← Product Code [FPA](/productcode/FPA) · K221518

# NP Medical nCompass Extension Set with Stabilizing Base (K221518)

_Np Medical · FPA · Nov 3, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221518

## Device Facts

- **Applicant:** Np Medical
- **Product Code:** [FPA](/productcode/FPA.md)
- **Decision Date:** Nov 3, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The NP Medical Extension Set with stabilizing base is a sterile single patient use device. The NP Medical Extension Set is intended for direct injection, intermittent infusion, continuous infusion or aspiration for all patients for which IV therapy is prescribed. The NP Medical Extension Set incorporates a normally closed bidirectional pressure-activated valve that prevents low pressure retrograde flow and is intended for power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second.

## Device Story

Sterile, single-use intravenous extension set; includes side-repositionable tube, connector hub, female luer, clamp, and stabilizing base. Hub contains bidirectional pressure-activated valve (PAV) to prevent low-pressure retrograde flow. Used by clinicians in clinical settings to facilitate IV therapy; connects to vascular access devices and needle-free connectors. Stabilizing base anchors set to patient skin. Supports power injection procedures up to 325 psi. Device enables fluid delivery or blood aspiration; PAV ensures valve closure when not in use, preventing retrograde flow. Benefits include secure vascular access and compatibility with high-pressure injection protocols.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including device flow rate (gravity/pressurized), retrograde pressure resistance, dynamic hemolysis (USP <788>), burst resistance, bond integrity, and biocompatibility (ISO 10993-1).

## Technological Characteristics

Materials: HDPE (caps), PC (luer/hub), PVC (tubing), POM (clamp), Si (PAV/fluid path), Polystyrene/Acrylic PSA/LDPE (stabilizer). Energy: User-operated. Connectivity: Standalone. Sterilization: Ethylene Oxide (SAL 10^-6). Features: Bidirectional PAV, rotating tube connection, stabilizing base. Complies with ISO 11135 and ISO 11607-1.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- Velano Vascular Velano ExT™ Extension Set ([K200439](/device/K200439.md))

## Reference Devices

- Nexus Medical, Nexus TKO®-6P, Luer Activated Device ([K130416](/device/K130416.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 3, 2022

NP Medical Krishna Govindarajan Global Quality Regulatory Affairs Manager 101 Union Street Clinton, Massachusetts 01510

### Re: K221518

Trade/Device Name: NP Medical nCompass Extension Set with Stabilizing Base Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: October 4, 2022 Received: October 5, 2022

### Dear Krishna Govindarajan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

David Wolloscheck, Ph.D. For Paval Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221518

Device Name

NP Medical nCompass Extension Set with Stabilizing Base

#### Indications for Use (Describe)

The NP Medical Extension Set with stabilizing base is a sterile single patient use device.

The NP Medical Extension Set is intended for direction, intermittent infusion, continuous infusion or aspiration for all patients for which IV therapy is prescribed.

The NP Medical Extension Set incorporates a normally closed bidirectional pressure-activated valve that prevents low pressure retrograde flow and is intended for power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second.

**Type of Use (Select one or both, as applicable)**

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for NP Medical. The logo consists of a blue abstract shape on the left, followed by the text "NP Medical" in a dark blue sans-serif font. Below the text, there is a tagline that reads "We Make Quality Patient Care Possible" in a smaller font.

## 510(k)Summary

The following summary is provided in accordance with 21 CFR 807.92:

## I. SUBMITTER

NP Medical, Inc., 101 Union Street Clinton, MA 01510 Phone: 978-368-6854

Contact Person: Krishna Govindaraian Date Prepared: November 3, 2022

## II. DEVICE

| Trade/Device Name:    | NP Medical nCompass™ Extension Set with Stabilizing Base |
|-----------------------|----------------------------------------------------------|
| Common or Usual Name: | Extension Set                                            |
| Regulation Number:    | 21 CFR 880.5440                                          |
| Regulation Name:      | Intravascular administration set                         |
| Regulatory Class:     | Class II                                                 |
| Product Code:         | FPA                                                      |

### III. PREDICATE DEVICE

#### Primary Predicate Device:

Velano Vascular Velano ExT™ Extension Set (K200439).

#### Reference Device:

Nexus Medical, Nexus TKO®-6P, Luer Activated Device (K130416).

