← Product Code [DWA](/productcode/DWA) · K221491

# Anivia SG1000 Pump Console (K221491)

_Apmtd, Inc. · DWA · Feb 3, 2023 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221491

## Device Facts

- **Applicant:** Apmtd, Inc.
- **Product Code:** [DWA](/productcode/DWA.md)
- **Decision Date:** Feb 3, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4380
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Anivia SG1000 Pump Console is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass (up to 6 hours).

## Device Story

The Anivia SG1000 Pump Console is an electromechanical system for driving and controlling centrifugal blood pumps in extracorporeal circuits. It consists of a display/control module, pump driver module, flow/bubble sensor, and backup battery. The device uses a brushless DC motor to provide non-contact, magnetically coupled pumping. It is operated by perfusionists, surgical nurses, or physician assistants in operating rooms, cardiac cath labs, or ICUs. The system monitors flow, speed, pressure, and temperature via external sensors and provides visual/auditory alarms for abnormal conditions. It does not contact blood directly. The console assists clinicians in maintaining blood flow during cardiopulmonary bypass procedures, providing stability and emergency backup power/manual hand-crank functionality.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including software verification/validation, functional design verification, electrical safety, EMI/EMC testing, interoperability evaluation, reliability testing, simulated use, and cleaning validation.

## Technological Characteristics

Electromechanical pump console; brushless DC motor; non-contact magnetic coupling. Modules: Display/Control (touchscreen/rotary knob), Pump Driver, Flow/Bubble Sensor (ultrasound clamp-on), Backup Battery (LiFePO4). Power: 90-264 VAC. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304, ISO 14971, IEC 62366, UL 2054, UN38.3. Non-sterile, non-patient contacting.

## Regulatory Identification

A cardiopulmonary bypass pump speed control is a device used that incorporates an electrical system or a mechanical system, or both, and is used to control the speed of blood pumps used in cardiopulmonary bypass surgery.

## Predicate Devices

- Bio-console® 560 Extracorporeal Blood Pumping Console ([K131964](/device/K131964.md))

## Reference Devices

- Rotaflow® Centrifugal Pump System ([K991864](/device/K991864.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 3, 2023

APMTD, Inc. John Sasso Director Of Projects and Programs 226 Lowell Street, Unit B16 Wilmington, Massachusetts 01887

Re: K221491

Trade/Device Name: Anivia SG1000 Pump Console Regulation Number: 21 CFR 870.4380 Regulation Name: Cardiopulmonary Bypass Pump Speed Control Regulatory Class: Class II Product Code: DWA Dated: January 3, 2023 Received: January 5, 2023

Dear John Sasso:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Nicole M. Gillette -S

Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221491

Device Name Anivia SG1000 Pump Console

#### Indications for Use (Describe)

The Anivia SG1000 Pump Console is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass (up to 6 hours).

Contraindications:

The Anivia SG1000 Pump Console is contraindicated as a cardiotomy suction device.

It is the responsibility of the physician to determine whether any physical impairment of the patient or associated equipment would contraindicate the use of this device.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for APMTD, which stands for Asia Pacific Medical Technology Development. The logo consists of a circular graphic on the left, followed by the text "APMTD" in a bold, teal font. Below the letters, the full name of the organization, "Asia Pacific Medical Technology Development," is written in a smaller font size and the same teal color.

## 510(k) Summary of Safety and Effectiveness as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which Substantial Equivalence is based

#### 1. 510(k) Owner's Information

| 510(k) Owner's Name: | APMTD Inc.                                                                                           |
|----------------------|------------------------------------------------------------------------------------------------------|
| Address:             | 226 Lowell Street, Unit B16<br>Wilmington, MA 01887<br>USA                                           |
| Telephone:           | 1-949-623-8403                                                                                       |
| Contact Information: | John Sasso<br>Director of Projects and Programs<br>Telephone: 781-962-6274<br>Email: jsasso@apmtd.us |
| Date Prepared:       | December 23, 2022                                                                                    |
| Device Name          |                                                                                                      |
| Trade Name:          | Anivia SG1000 Pump Console                                                                           |
| Common Name:         | Pump Console, Cardiopulmonary Bypass, Heart-Lung Machine                                             |
| Classification:      | 870.4380, Class II Cardiopulmonary bypass pump speed control                                         |
| Product Code:        | DWA                                                                                                  |

#### 3. Predicate Device:

2.

