Intended for facial stimulation and indicated for over-the-counter cosmetic use.
Device Story
Hand-held device for facial stimulation; applies low-level electrical impulses to strategic facial locations. User operates device by gliding two fixed, chrome-plated spherical probes over skin using conductive gel. Powered by 3.7V rechargeable battery; features adjustable intensity settings via up/down buttons. Device produces monophasic, alternating polarity rectangular pulses. Intended for home use by consumers for cosmetic purposes. Output provides stimulation to facial muscles; user monitors status via green LED indicator. Automatic 10-minute shut-off included. Benefits include cosmetic facial stimulation.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included oscilloscope tracings, mechanical strength, drop testing, and biocompatibility (cytotoxicity, sensitization, irritation per ISO 10993-5/10). Electrical safety and EMC testing confirmed compliance with ANSI/AAMI ES60601-1, IEC 60601-1-11, IEC 60601-2-10, and IEC 60601-1-2.
Technological Characteristics
Hand-held stimulator; thermoplastic resin case; chrome-plated spherical probes. Powered by 3.7V battery. Output: Monophasic, alternating polarity rectangular pulses; 1.37-1.73kHz frequency; 4μs pulse width. Compliance: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11. Biocompatibility: ISO 10993-5, ISO 10993-10. Type BF protection. Microprocessor-controlled with automatic shut-off.
Indications for Use
Indicated for over-the-counter cosmetic facial stimulation.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
October 21, 2022
Xtreem Pulse LLC Andrew Barile CEO 353 W. 29 St. Suite 3 New York, New York 10001
Re: K221443
Trade/Device Name: PureLift Pro Plus Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NFO Dated: September 22, 2022 Received: September 23, 2022
Dear Andrew Barile:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Rehabilitation Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221443
Device Name PureLift Pro Plus
Indications for Use (Describe) Intended for facial stimulation and indicated for-over-the counter cosmetic use.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| I. SUBMITTER | |
|---------------------------|-----------------------------------------------------------|
| Applicant: | XTREEM PULSE LLC |
| | 353 W. 29 ST. SUITE 3 |
| | NEW YORK, NY 10001 |
| | UNITED STATES |
| Contact Persons: | Jacqueline Schmainda |
| | Regulatory Consultant |
| | P: 763.269.2069 |
| | E: jaschmainda@gmail.com |
| | Andrew Barile, CEO |
| | Xtreem Pulse LLC |
| | P: 917.597.5436 |
| | E: andrew@xtreempulse.com |
| Date Prepared: | October 20, 2022 |
| II. DEVICE | |
| Name of Device: | PureLift Pro Plus |
| Classification Name: | Stimulator, Transcutaneous Electrical, Aesthetic Purposes |
| Regulatory Class: | Class II |
| Product Code: | NFO |
| Regulation Number: | 21 CFR 882.5890 |
| III. PREDICATE DEVICE | |
| Primary Predicate Device: | PureLift |
| Manufacturer: | Xtreem Pulse LLC |
| 510(k) Number: | K190269 |
| Reference Device: | Rejuvenique Model (RJV-10) |
| Manufacturer: | Salton, Inc. |
| 510(k) Number: | K011935 |
# IV. DEVICE DESCRIPTION
The PureLift Pro Plus is a hand-held device intended to apply electrical impulses to strategic locations on the face. The PureLift Pro Plus probes are designed for optimal contact with the face. The device continually alternates between the positive and negative probes and allows the user to adjust the settings for personalized comfort level by pressing the up/down button. The intensity starts at (10).
The device measures 20.7cm (H) x 4.8cm (W) x 4.5cm (D). Its outer case is injection molded of thermoplastic resin and the probes consist of chrome-plated spheres. The device, powered by a 3.7-volt battery, produces low-level current that is transmitted through the two fixed, smooth spherical probes. To turn the device on, the power button is pushed. Then the green LED light will illuminate indicating the unit is ready for use. Users then follow the instructions for use. The two probes gently glide over the skin to deliver low-level electrical impulses to strategic locations on the face with the use of conductive gel.
