← Product Code [HTN](/productcode/HTN) · K221485

# Arthrex FiberTape and TigerTape Cerclage Sutures (K221485)

_Arthrex, Inc. · HTN · Aug 22, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221485

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [HTN](/productcode/HTN.md)
- **Decision Date:** Aug 22, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Arthrex FiberTape and TigerTape cerclage sutures are intended for use in soft tissue approximation and or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft or autograft tissues are used for repair. When used as bone fixation cerclage the sutures are intended for: • Trochanteric reattachment after trochanteric osteotomy following total hip arthroplasty • Sternotomy indications including the “rewiring” of osteomized sternums • Trauma surgery indications including olecranon, ankle, patella and some shoulder fracture rewiring - Treatment of anterior glenoid bone loss using the Latarjet or bone block procedure (allograft or autograft) •Repair of long bone fractures due to trauma or reconstruction

## Device Story

Arthrex FiberTape and TigerTape are flat, braided cerclage sutures; constructed from UHMWPE and polyester polyblend; assembled in loop configuration on HDPE card or ABS loader. Used by surgeons in orthopedic/trauma procedures for soft tissue approximation, ligation, and bone fixation. Device functions as mechanical fixation construct; looped end tied as hitch over sheath. Provides structural support for bone/tissue repair. Clinical benefit includes stabilization of fractures and bone blocks during healing. No software or electronic components.

## Clinical Evidence

Bench testing only. Tensile and pressure distribution testing compared subject device to predicate to ensure mechanical integrity. Bacterial endotoxin testing performed per EP 2.6.14/USP <85>. Clinical literature provided to justify testing conditions and acceptance criteria.

## Technological Characteristics

Materials: UHMWPE and polyester polyblend. Design: Flat braided suture in loop configuration. Assembly: HDPE card or ABS loader. Sterilization: Not specified. Connectivity: None. Software: None.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Double ENDOBUTTON Fixation Device ([K183232](/device/K183232.md))

## Reference Devices

- Arthrex FiberTape Cerclage ([K170206](/device/K170206.md))
- Arthrex Suture ([K122374](/device/K122374.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 22, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Arthrex Inc. Kelsey Roberts Sr. Regulatory Affairs Specialist 1370 Creekside Boulevard Naples, Florida 34108-1945

Re: K221485

Trade/Device Name: Arthrex FiberTape and TigerTape Cerclage Sutures Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HTN, HWC, JDQ, GAT Dated: May 20, 2022 Received: May 24, 2022

Dear Kelsey Roberts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K221485

Device Name

Arthrex FiberTape and TigerTape cerclage sutures

#### Indications for Use (Describe)

Arthrex FiberTape and TigerTape cerclage sutures are intended for use in soft tissue approximation and or ligation. These sutures may be incorporated, as components, into surgeries where constructs including those with allograft or autograft tissues are used for repair.

When used as bone fixation cerclage the sutures are intended for:

- · Trochanteric reattachment after trochanteric osteotomy following total hip arthroplasty
- · Sternotomy indications including the "rewiring" of osteomized sternums
- · Trauma surgery indications including olecranon, ankle, patella and some shoulder fracture rewiring
- Treatment of anterior glenoid bone loss using the Latarjet or bone block procedure (allograft or autograft)

·Repair of long bone fractures due to trauma or reconstruction

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Date Prepared                                                                              | August 22, 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                                                                              | K221485                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Submitter                                                                                  | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Contact Person                                                                             | Kelsey N. Roberts<br>Sr. Regulatory Affairs Specialist<br>1-239-643-5553, ext. 72257<br>Kelsey.Roberts@arthrex.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Name of Device                                                                             | Arthrex FiberTape and TigerTape Cerclage Sutures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Common Name                                                                                | Bone Fixation Cerclage, Suture                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Product Code                                                                               | HTN<br>HWC<br>JDQ<br>GAT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification Name                                                                        | 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances and<br>accessories<br>21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener<br>21 CFR 888.3010: Bone Fixation Cerclage<br>21 CFR 878.5000: Nonabsorbable Poly(ethylene) Terephthalate Surgical Suture                                                                                                                                                                                                                                                                                                                                                           |
| Regulatory Class                                                                           | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Predicate Device                                                                           | K183232 Double ENDOBUTTON Fixation Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Reference Device                                                                           | K170206 Arthrex FiberTape Cerclage<br>K122374 Arthrex Suture                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Purpose of Submission                                                                      | This traditional 510(k) premarket notification is submitted to obtain the indication of<br>anterior glenoid bone loss using the Laterjet or bone block procedure (allograft or<br>autograft) for the Arthrex FiberTape and TigerTape Cerclage Sutures.                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Description                                                                         | The proposed Arthrex FiberTape and TigerTape Cerclage devices are available as a flat<br>braided suture assembled in a loop configuration. Cerclage is assembled on an HDPE<br>card or on an ABS loader. The devices are manufactured from a polyblend of Ultra<br>High Molecular Weight Polyethylene (UHMWPE) and polyester materials. These<br>materials are identical to those cleared in K170206. For the loop assembly, the<br>looped end of the suture is tied as a hitch over a sheath that secures a double loop.                                                                                                                                  |
| Comparison Summary<br>of Technological<br>Characteristics and<br>Modifications<br>Proposed | The proposed devices have the same technological characteristics (device design,<br>sterilization, and biocompatibility). The proposed device modification consists of the<br>additional indication of anterior glenoid bone loss using the Laterjet or bone block<br>procedure.<br><br>The Arthrex FiberTape and TigerTape cerclage sutures are substantially equivalent to<br>the predicate device in which the basic design features and intended uses are the<br>same. Any differences between the proposed device and the predicate device are<br>considered minor and do not raise new or different questions concerning safety or<br>effectiveness. |
| Indications for Use                                                                        | Arthrex FiberTape and TigerTape cerclage sutures are intended for use in soft tissue<br>approximation and or ligation. These sutures may be incorporated, as components,<br>into surgeries where constructs including those with allograft or autograft tissues are<br>used for repair.<br><br>When used as bone fixation cerclage the sutures are intended for:<br>• Trochanteric reattachment after trochanteric osteotomy following total hip<br>arthroplasty                                                                                                                                                                                           |
|                                                                                            | • Sternotomy indications including the “rewiring” of osteomized sternums • Trauma surgery indications including olecranon, ankle, patella and some shoulder fracture rewiring - Treatment of anterior glenoid bone loss using the Latarjet or bone block procedure (allograft or autograft) •Repair of long bone fractures due to trauma or reconstruction                                                                                                                                                                                                                                                                                                 |
| Performance Data                                                                           | Tensile and pressure distribution testing was performed on the subject device and compared to the predicate device to demonstrate that the modifications do not negatively impact mechanical strength. Bacterial endotoxin per EP 2.6.14/USP <85> was conducted to demonstrate that the device meets pyrogen limit specifications.                                                                                                                                                                                                                                                                                                                         |
| Clinical Literature                                                                        | Clinical literature was provided to justify the testing conditions and acceptance criteria of the subject Arthrex Cerclage Sutures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Conclusion                                                                                 | The Arthrex FiberTape and TigerTape Cerclage Sutures are substantially equivalent to the predicate device in which the basic design features and intended use are the same. Any differences between the Arthrex proposed device and the predicate device are considered minor and do not raised questions concerning safety and effectiveness. Based on the indications for use, technological characteristics, and the summary of data submitted, Arthrex Inc. has determined that the proposed device is substantially equivalent to the currently marketed predicate device.                                                                            |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K221485](https://fda.innolitics.com/device/K221485)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