- A. Bio-console® 560 Extracorporeal Blood Pumping Console, K131964, Medtronic, Inc. (DWA)
#### 4. Device Description

The Anivia SG1000 Pump Console is a device for pumping blood without direct contact in an extracorporeal blood circulation circuit. The device is prescribed (Rx) by specialists to be used in Cardiopulmonary Bypass (CPB) procedures. The Anivia SG1000 Pump Console consists of the following modules:

| Modules                   | Catalog Reference Number | Part Number       |
|---------------------------|--------------------------|-------------------|
| Display & Control Module  | Anivia SG1000-DCM-001    | CS-0510-ASY-00001 |
| Pump Driver Module        | Anivia SG1000-PCM-001    | CS-0210-ASY-00002 |
|                           | Anivia SG1000-PCM-002    | CS-0210-ASY-00010 |
| Flow Bubble Sensor Module | Anivia SG1000-FBS-001    | CS-0410-ASY-00001 |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for APMTD, which stands for Asia Pacific Medical Technology Development. The logo consists of a circular graphic on the left and the text "APMTD Asia Pacific Medical Technology Development" on the right. The graphic is a teal-colored circle with a white design inside. The text is also teal, with "APMTD" in a larger font than "Asia Pacific Medical Technology Development."

|                       | Anivia SG1000-FBS-002 | CS-0410-ASY-00002 |
|-----------------------|-----------------------|-------------------|
| Backup Battery Module | Anivia SG1000-BBM-001 | CS-0710-ASY-00001 |
| Cart Module           | Anivia SG1000-CRT-001 | CS-0810-ASY-00001 |

The Anivia SG1000 Pump Console is intended to be used in conjunction with other previously 510(k)-cleared devices and sterile accessories such as cannulas, catheters, oxygenators, luer connectors and single use centrifugal pump heads, supplied by hospital users or other manufacturers.

#### 5. Indications for Use

The Anivia SG1000 Pump Console is intended to pump blood through the extracorporeal bypass circuit for extracorporeal support for periods appropriate to cardiopulmonary bypass (up to 6 hours).

#### Contraindications:

The Anivia SG1000 Pump Console is contraindicated as a cardiotomy suction device.

It is the responsibility of the physician to determine whether any physical impairment of the patient or associated equipment would contraindicate the use of this device.

#### 6. Intended Users

The Anivia SG1000 Pump Console is designed for use by licensed physicians and specialists to perform cardiopulmonary bypass procedures. Trained perfusionists, surgical nurses and physician assistants may assist in the setup of the equipment and preparation of patients.

## 7. Intended Patient Population and Intended Part of Body

The device is intended to be used for patients needing cardiopulmonary bypass. Suitability of the device for specific patients is determined by licensed and trained physicians and specialists.

#### 8. Intended Use Environment

The Anivia SG1000 Pump Console is intended to be used in an operating room, a cardiac catheterization laboratory, in an intensive care unit, and placed outside a sterile field.

The device is supplied non-sterile.