The PureLift Pro Plus unit contains a power supply and rechargeable battery. The enclosure is made of medical grade biocompatible plastics and the output contacts (Probes) consist of chrome-plated spheres.
{4}------------------------------------------------
# V. INDICATIONS FOR USE
Intended for facial stimulation and indicated for over-the-counter cosmetic use.
| Elements of<br>Comparison | Subject Device<br>PureLift Pro Plus<br>K221443 | Predicate Device<br>PureLift<br>K190269 | Reference Device<br>Rejuvenique<br>K011935 |
|---------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Indications for Use | Intended for facial stimulation and indicated<br>for over-the-counter cosmetic use. | PureLift is intended for facial stimulation and<br>indicated for over-the-counter cosmetic use. | Rejuvenique System is<br>Indicated for cosmetic use. |
| Dimensions (HxWxD) | 20.7cm x 4.8cm x 4.5cm | 13.4in x 4.8in x 4.3in | 3in x 4½in x 1½in |
| Power Source | One 3.7V Battery | One 3.7V Battery | One 9V Battery |
| Number of output<br>modes | 2 | 1 | 1 |
| Number of output<br>channels | 1 output channel | 1 output channel | 1 output channel |
| Regulated current or<br>regulated voltage? | Regulated current | Regulated current | Regulated Voltage |
| Software/ Firmware/<br>Microprocessor<br>Control? | Yes | Yes | Yes |
| Automatic Shut off? | Yes | Yes | Yes |
| Patient override<br>control? | No | No | Yes |
| Indicator Display | Yes | Yes | Yes |
| Timer range | 10 minutes only | 10 minutes only | 16 min fixed |
| Type of protection | Type BF | Type BF | Not publicly available |
| On/off status | Yes | Yes | Yes |
| Standards Compliance | ANSI/AAMI ES60601-1 IEC<br>60601-1-2<br>IEC 60601-2-10<br>IEC 60601-1-11 | IEC 60601-1 IEC 60601-1-2<br>IEC 60601-2-10<br>IEC 60601-1-11 | Not publicly available |
| Biocompatibility | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | Not publicly available |
| Waveform | Pulses Monophasic,<br>alternating polarity | Pulses Monophasic,<br>alternating polarity | Pulsed biphasic |
| Shape | Rectangular Pulses | Rectangular Pulses | Rectangular (+phase)<br>Spike (-phase) |
| Maximum output<br>voltage | 20Vpp(@500Ω)<br>32Vpp(@2kΩ)<br>44Vpp(@10kΩ) | 15Vpp(@500Ω)<br>23Vpp(@2kΩ)<br>34Vpp(@10kΩ) | 18.8V @ 500Ω<br>24.8V @ 2kΩ<br>28.0V @ 10kΩ |
| Maximum output<br>current | 9mA(@500Ω)<br>4.4mA(@2kΩ)<br>1.2mA(@10kΩ) | 7mA(@500Ω)<br>3.2mA(@2kΩ)<br>0.9mA(@10kΩ) | 37.6mA @ 500Ω<br>12.4mA @ 2kΩ<br>2.8mA @ 10kΩ |
| Output tolerance | +/- 1mA | +/- 1mA | +/- 10% |
| Pulse Width | 4μs | 4μs | 300μs |
| Frequency (Hz) | 1.37kHz~1.73kHz | 1.37kHz~1.73kHz | 8Hz fixed |
| Symmetrical phases | Not multiphasic | Not multiphasic | No |
| Phase duration | 4μs | 4μs | 300µs (+phase)<br>124.7ms (-phase,<br>exponential) |
| Net Charge (µC per<br>pulse train) | 0µC per pulse train | 0µC per pulse train | 0μC @500Ω |
| Elements of<br>Comparison | Subject Device | Predicate Device | Reference Device |
| | PureLift Pro Plus | PureLift | Rejuvenique |
| | K221443 | K190269 | K011935 |
| Maximum Phase<br>Charge (μC) | 5.81μC@500Ω | 4.52μC@500Ω | 11.3μC @500Ω |
| Maximum current<br>Density (mA/cm²) | 8.8mA/cm²@500Ω | 6.8mA/cm²@500Ω | 46.4mA/cm² @500Ω |
| Maximum Power<br>Density | 39600μW/cm² | 23800μW/cm² | 2.31 mW/cm² |
| Pulse per burst | 30 pulses | 30 pulses | 160 pulses |
| Bursts per second | 2740 ~ 3460 | 2740 ~ 3460 | 1/240 (per electrode group) |
| Burst duration | 230µs | 230µs | 20 seconds |
| Duty cycle | 0.63 ~ 0.80 | 0.63 ~ 0.80 | 1/12 |
| ON Time (seconds) | Constant | Constant | 20 seconds/electrode group |
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
{5}------------------------------------------------
The following technological differences exist between the subject and predicate devices:
- Dimensions: As compared to the predicate PureLift device, the PureLift Pro Plus device has a increased height and depth; however, the shape and dimensions of the two fixed, smooth spherical probes located at the distal end of the device remain unchanged. These spheres represent the active part of the device that provides facial stimulation. The device's overall dimensional differences do not alter the manner in which the device is intended to be used nor does it introduce a usability issue. The device's hourglass shape is intended to aid in holding the device as opposed to the predicate PureLift device which was designed with a more oval shape. As such, the dimensional modification does not adversely impact the safety or effectiveness of the device.
- . Maximum output voltage/ current: The maximum output voltage and maximum output current values for the PureLift Pro Plus device are slightly higher than the predicate PureLift device; however, the energy per second, which is calculated by the output voltage and current for the PureLift Pro Plus device is much lower than the reference device as well as the value specified by the IEC standards. Additionally, the energy for total 10 minutes for the PureLift is slightly higher than the PureLift predicate device but much lower than the reference device. The difference of the maximum output voltage and maximum output current between the PureLift Pro Plus and both the predicate and reference devices does not adversely impact the safety or effectiveness of the device.
- Maximum phase charge (μC): Maximum phase charge is slightly different between the PureLift Pro Plus device and the predicate PureLift device; however, the value is much lower than the reference device indicating substantial equivalence. The difference in value does not increase the risk of skin burn or other injury nor does it raise new questions safety or effectiveness.
- Maximum current density (mA/cm²): The maximum current density of the PureLift Pro Plus device is much lower than the reference device and the difference does not adversely influence safety or effectiveness.
- . Maximum power density (μW/cm²): The maximum power density of the PureLift Pro Plus device is slightly larger than the PureLift predicate and reference devices. However, this value is slightly lower than 0.25W/cm² which is described as a threshold for reducing the risk of burn in FDA's guidance document for powered muscle stimulators. As such, there is no increased risk of injury or new questions of safety or effectiveness.
{6}------------------------------------------------
#### VII. PERFORMANCE DATA
Oscilloscope tracings, vibration, temperature, push, mold stress, markings, mechanical strength, drop, ball, acoustic, and accessible parts testing was conducted. The test results confirmed that the device met established design specifications.
#### Biocompatibility testing
Biocompatibility evaluations for the subject device were conducted in accordance with ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA, and included:
- Cytotoxicity
- Sensitization
- Irritation
The subject device is considered tissue contacting for a duration of less than 24 hours. The device passed each test, and was found not be cytotoxic, sensitizing, orirritating.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the subject device. It was found to comply with the ANSI/AAMI ES60601-1, IEC 60601-1-11 and IEC 60601-2-10 standards for safety and the IEC 60601-1-2 standard for EMC.
## Software Verification and Validation Testing
Software verification and validation testing was conducted and confirmed that the device met established design specifications and conforms to the user needs. The software was assigned a "moderate" level of concern.
#### VIII. CONCLUSIONS
The subject device is identical to the predicate in terms of intended use. The technological differences between the subject and predicate device were addressed using performance data and confirmed that differences did not raise new questions of safety or effectiveness. Thus, it is concluded that the PureLift Pro Plus is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.