K221476 · X6 Innovations · HIR · Feb 10, 2023 · Obstetrics/Gynecology
Device Facts
Record ID
K221476
Device Name
Perifit
Applicant
X6 Innovations
Product Code
HIR · Obstetrics/Gynecology
Decision Date
Feb 10, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1425
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Perifit is a perineometer designed to treat stress, mild-moderate urge and mixed urinary incontinence in women, by strengthening of the pelvic floor muscles through exercise. This device provides biofeedback via smart phone technology.
Device Story
Perifit is a reusable, single-patient intravaginal perineometer for home use. Device consists of a rigid PC/ABS probe encased in medical-grade silicone, containing force-sensing resistors and a Bluetooth transmitter. User inserts probe into vagina; sensors measure contraction force of pubococcygeus and puborectalis muscles. Data transmits wirelessly to a smartphone application, providing real-time visual biofeedback to guide voluntary Kegel exercises. Output allows users to monitor muscle strength and progress over time. Device is powered by a non-rechargeable CR2032 battery. Intended to treat urinary incontinence by facilitating pelvic floor muscle strengthening.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-5, -10, -11), electrical safety/EMC (IEC 60601-1, -1-2, -1-11), software verification/validation (IEC 62304), and mechanical testing (drop, durability, sensor behavior). User questionnaire confirmed ability to understand labeling and operate the device.
Technological Characteristics
Intravaginal probe with rigid PC/ABS structure and medical-grade silicone sheath. Sensing via force-sensing resistors. Connectivity via Bluetooth to smartphone app. Power: 3.0VDC CR2032 battery. Dimensions: 90mm shaft length. Non-sterile. Biocompatibility per ISO 10993. Electrical safety per IEC 60601-1/1-2/1-11.
Indications for Use
Indicated for adult females with stress, mild-moderate urge, or mixed urinary incontinence to strengthen pelvic floor muscles via exercise-based biofeedback.
Regulatory Classification
Identification
A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.
February 10, 2023
X6 Innovations % Lina Kontos Partner Hogan Lovells US LLP 555 13th Street NW Washington, DC 20016
Re: K221476 Trade/Device Name: Perifit Regulation Number: 21 CFR§ 884.1425 Regulation Name: Perineometer Regulatory Class: II Product Code: HIR Dated: January 9, 2023 Received: January 9, 2023
Dear Lina Kontos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
for Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number: K221476
K221476
Device Name
Perifit
Indications for Use (Describe)
The Perifit is a perineometer designed to treat stress, mild-moderate urge and mixed urinary incontinence in women, by strengthening of the pelvic floor muscles through exercise. This device provides biofeedback via smart phone technology.
Perifit is indicated for an adult female.
Type of Use (Select one or both, as applicable)
□ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY X6 Innovations' Perifit
## Submitter
X6 Innovations 128 rue de la Boétie, 75008 Paris, France Phone: +33 6 51 66 55 94 Contact Person: Artem Rodionov Date Prepared: May 20, 2022
Name of Device: Perifit
Common or Usual Name: Perineometer
Classification Name: (21CFR 884.1425) Perineometer
Regulatory Class: Class II
Product Code: HIR
## Predicate Devices:
Analytica Ltd.'s PeriCoach® OTC (K160758)
#### Intended Use / Indications for Use
The Perifit is a perineometer designed to treat stress, mild-moderate urge and mixed urinary incontinence in women, by strengthening of the pelvic floor muscles through exercise. This device provides biofeedback via smart phone technology.
Perifit is indicated for an adult female.
#### Device Description
The Perifit device consists of a rigid probe covered in a silicone sheath that is temporarily inserted into the vagina. Sensors located under the sheath measure the strength of the user's pelvic floor muscles. This information is then transmitted wirelessly to a smartphone application in order to provide real-time feedback to the user. It is a single patient, reusable device to be supplied over-thecounter.
#### Summary of Technological Characteristics
Pelvic floor muscle contraction and instantaneous feedback to the user's smartphone is the technological principle for both the subject and predicate devices. An intravaginal device with embedded force sensors is used to monitor the contraction force of the user's pelvic floor muscles. This information is transmitted to the patient's smartphone via bluetooth and is displayed on the screen. The subject and predicate devices are based on the following same technological elements:
{4}------------------------------------------------
- an intravaginal probe with external silicone shell
- force sensors and bluetooth transmitter embedded inside the probe
- a smartphone with a dedicated App
The minor technological differences between the Perifit and the predicate device do not raise different questions of safety or effectiveness. A table comparing the key features of the subject and predicate devices is provided below.