#### 9. Technological Characteristics

The intended use and technological characteristics are compared with one predicate, legally marketed device and one reference device in the table below.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for APMTD Asia Pacific Medical Technology Development. The logo consists of a circular graphic on the left and the text "APMTD Asia Pacific Medical Technology Development" on the right. The graphic is a stylized representation of a medical symbol. The text is in a dark teal color.

| Product Name<br>and Model                                     | Anivia SG1000 Pump<br>Console                                                                                                                                                                                                                                   | Bio-Console® 560<br>Extracorporeal Blood<br>Pump Console<br>(Predicate)                                                                                                                                                                                      | Rotaflow® Centrifugal<br>Pump System<br>(Reference Device)                                                                                                                                                               |
|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                                  | APMTD Inc.<br>USA                                                                                                                                                                                                                                               | Medtronic Inc.<br>USA                                                                                                                                                                                                                                        | Getinge Maquet<br>Cardiopulmonary AG,<br>Germany                                                                                                                                                                         |
| FDA 510(k)<br>number                                          | K221491                                                                                                                                                                                                                                                         | K131964                                                                                                                                                                                                                                                      | K991864                                                                                                                                                                                                                  |
| Indications for<br>Use                                        | The Anivia SG1000<br>Pump Console is<br>intended to pump blood<br>through the<br>extracorporeal bypass<br>circuit for extracorporeal<br>support for periods<br>appropriate to<br>cardiopulmonary bypass<br>(up to 6 hours).                                     | The Medtronic<br>centrifugal blood<br>pumping system is<br>intended to pump blood<br>through the<br>extracorporeal bypass<br>circuit for extracorporeal<br>support for periods<br>appropriate to<br>cardiopulmonary bypass<br>procedures (up to 6<br>hours). | The RotaFlow Centrifugal<br>Pump System is intended<br>for use in an<br>extracorporeal perfusion<br>circuit to pump blood<br>during short duration<br>cardiopulmonary bypass<br>procedures lasting six<br>hours or less. |
| FDA<br>Classification<br>Codes                                | Class II, CFR 870.4380,<br>DWA,<br>Cardiopulmonary<br>bypass pump speed<br>control                                                                                                                                                                              | Class II, CFR 870.4380,<br>DWA<br>Cardiopulmonary bypass<br>pump speed control                                                                                                                                                                               | Class II, CFR 870.4360,<br>KFM<br>Pump, Blood,<br>Cardiopulmonary Bypass,<br>Non-Roller Type                                                                                                                             |
| Duration of Use                                               | Up to 6 hours (limited<br>by disposables, not<br>limited by Console<br>electro-mechanical<br>modules)                                                                                                                                                           | Up to 6 hours                                                                                                                                                                                                                                                | Up to 6 hours                                                                                                                                                                                                            |
| Intended Use                                                  | Pump Speed Control,<br>CPB Machine Console<br>(not including sterile,<br>blood-contacting<br>accessories)                                                                                                                                                       | Pump Speed Control,<br>CPB Machine Console<br>(not including sterile,<br>blood-contacting<br>accessories)                                                                                                                                                    | Pump Speed Control, CPB<br>Machine Console<br>(bundled with sterile,<br>blood-contacting<br>accessories, outside the<br>scope of this application)                                                                       |
| Product Name<br>and Model                                     | Anivia SG1000 Pump<br>Console                                                                                                                                                                                                                                   | Bio-Console® 560<br>Extracorporeal Blood<br>Pump Console<br>(Predicate)                                                                                                                                                                                      | Rotaflow® Centrifugal<br>Pump System<br>(Reference Device)                                                                                                                                                               |
| Intended Users                                                | Cardiopulmonary<br>physicians,<br>perfusionists, trained<br>physician assistants                                                                                                                                                                                | Cardiopulmonary<br>physicians, perfusionists,<br>trained physician<br>assistants                                                                                                                                                                             | Cardiopulmonary<br>physicians, perfusionists,<br>trained physician<br>assistants                                                                                                                                         |
| Intended Use<br>Environment                                   | Cardiopulmonary<br>procedure rooms,<br>intensive care units                                                                                                                                                                                                     | Cardiopulmonary<br>procedure rooms,<br>intensive care units                                                                                                                                                                                                  | Cardiopulmonary<br>procedure rooms,<br>intensive care units                                                                                                                                                              |
| Intended<br>Patients                                          | As prescribed by<br>cardiopulmonary<br>specialists                                                                                                                                                                                                              | As prescribed by<br>cardiopulmonary<br>specialists                                                                                                                                                                                                           | As prescribed by<br>cardiopulmonary<br>specialists                                                                                                                                                                       |
| Compatible<br>Disposable<br>Centrifugal<br>Pump Heads         | Medtronic BPX-80, BP-<br>50 centrifugal pump<br>heads (w/ Pump Driver<br>Module SG1000-PDM-<br>001)<br><br>RotaFlow® RF-32<br>centrifugal pump heads<br>(w/ Pump Driver<br>Module SG1000-PDM-<br>002)                                                           | Medtronic Bio-Pump®<br>or Affinity® Series<br>centrifugal pump heads,<br>including BPX-80, BP-<br>50                                                                                                                                                         | RotaFlow® RF-32<br>centrifugal pump heads                                                                                                                                                                                |