#### IV. DEVICE DESCRIPTION

The NP Medical Extension Set with Stabilizing Base is a sterile, single use, nonpyrogenic intravenous administration extension set comprised of a side repositionable tube bonded to a connector hub on one end and a female luer on the other end with a clamp in between. The connector hub has a locking male luer and a rotating tube connection joint. The hub houses a bidirectional Pressure-Activated Valve (PAV) and an integral stabilizer under the connector hub. The stabilizing base component (along with IV tape / adhesive) is intended to help anchor the extension set to the patient. This device is not made with plasticizers Diethylhexylphthalate (DEHP). The NP Medical Extension Set with Stabilizing Base may be used with compatible Vascular Access Devices (VAD), Needle Free Connectors, and other commercially available devices.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for NPMedical. The logo consists of a stylized letter "D" in two shades of blue, followed by the text "NPMedical" in a dark blue sans-serif font. Below the logo is the tagline "We Make Quality Patient Care Possible" in a smaller, lighter font.

## 510(k)Summary

The following summary is provided in accordance with 21 CFR 807.92:

## V. INDICATIONS FOR USE

The NP Medical Extension Set with stabilizing base is a sterile single patient use device.

The NP Medical Extension Set is intended for direct injection, intermittent infusion, continuous infusion or aspiration for all patients for which IV therapy is prescribed.

The NP Medical Extension Set incorporates a normally closed bidirectional pressure-activated valve that prevents low pressure retrograde flow and is intended for power injector procedures to a maximum pressure of 325 psi at a flow rate of 10ml per second.

## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

#### Primary Predicate Device:

The proposed device and primary predicate device are identical with regards to intended use and are substantially equivalent with regards to design, technological characteristics, performance, safety, and effectiveness. Both devices also have the same indications for use as extension set for direct injection, intermittent infusion, continuous infusion, or aspiration for which IV therapy is prescribed.

At a high level, the proposed and primary predicate devices are based on the following same technological elements:

- Intravascular administration set/extension set
The following technological differences exist between the subject and predicate devices:

- Use of rotating tube connection joint for repositionable extension tube ●
- Use of integrated Pressure Activated Valve (PAV) for retrograde flow control ●

#### Reference Device:

The proposed device integrated Pressure Activated Valve (PAV) for retrograde flow control functionality and reference device bidirectional pressure activated valve functionality are identical with regards to operating principle, technological characteristics, performance, safety, and effectiveness. Both the proposed device and reference device include a normally closed bidirectional pressure activated valve that opens in the forward direction enabling infusion, and in the reverse direction enabling blood aspiration.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for NP Medical. The logo consists of a blue icon on the left and the text "NPMedical" on the right. Below the logo, the text "We Make Quality Patient Care Possible" is displayed.

## 510(k)Summary

The following summary is provided in accordance with 21 CFR 807.92:

The following table provides a comparison of technological characteristics between the subject and predicate device:

