K221576 · Ameda, Inc. · HGX · Oct 12, 2022 · Obstetrics/Gynecology
Device Facts
Record ID
K221576
Device Name
Ameda Pearl Electric Breast Pump
Applicant
Ameda, Inc.
Product Code
HGX · Obstetrics/Gynecology
Decision Date
Oct 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5160
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ameda Pearl Electric Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Ameda Pearl Electric Breast Pump is intended for multiple users in a hospital setting. It is also intended for home use by a single user.
Device Story
Ameda Pearl Electric Breast Pump is a powered breast pump for lactating women; used in hospital or home settings. Device utilizes DC-powered motor driving diaphragm-type vacuum pump and electromechanical solenoid. User selects vacuum levels and cycle speeds via soft-touch buttons; backlit LCD displays mode, suction, cycle speed, timer, and battery status. Vacuum sensor on internal PCB monitors performance. HygieniKit pumping kit connects to pump; diaphragm in flange assembly isolates pump/tubing from milk to prevent contamination. Device operates via AC adapter or internal rechargeable lithium-ion battery. Healthcare providers or patients use device to facilitate milk expression; output is expressed breast milk collected in kit. Benefits include efficient milk collection and protection against backflow/overflow.
Clinical Evidence
Bench testing only. Performance testing verified vacuum pressure and cycle speed specifications across all user-selectable settings in stimulation and expression modes, for both single and double pumping, using AC and battery power. Testing also confirmed device life, battery operating time, battery indicator accuracy, and backflow/overflow protection.
Technological Characteristics
Powered breast pump; diaphragm-type vacuum pump; electromechanical solenoid. Powered by 15V, 2A AC/DC adapter or 10.8V 2600 mAh lithium-ion battery. Features backlit LCD, 8 soft-touch buttons. Vacuum range 30–250 mm Hg; cycle speeds 30–120 CPM. Non-sterile. Software level of concern: Moderate. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11.
Indications for Use
Indicated for lactating women to express and collect breast milk. Intended for multiple-user hospital settings and single-user home environments.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
October 12, 2022
Ameda, Inc. Sean Pettibone Coo 485 Half Day Road; Suite 320 Buffalo Grove. Illinois 60089
Re: K221576
> Trade/Device Name: Ameda Pearl Electric Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: May 30, 2022 Received: June 1, 2022
Dear Sean Pettibone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221576
Device Name Ameda Pearl Electric Breast Pump
Indications for Use (Describe)
The Ameda Pearl Electric Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Ameda Pearl Electric Breast Pump is intended for multiple users in a hospital setting. It is also intended for home use by a single user.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY - K221576
In accordance with 21 CFR 807.92(a) the following summary is provided:
### SUBMITTER:
Ameda, Inc. 485 Half Day Road, Suite 320 Buffalo Grove, IL 60089 Phone: 847-964-2620
#### PRIMARY CONTACT PERSON:
Carolin Archibald President and CEO Ameda, Inc. 847-964-2620 carolin.archibald@ameda.com
DATE PREPARED: October 12, 2022
### Device information
Trade Name: Ameda Pearl Electric Breast Pump Common Name: Powered breast pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered breast pump Product Code: HGX (Pump, Breast, Powered) Classification Panel: Obstetrics/Gynecology Regulatory Class: II
### Predicate Device Information
Spectra S3 Plus: K181784
The predicate device has not been subject to a design-related recall.
### Device Description
The Ameda Pearl Electric Breast Pump is an electric breast pump powered by an external AC-DC power supply or its internal rechargeable lithium-ion battery. The device is provided non-sterile.
{4}------------------------------------------------
The device is intended to be used by lactating women to express and collect milk from their breasts. Pumping can be performed on either one breast (single pumping) or both breasts at the same time (double pumping).
The Ameda Pearl Electric Breast Pump utilizes a DC-powered motor driving a diaphragm-type vacuum pump and an electromechanical solenoid which are controlled electronically to provide a range of user-selectable vacuum (suction) levels and pumping (cycle) speeds. Vacuum is monitored continuously by a sensor mounted on an internal printed circuit board.
The Ameda Pearl Electric Breast Pump has a backlit LCD display, which shows pumping mode, suction level, cycle speed level, timer, and battery charge level. Surrounding the front panel display are eight soft-touch buttons allowing the user to power the device on/off, switch between stimulation and expression pumping modes, control vacuum strength and cycle speed, and adjust the lighting intensity of the LCD display and of a nightlight which shines from the underside of the pump handle.
There are 6 levels of suction strength in stimulation mode and 12 levels in expression mode. While the user can select any of 6 available cycle speeds at each of the 12 vacuum levels in expression mode, the cycle speed in stimulation mode is pre-set at each of the 6 vacuum levels available in that mode. At minimum vacuum, level 1, the cycle speed is highest, 120 CPM (cycles per minute). Medium vacuum levels 2 and 3 are cycled at 100 CPM, and the highest 3 vacuum levels are delivered at 80 CPM.
The Ameda Pearl Electric Breast Pump is primarily intended for multiple users in a hospital environment. Use is also expected in the home environment. When properly connected, the Ameda Hygienikit pumping kit transfers the vacuum generated by the powered pump to the breast enabling expression and collection of milk. A diaphragm in the breast flange assembly physically isolates pump and tubing from the air space where milk is expressed and collected, protecting the breast milk from contamination.