| Centrifugal<br>Pump Driver                                    | Non-contact,<br>magnetically coupled                                                                                                                                                                                                                            | Non-contact,<br>magnetically coupled                                                                                                                                                                                                                         | Non-contact, magnetically<br>coupled                                                                                                                                                                                     |
| Control<br>Function                                           | • Speed Control<br>• Flow Control<br>(Stability)<br>Mechanical Knob and<br>Touch Display                                                                                                                                                                        | • Speed Control<br>• Coast Mode (low set<br>speed)<br>Mechanical Knob                                                                                                                                                                                        | • Speed Control<br>• Flow Control (Stability)<br>• Mechanical Knob                                                                                                                                                       |
| Product Name<br>and Model                                     | Anivia SG1000 Pump<br>Console                                                                                                                                                                                                                                   | Bio-Console® 560<br>Extracorporeal Blood<br>Pump Console<br>(Predicate)                                                                                                                                                                                      | Rotaflow® Centrifugal<br>Pump System<br>(Reference Device)                                                                                                                                                               |
| Components                                                    | • Pump Driver Module<br>• Touch screen control<br>and display panel<br>• Flow Bubble Sensor<br>• Backup Battery<br>Module<br>• Mobile cart with<br>adjustable Support<br>Arm (optional<br>accessory)<br>• Backup Pump Driver<br>Module<br>• Emergency Handcrank | • Bio-Console 560 Pump<br>Speed Controller<br>• Flow Transducer<br>• Backup battery<br>• 540T External Drive<br>(Pump) Unit<br>• PBS Cabinet<br>• Adjustable Support<br>Arm<br>• Emergency Handcrank                                                         | • Integrated flow and<br>bubble sensors in the<br>drive unit<br>• Integrated battery back-<br>up<br>• Rotaflow Console<br>• Backup battery<br>• Adjustable Support Arm<br>• Emergency Handcrank                          |
| Pump Speed<br>(RPM)                                           | 0 – 4500 for CS-0210-<br>ASY-00002 (Medtronic<br>BPX-80 compatible<br>Pump Driver Module)<br>0 – 5000 for CS-0210-<br>ASY-00010 (Maquet<br>Rotaflow RF-32<br>compatible Pump Driver<br>Module)                                                                  | 0 - 4500                                                                                                                                                                                                                                                     | 0 - 5000                                                                                                                                                                                                                 |
|                                                               | Same as predicate.                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                          |
| Visual and<br>Auditory<br>Alarms on<br>Abnormal<br>Conditions | Yes, preset limits:<br>Speed, Flow Rate, Back<br>Flow, Bubble, Pressure,<br>Temperature                                                                                                                                                                         | Yes, preset limits:<br>Speed, Flow Rate, Back<br>Flow, Bubble, Pressure,<br>Temperature                                                                                                                                                                      | Yes, preset limits                                                                                                                                                                                                       |
|                                                               | Same as predicate                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                          |
| Blood Flow<br>Rate (L/min)                                    | Dependent on external<br>circuit, up to 0 – 9.9<br>L/min                                                                                                                                                                                                        | Dependent on external<br>circuit, up to 0 – 9.9<br>L/min                                                                                                                                                                                                     | Dependent on external<br>circuit, up to 0 – 9.9<br>L/min                                                                                                                                                                 |
|                                                               | Same as predicate                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                          |
| Interface to<br>Blood Flow<br>Sensor                          | Yes (Qty 1)                                                                                                                                                                                                                                                     | Yes (Qty 1)                                                                                                                                                                                                                                                  | Yes (Qty 1)                                                                                                                                                                                                              |
|                                                               | Same as predicate                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                          |
| Air Bubble<br>Detector                                        | Yes (Qty 1), integrated<br>with Blood Flow Sensor                                                                                                                                                                                                               | Yes (Qty 1)                                                                                                                                                                                                                                                  | Yes (Qty 1)                                                                                                                                                                                                              |
|                                                               | Same as predicate                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                          |
| Product Name and Model                                        | Anivia SG1000 Pump Console                                                                                                                                                                                                                                      | Bio-Console® 560 Extracorporeal Blood Pump Console (Predicate)                                                                                                                                                                                               | Rotaflow® Centrifugal Pump System (Reference Device)                                                                                                                                                                     |
| Blood Flow and Bubble Detector Sensor Technology              | Non-contact, ultrasound Clamp-On around blood tube                                                                                                                                                                                                              | Non-contact, electrical conductivity, Clamp-On around blood tube                                                                                                                                                                                             | Non-contact, ultrasound Clamp-On around blood tube                                                                                                                                                                       |