| | X6 Perifit | Analytica Ltd.<br>PeriCoach® OTC<br>(K160758) | Comments |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | | | |
| Indications for<br>Use | The Perifit is a<br>perineometer designed to<br>treat stress, mild-moderate<br>urge and mixed urinary<br>incontinence in women, by<br>strengthening of the pelvic<br>floor muscles through<br>exercise. This device<br>provides biofeedback via<br>smart phone technology. | The PeriCoach® OTC is a<br>perineometer designed to<br>treat stress, mild-moderate<br>urge and mixed urinary<br>incontinence in women, by<br>strengthening of the pelvic<br>floor muscles through<br>exercise. This device<br>provides biofeedback via<br>smart phone technology. | Identical |
| User Population | Adult females with urinary<br>incontinence; available<br>over-the-counter | Adult females with urinary<br>incontinence; available by<br>prescription and over-the-<br>counter | Identical |
| Technological Characteristics | | | |
| Mode of Use | Reusable for single patient | Reusable for single patient | Identical |
| Principle of<br>Operation | A probe inserted into the<br>vagina to determine the<br>strength of the pelvic floor<br>muscles. Probe sends<br>signals to external device<br>to indicate muscle<br>contraction strength to<br>encourage and assist user<br>with voluntary kegel<br>exercises. | A probe inserted into the<br>vagina to determine the<br>strength of the pelvic floor<br>muscles. Probe sends<br>signals to external device<br>to indicate muscle<br>contraction strength to<br>encourage and assist user<br>with voluntary kegel<br>exercises. | Identical |
| Sensing method | Output from force sensing<br>resistors (wireless). | Output from force sensing<br>resistors. | Identical |
| | X6 Perifit | Analytica Ltd.<br>PeriCoach® OTC<br>(K160758) | Comments |
| Sensor's<br>placement | Inside the rigid plastic<br>enclosure | On the exterior of the<br>plastic enclosure | |
| Materials | Rigid plastic (PC/ABS)<br>structure enclosed within a<br>medical grade silicone<br>outer layer | Rigid polymer structure<br>enclosed within a medical<br>grade silicone outer layer | Similar |
| Parameter<br>monitored | Analogue to digital output<br>of uncalibrated force<br>exerted against external<br>walls of device by<br>pubococcygeus and<br>puborectalis muscles. | Analogue to digital output<br>of uncalibrated force<br>exerted against external<br>walls of device by<br>pubococcygeus and<br>puborectalis muscles. | Identical |
| User Interface | Smartphone GUI | Smartphone GUI | Identical |
| Anatomical<br>Sites | Female Pubococcygeus<br>muscle area | Female Pubococcygeus<br>muscle area | Identical |
| External shape | Two egg shaped sensing<br>areas | One sensing area | |
| Accessories | None | LiPo charger | |
| Shaft length | 90 mm | 76 mm | |
| Weight | Probe weight: 54g | Probe weight: 50g | |
| Power Source | Non-rechargeable CR2032<br>Panasonic batteries,<br>Voltage 3.0VDC | Rechargeable Lithium-<br>polymer: Voltage 3.7 VDC | |
| Safety Features | Electronics and internal<br>parts sealed in a medical<br>grade silicone shell | Electronics and internal<br>parts sealed in a medical<br>grade silicone shell | Identical |
| Electrical Safety | Tested in accordance with<br>IEC 60601-1-2 and IEC<br>60601-1 | Tested in accordance with<br>IEC 60601-1-2 and IEC<br>60601-1 | Identical |
| Biocompatibility | Biocompatible - tested in<br>accordance with ISO10993<br>standards | Probe outer surface<br>constructed of chemically<br>inert materials and tested in<br>accordance with ISO10993 | Identical |
| Software | Smartphone app<br>compatible for iOS and<br>Android | Smartphone app | Identical |
| Sterilization | Non-sterile device | Non-sterile device | Identical |
# Table 1: Substantial Equivalence Table
{5}------------------------------------------------
{6}------------------------------------------------
# Performance Data
The patient contacting materials in the Perifit have been tested in accordance with ISO 10993 standards and was found to be safe for the intended purpose. Biocompatibility testing included Cytotoxicity (ISO 10993-5, 2009), Sensitization (ISO 10993-10, 2010), Vaginal Irritation (ISO 10993-10, 2010), and Systemic Toxicity (ISO 10993-11, 2006). Electrical safety and electromagnetic compatibility testing have been conducted in accordance with IEC 60601-1-2, and IEC 60601-1-11 to establish the safety of the device. Software verification and validation testing has also been conducted in accordance with FDA guidance (Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices) and IEC 62304:2006. In addition, various mechanical tests have been conducted which support substantial equivalence including mechanical drop testing, durability testing, and sensor behavior testing through a questionnaire supports that users understand the key labeling provisions and how to operate the device.
#### Conclusions
The Perifit is as safe and effective as the PeriCoach® OTC. The Perifit has the same intended use and similar indications for use, technological characteristics, and principles of operation as the predicate device. In addition, the minor technological differences between the Perifit and the predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Perifit is as safe and effective as the PeriCoach® OTC. Thus, the Perifit is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.