The Ameda Pearl Electric Breast Pump is supplied with an AC adapter. HygieniKit is not included.
# Indications for Use
The Ameda Pearl Electric Breast Pump is a powered breast pump to be used by lactating women to express and collect milk from their breasts. The Ameda Pearl Electric Breast Pump is intended for multiple users in a hospital setting. It is also intended for home use by a single user.
{5}------------------------------------------------
# Comparison of Intended Use and Technological Characteristics of the Subject and Predicate Device
The intended use and key technological characteristics of the Ameda Pearl Electric Breast Pump and the predicate device are compared side-by-side in the table below.
| | Subject Device | Predicate Device |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Ameda Pearl Electric Breast Pump | Spectra S3 Plus Breast Pump |
| 510(k) Number | K221576 | K181784 |
| Manufacturer | Ameda, Inc. | Uzinmedicare Co. |
| Product Code | HGX | HGX |
| Device Class | 2 | 2 |
| Indications for<br>Use | The Ameda Pearl Electric Breast Pump<br>is a powered breast pump to be used<br>by lactating women to express and<br>collect milk from their breasts. The<br>Ameda Pearl Electric Breast Pump is<br>intended for multiple users in a<br>hospital setting. It is also intended for<br>home use by a single user. | The Spectra 3 Plus Breast Pump is a<br>powered breast pump to be used by<br>lactating women to express and collect<br>milk from their breast. The Spectra 3<br>Plus Breast Pump is intended for<br>multiple users in a hospital setting. It is<br>also intended for home use by a single<br>user. |
| Intended Use<br>Environment | Hospital and Home Healthcare | Hospital and Home Healthcare |
| POWER | | |
| Power Source<br>(external) | AC/DC adapter 15V, 2A DC | AC/DC adapter 12V, 2A DC |
| Power Source<br>(internal) | Rechargeable lithium-ion battery,<br>10.8V, 2600 mAh | Rechargeable lithium-ion battery,<br>11.1V, 2000 mAh |
| Battery<br>performance | 4 hours on full charge | 3 hours on full charge |
| Auto Power-off | After 60 minutes in expression mode | After 30 minutes in expression mode |
| VACUUM<br>PERFORMANCE | | |
| Pump Type | Diaphragm | Diaphragm |
| Modes | Stimulation and Expression | Massage and Expression |
| Vacuum levels | 6 in stimulation mode<br>12 in expression mode | 5 in massage mode<br>12 in expression mode |
| Vacuum range | 30 – 250 mm Hg | 50 – 270 mm Hg |
{6}------------------------------------------------
| | Subject Device | Predicate Device |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Cycle speed<br>range | 80, 100, 120 cycles/minute<br>(stimulation mode) | 70 cycles/minute (massage mode) |
| | 30 – 48 cycles/minute (expression mode) | 38 – 54 cycles/minute (expression mode) |
| USER INTERFACE | | |
| LCD Display | Mode, Time, Vacuum Level, Battery Status, Cycle speed | Mode, Time, Vacuum Level, Battery Status, Cycle speed |
| LCD brightness | 3-level adjustable | Non-adjustable |
| Night light<br>brightness | 3-level adjustable | 2-level adjustable |
| Button controls | 8: Power, Mode, Suction increase and decrease, Cycle Speed increase and decrease, LCD brightness, Night light brightness | 7: Power, Mode, Suction increase and decrease, Cycle Speed increase and decrease, Night light brightness |
The subject and predicate device have identical indications for use statement and have the same intended use. The subject and predicate device have different technological features, including the user interface, vacuum range, cycle speed range, and power sources. These technological differences do not raise different questions of safety or effectiveness.
# Summary Of Non-Clinical Testing
The following data were provided in support of the substantial equivalence determination.
Risk Analysis in accordance with ISO 14971:2019
Electrical safety testing per IEC 60601-1:2005 (3rd Edition) with US deviations per AAMI/ANSI ES60601-1:2005/(R)2012 and A1:2012
Electromagnetic compatibility testing per IEC 60601-1-2: 2014+AMD1:2020, edition 4.1
# Use in the home healthcare environment per IEC 60601-1-11:2015
Software verification and validation was conducted in accordance with the 2005 FDA quidance document "Guidance for the Content of Premarket Submission for Software Contained in Medical Devices." The software for this device was considered a Moderate level of concern.
# Bench Performance Testing
{7}------------------------------------------------
The Ameda Pearl Electric Breast Pump was tested to demonstrate it meets stated performance specifications (vacuum pressure and cycle speed). Testing involved measurement of vacuum at user-selectable settings in both stimulation and expression modes for pumping at a single breast (single pumping) or both breasts simultaneously (double pumping). Testing was conducted separately under two states of power: (1) externally supplied by an AC/DC adapter and (2) internally supplied from a rechargeable lithiumion battery. Specifications were met under all conditions.
Testing confirmed device life and battery operating time, battery indicator accuracy, and protection against backflow and overflow.
### Conclusion
The results of the performance testing described above demonstrate that the Ameda Pearl Electric Breast Pump is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.