| Comparison<br>Element                     | Subject Device -<br>NP Medical Extension<br>Set with stabilizing<br>base                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Device -<br>Velano ExT™<br>Extension Set<br>(K200439)                                                                                                                                                                                                                                                                                                                                                                                          | Analysis of Differences                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code                              | FPA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | FPA                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Regulation Number                         | 880.5440                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 880.5440                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Class                                     | ll                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | ll                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Intended Use                              | Intravascular Extension<br>Set                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Intravascular<br>Extension Set                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for Use                       | The NP Medical<br>Extension Set with<br>stabilizing base is a<br>sterile single patient use<br>device.<br>The NP Medical<br>Extension Set is intended<br>for direct injection,<br>intermittent infusion,<br>continuous infusion or<br>aspiration for all patients<br>for which IV therapy is<br>prescribed.<br>The NP Medical<br>Extension Set<br>incorporates a normally<br>closed bidirectional<br>pressure-activated valve<br>that prevents low<br>pressure retrograde flow<br>and is intended for power<br>injector procedures to a<br>maximum pressure of<br>325 psi at a flow rate of<br>10ml per second. | The Velano ExT™<br>Extension Set with<br>needle free connector<br>is for single use only.<br>Each port of the<br>Velano ExT™<br>Extension Set may be<br>used for direct<br>injection, intermittent<br>infusion, continuous<br>infusion or aspiration.<br>The pressure rated<br>port of the Velano<br>ExT™ Extension Set<br>may be used with<br>power injector<br>procedures to a<br>maximum pressure of<br>325 psi at a flow rate<br>of 10ml per second. | Different. The subject<br>device does not<br>reference elements<br>associated with the<br>needle free connector of<br>the predicate device.<br>These excluded features<br>are intended for Velano's<br>PIVO™ compatibility for<br>which the subject device<br>does not seek to claim.<br>Added the incorporation<br>of a normally closed<br>bidirectional pressure-<br>activated valve similar to<br>reference device<br>K130416. (Performance<br>tests leveraged from the<br>reference device:<br>Retrograde pressure<br>resistance and PAV<br>robustness)<br>Device Flow Rate (gravity<br>and pressurized) and<br>Dynamic Hemolysis (per<br>USP <788>) testing<br>performed to demonstrate<br>substantial equivalence. |
|                                           | The following summary is provided in accordance with 21 CFR 807.92:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Comparison<br>Element                     | Subject Device –<br>NP Medical Extension<br>Set with stabilizing<br>base                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Device –<br>Velano ExT™<br>Extension Set<br>(K200439)                                                                                                                                                                                                                                                                                                                                                                                          | Analysis of Differences                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Extension Set<br>Components<br>/Materials | Male Luer cap/HDPE<br>Male Luer/PC<br>Male Luer spin lock<br>connector/PC<br>Tubing/PVC<br>Pinch clamp/POM<br>Female Luer/PC<br>Female Luer Cap/HDPE<br>Fluid Path Body/Si, PC<br>Stabilizer/ Polystyrene,<br>Acrylic PSA, Polycoated<br>LDPE<br>Pressure activated<br>valve/Si                                                                                                                                                                                                                                                                                                                                 | Male Luer<br>cap/HDPE Male<br>Luer/PC<br>Male Luer spin lock<br>connector/PC<br>Tubing/PVC<br>Pinch clamp/POM<br>Female Luer/PC<br>Female Luer<br>Cap/HDPE<br>Fluid Path Body/Si, PC<br>Stabilizer/ Polystyrene,<br>Acrylic PSA,<br>Polycoated LDPE<br>Pressure activated<br>valve/Si                                                                                                                                                                    | Different. The subject<br>stabilizer uses a<br>polystyrene fabric with<br>acrylic PSA and<br>Polycoated LDPE liner<br>instead of TPE on the<br>predicate.<br>There is no T-Port<br>connector on subject<br>device since PIVO™<br>compatibility is not<br>required. The subject<br>device adds a PAV. ISO<br>10993 biocompatibility<br>testing was conducted to<br>demonstrate these<br>changes did not affect the<br>product performance and<br>did not raise any questions<br>of safety or effectiveness.                                                                                                                                                                                                               |
| Overall Length                            | Overall Length:<br>3.7 inches (9.3cm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Overall Length:<br>5.0 inches (12.7cm)                                                                                                                                                                                                                                                                                                                                                                                                                   | Different. The subject<br>device has a shorter<br>overall length.<br>Priming Volume, Device<br>Flow Rate (gravity and<br>pressurized) and Dynamic<br>Hemolysis (per USP<br><788>) testing conducted<br>did not raise any questions<br>of safety or effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Priming Volume                            | < 1 mL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | < 1 mL                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Particulate                               | USP 788                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | USP 788                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device flow rate<br>(gravity)             | Equal to or greater than<br>3 L/hr.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Equal to or greater<br>than 3 L/hr                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device flow rate<br>(pressurized)         | Equal to or greater than<br>10ml/second from the<br>device distal end at a<br>pressure at or below<br>325psi.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Use with low pressure<br>power injectors up to<br>325psi and maximum<br>flow rate of 10mL/sec                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Energy Source                             | User Operated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | User Operated                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Sterilization<br>Method & Minimum<br>SAL  | Ethylene Oxide<br>Minimum SAL 1X10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Ethylene Oxide<br>Minimum SAL 1X10-6                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Comparison<br>Element                     | Subject Device -<br>NP Medical Extension<br>Set with stabilizing<br>base                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Predicate Device -<br>Velano ExT™<br>Extension Set<br>(K200439)                                                                                                                                                                                                                                                                                                                                                                                          | Analysis of Differences                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Packaging                                 | Reinforced paper /<br>formed blister (PP/LDPE)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Reinforced paper /<br>formed blister (nylon)                                                                                                                                                                                                                                                                                                                                                                                                             | Different. The subject uses<br>a Tyvek + PP/LDPE<br>package that meets ISO<br>11135 and ISO 11607-1<br>requirements.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Disposable or<br>Reusable                 | Disposable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Disposable                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for NP Medical. The logo consists of a stylized blue symbol on the left, followed by the text "NPMedical" in a sans-serif font. The symbol is a geometric shape with rounded corners, and the text is in a dark blue color.

