K221318 · Vapotherm, Inc. · QAV · Dec 13, 2022 · Anesthesiology
Device Facts
Record ID
K221318
Device Name
HVT 2.0
Applicant
Vapotherm, Inc.
Product Code
QAV · Anesthesiology
Decision Date
Dec 13, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5454
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The HVT 2.0 system provides high velocity nasal insufflation (HVNI) with simultaneous warmed and humidified respiratory gas delivery to augment breathing of spontaneously breathing adult and pediatric patients (5 kg and up) suffering from respiratory distress and/or hypoxemia in the hospital setting, via a small bore nasal cannula. HVT 2.0 is not intended to provide total ventilatory requirements of the patient and not for use during field transport. The flow rates may be from 5 to 45 liters per minute (BTPS).
Device Story
HVT 2.0 delivers high velocity nasal insufflation (HVNI) to spontaneously breathing patients in hospital settings. System inputs include room air (via internal blower) and supplemental oxygen (wall/tank); device blends gases and monitors flow/FiO2. Gas is warmed and humidified via semi-permeable polymer tubing before delivery through small-bore nasal cannula. Operated by clinicians; device provides continuous therapy to augment breathing. Integrated sensors and electronic blender maintain set flow (5-45 L/min) and FiO2 (21-100%). Output includes real-time monitoring of therapy parameters and extensive audible/visual alarms for safety. Benefits include CO2 flush from extrathoracic dead space and respiratory support without total ventilation. Device includes internal battery for backup power.
Clinical Evidence
Bench testing only. Performance testing included oxygen accuracy, blender performance, temperature accuracy/stability, flow rate accuracy, humidification output (ISO 80601-2-74), thermal safety, and usability/human factors. Biocompatibility testing followed ISO 10993-1. Electrical safety and EMC testing conducted per IEC 60601-1 and IEC 60601-1-2. Alarms tested per IEC 60601-1-8.
Technological Characteristics
System uses internal blower for air and electronic blender for oxygen. Humidification via semi-permeable polymer tubing. Flow range 5-45 L/min (BTPS); FiO2 0.21-1.00. Powered by 110-240VAC with Li-Ion battery backup. Dimensions/form factor: hardware unit with disposable patient circuit. Connectivity: nurse call compatibility. Software-controlled. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, and ISO 80601-2-74.
Indications for Use
Indicated for spontaneously breathing adult and pediatric patients (≥5 kg) in hospital settings suffering from respiratory distress and/or hypoxemia. Contraindicated for patients not spontaneously breathing, unable to protect airway, or with nasal pathway blockage; not for OSA/snoring, field transport, MRI environments, or use with oxygen concentrators.
Regulatory Classification
Identification
A high flow humidified oxygen delivery device is a prescription device that delivers high flow oxygen with humidification for patients who are suffering from respiratory distress and/or hypoxemia.
Special Controls
In combination with the general controls of the FD&C Act, the high flow humidified oxygen delivery device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions for use, including the following:
(i) Alarm testing must be performed;
(ii) Continuous use thermal stability testing must be performed;
(iii) Humidity output testing must be performed; and
(iv) Blender performance testing must evaluate fraction of inspired oxygen (
*Fi* O2 ) blending accuracy.(3) Performance data must validate cleaning instructions for any reusable components of the device.
(4) Electrical safety, thermal safety, mechanical safety, electromagnetic compatibility, and radiofrequency identification testing must be performed.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Labeling must include:
(i) A description of available
*Fi* O2 ranges for different flowrates and inlet gas pressures;(ii) Instructions for applicable flowrates for all intended populations;
(iii) A warning that patients on high flow oxygen are acute and require appropriate monitoring, to include pulse oximetry;
(iv) A warning regarding the risk of condensation at low set temperatures and certain flows; and
(v) A description of all alarms and their functions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" being larger and bolder than the rest of the text.
December 13, 2022
Vapotherm Inc. Michael Webb VP of Regulatory Affairs 100 Domain Drive Exeter, New Hampshire 03833
Re: K221318
Trade/Device Name: HVT 2.0 Regulation Number: 21 CFR 868.5454 Regulation Name: High Flow Humidified Oxygen Delivery Device Regulatory Class: Class II Product Code: QAV Dated: November 10, 2022 Received: November 14, 2022
Dear Michael Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ethan L. Nyberg -S
for James Lee, Ph.D. Division Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221318
Device Name HVT 2.0
#### Indications for Use (Describe)
The HVT 2.0 system provides high velocity nasal insufflation (HVNI) with simultaneous warmed and humidified respiratory gas delivery to augment breathing of spontaneously breathing adult and pediatric patients (5 kg and up) suffering from respiratory distress and/or hypoxemia in the hospital setting, via a small bore nasal cannula. HVT 2.0 is not intended to provide total ventilatory requirements of the patient and not for use during field transport. The flow rates may be from 5 to 45 liters per minute (BTPS).
