaneous reactivity, acute/subchronic systemic toxicity, pyrogenicity, bone implantation, degradation, and genotoxicity (ISO 10993-3, 6, 10, 11, 13). All tests passed.
Indicated for fixation of cranial bone flaps following craniotomy in patients undergoing such procedures.
Regulatory Classification
Identification
A burr hole cover is a plastic or metal device used to cover or plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery.
Predicate Devices
RapidFlap LS Cranial Fixation Flap System (K130309)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date November 21, 2023. The text is written in a clear, sans-serif font. The date is presented in a standard format, with the month, day, and year clearly indicated. The text is horizontally aligned and evenly spaced.
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
Chendu MedArt Medical Scientific Co., Ltd. % Joyce Yang Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square, Nanshan District Shenzhen, Guangdong 518100 China
Re: K221606
Trade/Device Name: Absorbable Cranial Flap Fixation System Regulation Number: 21 CFR 882.5250 Regulation Name: Burr Hole Cover Regulatory Class: Class II Product Code: GXR Dated: October 20, 2023 Received: October 23, 2023
Dear Joyce Yang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/8 description: The image shows the name "Adam D. Pierce -S" in a large font on the left side of the image. On the right side of the image, the text "Digitally signed by Adam D. Pierce -S" is present. Below this, the date "2023.11.21" and time "09:27:14-05'00'" are listed.
Adam D. Pierce. Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices
{2}------------------------------------------------
OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221606
Device Name Absorbable Cranial Flap Fixation System
Indications for Use (Describe)
The Absorbable Cranial Flap Fixation System is intended to be used for fixation of cranial bone flaps following craniotomy.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# Date prepared: November 15, 2023
# 1. Submitter
| 1. Submitter | |
|--------------------------|---------------------------------------------------------------------------------------------------|
| Applicant Name | Chengdu MedArt Medical Scientific Co., Ltd. |
| Address | No.102, Tianying Road, Chengdu High-tech<br>Industrial Development Zone, 611731 Sichuan,<br>China |
| Contact person | Huaping Ran |
| Phone | 086-28-87958626 |
| E-mail | yangm@biomedart.cn |
| 2. Device Identification | |
| Trade Name | Absorbable Cranial Flap Fixation System |
| Regulation Number | 21 CFR § 882.5250 |
| Device Class | II |
| Regulation Description | Burr hole cover |
| Product Code | GXR |
| 3. Predicate Device | |
| Trade Name | RapidFlap LS Cranial Fixation Flap System |
| Submission Number | K130309 |
| Regulation Number | 21 CFR § 882.5250 |
| Device Class | II |
| Regulation Description | Burr hole cover |
| Product Code | GXR |
| Manufacturer | Biomet Microfixation |
{5}------------------------------------------------
#### 4. Device Description
The Absorbable Cranial Flap Fixation System is comprised of a lower disc, an upper disc, a connecting rod and a rotating lock. The lower disc consists of a circular disk through which the connecting rod is centrally extruded outward. The upper disc consists of a circular disk with a threaded hole to accept the connecting rod. The upper disc is driven to move axially along the connecting rod by rotating the rotating lock so that the discs tightly grip the bone flap and provide rigid attachment and coplanar alignment to the surrounding bone.
The Absorbable Cranial Flap Fixation System is made from poly-L-lactic acid. The device is supplied sterile and is intended for single use.
The Heat/Contouring Pen is an accessory used to cut off the excess connecting rod of the Absorbable Cranial Flap Fixation System. The Heat/Contouring Pen is a disposable, lithium battery powered handheld device that includes a cutting heating head and a smoothing head. The Heat/Contouring Pen is supplied sterile and is intended for single use.
#### 5. Indications for Use
The Absorbable Cranial Flap Fixation System is intended to be used for fixation of cranial bone flaps following craniotomy.
| Specification/<br>Characterization | Subject Device:<br>Absorbable Cranial Flap<br>Fixation System (K221606) | Predicate Device:<br>RapidFlap LS Cranial Fixation<br>Flap System (K130309) | Comments |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Indication for Use | The Absorbable Cranial<br>Flap Fixation System is<br>intended to be used for<br>fixation of cranial bone<br>flaps following craniotomy. | The RapidFlap LS Cranial<br>Fixation Flap System is indicated<br>for use in rigid fixation of a<br>craniotomy in adult and pediatric<br>populations. | Similar |
| Material | Poly-L-lactic acid | Polylactic/polyglycolic acid<br>copolymer | Similar |
| Material property | Resorbable | Resorbable | Same |
| Sterilization | Sterile by EO | Sterile by EO | Same |
## 6. Tabular Comparison of the Subject Device and Predicate Device
{6}------------------------------------------------
| Specification/<br>Characterization | Subject Device:<br>Absorbable Cranial Flap<br>Fixation System (K221606) | Predicate Device:<br>RapidFlap LS Cranial Fixation<br>Flap System (K130309) | Comments |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------|
| Component<br>description | Lower disc and upper disc<br>(each consisting of a circular<br>disk with a threaded hole)<br>connected by a connecting<br>rod | Inner and outer plates (each<br>consisting of a circular disk with a<br>threaded hole) connected by a<br>threaded central post | Similar |
| Accessories | Battery operated<br>Heat/Contouring Pen with<br>cutting heating head and<br>smoothing heating head | LactoSorb SE Heat Pen with<br>cutting tip and contouring tip | Different |
# 7. Summary of Non-clinical Testing
The Absorbable Cranial Flap Fixation System is made from poly-L-lactic acid.
