← Product Code [DQX](/productcode/DQX) · K221575

# Pre-Formed Blue (K221575)

_Lake Region Medical · DQX · Oct 7, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221575

## Device Facts

- **Applicant:** Lake Region Medical
- **Product Code:** [DQX](/productcode/DQX.md)
- **Decision Date:** Oct 7, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Pre-Formed Blue guidewire is intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used during transcatheter aortic valve procedures.

## Device Story

Pre-Formed Blue is a single-use catheter guidewire designed to facilitate device placement during TAVR procedures. It features a unique pre-shaped spiral distal tip to ensure placement within heart chambers. The device consists of a stainless-steel core and a stainless-steel coil, the latter coated with blue PTFE for market differentiation. It is 0.035" in diameter and 275cm in length, with distal ends available in extra-small and small sizes. Used in clinical settings by physicians, the guidewire acts as a track for interventional devices. The device is sterilized via ethylene oxide.

## Clinical Evidence

Non-clinical bench testing included thermal conditioning (ASTM D4169, ISTA 2A), packaging integrity (ASTM F88-15, F1929-15, F2096-11), dimensional analysis, lubricity, joint pull strength, 3-point bend, coating adhesion, radiopacity, corrosion, particulate residue, fracture, flex, torque fatigue, kink resistance, and tip shape retention (EN ISO 11070). Biocompatibility was verified per ISO 10993-1. Animal study performed under GLP (21 CFR Part 58) demonstrated performance equivalent to the predicate.

## Technological Characteristics

Stainless steel core and coil; blue PTFE coating. 0.035" diameter, 275cm length. Pre-shaped spiral distal tip (extra-small/small). Sterilized via ethylene oxide. Complies with EN ISO 11070, ISO 10993-1, ASTM D4169, ISTA 2A, ASTM F88-15, ASTM F1929-15, and ASTM F2096-11.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Pre-Formed Guidewire ([K151244](/device/K151244.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

October 7, 2022

Lake Region Medical Deko Hussein Sr. Regulatory Specialist 340 Lake Hazeltine Drive Chaska, Minnesota 55318

Re: K221575

Trade/Device Name: Pre-Formed Blue Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DOX Dated: September 9, 2022 Received: September 9, 2022

Dear Deko Hussein:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Jaime Raben Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221575

Device Name Pre-Formed Blue

Indications for Use (Describe)

The Pre-Formed Blue guidewire is intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used during transcatheter aortic valve procedures.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K221575 510(k) Summary

Submitter Address:

Establishment Registration Number: Primary Contact Person: Email: Phone: Date Prepared

## Subject Device

Name of Device (Trade): Common or Usual Name: Classification: Regulatory Class: Product Code:

# Predicate Device

- Device Name: Manufacturer: 510(k) Number: Regulatory Class: Product Code:
Lake Region Medical 340 Lake Hazeltine Drive Chaska, MN 55318

2126666 Deko Hussein, Sr. Regulatory Specialist deko.hussein@integer.net (952) 641-8302 May 31, 2022

Pre-Formed Blue Catheter Guidewire Catheter Guidewire (21 CFR 870.1330) Class II DQX

Pre-Formed Guidewire Lake Region Medical K151244 (Cleared June 11, 2015) Class II per 21 CFR 870.1330 DOX

The identified predicate device, Pre-Formed guidewire (K151244), has not been subject to a design related recall or any recall since clearance. This Traditional 510K is submitted to modify the coating applied to the device from a green PTFE to blue PTFE for market differentiation purposes, this is the primary change to the device requiring the submission of a pre-market notification.

The Pre-Formed Blue guidewire is designed to facilitate device placement during TAVR procedures and has a unique pre-shaped spiral distal tip to ensure placement within the heart. The guidewire is intended for single use.

The guidewire has a 0.035" diameter and is 275cm in length. The guidewire composed of two primary components: a core, and a coil. Both components are made of stainless steel. The core wire component is a piece of stainless-steel wire which is ground on the distal end to fit into the coil and provide flexibility. The coil component is fitted over the distal end of the core and is a stainless-steel coiled wire coated with blue PTFE coating. The coil and core components are secured together using a weld on the distal end, forming the guidewire. The pre-shaped distal end of the guidewire is available in two sizes (extra-small and small).

The Pre-Formed Blue guidewire is sterilized using ethylene oxide. There are no accessories packaged with the Pre-Formed Blue guidewire.

## Indications for Use:

{4}------------------------------------------------

Pre-Formed Blue guidewire is intended to facilitate the introduction and placement of interventional devices within the chambers of the heart. including those used during transcatheter aortic valve procedures.

