K221407 · Cutera, Inc. · GEI · Jul 13, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221407
Device Name
truSculpt iD
Applicant
Cutera, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The truSculpt iD device is intended to generate heat within body tissues for the treatment of selected medical conditions, such as the relief of minor aches, pain, and muscle spasms; an increase in local circulation; a reduction in circumference of the abdomen; and non-invasive lipolysis (breakdown of fat) of the abdomen. The truGlide roller is intended to provide temporary reduction in the appearance of cellulite.
Device Story
System consists of console, RF handpieces, and truGlide massage roller; console houses RF generator, control electronics, and touchscreen interface. Handpieces deliver 1MHz or 2MHz RF energy to subcutaneous tissue; system monitors epidermal temperature to maintain therapeutic range (40-45°C). Used in clinical settings by trained operators. Output is controlled heating for pain relief, circulation, abdominal circumference reduction, and lipolysis; massage roller provides temporary cellulite appearance reduction. Adhesive decals and silicone belts secure hands-free handpieces to patient. Real-time temperature monitoring ensures safety by preventing skin temperatures from exceeding 45°C; clinical decision-making supported by controlled thermal delivery.
Clinical Evidence
No clinical data. Bench testing only: includes electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), and high-frequency surgical equipment safety (IEC 60601-2-2). Biocompatibility testing per ISO 10993-1. Skin temperature testing on 3 human subjects confirmed surface temperatures remain within 40-45°C and do not exceed 45°C.
Technological Characteristics
Console-based RF generator; 1 MHz and 2 MHz output. Materials: PVDF (handpieces), liquid resin adhesive, silicone (cummerbund), hydrogel (electrodes), stainless steel (roller). Biocompatibility per ISO 10993-1. Active temperature sensing and control. Touchscreen UI with microprocessor control. Sterilization not specified. Connectivity: Standalone console.
Indications for Use
Indicated for topical heating to elevate tissue temperature for relief of pain, muscle spasms, and increased local circulation. The 2MHz setting for the 40cm2 handpiece is indicated for reduction in circumference of the abdomen and non-invasive lipolysis of the abdomen. The massage device is indicated for temporary reduction in the appearance of cellulite.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
truSculpt RF Device; truSculpt; truSculpt 3D (K180709)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
July 13, 2022
Cutera Inc. Mr. Amogh Kothare, M.S. VP Clinical and Regulatory Affairs 3240 Bayshore Blvd. Brisbane, California 94005
Re: K221407
Trade/Device Name: truSculpt iD Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, PBX Dated: May 31, 2022 Received: June 2, 2022
Dear Mr. Kothare:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221407
Device Name truSculpt iD
#### Indications for Use (Describe)
The trusculpt RF energy is intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions, such as relief of pain and muscle spasms and increase in local circulation.
Additionally, the 2MHz setting for the 40cm2 handpiece is indicated for reduction in circumference of the abdomen and non-invasive lipolysis (breakdown of fat) of the abdomen.
The trusculpt massage device is intended to provide a temporary reduction in the appearance of cellulite.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
This 510(k) Summary of safety and effectiveness for the truSculpt RF device is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(k) Summary.
| Applicant: | Cutera, Inc. |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 3240 Bayshore Blvd<br>Brisbane, CA 94005 |
| Contact Person: | Mr. Amogh Kothare, MS<br>VP Clinical and Regulatory Affairs<br>3240 Bayshore Blvd<br>Brisbane, CA 94005<br>akothare@cutera.com<br>m. (415) 690-6796<br>o. (415) 657-5730<br>f. (415) 715-3730 |
| Preparation Date: | July 12, 2022 |
| Device Trade Name: | truSculpt iD |
| Common Name: | Massager, Vacuum, Radio Frequency Induced Heat |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories<br>GEI, PBX, 21 CFR 878.4400 |
| Primary Predicate<br>Device: | truSculpt RF Device (K172004) |
| Secondary<br>Predicate Device: | truSculpt RF Device; truSculpt; truSculpt 3D (K180709) |
| Intended Use: | The truSculpt iD device is intended to generate heat within body<br>tissues for the treatment of selected medical conditions, such as the<br>relief of minor aches, pain, and muscle spasms; an increase in local<br>circulation; a reduction in circumference of the abdomen; and non-<br>invasive lipolysis (breakdown of fat) of the abdomen. The truGlide<br>roller is intended to provide temporary reduction in the appearance of<br>cellulite. |
