K221487 · Gunze Limited · OXC · Aug 26, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221487
Device Name
NEOVEIL Staple Line Reinforcement
Applicant
Gunze Limited
Product Code
OXC · General, Plastic Surgery
Decision Date
Aug 26, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
NEOVEIL® Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. NEOVEIL® Staple Line Reinforcement of staple lines during lung resection, bronchial, bariatric, colon, colorectal, esophagus, gastric, mesentery, pancreas, and small bowel procedures.
Device Story
NEOVEIL® Staple Line Reinforcement is a biodegradable, nonwoven polyglycolic acid (PGA) mesh sleeve used to reinforce staple lines during surgical procedures. The device is applied by sliding sleeve components over the jaws of a mechanical surgical stapler; the mesh is held in place by non-absorbable elastic knit (polyurethane/nylon) and PGA tacking threads. After the stapler is fired, the elastic knit and tacking threads are removed and discarded, leaving the PGA mesh at the surgical site. Used in clinical settings (e.g., OR) by surgeons, the device provides mechanical reinforcement to tissue transections. The modification in this submission involves longer tacking threads to improve compatibility with various endoscopic stapler guns. The device is intended to reduce potential leaks or bleeding at staple lines, benefiting patients by supporting tissue integrity during healing.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical performance testing, including simulated usability testing, stapler insertion/removal force analysis, firing force testing, and staple formation assessment conducted per internal GUNZE methods and relevant standards (IEC 62366-1, ISO 14971).
Technological Characteristics
Material: Biodegradable polyglycolic acid (PGA) nonwoven mesh; non-absorbable elastic knit (polyurethane and nylon) used for application. Form factor: Sleeve/tube configuration. Energy source: None (mechanical application). Sterilization: Not specified. Standards: IEC 62366-1 (usability), ISO 14971 (risk management).
Indications for Use
Indicated for patients undergoing surgical procedures requiring soft tissue transection or resection, specifically lung, bronchial, bariatric, colon, colorectal, esophageal, gastric, mesenteric, pancreatic, and small bowel procedures where staple line reinforcement is required.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
Predicate Devices
NEOVEIL™ Tube/Sheet Type Suture and Staple Line Reinforcement Material (K130997)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
August 26, 2022
Gunze Limited % Stuart Goldman Senior Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
Re: K221487
Trade/Device Name: NEOVEIL Staple Line Reinforcement Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OXC Dated: July 27, 2022 Received: July 27, 2022
Dear Stuart Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Deborah Fellhauer Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221487
Device Name NEOVEIL® Staple Line Reinforcement
#### Indications for Use (Describe)
NEOVEIL® Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. NEOVEIL® Staple Line Reinforcement of staple lines during lung resection, bronchial, bariatric, colon, colorectal, esophagus, gastric, mesentery, pancreas, and small bowel procedures.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Special 510(k) Summary NEOVEIL® Staple Line Reinforcement
## 1. Submission Sponsor
GUNZE LIMITED Medical Division 46 Natsumegaichi, Aono Ayabe, Kyoto Japan 623-8513 Contact: Mr. Hidenori Nishioka Title: Regulatory Affairs
## 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Contact: Mr. Stuart R. Goldman Title: Sr. Consultant RA/QA
## 3. Date Prepared
August 23, 2022
## 4. Device Identification
| Type of 510(k): | Special 510(k) |
|-------------------------|-------------------------------------------------------|
| Trade Name: | NEOVEIL® Staple Line Reinforcement |
| Product Code: | OXC |
| Classification Name: | Mesh, Surgical, Absorbable, Staple Line Reinforcement |
| Regulation Number: | 21 CFR 878.3300 |
| Regulation Description: | Surgical Mesh |
| Device Class: | 2 |
| Review Panel: | General & Plastic Surgery |
### 5. Legally Marketed Predicate Device
| Trade Name: | NEOVEIL™ Tube/Sheet Type Suture and Staple Line Reinforcement Material |
|---------------|------------------------------------------------------------------------|
| 510(k) No.: | K130997 |
| Manufacturer: | GUNZE LIMITED |
The predicate device has not been subject to a design related recall.