# Comparison of Technological Characteristics with the Predicate Devices

The Pre-Formed Blue is substantially equivalent to the selected predicate. The Pre-Formed Blue has the same indication for use/intended use and principles of operation as the legally marketed Pre-Formed guidewire (K151244). The functional characteristics of the Pre-Formed Blue guidewire are substantially equivalent to the Pre-Formed guidewire with regards to material, dimensions and method of construction. The design modifications to the new device (Pre-Formed Blue) include differences in the PTFE color used to coat the device and dimensional changes to the distal spiral to meet customer requirements. Performance and safety testing have shown that these modifications have not raised any new questions of safety or efficacy and the device continues to meet its intended use.

# Summary of Testing

The determination of substantial equivalence includes an assessment of non-clinical (bench) performance testing. This testing was performed in order to demonstrate that the Pre-Formed Blue guidewire met applicable design and performance requirements and to support a determination of substantial equivalence. The following testing was conducted, according to established procedures, and samples were analyzed according to predetermined acceptance criteria utilized by the predicate device, Pre-Formed. In addition, FDA guidance Coronary, Peripheral and Neurovascular Guidewires Performance Tests and Recommended Labeling (October 10, 2019) was utilized to support demonstrating substantial equivalence.

- . Thermal Conditional and Packaging Distribution - Tested in accordance with ASTM D4169 (Distribution simulation) and ISTA 2A (environmental conditioning). The predetermined acceptance criteria were met.
- . FDA Preconditioning - Tested in accordance with FDA guidance. The predetermined acceptance criteria were met.
- Visual Inspection Tested in accordance with EN ISO 11070 & FDA Guidance. The . predetermined acceptance criteria were met.
- Pouch Peel Tear Tested in accordance with ASTM F88-15. The predetermined acceptance . criteria were met.
- . Seal Integrity - Tested in accordance with ASTM F1929-15. The predetermined acceptance criteria were met.
- . Bubble Leak Testing - Tested in accordance with ASTM F2096-11(2019). The predetermined acceptance criteria were met.
- Dimensional – Length, Outer Diameter – Tested in accordance with EN ISO 11070 & FDA Guidance. The predetermined acceptance criteria were met.
- . Dimensional - Curve - Tested in accordance with internal procedures. The predetermined acceptance criteria were met.
- . Lubricity - Tested in accordance with FDA Guidance. The predetermined acceptance criteria were met.
- . Guidewire Distal & Proximal Joint Pull Test - Tested in accordance with EN ISO 11070 & FDA Guidance. The predetermined acceptance criteria were met.
- . 3 Point Bend - Tested in accordance with internal procedures. The predetermined acceptance criteria were met.

{5}------------------------------------------------

- . Coating Adhesion - Tested in accordance with FDA Guidance. The predetermined acceptance criteria were met.
- . Radiopacity - Tested in accordance with EN ISO 11070 & FDA Guidance. The predetermined acceptance criteria were met.
- . ISO Corrosion Test - Tested in accordance with EN ISO 11070 & FDA Guidance. The predetermined acceptance criteria were met.
- . Particulate Residue Test - Tested in accordance with FDA Guidance. The predetermined acceptance criteria were met.
- . ISO Fracture Test - Tested in accordance with EN ISO 11070. The predetermined acceptance criteria were met.
- . ISO Flex Test - Tested in accordance with EN ISO 11070. The predetermined acceptance criteria were met.
- . Torque Fatigue - Tested in accordance with FDA Guidance. The predetermined acceptance criteria were met.
- . Kink Resistance - Tested in accordance with FDA guidance. The predetermined acceptance criteria were met.
- . Tip Shape Retention - Tested in accordance with internal procedures. The predetermined acceptance criteria were met.
- . Spiral (Tip) Compression - Tested in accordance with internal procedures. The predetermined acceptance criteria were met.
- . Biocompatibility - Tested in accordance with ISO 10993-1 and all applicable parts of this standard series. The predetermined acceptance criteria were met.

# Performance Data (Animal)

Testing was performed in accordance with 21 CFR Part 58 for Good Laboratory Practice (GLP) for Non-Clinical Laboratory studies. The subject device Pre-Formed Blue guidewire shares the same intended use and fundamental scientific technology as the predicate and demonstrated all safety and performance objectives as the device performed equivalent to the predicate, Pre-Formed guidewire, and demonstrated all Customer Requirement Specifications as measured by post procedure feedback forms.

Conclusion: The Pre-Formed Blue guidewire, described in this submission, met all design input requirements based on the intended use. The results of these tests support a determination of substantial equivalence to the predicate device, the Pre-Formed Guidewire (K151244).

---

**Source:** [https://fda.innolitics.com/device/K221575](https://fda.innolitics.com/device/K221575)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
