K221479 · Zimmer, Inc. · JWH · Nov 17, 2022 · Orthopedic
Device Facts
Record ID
K221479
Device Name
Persona Personalized Knee System
Applicant
Zimmer, Inc.
Product Code
JWH · Orthopedic
Decision Date
Nov 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When a mechanical alignment approach is utilized, this device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Collagen disorders, and/or avascular necrosis of the femoral condyle. - Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. - Moderate valgus, varus, or flexion deformities. - The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery. When a kinematic alignment approach is utilized, this device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. - Collagen disorders, and/or avascular necrosis of the femoral condyle. - Moderate valgus, varus, or flexion deformities. The kinematic alignment (KA) surgical technique may only be used with Persona porous and cemented non-porous CR femoral components, Persona CR or UC articular surface components, and the cemented nonporous Persona Natural cemented 5 degree tibial components without a stem extension, Porous coated components may be used cemented (biological fixation), except for the Persona porous 0 degree OsseoTi Spiked Keel Tibia which is for uncemented use only. All other femoral, tibial baseplate and all-polyethylene (UHMWPE and VEHXPE) patella components are indicated for cemented use only.
Device Story
Persona Personalized Knee System is a semi-constrained modular knee prosthesis for total knee arthroplasty; resurfaces articulating surfaces of femoral, tibial, and patellar bones. System includes new components: non-porous cemented stemmed tibial baseplate, porous OsseoTi non-cemented tibial baseplate, and porous plasma-sprayed femoral component. Components provided sterile, single-use; implanted by orthopedic surgeons in clinical settings. Device restores joint function, reduces pain, and improves stability for patients with severe knee pathology. System utilizes mechanical or kinematic alignment surgical techniques. Reusable instruments aid implantation.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including posterior condyle and anterior flange fatigue testing, porous plasma spray and OsseoTi structure evaluation, fatigue strength testing per ASTM F1800 and ISO 14879-1, micromotion and fixation performance assessments, drill perforation analysis, MR compatibility, and bacterial endotoxin testing per ANSI/AAMI ST 72:2011.
Technological Characteristics
Semi-constrained modular knee prosthesis. Materials: Metal/polymer/metal. Features: Porous plasma-sprayed femoral components, cemented non-porous stemmed tibial baseplates, and porous OsseoTi stemmed tibial baseplates. Fatigue testing per ASTM F1800 and ISO 14879-1. Sterilization: Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for patients with severe knee pain and disability due to rheumatoid, osteoarthritis, traumatic, or polyarthritis; collagen disorders; avascular necrosis of femoral condyle; post-traumatic joint loss; or moderate valgus, varus, or flexion deformities. Includes salvage of failed surgical attempts or unstable knees.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 17, 2022
Zimmer, Inc. Gregory Foster Senior Regulatory Affairs Specialist 1800 W. Center Street Warsaw, Indiana 46580
Re: K221479
Trade/Device Name: Persona® Personalized Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH. MBH. OIY Dated: October 17, 2022 Received: October 18, 2022
Dear Gregory Foster:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ting Song -S
Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221479
#### Device Name
Zimmer® Persona™ Personalized Knee System
Indications for Use (Describe)
When a mechanical alignment approach is utilized, this device is indicated for patients with severe knee pain and disability due to:
- · Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- · Collagen disorders, and/or avascular necrosis of the femoral condyle.
- · Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- · Moderate valgus, varus, or flexion deformities.
· The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
When a kinematic alignment approach is utilized, this device is indicated for patients with severe knee pain and disability due to:
- · Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- · Collagen disorders, and/or avascular necrosis of the femoral condyle.
- · Moderate valgus, varus, or flexion deformities.