We Make Quality Patient Care Possible™

## 510(k)Summary

with 21 CFR 807 92: The following

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for NP Medical. The logo consists of a blue abstract shape on the left, followed by the text "NPMedical" in a blue sans-serif font. The "NP" part of the text is in a darker shade of blue compared to the "Medical" part.

We Make Quality Patient Care Possible

#### 510(k)Summarv

# 510(k)Summary

The following summary is provided in accordance with 21 CFR 807.92:

# VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

# 1. Performance Testing - Bench

The functional performance of the proposed device has been evaluated using the appropriate methodology as specified in the FDA recognized consensus standards and in conjunction with NP Medical established test protocols in accordance with the recommendations provided in the FDA quidance document titled "Intravascular Administration Sets Premarket Notification Submissions [510(k)]".

The following functional performance tests were conducted as part of the Design Verification and Validation activities:

- Device flowrate (gravity and pressurized) ●
- Retrograde pressure resistance
- Dynamic hemolysis ●
- Positive pressure ●
- 325psi burst resistance (for rotating tube connection and other joints) ●
- Extension tube bond integrity ●
- Stabilization pad bond integrity ●
- Extension tubing port (rotating connection) interfaces, sub-atm. pressure air leakage ●
- Resistance to overriding ●
- ESC resistance to lipids ●
- PAV robustness ●
- Particulate contamination, sub-visible particles .

The results of the functional performance testing demonstrated that the proposed device met all requirements and the differences in design, such as rotating tube connection joint and intecrated bidirectional Pressure-Activated Valve (PAV) feature do not affect the final product performance and the safety or effectiveness of the proposed device.

# 2. Biocompatibility

The biocompatibility evaluation was conducted in accordance with the FDA Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", as recognized by FDA.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for NP Medical. The logo consists of a blue icon on the left and the text "NP Medical" on the right. The icon is a stylized shape that resembles a rounded square with an arrow pointing to the right. The text "NP Medical" is written in a sans-serif font, with "NP" in a darker blue and "Medical" in a lighter blue.

## 510(k)Summarv

The following summary is provided in accordance with 21 CFR 807.92: Biocompatibility of intact skin surface device (Stabilizing Base) and indirect blood path external communicating device (Fluid Pathway) with prolonged contact duration were confirmed by evaluating the following ISO 10993-1 recommended biocompatibility evaluation endpoints.

Stabilizing Base

- Cytotoxicity .
- Sensitization ●
- Irritation or Intracutaneous Reactivity ●

Extension Set Fluid Pathway

- Chemical Characterization .
- Establishment of Allowable Limits ●
- Cytotoxicity ●
- Sensitization ●
- Irritation or Intracutaneous Reactivity
- Acute Systemic Toxicity
- Material Mediated Pyrogenicity ●
- Hemocompatibility ●

All the Biocompatibility study results satisfy the acceptance criteria specified by the above applicable standards and proposed device has been found biocompatible for its intended use/indications for use.

#### 3. Sterilization and Shelf Life

The sterilization validation activities have been performed using Ethylene Oxide sterilization method in accordance with ISO 11135-2014 Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices to state the Sterility Assurance Level (SAL) of 10-9 for the proposed device labeled as sterile.

The stability of the package in maintaining sterility has been assessed through the transit and storage life cycles in accordance with the ISO 11607-1:2019 Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems.

#### VIII. CLINICAL DATA

Not applicable.

#### IX. CONCLUSIONS

NP Medical believes that the NP Medical Extension Set with Stabilizing Base is to be as safe, as effective, and substantially equivalent in performance to the above identified predicate device.

---

**Source:** [https://fda.innolitics.com/device/K221518](https://fda.innolitics.com/device/K221518)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
