K221530 · Sibel, Inc. · DRG · Nov 22, 2022 · Cardiovascular
Device Facts
Record ID
K221530
Device Name
ANNE Pediatric
Applicant
Sibel, Inc.
Product Code
DRG · Cardiovascular
Decision Date
Nov 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
ANNE Pediatric is a wireless vital signs and physiological data monitoring platform indicated for the measurement of heart rate, skin temperature, and body temperature by qualified healthcare professionals in healthcare settings. The device is intended for use on neonatal patients of any gestational age up to infants 2 years of age as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The device is not intended to monitor critically ill patients. The device is not intended to monitor or measure heart rate during motion. The device is not an apnea alarm.
Device Story
ANNE Pediatric is a wireless vital signs monitoring platform for neonates and infants up to 2 years old. System consists of two skin-mounted, bio-integrated sensors (chest and foot) and the ANNE View mobile application. Sensors collect ECG and temperature biosignals; data transmitted via Bluetooth to mobile device for real-time display. Used by healthcare professionals in clinical settings to aid diagnosis and treatment. Device provides continuous physiological information; does not provide alarms or monitor heart rate during motion. Benefits include non-invasive, continuous monitoring of vital signs in non-critically ill pediatric patients.
Clinical Evidence
Clinical study conducted at a tertiary facility with n=137 neonates (median gestational age 38 weeks). HR performance compared against Masimo Rad-97 in one open-label and three closed-label rounds (n=70). Final closed-label round (n=20) demonstrated a normalized root mean square deviation (RMSD) of 1.2%, meeting the ±10% or ±5bpm accuracy requirement per IEC 60601-2-27. No adverse events or skin injuries reported.
Technological Characteristics
Wireless vital signs monitor using Bluetooth connectivity. Sensors are skin-mounted and bio-integrated. Biocompatibility per ISO 10993-1, -5, -10. Electrical safety per ANSI/AAMI ES60601-1 and IEC 60601-1-2. Wireless coexistence per ANSI IEEE C63.27. Temperature measurement per ISO 80601-2-56. Heart rate measurement per IEC 60601-2-27. Software developed per IEC 62304.
Indications for Use
Indicated for neonatal patients of any gestational age up to infants 2 years of age. Used by healthcare professionals in healthcare settings for continuous monitoring of heart rate, skin temperature, and body temperature. Not for critically ill patients, not for heart rate monitoring during motion, and not an apnea alarm.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2022
Sibel Inc. Sarah Coughlin OARA Engineer 6650 W. Touhy Ave. Niles, Illinois 60714
Re: K221530
Trade/Device Name: ANNE Pediatric Regulation Number: 21 CFR 870.29870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II Product Code: DRG, MWI, FLL Dated: Mav 23, 2022 Received: May 26, 2022
Dear Sarah Coughlin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K221530
Device Name ANNE Pediatric
Indications for Use (Describe)
ANNE Pediatric is a wireless vital signs and physiological data monitoring platform indicated for the measurement of heart rate, skin temperature by qualified healthcare professionals in healthcare settings. The device is intended for use on neonatal patients of any gestational age up to infants 2 years of age as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The device is not intended to monitor critically ill patients. The device is not intended to monitor or measure heart rate during motion. The device is not an apnea alarm.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{3}------------------------------------------------
# 510(K) Summary
#### -Submitter:
Sibel Inc. 6650 W. Touhy Avenue, Niles, IL 60714 Tel: (224) 251-8859
Official Correspondent: Sarah Coughlin, Senior Regulatory Affairs and Quality Assurance Engineer 6650 W. Touhy Avenue, Niles, IL 60714 Tel: (224) 251-8859
Date Prepared: 10/7/2022
#### II. Device Information
Name of Device: ANNE Pediatric
510(k) Number: K221530
Common or Usual Name: Wireless Remote Monitoring System
Classification Name: Radiofrequency Physiological Signal Transmitter and Receiver
Review Panel: Cardiovascular
Regulation: 21 CFR §870.2910
Regulatory Class: Class II
Product Classification Code: DRG, MWI, FLL
#### . Predicate Device
Trade Name: ANNE One Predicate 510(k): K211305 Predicate device manufacturer: Sibel Inc.
#### IV. Device Description
ANNE Pediatric is a wireless vital signs and physiological data monitoring platform that collects electrocardiography (ECG) and temperature biosignals to measure vital signs such as heart rate (HR), skin temperature, and body temperature. The system features two skin-mounted, bio-integrated sensors that pair with the ANNE View application for the display of real-time vital measurements. Data is transmitted wirelessly from the sensors via Bluetooth to the mobile device.
