Cure Ultra Male, Cure Ultra Female, Cure Ultra Plus
K221593 · Convatec, Inc. · EZD · Nov 17, 2022 · Gastroenterology, Urology
Device Facts
Record ID
K221593
Device Name
Cure Ultra Male, Cure Ultra Female, Cure Ultra Plus
Applicant
Convatec, Inc.
Product Code
EZD · Gastroenterology, Urology
Decision Date
Nov 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cure Ultra® is an intermittent urinary catheter that is inserted through the urethra and indicated for the purpose of bladder drainage for males and females. The urinary catheter comes in a variety of sizes and is packaged sterile for single-use.
Device Story
Cure Ultra® is an intermittent urinary catheter for bladder drainage; used by adult patients (self-catheterization) or caregivers/clinicians. Device is a pre-lubricated, open-system catheter (no collection bag); urine flows directly into a toilet or external collection system. Input is manual insertion through the urethra; output is urine drainage. Designed with a trauma-reducing tip; available in various French sizes and lengths per ISO 20696:2018. Benefits include convenience for patients away from home. Operated manually; no electronic or software components.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), sterilization validation (AAMI/ISO 11135-1), packaging integrity (ASTM F2096-11), and mechanical performance (ISO 20696:2018) including diameter, shaft length, flow rate, tensile strength, and kink stability. No clinical data presented.
Technological Characteristics
Materials: Polyvinyl chloride (PVC) catheter tube and funnel; Lubrajel MG lubricant. Design: Pre-lubricated, open-system intermittent catheter. Standards: ISO 20696:2018 (dimensions, flow, strength), ISO 10993-1:2018 (biocompatibility), AAMI/ISO 11135-1:2014 (sterilization). Sterile, single-use. No electronic components or software.
Indications for Use
Indicated for bladder drainage in adult males and females with urinary system dysfunction. Suitable for self-catheterization or use by healthcare professionals, nurses, caregivers, or family members. No specific contraindications listed.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
November 17, 2022
ConvaTec, Inc. % Dawn Norman Partner MRC Global, LLC 9085 E Mineral Circle, Suite 110 Centennial, CO 80112
Re: K221593
> Trade/Device Name: Cure Ultra® Female, Cure Ultra® Male, Cure Ultra® Plus Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZD Dated: October 6, 2022 Received: October 18, 2022
Dear Dawn Norman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
# Angel A. Soler-garcia -S
Jessica K. Nguyen, Ph.D.
Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221593
#### Device Name
Cure Ultra® Female, Cure Ultra® Male, Cure Ultra® Plus
Indications for Use (Describe)
The Cure Ultra® is an intermittent urinary catheter that is inserted through the urethra and indicated for the purpose of bladder drainage for males. The urinary catheter comes in a variety of sizes and is packaged sterile for singleuse.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
Cure Ultra® November 17, 2022
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92 (c).
| Company: | ConvaTec, Inc.<br>7815 National Service Road, Suite 600<br>Greensboro, NC 27409 |
|-------------------------------|------------------------------------------------------------------------------------------------------|
| Primary Contact: | Dawn Norman<br>Partner, MRC Global<br>Phone: 618-604-3064<br>Dawn.Norman@askmrcglobal.com |
| Company/Secondary<br>Contact: | Courtney Smith<br>Director, Regulatory Affairs<br>Phone: 239-776-2947<br>Courtney.Smith@ConvaTec.com |
| Trade Name: | Cure Ultra® Female, Cure Ultra® Male, Cure Ultra® Plus |
| Common Name: | Catheter, Urological |
| Regulation Name: | Urological Catheter and Accessories |
| Classification: | Class II |
| Regulation Number: | 21 CFR 876.5130 |
| Panel: | Gastroenterology/Urology |
| Product Code: | EZD |
| Predicate Device | Cure Catheter® Closed System, K080881 |
### Description:
The Cure Ultra® catheters, which include the Cure Ultra® Male, Cure Ultra® Female and Cure Ultra® Plus products are intermittent urinary catheters intended to be used by males and females for the purpose of bladder drainage. They are manufactured with conventional medical
{4}------------------------------------------------
grade, biocompatible materials that are not made with natural rubber latex. The tip has been designed to eliminate trauma to the urethra and is provided in various sizes in easy-to-open, sterile, single-use packages.
#### Indications for Use:
The Cure Ultra® is an intermittent urinary catheter that is inserted through the urethra and indicated for the purpose of bladder drainage for males and females. The urinary catheter comes in a variety of sizes and is packaged sterile for single-use.
#### The Intended Use Population
Adults who need assistance with bladder drainage due to conditions causing dysfunction of the urinary system.
The target population for Cure Ultra® is male and female adults. The selection of the CH size of the catheter is based on the size of the patient and is prescribed by a health care professional. The length of the catheter meets the requirements of ISO 20696:2018 for the minimum effective shaft length (275 mm) for males and minimum effective shaft length (60 mm) for females.
Additional users are healthcare professionals including nurses, and/or caregivers, and family members who help patients with catheterization.
Cure Ultra® target users will be those who are able self-catheterize and have no or minimal dexterity impairment. However, users with dexterity impairments are not excluded from the intended users should they wish to use the device with assistance from others.
#### Basis of Substantial Equivalence:
The subject Cure Ultra® is substantially equivalent to the predicate Cure Catheter® Closed System (K080881).
The subject catheter is a modified version of the predicate Cure Catheter® Closed System (K080881). The subject device is a pre-lubricated, open catheter system that allows for patient convenience when catheterization is required while away from home. It does not include a collection bag and instead should be used to allow urine to flow directly into a toilet or other collection system (similar to the previously cleared Cure Catheter® (K072539)). Modifications to French size and length have been validated, the addition of male coude tip type, addition of the male gripper sleeve improved handling, and the lack of the collection bag does not present new safety or effectiveness questions or new risk to patients. The catheter tube and funnel are manufactured from Polyvinyl chloride (PVC) and pre-lubricated with Lubrajel MG, identical to the predicate device.
Thus, it can be concluded that the subject device modifications do not raise new questions about safety and effectiveness.
{5}------------------------------------------------
## Performance Testing:
Performance testing for the Cure Ultra® device system was conducted per applicable sections of voluntary and FDA consensus standards:
- Sterilization validation was adopted into an existing terminal cycle validation per AAMI/ISO 11135-1:2014/AMD 1:2018 and ISO 10993-7:1995
- Biocompatibility testing (cytoxicity, irritation, sensitization, and chemical characterization) according to ISO 10993-1:2018 and FDA Guidance "Use of International Standard ISO 10993-1" (2020)
- Sterile packaging in accordance with ISO 11607-1:2006 and ISO 11607-2:2006
- Real time aged shelf-life testing according to ISO 11607-1:2006
- Packaging integrity testing according to ASTM F2096-11 (2019)
- Diameter and effective shaft length testing performed according to ISO 20696:2018
- Flow rate, tensile, strength, and kink stability testing performed in accordance with ISO 20696:2018
## Conclusion:
The data and information provided in this submission support the conclusion that the Cure Ultra® device system (subject device) is substantially equivalent to its predicate device (K080881) and the proposed modification does not significantly affect the safety or effectiveness of the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.