The kinematic alignment (KA) surgical technique may only be used with Persona porous and cemented non-porous CR femoral components, Persona CR or UC articular surface components, and the cemented nonporous Persona Natural cemented 5 degree tibial components without a stem extension,
Porous coated components may be used cemented (biological fixation), except for the Persona porous 0 degree OsseoTi Spiked Keel Tibia which is for uncemented use only. All other femoral, tibial baseplate and allpolyethylene (UHMWPE and VEHXPE) patella components are indicated for cemented use only.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Sponsor: | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708<br>Establishment Registration Number: 1822565 | | | | | | | | | | | | | | | | | | |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Contact Person: | Gregory Foster<br>Sr. Regulatory Specialist<br>Telephone: (574) 371-0519<br>Fax: (574) 377-3718<br>Gregory.foster@zimmerbiomet.com | | | | | | | | | | | | | | | | | | |
| Date: | 14-Nov-2022 | | | | | | | | | | | | | | | | | | |
| Subject Device: | Trade Name: Persona® Personalized Knee System<br>Common Name: Knee Prosthesis | | | | | | | | | | | | | | | | | | |
| | Classification Name:<br>• JWH – Knee joint patellofemorotibial polymer/metal/polymer<br>semi-constrained cemented prosthesis (21 CFR 888.3560)<br>• MBH - Knee joint patellofemorotibial metal/polymer porous-<br>coated uncemented prosthesis (21 CFR 888.3565)<br>• OIY- Knee joint patellofemorotibial polymer/metal/polymer<br>semi-constrained cemented prosthesis (21 CFR 888.3560) | | | | | | | | | | | | | | | | | | |
| Predicate Device(s): | K113369 Zimmer®<br>Persona™ Knee<br>System Zimmer, Inc. K122745 Zimmer®<br>Persona™ Knee<br>System Zimmer, Inc. K173417 Zimmer®<br>Persona™ Knee<br>System Zimmer, Inc. K113550 Vanguard<br>Complete Knee<br>System Biomet, Inc. K140669 G7 OsseoTi™<br>Acetabular Shells Biomet, Inc. K936159 Natural Knee II<br>System Zimmer, Inc. | | | | | | | | | | | | | | | | | | |
{4}------------------------------------------------
| Device Description: | The purpose of this submission is to add three new components to the<br>Persona Personalized Knee System. These components consist of a non-<br>porous cemented stemmed tibial baseplate, a porous OsseoTiTM non-<br>cemented tibial baseplate, and a porous plasma sprayed femur. The<br>addition of these components do not change the intended use or<br>fundamental scientific technology of the device system. |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The Persona Personalized Knee System is a semiconstrained modular<br>knee prosthesis designed to resurface the articulating surface of the<br>femoral, tibial, and patellar bones. With this submission a new porous<br>plasma sprayed femoral component will be added to the system. These<br>femoral components articulate against tibial and patellar articular surface<br>to form a total knee system. In addition, two new tibial plates will be<br>added, a cemented non-porous stem tibial baseplate and a porous<br>OsseoTiTM stemmed tibial baseplate. All of the new components come in<br>a variety of sizes to match the needs of a patient's anatomy when<br>performing total knee arthroplasty. These components are provided<br>sterile and single use. To aid the implantation of these devices several<br>new reusable instruments are also introduced. |
| Indications for Use: | When a mechanical alignment approach is utilized, this device is<br>indicated for patients with severe knee pain and disability due to:<br>- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.<br>- Collagen disorders, and/or avascular necrosis of the femoral condyle.<br>- Post-traumatic loss of joint configuration, particularly when there is<br>patellofemoral erosion, dysfunction or prior patellectomy.<br>- Moderate valgus, varus, or flexion deformities.<br>- The salvage of previously failed surgical attempts or for a knee in which<br>satisfactory stability in flexion cannot be obtained at the time of surgery. |
| | When a kinematic alignment approach is utilized, this device is indicated<br>for patients with severe knee pain and disability due to:<br>- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.<br>- Collagen disorders, and/or avascular necrosis of the femoral condyle.<br>- Moderate valgus, varus, or flexion deformities. |
| | The kinematic alignment (KA) surgical technique may only be used with<br>Persona porous and cemented non-porous CR femoral components,<br>Persona CR or UC articular surface components, and the cemented<br>nonporous Persona Natural cemented 5 degree tibial components without<br>a stem extension, |
| | Porous coated components may be used cemented or uncemented<br>(biological fixation), except for the Persona porous 0 degree OsseoTi<br>Spiked Keel Tibia which is for uncemented use only. All other femoral, |
{5}------------------------------------------------
tibial baseplate and all-polyethylene (UHMWPE and VEHXPE) patella components are indicated for cemented use only.
### Summary of Technological Characteristics:
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use: Identical to the predicate .
- Indications for Use: Similar to the predicate ●
- Materials: Identical to the predicate
- Design Features: Identical and similar to the predicate ●
- Sterilization: Identical to the predicate ●
## Summary of Performance Data (Nonclinical and/or Clinical)
- Non-Clinical Tests:
- Posterior condyle fatigue testing per internal test methods. O
- Anterior flange fatigue testing per an internal test method. O
- Evaluation of porous plasma spray (PPS) coating. O
- Evaluation of OsseoTi porous structure. O
- Fatigue strength of the Persona 0° Cemented Keel Tibial o Baseplate per ASTM F1800 and ISO 14879-1.
- Fatigue strength of the Persona 0° Spiked Keel Tibial Baseplate O per ASTM F1800 and ISO 14879-1.
- Evaluation of the level of micromotion of the Persona 0° O OsseoTi tibial baseplate per an internal test method.
- Fixation performance of the Persona 0° Cemented Keel Tibia per о an internal test method.
- Analysis of keel and peg drill perforation rate. O
- Assessment of MR compatibility. O
- O Bacterial Endotoxin Test (BET) per ANSI/AAMI ST 72:2011as part of cleaning validation demonstrating implants meet the limit of <20 Endotoxin units (EU)/Device per USP41-NF36 Chapter <161> Medical Devices - Bacterial Endotoxin and Pyrogen Tests.
{6}------------------------------------------------
#### Clinical Tests: ●
- 0 None provided
## Substantial Equivalence Conclusion
Based on the information contained within this submission, it is concluded that the Persona® Personalized Knee System are substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.