| Pressure Sensors                                              | Yes (Qty 2, external), previously 510(k) cleared accessories manufactured by third parties                                                                                                                                                                      | Yes (Qty 2, external), previously 510(k) cleared accessories manufactured by third parties                                                                                                                                                                   | N/A                                                                                                                                                                                                                      |
| Temperature Sensors                                           | Yes (Qty 2, external), previously 510(k) cleared accessories manufactured by third parties                                                                                                                                                                      | Yes (Qty 2, external), previously 510(k) cleared accessories manufactured by third parties                                                                                                                                                                   | N/A                                                                                                                                                                                                                      |
| Power Input                                                   | Universal<br>90-264 VAC/50-60 Hz<br>2.5A / 1.3A<br>up to 250 W<br>Same as predicate – single universal power supply, instead of different models, no effect on safety and effectiveness.                                                                        | Three models:<br>100 VAC, 50-60 Hz, 3 amps<br>110-120 VAC, 50-60 Hz, 3 amps<br>220-240 VAC, 50-60 Hz, 1.5 amps                                                                                                                                               | Factory Set<br>100/115/230/240 V AC<br>2A / 1A                                                                                                                                                                           |
| Backup Battery                                                | LiFePO4<br>Improvement in technology does not raise question on safety and effectiveness.                                                                                                                                                                       | Lead acid                                                                                                                                                                                                                                                    | NiCad                                                                                                                                                                                                                    |
| Backup Battery Capacity                                       | 25.6 VDC, 12 AH, 307 WH,<br>minimum 1 hour, up to 3 hours depending on speed and flow<br>Same as predicate                                                                                                                                                      | 24 VDC, capacity not listed                                                                                                                                                                                                                                  | 24 VDC, 5 AH, 120 WH                                                                                                                                                                                                     |
| Product Name<br>and Model                                     | Anivia SG1000 Pump<br>Console                                                                                                                                                                                                                                   | Bio-Console® 560<br>Extracorporeal Blood<br>Pump Console<br>(Predicate)                                                                                                                                                                                      | Rotaflow® Centrifugal<br>Pump System<br>(Reference Device)                                                                                                                                                               |
| Backup Pump                                                   | Yes, backup electrical<br>Pump Driver Module on<br>standby, and hand-crank<br><br>Same as predicate                                                                                                                                                             | Yes, hand-crank                                                                                                                                                                                                                                              | Yes, hand-crank                                                                                                                                                                                                          |
| Pump Motor<br>Technology                                      | Brushless DC motor<br><br>Same as predicate                                                                                                                                                                                                                     | Brushless DC motor                                                                                                                                                                                                                                           | Brushless DC motor                                                                                                                                                                                                       |
| Display Screen                                                | 31 cm (12.1")<br><br>Difference has no<br>impact on safety and<br>effectiveness                                                                                                                                                                                 | 26 cm (10.4")                                                                                                                                                                                                                                                | None                                                                                                                                                                                                                     |
| Dimensions,<br>Display &<br>Control<br>Module                 | 14 cm W x 15 cm H x 9<br>cm D<br><br>Difference has no<br>impact on safety and<br>effectiveness                                                                                                                                                                 | 31.88 cm (12.55 in) high<br>by 22.83 cm (8.99 in)<br>wide by 43.02 cm (16.9<br>in) long                                                                                                                                                                      | 18 x 39 x 24 cm                                                                                                                                                                                                          |
| Weight,<br>Display &<br>Control<br>Module                     | 3.3 kg – Display &<br>Control Module<br>4 kg – Backup Battery<br>Module<br><br>Difference (mostly due<br>to newer technology)<br>has no impact on safety<br>and effectiveness.                                                                                  | 17.2 kg<br>including backup<br>batteries                                                                                                                                                                                                                     | 14.4 kg including backup<br>battery                                                                                                                                                                                      |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for APMTD Asia Pacific Medical Technology Development. The logo consists of a circular graphic on the left and the text "APMTD Asia Pacific Medical Technology Development" on the right. The graphic is a stylized representation of a medical symbol. The text is in a dark teal color.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for APMTD Asia Pacific Medical Technology Development. The logo consists of a circular graphic on the left and the text "APMTD Asia Pacific Medical Technology Development" on the right. The graphic is a stylized representation of a medical symbol. The text is in a dark teal color.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for APMTD, which stands for Asia Pacific Medical Technology Development. The logo consists of a circular graphic on the left and the text "APMTD" in a bold, sans-serif font on the right. Below the text "APMTD" is the full name of the organization, "Asia Pacific Medical Technology Development", in a smaller font. The circular graphic is a stylized representation of interconnected shapes, possibly symbolizing collaboration or innovation.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for APMTD, which stands for Asia Pacific Medical Technology Development. The logo consists of a circular graphic on the left, followed by the text "APMTD" in a large, bold font. Below the text "APMTD" is the full name of the organization, "Asia Pacific Medical Technology Development", in a smaller font. The logo is teal in color.

