Miro3D Wound Matrix

K221520 · Reprise Biomedical, Inc. · KGN · Aug 18, 2022 · SU

Device Facts

Record IDK221520
Device NameMiro3D Wound Matrix
ApplicantReprise Biomedical, Inc.
Product CodeKGN · SU
Decision DateAug 18, 2022
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Intended Use

The Miro3D Wound Matrix is intended for the management of wounds including: - Partial and full thickness wounds - Pressure ulcers - Venous ulcers - Chronic vascular ulcers - Diabetic ulcers - Tunneled, undermined wounds - Trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) - Draining wounds - Surgical wounds (donor sites/grafts, post-Mohs' surgery, post-laser surgery, podiatric, wound dehiscence)

Device Story

Miro3D Wound Matrix is a sterile, single-use, non-crosslinked acellular collagen scaffold derived from porcine liver via perfusion decellularization. Device is supplied dry and must be rehydrated with sterile saline or lactated Ringer's solution before application. Used in surgical suites, hospitals, or clinics by trained clinicians to provide a protective environment for wound healing. Acts as a resorbable matrix; supports tissue management. Configuration is a 3D porous scaffold (2 cm thickness) compared to predicate's flat sheet. Output is a physical wound dressing; clinical decision-making involves selecting appropriate size and applying to wound site to facilitate healing.

Clinical Evidence

No human clinical data. Evidence includes extensive bench testing (biocompatibility per ISO 10993, collagen analysis, residual DNA/detergent, endotoxin, viral inactivation) and a GLP-compliant animal study using a porcine full-thickness wound model. The animal study demonstrated that Miro3D performed comparably to the predicate MiroDerm in all evaluated aspects.

Technological Characteristics

Material: Perfusion-decellularized porcine liver collagen. Form: 3D porous scaffold, 2 cm thickness. Energy: None. Sterilization: E-beam irradiation (SAL 10^-6). Packaging: PETG tray, Tyvek lid, aluminum laminate pouch. Standards: ISO 10993 (biocompatibility), ASTM D3418 (denaturation temp), ST72:2019 (endotoxin), ASTM F988, F2096, F2825, D4169 (packaging).