The main performance of the product is tripping force, compression force and torque force and the test results showed no statistical difference between the subject device and predicate device.
| Test | Test method summary | Results |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Tripping Force | The purpose of this test is to determine the locking function of the device is comparable to the predicate device.<br><br>1) The devices were fastened with the fixture of the tensile testing machine.<br><br>2) Apply tension at the stretching speed of 20 mm/min until the upper disc is released and record the tension force<br><br>3) Compared the test result of the subject and predicate device. | Pass<br>No statistical difference between the predicate and subject device. |
| Compression Force | 1) Place the upper/lower disk on the flat test bench.<br><br>2) Apply pressure to the disk with the microcomputer control electronic universal tester and record the maximum force required for the device to deform or crack. Compare the | Pass<br>The acceptance criteria were met, and the results showed no statistical difference |
{7}------------------------------------------------
| Test | Test method summary | Results |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| | results with the predicate and ensure that the<br>acceptance criteria is met. | between the subject<br>and predicate<br>device. |
| Torque Force | Clamp and fix both end of the connecting rod and<br>rotate the connecting plate at a fixed rotational<br>speed until the connecting rod is damaged or<br>stops rotating and record the maximum force<br>required for the connecting rod to break. | Pass<br>No statistical<br>difference between<br>the subject and<br>predicate device. |
The biocompatibility evaluation for the Absorbable Cranial Flap Fixation System was conducted in accordance with the FDA guidance document "Use of International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process"".
| Test | Results | Conclusion |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| In Vitro Cytotoxicity Test<br>(ISO 10993-5:2009) | The percent cell viability of the<br>subject device was 94.7% (>70%). | Pass<br>Non - Cytotoxic |
| Skin Sensitization Test<br>(ISO 10993-10:2010) | Under the conditions of the study no<br>visible change was observed in the<br>test article groups. | Pass<br>Non - Sensitizer |
| Intracutaneous<br>Reactivity Test<br>(ISO 10993-10:2010) | Under the conditions of the study the<br>overall<br>mean<br>score<br>difference<br>between the test article and vehicle<br>was 0.56, less than 1.0. | Pass<br>Non - Irritant |
| Acute Systemic Test<br>(ISO 10993-11:2017) | Under the conditions of the study the<br>no toxic response was observed in<br>the sample group. | Pass<br>Not Systemically<br>Toxic |
| Material-Mediated<br>Pyrogenicity<br>(ISO 10993-11:2017) | The results showed that the body<br>temperature of the three rabbits did<br>not increase more than 0.5 °C. | Pass<br>Not Pyrogenic |
{8}------------------------------------------------
| Test | Results | Conclusion |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Bone Implantation Test<br>(ISO 10993-6:2016) | Under the conditions of the study the<br>histopathological response of the test<br>article was similar with the control<br>sample. | Pass |
| Subchronic Systemic<br>Toxicity<br>(ISO 10993-11:2017) | Under the conditions of the study no<br>toxic response was observed. | Pass<br>No evidence of<br>systemic<br>toxicity |
| Degradation Test<br>(ISO 10993-13:2010) | Comparable results to the control<br>group for blood lactate concentration,<br>cerebrospinal fluid lactate<br>concentration, pH, and<br>histopathological. | Degradable and no<br>toxicity |
| Genotoxicity<br>(ISO 10993-3:2014) | - The result of Ames test for the test<br>article was negative.<br>- The results indicated that the test<br>sample could not induce<br>chromosome aberrations.<br>- The results showed that the test<br>article did not induce gene mutation<br>at TK locus. | Pass<br>Non - genotoxic |
The LAL endotoxin testing demonstrates that the sterile Absorbable Cranial Flap Fixation System conforms to the required ≤2.15 EU/device.
## 8. Conclusion
The subject device, Absorbable Cranial Flap Fixation System, has the same intended use, design, function, and is composed of a similar material to the predicate device, RapidFlap LS Cranial Fixation Flap System. Based on the comparison of the technological and functions characteristics, intended use and performance data, the Absorbable Cranial Flap Fixation System is substantially equivalent to the predicate device, RapidFlap LS Cranial Fixation Flap System.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.