The technological characteristics of the Anivia SG1000 Pump Console device are substantially the same as predicate device and where there are minor differences they do not raise different questions of safety and effectiveness.

The geometry and design parameters are consistent with the device's intended use as an electromechanical pump driver and controller in cardiopulmonary bypass procedures.

- Pump Driver Module is based on a brushless DC motor control electronics, and a A. contactless magnetic coupler driving a detachable one-time use centrifugal pump.

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for APMTD, which stands for Asia Pacific Medical Technology Development. The logo consists of a circular graphic on the left, followed by the text "APMTD" in a large, bold font. Below the text "APMTD" is the full name of the organization, "Asia Pacific Medical Technology Development", in a smaller font. The circular graphic is a stylized design, possibly representing a medical or technological concept.

- Display & Control Module is based on a video display, a touch screen, a rotary knob, control B. electronics, isolated electronic signal interface for external sensors (accessories), and a medical grade AC-to-DC power supply.
- C. Blood Flow Bubble Sensor Module is clamp-on non-contact flow meter, compatible with commonly used plastic blood tubes of specified diameters and material.
- D. Backup Battery Module is based on lithium iron phosphate (LiFePO4) chemistry that can provide at least 60 minutes of backup power in case of brief power outage of the AC power line, or for supporting transport within a hospital.