Type of Use (Select one or both, as applicable)
| <span> <b> <span style="font-size:16px;">☑</span> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|---------------------------------------------------------------------------------------------------------------------------|
| <span> <b> <span style="font-size:16px;">☐</span> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
| Date: | 12-Dec-2022 |
|----------------------------|-------------------------------------------------------------|
| Company: | Vapotherm, Inc.<br>100 Domain Drive<br>Exeter, NH 03833 USA |
| Official Contact: | Michael J. Webb<br>Tel: 866-410-9986 |
| Proprietary or Trade Name: | HVT 2.0 |
| Common/Usual Name: | High Flow Humidified Oxygen Delivery Device |
| Classification: | 21 CFR 868.5454, Class II<br>Product Code: QAV |
| Predicate Device: | DEN170001 Vapotherm Precision Flow HVNI |
| Reference Device: | K203357 Vapotherm HVT 2.0 |
#### Device Description:
The HVT 2.0 System is comprised of hardware containing software, and single use disposables that are designed to deliver the same High Velocity Nasal Insufflation (HVNI) as the Vapotherm Precision Flow HVNI.
The HVT 2.0 system consists of the HVT 2.0 hardware containing software, a Disposable Patient Circuit (DPC), and a patient interface (nasal cannula). The HVT 2.0 system utilizes an integrated internal blower to deliver warmed and humidified breathing gas at flows 5 to 45 L/min (BTPS) to spontaneously breathing patients, without the need of wall air or any pressurized air source. The addition of an external oxygen source (wall or tank for QAV indication) enables FiO2 delivery from 21% to 100%. The device incorporates a blender and flow sensors that allow the oxygen percentage and total gas flow to be set independently.
The HVT 2.0 main hardware device contains all the electrical and electronic components including the electronic blender and flow controllers, and sensors to monitor the disposable patient circuit. The main hardware device has no water pathways and the gas pathway only contains dry gas at room temperature, and therefore does not require internal cleaning or disinfection.
{4}------------------------------------------------
### Indications for Use:
The HVT 2.0 system provides high velocity nasal insufflation (HVNI) with simultaneous warmed and humidified respiratory gas delivery to augment breathing of spontaneously breathing adult and pediatric patients (5 kg and up) suffering from respiratory distress and/or hypoxemia in the hospital setting, via a small bore nasal cannula. HVT 2.0 is not intended to provide total ventilatory requirements of the patient and not for use during field transport. The flow rates may be from 5 to 45 liters per minute (BTPS).
## Patient Population:
#### For pediatric and adult patients
Note: Vapotherm follows the definition of pediatric patients and pediatric subpopulations discussed in Guidance for Industry and Food and Drug Administration Staff "Providing Information about Pediatric Uses of Medical Devices".
#### Environment of Use:
Hospital settings
### Contraindications:
- Not appropriate for patients who are not spontaneously breathing, are unable ● to protect their airway, or have anatomic or injury induced blockage of the nasal pathway to the nasopharyngeal space.
- Not for treating OSA and snoring.
- HVT 2.0 is not for field transport.
- HVT 2.0 is MRI unsafe. Do not use it in an MR environment. ●
- Not for use with an Oxygen Concentrator .
#### Substantial Equivalence:
The technological characteristics that define HVNI and achieve the flush of expired CO2 from the accessible extrathoracic anatomic dead space are equivalent for HVT 2.0 and the predicate Precision Flow HVNI DEN170001. Therefore HVT 2.0 possesses the same clinical utility as Precision Flow HVNI (Product Code QAV). Bench testing has further demonstrated substantial equivalence of the devices.