{4}------------------------------------------------
#### V. Indications for Use
ANNE Pediatric is a wireless vital signs and physiological data monitoring platform indicated for the measurement of heart rate, skin temperature, and body temperature by qualified healthcare professionals in healthcare settings. The device is intended for use on neonatal patients of any gestational age up to infants 2 years of age as a general patient monitor to provide continuous physiological information as an aid to diagnosis and treatment. The device is not intended to monitor critically ill patients. The device is not intended to monitor or measure heart rate during motion. The device is not an apnea alarm.
| | Subject device<br>ANNE Pediatric | Predicate device<br>Sibel Inc. | Variances/Equivalence |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Trade Name | ANNE Pediatric | ANNE One | |
| Class | II | II | Equivalent |
| Product Code | DRG, MWI, FLL | DRG, MWI, FLL | Equivalent |
| Regulation<br>Number and Name | 870.2910<br>Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | 870.2910<br>Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | Equivalent |
| Indications for Use | ANNE Pediatric is a<br>wireless vital signs and<br>physiological data<br>monitoring platform<br>indicated for the<br>measurement of heart<br>rate, skin temperature,<br>and body temperature by<br>qualified healthcare<br>professionals in<br>healthcare settings. The<br>device is intended for<br>use on neonatal patients<br>of any gestational age up<br>to infants 2 years of age<br>as a general patient<br>monitor to provide<br>continuous physiological<br>information as an aid to<br>diagnosis and treatment.<br>The device is not<br>intended to monitor<br>critically ill patients. The<br>device is not intended to<br>monitor or measure<br>heart rate during motion.<br>The device is not an<br>apnea alarm. | ANNE One is a wireless<br>vital signs and<br>physiological data<br>monitoring platform<br>indicated for the<br>measurement of heart<br>rate, respiratory rate,<br>step count, fall count,<br>skin temperature, and<br>body temperature by<br>qualified healthcare<br>professionals in<br>healthcare settings. The<br>device is intended for<br>use on general care<br>patients who are 18<br>years of age or older as<br>a general patient monitor<br>to provide continuous<br>physiological information<br>as an aid to diagnosis<br>and treatment. The<br>device is not intended for<br>use on critical care<br>patients. The device is<br>not intended to monitor<br>or measure respiratory<br>rate or heart rate on<br>ambulatory patients. | Substantially<br>Equivalent |
| Target Population | Neonates and infants up<br>to 2 years of age | Adults, 18 years of age<br>and older | Different |
| Use Environment | Healthcare Settings | Healthcare Settings | Equivalent |
| Sensor Placement | Chest and Foot | Chest and Finger | Substantially<br>Equivalent |
| Data | Data transmitted<br>wirelessly via Bluetooth<br>from sensors to mobile<br>device | Data transmitted<br>wirelessly via Bluetooth<br>from sensors to mobile<br>device | Equivalent |
| Notification | No notification ability | No notification ability | Equivalent |
| Heart Rate | Yes | Yes | Equivalent |
| Skin Temperature | Yes | Yes | Equivalent |
| Body Temperature | Yes | Yes | Equivalent |
| Respiratory Rate | No | Yes | Different |
| Step Count | No | Yes | Different |
| Fall Count | No | Yes | Different |
{5}------------------------------------------------
#### VI. Clinical Data
The functionality and performance of the ANNE Pediatric system was assessed in n=137 neonates at Aga Khan University, a tertiary healthcare facility in Nairobi, Kenya. The study compared the performance of HR measurements with ANNE Pediatric against the Masimo Rad-97 in one open label and three closed-label rounds with n=70 neonates. Median gestational age of neonates included in HR analysis was 38 (range 25 to 42) weeks. The HR analysis cohort was 44.7% female, 52.9% male, and 1.4% intersex. The normalized root mean square deviation (RMSD) in the final closed-label round of testing with n=20 neonates was 1.2%, meeting the accuracy requirement of ±10% or ±5bpm, whichever is greater, defined in IEC 60601-2-27. No adverse events or skin injuries were reported during the study.
#### VII. Performance Data
The following consensus standards and bench testing were used to evaluate the safety and effectiveness of ANNE Pediatric:
- Electrical safety and electromagnetic compatibility testing according to ANSI/AAMI ES60601-1:2005/(R)2012 and IEC 60601-1-2:2014 standards.
- . Biocompatibility testing according to ISO 10993-1:2018, ISO 10993-5:2009, and ISO 10993-10:2010 for patient contacting materials.
- Wireless coexistence testing according to ANSI IEEE C63.27-2017. ●
- Defibrillation testing according to Section 8.5.5 of ANSI/AAMI ES60601-1:2005/(R)2012 ●
{6}------------------------------------------------
- . Software verification and validation testing according to IEC 62304:2015 and the FDA guidance document, Content of Premarket Submissions for Software Contained in Medical Devices.
- Shelf life testing of the adhesive to demonstrate safe and effective performance over the . intended device life cycle.
- . Bench testing to demonstrate the mechanical durability of the sensors.
- o Usability testing in accordance with the FDA guidance document, Applying Human Factors and Usability Engineering to Medical Devices
- Performance testing of body temperature measurements according to ISO 80601-2-56:2017.
- 0 Performance testing of heart rate measurements according to Sections 201.12.1.101.15 and 201.12.1.101.17 of IEC 60601-2-27:2011.
- Cybersecurity evaluation according to the requirements of the FDA draft guidance . document, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
- Assessment of Software of Unknown Provenance per the FDA guidance document, Off-The-Shelf Software Use in Medical Devices
#### VIII. Conclusion
The results of the substantial equivalence assessment, taken together with the clinical and performance testing data, demonstrate that the ANNE Pediatric system's performance characteristics are substantially equivalent to the predicate device in both technology and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.