Not included in this 510(k) application are the following accessories:

- Cart Module (optional, supplied by Company, Product Code BZN, Class I, 510(K) Exempt) E. is a hand cart for mounting the Display & Control Module, the Pump Driver Module, the Flow Bubble Sensor Module, and the Backup Battery Module. The cart also provides a simple work surface, a storage bin, two IV poles, a holder for a gas cylinder, and locking caster wheels.
- F. The Anivia SG1000 Pump Console may be used with the following previously 510(k) cleared devices, supplied by hospital users:
	- Blood pressure sensors, supplied sterile, resistive bridge type that complies with a. ANSI/AAMI BP-22, 1994 and IEC 60601-2-34:Ed.3.0 standards, as listed in the Instructions for Use
	- b. Temperature Sensors, supplied sterile, that meets YSI-400 and ISO 80601-2-56:2017 standards, as listed in the Instructions for Use
	- One time use centrifugal pump heads, supplied sterile, as listed in the Instructions for Use c.

## 9.1. Biocompatibility

Not applicable. The Anivia SG1000 Pump Console does not contain any blood contacting or patient contacting devices.

## 9.2. Sterility and Shelf-Life

Not applicable. The Anivia SG1000 Pump Console and accessories are provided non-sterile and have been tested with sterile, one-time use accessories specified in the Instructions for Use.

## 9.3. Non-Clinical Tests

The Anivia SG1000 Pump Console performance characteristics were demonstrated through system bench testing, mechanical testing, electrical safety and electromagnetic interference and compatibility testing, software testing, reliability, and usability testing.

The following performance tests were conducted on the Anivia SG1000 Pump Console System to support the determination of substantial equivalence:

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for APMTD, which stands for Asia Pacific Medical Technology Development. The logo consists of a circular graphic on the left, followed by the text "APMTD" in a bold, serif font. Below the text "APMTD" is the full name of the organization, "Asia Pacific Medical Technology Development", in a smaller font size. The logo is teal in color.

- Software verification and validation testing .
- . Functional design verification and validation testing
- . Electrical safety, electromagnetic interference and compatibility (EMI/EMC) testing
- . Interoperability evaluation with specified accessories
- . Reliability testing
- . Simulated use testing
- . Cleaning validation
- Packaging and ship testing

All testing met predetermined acceptance criteria.

The Anivia SG1000 Pump Console was also tested and certified by accredited third party laboratories to meet the following consensus standards:

- IEC 60601-1:2005/AMD1:2012 ●
- IEC 60601-1-2:2014 ●
- IEC 60601-1-6:2010/AMD1:2013 ●
- IEC 60601-1-8:2006/AMD1:2012 ●
- ISO 80601-2-56:2017/AMD1:2018 ●
- IEC 60601-2-34:2011
- IEC 80601-2-49:2018 ●
- ISO 14971:2019
- IEC 62304:2006/AMD 1: 2015 ●
- IEC 62366:2015
- UL 2054:2004 R9.11
- IEC 62133-2:2017/AMD1:2021 ●
- UN38.3, Seventh Edition ●
- ASTM D4169:2022, DC13, Level I ●
- ISO 17664-2:2021 ●

## 9.4. Clinical Tests

Not Applicable

#### 9.5. Labeling

The labeling includes instructions on circuit setup, explanation of the hardware and software user interface, visual and auditory alarms, maintenance during a procedure, precautions and warnings, trouble shooting guide, and performance characteristics relevant to compatibility among different devices and accessories in the circuit.

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image is a logo for APMTD, which stands for Asia Pacific Medical Technology Development. The logo consists of a circular graphic on the left and the text "APMTD" in large, bold letters on the right. Below the letters is the full name of the organization in smaller letters. The graphic is a stylized representation of a medical symbol.

#### 10. Conclusions

Based on the above comparisons of intended use, intended users, use environment, indications for use, operating principles, technological characteristics, performance test data, and compliance with the listed consensus standards, the Anivia SG1000 Pump Console is as safe and effective as the predicate device, the Medtronic BioConsole (K131964).

---

**Source:** [https://fda.innolitics.com/device/K221491](https://fda.innolitics.com/device/K221491)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