| Indications for Use: | The truSculpt RF energy is intended to provide topical heating for the<br>purpose of elevating tissue temperature for the treatment of selected<br>medical conditions, such as relief of pain and muscle spasms and<br>increase in local circulation.<br><br>Additionally, the 2 MHz setting for the 40 cm² handpiece is indicated<br>for reduction in circumference of the abdomen and non-invasive<br>lipolysis (breakdown of fat) of the abdomen.<br><br>The truSculpt massage device is intended to provide a temporary<br>reduction in the appearance of cellulite. |
| Device Description: | The truSculpt iD device consists of a console, one or more RF handpieces that connect to the console with umbilical cables, and a truGlide massage roller. All system functions are controlled through the console. The handpieces deliver RF energy to generate a heating profile that produces a moderate temperature rise in the subcutaneous tissue, while monitoring epidermal temperature. The truGlide is a separate mechanical roller that can be used as a massager. |
| Summary of<br>Technological<br>Characteristics: | The truSculpt iD device has similar technological characteristics as the primary and secondary predicate devices. The truSculpt RF Device, truSculpt, truSculpt 3D, and truSculpt iD are all comprised of a console and RF applicators. The consoles for all devices consist of a mechanical enclosure, an RF generator, control electronics, a touchscreen user interface, and a control microprocessor.<br><br>The shape and dimensions of the truSculpt iD RF handpieces are identical to those of the primary predicate device. The handpieces are now held in place by adhesive decals and a silicone belt, which are manufactured from biocompatible materials.<br><br>The truSculpt iD device heats tissue through delivery of RF energy at 1 MHz and 2 MHz, which is identical to the primary and secondary predicate devices. |
| Performance Data: | IEC 60601-1/A1:2012 Medical Electrical Equipment – Part 1: General Requirements for Safety IEC 60601-2-2:2009 Medical Electrical Equipment – Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories IEC 60601-1-6:2010/A1:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests Software Verification and Validation Testing Biocompatibility testing of patient-contact materials according to ISO-10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process Skin temperature testing on 3 human subjects to demonstrate the skin temperatures do not exceed 45°C during the course of the treatment |
| Clinical Data: | None |
{4}------------------------------------------------
{5}------------------------------------------------
Technical Comparison to Predicate Devices:
| | truSculpt iD Device<br>(current submission) | Primary Predicate Device<br>Cutera truSculpt RF Device<br>(K172004) | Secondary Predicate Device<br>Cutera truSculpt RF Device<br>(K180709) |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Energy type | Radiofrequency | Radiofrequency | Radiofrequency |
| Massage | Yes | Yes | Yes |
| Temperature sensing | Yes | Yes | Yes |
| Temperature sensing<br>active control | Yes | Yes | Yes |
| Treatment activation | Control screen button or fingerswitch | Control screen button or fingerswitch | Fingerswitch |
| Area treated | Handheld: 16 and 40 cm²<br>Hands-free: 40 to 240 cm² | Handheld: 16 and 40 cm²<br>Hands-free: 40 to 240 cm² | Handheld: 16 and 40 cm² |
| Electrode shape | Square or Rectangle | Square or Rectangle | Square or Rectangle |
| RF frequency | 1 MHz and 2 MHz | 1 MHz and 2 MHz | 1 MHz and 2 MHz |
| RF type | Bipolar/Monopolar | Bipolar/Monopolar | Bipolar/Monopolar |
| Max RF power | 300 W | 300 W | 300 W |
| Patient contact<br>material | · Polyvinylidene fluoride (PVDF) for<br>handheld handpieces<br>· Liquid resin adhesive for patient<br>decals (for hands-free handpieces)<br>· Silicone for cummerbund (for hands-<br>free handpieces)<br>• Hydrogel for neutral electrode pads<br>· Stainless steel for massage roller | · Polyethylene (3M Tegaderm) for<br>handheld and hands-free<br>handpieces<br>• Hydrogel for neutral electrode pads<br>· Stainless steel for massage roller | · Polyethylene (3M Tegaderm) for<br>handheld handpieces<br>· Hydrogel for neutral electrode pads<br>· Stainless steel for massage roller |
| Attachment of hands-<br>free handpieces to<br>patient | Adhesive patient decals and<br>cummerbund | Cloth belt | Not Applicable |
{6}------------------------------------------------
The truSculpt iD device has equivalent technological characteristics and Conclusion: principles of operation as the primary and secondary predicate devices.
> Performance data demonstrates that any differences between the truSculpt iD and its predicate devices do not raise new safety or effectiveness questions. Skin temperature testing on three subjects further demonstrates that the truSculpt iD device can maintain therapeutic temperatures of 40°-45°C on the surface of the human skin and that maximum skin temperatures do not exceed 45°C.
The truSculpt iD device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.