### 6. Indications for Use
NEOVEIL® Staple Line Reinforcement is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. NEOVEIL® Staple Line Reinforcement can be used for
{4}------------------------------------------------
reinforcement of staple lines during lung resection, bronchial, bariatric, colon, colorectal, esophagus, gastric, mesentery, pancreas, and small bowel procedures.
# 7. Device Description
NEOVEIL® Staple Line Reinforcement is composed of a biodegradable synthetic polymer, polyglycolic acid (PGA) and is offered as a nonwoven surgical mesh configured into sleeves. The device is applied to the surgical site via a mechanical stapler with two jaws, where one piece of the sleeve is slid over each jaw of the stapler. This is accomplished by attaching a similar-sized piece of nonabsorbable elastic knit to the PGA felt that is held together by means of PGA tacking threads. After deployment of the tube type reinforcement material, the non-degradable elastic knits, comprised of polyurethane and nylon, are removed and discarded along with the PGA tacking sutures. The PGA material is dyed with D&C Green No.6.
## 8. Device Changes
This Special 510(k) is being made by GUNZE to modify one of their existing NEOVEIL™ Tube/Sheet Type Suture and Staple Line Reinforcement Material model devices cleared under K130997. The modification gives that device longer tacking threads to better facilitate its use during certain endoscopic procedures when used with certain stapler guns. A new catalogue number has also been given to this new model. In addition, the trade name of the subject device is also being changed to NEOVEIL® Staple Line Reinforcement to better reflect its intended use.
## 9. Substantial Equivalence Discussion
Except for the minor differences in the size of the absorbable PGA felt and elastic knit and the length of the PGA tacking threads found in NEOVEIL® Staple Line Reinforcement as described above when compared to those same components found in NEOVELL™ Tube/Sheet Type Suture and Staple Line Reinforcement Material (K130997), the subject device is equivalent to the predicate device in intended use, indications for use, technological characteristics and performance.
## 10. Non-Clinical Performance Testing
As part of demonstrating substantial equivalence of the subject device to the predicate device, GUNZE conducted side-by-side non-clinical performance testing in the form of a simulated usability testing on final finished versions of NEOVEIL®™ using a relevant mechanical stapler that received FDA 510(k) clearance. The non-clinical performance testing was done to internal GUNZE test methods. The test methods used were selected based on the 510(k) Premarket Notification of the predicate device (K130997). The performance testing validated that NEOVEIL® Staple Line Reinforcement meets its product specification and performs as intended. Results confirm that the specification requirements for the subject device have been met.
- . Non-Clinical Performance Testing:
- GUNZE internal test method for stapler insertion and removal forces with NEOVEIL®™
- GUNZE internal test method for stapler firing force with NEOVEIL®™ o
- GUNZE internal test method for staple formation with NEOVEIL®™ o
- Usability: ●
- O IEC 62366-1:2015, Application of Usability Engineering to Medical Devices
- Risk Analysis: ●
{5}------------------------------------------------
- o ISO 14971:2019, Application of Risk Management to Medical Devices
## 11. Clinical Performance Data
Not applicable to this submission.
## 12. Substantial Equivalence Conclusion
NEOVEIL® Staple Line Reinforcement has the same intended use and indications for use as the previously cleared predicate device, NEOVEIL™ Tube/Sheet Type Suture and Staple Line Reinforcement Material. The addition of the new size of the subject device and increase in length of the tacking threads for use with other brands of mechanical staplers has been addressed with non-clinical performance testing to well established GUNZE test methods. The test results and analysis support a determination of substantial equivalence of NEOVEIL® Staple Line Reinforcement to the predicate device in terms of safety, effectiveness and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.