IPL hair removal

K221451 · Shenzhen Lizbella Technology Co.,Ltd · OHT · Aug 1, 2022 · General, Plastic Surgery

Device Facts

Record IDK221451
Device NameIPL hair removal
ApplicantShenzhen Lizbella Technology Co.,Ltd
Product CodeOHT · General, Plastic Surgery
Decision DateAug 1, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

IPL hair removal is an over-the-counter device intended for removal of unwanted body hair.

Device Story

Personal light-based hair reduction device; utilizes Intense Pulsed Light (IPL) via Xenon arc flashlamp; operates below skin surface to reduce hair growth. Device powered by external adapter; activation via finger switch. Includes integrated skin sensor to ensure full skin contact before pulse emission; prevents accidental discharge. Intended for over-the-counter use by end-users for body hair removal. Healthcare providers do not operate device; output is non-visual (light pulse). Benefit is reduction of unwanted hair with minimal pain. Models (FZ-610, FZ-609, FZ-612, FZ-611, FZ-613, FZ-608) vary by appearance and treatment window spot size.

Clinical Evidence

No clinical data provided. Evidence consists of bench testing, including biocompatibility (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57), and photobiological eye safety (IEC 62471). Software verification and validation were performed for a moderate level of concern.

Technological Characteristics

Materials: ABS plastic (ISO 10993 compliant). Energy source: External power adapter, Xenon arc flashlamp. Wavelength: 530-1100nm. Energy density: 2.1~5.3J/cm². Pulse duration: 8~10ms. Spot size: 3.4-4.1cm². Connectivity: None. Software: Moderate level of concern. Safety features: Skin contact sensor.

Indications for Use

Indicated for removal of unwanted body hair in over-the-counter home use settings.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 1, 2022 Shenzhen Lizbella Technology Co.,Ltd % Rain Yip Registration Engineer Feiying Drug & Medical Consulting Technical Service Group Contact Address Re: K221451 Trade/Device Name: IPL hair removal Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: May 9, 2022 Received: May 18, 2022 Dear Rain Yip: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221451 Device Name IPL hair removal Model(s): FZ-610, FZ-609, FZ-612, FZ-611, FZ-613, FZ-608 Indications for Use (Describe) IPL hair removal is an over-the-counter device intended for removal of unwanted body hair. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary ### "510(k) Summary" as required by 21 CFR Part 807.92. #### Date: 2022-05-09 #### I. Submitter Shenzhen Lizbella Technology Co.,Ltd No.1202, C Block, 3rd Building, Huaqiang Creative Industrial Park, Biyan Community, Guangming District, Shenzhen City, Guangdong Province, China Post code: 518000 Tel.: +86 0755 2319 4813 Shiwei Zhang Deputy General Manager Tel: +86 0755 2319 4813 Email: fasiz04@lizbella.cn #### II. Device Name of Device: IPL hair removal Model(s): FZ-610, FZ-609, FZ-612, FZ-611, FZ-613, FZ-608 Common or Usual Name: Light Based Over-The-Counter For Hair Removal Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: OHT Regulation Number: 21 CFR 878.4810 ### III. Predicate Device Primary predicate device: 510(k) number: K192432 Manufacturer: Shenzhen Bosidi Technology Co.,Ltd. Trade name: IPL Home Use Hair Removal Device, D-1103 Product code: OHT Approval date: 11/08/2019 Reference device: 510(k) number: K160968 Manufacturer: CyDen Limited. Trade name: iPulse SmoothSkin Gold Hair Removal System Product code: OHT, GEX Approval date: 14/04/2016 ### IV. Device Description {4}------------------------------------------------ The IPL hair removal is a personal, light-based, hair reduction device intended to be sold overthe-counter directly to the end user. The device provides hair reduction using Intense Pulsed Light (IPL) technology, and it works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The IPL hair removal is only powered by the external power adapter and its IPL emission activation is by finger switch. The device contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not in full contact with the skin, the device cannot emit the light pulse. The IPL hair removal includes FZ-610, FZ-612, FZ-611, FZ-613, and FZ-608 six models. Their intended use, performance, structure design and operation are basically identical, with the differences mainly contains product appearance and its spot size of the treatment window, but these differences do not affect or change the intended use of the device. # V. Indications for Use IPL hair removal is an over-the-counter device intended for removal of unwanted body hair. # VI. Comparison of Technological Characteristics With the Predicate Device The subject device IPL hair removal has the same intended use, mode of action and similar operational characteristics as the predicate device. Any minor differences between the subject device and the listed predicate device do no raise any issues of safety or efficacy. Performance data supports that the device is safe and as effective as the predicate device for its intended use. Therefore, the subject device may be found substantially equivalent to its predicate device. The subject device is compared with the following Predicate Device in terms of intended use, design, material, specifications, and performance: | Comparison<br>Elements | Subject Device | Primary predicate device<br>K192432 | |------------------------------------|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------| | K Number | Pending | K192432 | | Trade name | IPL hair removal/ FZ-610, FZ-609,<br>FZ-612, FZ-611, FZ-613, FZ-608 | IPL Home Use Hair Removal<br>Device/D-1103 | | Wavelength range | 530-1100nm | 510-1100nm | | Energy medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | | Energy density | 2.1~5.3J/cm² | 2.5~4.5J/cm² | | Spot size | 4.1cm², 3.4cm² | 3.0cm² | | Pulse duration | 8~10ms | 7.5~14ms | | Pulsing control | Finger switch | Finger switch | | Delivery device | Direct illumination to tissue | Direct illumination to tissue | | Indication for<br>use/Intended use | IPL hair removal is an over-the-<br>counter device intended for<br>removal of unwanted body hair. | IPL Home Use Hair Removal<br>Device is an over-the-counter<br>device intended for removal of<br>unwanted body and/or facial hair. | | Location for use | OTC | OTC | {5}------------------------------------------------ # VII.Performance Data The following performance data were provided in support of the substantial equivalence determination. # 1) Biocompatibility Safety The materials of the patient-directly contacting components of the subject device is ABS plastic and performed the biocompatibility evaluation in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document issued on Sep. 4, 2020", as recognized by FDA. The batterv of testing was performed to, and passed, including: - > ISO 10993-5 Biological Evaluation of Medical Devices -Par t 5: Tests for In Vitro Cytotoxicity - > ISO 10993-10 Biological Evaluation of Medical Devices –Par t 10: Tests for Irritation and Skin Sensitization # 2) Electrical Safety and Eye Safety Electrical safety and Eye safety testing was performed to, and passed, the following standards: - > IEC 60601-1-2 Medical electrical equipment -Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility - > IEC 60601-1 Medical electrical equipment -Part 1: General requirements for basic safety and essential performance - > IEC 60601-1-11 Medical Electrical Equipment -Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment - > IEC 60601-2-57 Medical electrical equipment -- Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use ### 3) Eve Safetv - > IEC 62471 Photobiological safety of lamps and lamp systems # 4) Software Verification and Validation Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. Summarv Based on the above performance as documented in this application, the subject device IPL hair removal was found to have a safety and effectiveness profile that is similar to the predicate device. #### VIII. Conclusions In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the subject device IPL hair removal is to be concluded substantial equivalent to its predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...