Indications for Use

Indicated for management of partial/full thickness wounds, pressure/venous/chronic vascular/diabetic ulcers, tunneled/undermined wounds, trauma wounds (abrasions, lacerations, second-degree burns, skin tears), draining wounds, and surgical wounds (donor sites/grafts, post-Mohs'/laser surgery, podiatric, wound dehiscence).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. August 25, 2022 Reprise Biomedical, Inc. Kathy Herzog Regulatory Consultant 17400 Medina Road Suite 100 Plymouth, Minnesota 55447 Re: K221520 Trade/Device Name: Miro3D Wound Matrix Regulatory Class: Not Classified Product Code: KGN Dear Kathy Herzog: The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated August 18, 2022. Specifically, FDA is updating this SE Letter as an administrative correction to include the 510k Summary document for K221520, as it was inadvertently excluded from our August 18, 2022 SE Letter. Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-247-6328 or at julie.morabito@fda.hhs.gov. Sincerely, Julie A. Morabito -S Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {1}------------------------------------------------ Image /page/1/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. August 18, 2022 Reprise Biomedical, Inc. Kathy Herzog Regulatory Consultant 17400 Medina Road Suite 100 Plymouth, Minnesota 55447 Re: K221520 Trade/Device Name: Miro3D Wound Matrix Regulatory Class: Not Classified Product Code: KGN Dated: Mav 24, 2022 Received: May 25, 2022 Dear Kathy Herzog: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {2}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K221520 Device Name Miro3D Wound Matrix ### Indications for Use (Describe) The Miro3D Wound Matrix is intended for the management of wounds including: - · Partial and full thickness wounds - · Pressure ulcers - Venous ulcers - Chronic vascular ulcers - · Diabetic ulcers - Tunneled, undermined wounds - · Trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) - · Draining wounds - · Surgical wounds (donor sites/grafts, post-Mohs' surgery, post-laser surgery, podiatric, wound dehiscence) Type of Use (Select one or both, as applicable) | Over-The-Counter Use (21 CFR 801 Subpart C) X Prescription Use (Part 21 CFR 801 Subpart D) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92: #### I. SUBMITTER Reprise Biomedical, Inc. 17400 Medina Road, Suite 100 Plymouth, MN 55447 763-284-6795 Contact Person: Carrie Powers Date Prepared: August 18, 2022 #### II. DEVICE | Trade/Proprietary Names: | Miro3D Wound Matrix | |--------------------------|---------------------------------------------------------| | Common Name: | Animal-derived, extracellular matrix wound care product | | Regulation Number: | Unclassified | | Regulation Name: | NA | | Device Class: | Unclassified | | Product Code: | KGN | | Panel: | General & Plastic Surgery | #### III. PREDICATE DEVICE MiroDerm Wound Matrix (also known as Miromatrix Wound Matrix), K140510 This predicate has not been subject to a design-related recall. No reference devices were used in this submission. #### IV. DEVICE DESCRIPTION The Reprise Miro3D Wound Matrix is a sterile, single use, non-crosslinked acellular wound dressing that is derived from porcine liver tissue. The liver is perfusion decellularized resulting in a collagen matrix that is dried and cut to defined sizes. The Miro3D porous scaffold provides a protective environment for wound healing. The device is packaged dry, terminally sterilized in its packaging by e-beam irradiation and is rehydrated with sterile saline or lactated Ringer's solution prior to use. The Miro3D Wound Matrix is provided in four sizes that may be cut to fit a wound size prior to application. {5}------------------------------------------------ #### V. INDICATIONS FOR USE The Miro3D Wound Matrix is indicated for the following: The Miro3D Wound Matrix is intended for the management of wounds including: - Partial and full thickness wounds o - o Pressure ulcers - Venous ulcers o - Chronic vascular ulcers ● - Diabetic ulcers ● - Tunneled, undermined wounds ● - Trauma wounds (abrasions, lacerations, second-degree burns, and skin tears) - Draining wounds o - . Surgical wounds (donor sites/grafts, post-Mohs' surgery, post-laser surgery, podiatric, wound dehiscence) ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The Miro3D Wound Matrix is the dry, uncompressed version of MiroDerm which is supplied as a wet, compressed flat sheet. The subject and predicate devices have the same technological characteristics as the change in configuration does not change the Intended Use or fundamental scientific technology of the collagen matrix for wound management. No new device materials or manufacturing materials are introduced with the Miro3D as compared to MiroDerm. The devices have the same material/chemical composition, principle of operation, clinical use, biocompatibility, and sterilization. The device packaging materials for Miro3D (PETG and Tyvek) are commonly used in the medical device industry with a long history of safe use. | Feature | Miro3D<br>Wound Matrix<br>(Subject Device) | MiroDerm<br>Wound Matrix<br>(Predicate Device) | |-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | K Number | K221520 | K140510 | | Classification | Unclassified (pre-amendment) | Unclassified (pre-amendment) | | Product Code | KGN | KGN | | Class | II | II | | Intended Use | Wound management | Wound management | | Indications For<br>Use | The Miro3D Wound Matrix is<br>intended for the management of<br>wounds including: | The Miromatrix Wound Matrix<br>is intended for the management<br>of wounds including: | | Feature | Miro3D<br>Wound Matrix<br>(Subject Device) | MiroDerm<br>Wound Matrix<br>(Predicate Device) | | | • Partial and full thickness<br>wounds<br>• Pressure ulcers<br>• Venous ulcers<br>• Chronic vascular ulcers<br>• Diabetic ulcers<br>• Tunneled, undermined wounds<br>• Trauma wounds (abrasions,<br>lacerations, second-degree<br>burns, and skin tears)<br>• Draining wounds<br>• Surgical wounds (donor<br>sites/grafts, post-Mohs'<br>surgery, post-laser surgery,<br>podiatric, wound dehiscence) | • Partial and full thickness<br>wounds<br>• Pressure ulcers<br>• Venous ulcers<br>• Chronic vascular ulcers<br>• Diabetic ulcers<br>• Tunneled, undermined wounds<br>• Trauma wounds (abrasions,<br>lacerations, second-degree<br>burns, and skin tears)<br>• Draining wounds<br>• Surgical wounds (donor<br>sites/grafts, post-Mohs'<br>surgery, post-laser surgery,<br>podiatric, wound dehiscence) | | Type of Use | Wound management | Wound management | | User | Physician or other clinician<br>trained in wound care | Physician or other clinician<br>trained in wound care | | Intended Use<br>Environment | Surgical suite, hospital,<br>ambulatory surgery center or<br>out-patient clinic | Surgical suite, hospital,<br>ambulatory surgery center or<br>out-patient clinic | | Description | Animal-sourced, non-<br>crosslinked, acellular collagen<br>tissue matrix | Animal-sourced, non-<br>crosslinked, acellular collagen<br>tissue matrix | | Principle of<br>Operation | Provide a protective<br>environment for wound<br>healing | Provide a protective<br>environment for wound<br>healing | | Material | Perfusion-decellularized<br>porcine liver | Perfusion-decellularized<br>porcine liver | | Resorbable | Yes | Yes | | Configuration | Three-dimensional collagen<br>scaffold provided in four sizes<br>(W x L), all 2 cm thickness<br>(Model Number)<br>• 2 cm x 2 cm (3000)<br>• 3 cm x 3 cm (3005)<br>• 5 cm x 5 cm (3010)<br>• 10 cm x 5 cm (3015) | Flat collagen sheet provided in<br>nine sizes (W x L), thickness<br>ranging from 0.3 to 1.5 mm<br>(Model Number):<br>• 2 cm x 2 cm (BLM-200-02-<br>0202)<br>• 2 cm x 3 cm (BLM-200-02-<br>0203)<br>• 3 cm x 3 cm (BLM-200-02-<br>0303) | | Feature | Miro3D<br>Wound Matrix<br>(Subject Device) | MiroDerm<br>Wound Matrix<br>(Predicate Device) | | | | 4 cm x 4 cm (BLM-200-02-0404) 3 cm x 7 cm (BLM-200-02-0307) 5 cm x 5 cm (BLM-200-02-0505) 8 cm x 8 cm (BLM-200-02-0808) 7 cm x 10 cm (BLM-200-02-0710) 8 cm x 15 cm (BLM-200-02-0815) | | Wound Matrix<br>Preparation | Rehydrate a minimum of five<br>minutes in either sterile saline or<br>lactated Ringer's solution;<br>cut/trim the wound matrix to fit<br>wound | Soak for two minutes in either<br>sterile saline or lactated Ringer's<br>solution; cut/trim wound matrix<br>to fit wound | | Single Use or<br>Reusable | Single Use | Single Use | | Sterilization<br>Method | Electron beam irradiation | Electron beam irradiation | | Sterilization<br>Assurance Level<br>(SAL) | $10^{-6}$ | $10^{-6}$ | | Packaging | Device package: Packaged dry in a PETG plastic tray and snap-on lid with Tyvek lid seal Sterile barrier: Aluminum laminate foil pouch Shelf box: Cardboard | Device package: Packaged wet in an aluminum laminate foil pouch Sterile barrier: Aluminum laminate foil pouch Shelf box: Cardboard | | Shelf Life | 18 months (as of this submission<br>date; real-time aging will<br>continue to support 3 year shelf<br>life) | 3 years | | Storage<br>Conditions | No special storage conditions<br>required | No special storage conditions<br>required | | Biocompatibility | Biocompatibility testing or<br>justification for all applicable | Biocompatibility testing or<br>justification for all applicable | | Feature | Miro3D<br>Wound Matrix<br>(Subject Device) | MiroDerm<br>Wound Matrix<br>(Predicate Device) | | | biological endpoints per ISO<br>10993-1:2018 were completed | biological endpoints per ISO<br>10993-1:2009 were completed | | Viral<br>Inactivation | Manufacturing process is<br>capable of inactivating four<br>viruses | Manufacturing process is<br>capable of inactivating four<br>viruses | | Bench Testing | The following tests were<br>conducted:<br>● Residual DNA<br>● Collagen analysis<br>● Endotoxin<br>● Expiration dating<br>● Residual detergent<br>● Dimensions<br>● Rehydration | The following tests were<br>conducted:<br>● Residual DNA<br>● Collagen analysis<br>● Endotoxin<br>● Expiration dating<br>● Residual detergent | | Animal Wound<br>Healing Study | GLP Study to evaluate Miro3D<br>compared to MiroDerm as a<br>control article in a porcine full<br>thickness wound healing model | None completed | Table 1: Subject vs. Predicate Wound Matrix Comparison {6}------------------------------------------------ {7}------------------------------------------------ {8}------------------------------------------------ #### VII. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination: ## 1. Biocompatibility Testing - Cytotoxicity, ISO 10993-5:2009 a. - b. Sensitization, ISO 10993-10:2010 - c. Intracutaneous Reactivity, ISO 10993-10:2010 - d. Acute Systemic Toxicity, ISO 10993-11: 2017 - e. Material Mediated Pyrogenicity, ISO 10993-11:2017 - f. Implantation, ISO 10993-6:2016 - Subacute Systemic Toxicity, ISO 10993-11:2017 g. - h. Subchronic Systemic Toxicity, ISO 10993-11:2017 - Genotoxicity (Ames Bacterial Reverse Mutation and Mouse Lymphoma), ISO i. 10993-3:2014 {9}------------------------------------------------ ## 2. Bench Testing - Package Stability Testing, ASTM F988-09, ASTM F2096, ISO 11607-1:2019, a. ASTM F2825-18, ASTM D4169-16, and ASTM F2096-11 - b. Product Testing - 1. Collagen denaturation temperature, ASTM D3418 - 2. Mechanical testing Dimensional - 3. Rehvdration - 4. Residual detergent - 5. Residual DNA - 6. Bacterial endotoxin testing, ST72:2019 - 7. Viral inactivation In addition, Reprise Biomedical sponsored a GLP compliant animal study to evaluate Miro3D compared to MiroDerm on healing of porcine full thickness wounds. The study results indicated that Miro3D performed comparably to MiroDerm in all aspects evaluated. MR compatibility was not evaluated. #### VIII. CONCLUSIONS The subject Miro3D Wound Matrix has the same Intended Use as the predicate MiroDerm Wound Matrix to provide a protective environment for wound healing. The subject and predicate devices have the same technological characteristics as the change in configuration (dry, uncompressed for Miro3D and wet, compressed for MiroDerm) does not change the Intended Use, chemical composition, or fundamental scientific technology of the collagen matrix for wound management. The Miro3D device packaging materials are common in the industry and do not introduce any new hazards. These modifications in configuration and device packaging materials do not raise different questions of safety and effectiveness compared to the predicate device. Performance testing provides evidence the Miro3D device performs as intended and is as safe and effective as the predicate device. Therefore, the subject Miro3D device is substantially equivalent to the predicate MiroDerm device (K140510).
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%