{5}------------------------------------------------
| | Predicate: | Subject Device: |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Precision Flow HVNI<br>DEN170001 | HVT 2.0 |
| Indications for<br>Use | Precision Flow® HVNI is intended for<br>use to add warm moisture to breathing<br>gases from an external source for<br>administration to a neonate/infant,<br>pediatric and adult patients in the<br>hospital and subacute institutions<br>settings. It adds heat and moisture to a<br>blended medical air/ oxygen mixture<br>and assures the integrity of the precise<br>air/oxygen mixture via an integral<br>oxygen analyzer. The flow rates may<br>be from 1 to 40 liters per minute via<br>nasal cannula.<br>Precision Flow® HVNI provides high<br>velocity nasal insufflation (HVNI)<br>with simultaneous oxygen delivery to<br>augment breathing of spontaneously<br>breathing patients suffering from<br>respiratory distress and/or hypoxemia<br>in the hospital setting. Precision<br>Flow® HVNI is not intended to<br>provide total ventilatory requirements<br>of the patient and not for use during<br>field transport. | SIMILAR<br>QAV Indications:<br>The HVT 2.0 system provides high<br>velocity nasal insufflation (HVNI) with<br>simultaneous warmed and humidified<br>respiratory gas delivery to augment<br>breathing of spontaneously breathing adult<br>and pediatric patients (5 kg and up)<br>suffering from respiratory distress and/or<br>hypoxemia in the hospital setting, via a<br>small bore nasal cannula. HVT 2.0 is not<br>intended to provide total ventilatory<br>requirements of the patient and not for use<br>during field transport. The flow rates may<br>be from 5 to 45 liters per minute (BTPS). |
| Patient<br>Population | Pediatric, adult | DIFFERENT<br>Pediatric (>5kg), adult |
| Environment of<br>Use for QAV | Hospital setting | SAME<br>Hospital setting |
| Duration of<br>Use | Disposable can be used for up to 30<br>days single patient use. | SAME<br>Disposable can be used for up to 30 days<br>single patient use. |
| Prescriptive | Rx Only | SAME<br>Rx Only |
| Patient<br>Interface for<br>QAV | Small bore nasal cannula | SAME<br>Small bore nasal cannula |
| Flow Range | 1 to 40 L/Min (SLPM), converted as:<br>1.1 to 45 L/Min (BTPS) | DIFFERENT<br>5-45 L/Min (BTPS) |
| Patient<br>contacting<br>materials | Externally communicating, tissue,<br>permanent duration | SAME<br>Externally communicating, tissue,<br>permanent duration |
| Temperature<br>Range | 33°C to 39°C | SAME<br>33°C to 39°C |
| Temperature<br>Alarm | Alarms if 2°C above set point. | SAME<br>Alarms if 2°C above set point. |
| Temperature | ±2°C | SAME<br>±2°C |
| Characteristic | Predicate:<br>Precision Flow HVNI<br>DEN170001 | Subject Device:<br>HVT 2.0 |
| FiO2 Range | 0.21 to 1.00 | SAME<br>0.21 to 1.00 |
| Oxygen<br>Accuracy | ±2% Wall/tank | SAME<br>±2% Wall/tank |
| Operating<br>Principle | Air flow via a compressor and<br>humidification via semi-permeable<br>polymer technology which allows for<br>delivery of entrained, humidified<br>gases at constant flow to the patient | DIFFERENT<br>Air flow via a blower and<br>humidification via semi-permeable<br>polymer technology which allows for<br>delivery of entrained, humidified gases at<br>constant flow to the patient |
| Alarms | Extensive audible and visual alarms to<br>ensure essential performance of the<br>device is maintained. | SAME<br>Extensive audible and visual alarms to<br>ensure essential performance of the device<br>is maintained. |
| User interface<br>settings | User sets flow rate, temperature, and<br>oxygen fraction F;O2 | SAME<br>User sets flow rate, temperature, and<br>oxygen fraction F;O2 |
| Control | Software control | SAME<br>Software control |
| Modes of<br>operation | Sleep, Standby and Run Modes | SAME<br>Sleep, Standby and Run Modes |
| Gas/Air input | Standard DISS non-interchangeable<br>fitting for oxygen, room air via wall<br>source or compressor | DIFFERENT<br>Standard DISS non-interchangeable fitting<br>for oxygen, room air via blower |
| Humidification | Via semi-permeable small-bore tubing | SAME<br>Via semi-permeable small-bore tubing |
| Power<br>Requirements | 100-240 VAC, 50-60Hz<br>Back-up power: 4.8V nickel-metal<br>hydride battery pack<br><br>Back-up power with Vapotherm<br>Transfer Unit (VTU): Medipower™<br>unit, one hour limited use | DIFFERENT<br>110 to 240VAC, 50 to 60Hz<br>Back-up power (Safety Battery): Li-Ion<br>Battery 14.4V, 2900mAh, 99.Wh.<br><br>Back-up power with Transfer Battery:<br>Li-Ion Battery Pack consisting of 2X<br>(14.4V, 2900mAh, 99.4Wh), one hour<br>limited use |
{6}------------------------------------------------
From the comparison above, the subject device and predicate device have the same intended use, are both prescription use, and have the same operating principle, gas input and method of humidification. Any differences do not raise different questions of safety or effectiveness.
## Non-clinical performance testing Biocompatibility / Materials:
Biocompatibility testing is leveraged from the prior testing performed as part of K203357, whereby the testing was conducted in accordance with FDA guidance, Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". Testing of the patient-contacting parts of the HVT 2.0 System demonstrates an appropriate biocompatibility profile for the device.
{7}------------------------------------------------
## Electrical Safetv, Alarms, RFID:
Electrical safety and electromagnetic compatibility testing were conducted in accordance with IEC 60601-1:2012 and IEC 60601-1-2: 2014 Ed.4 to demonstrate the basic safety, essential performance and emissions and immunity characteristics of the device. The testing demonstrated the appropriate electrical safety and electromagnetic compatibility profile for the device. Alarms testing was performed compliant with IEC 60601-1-8:2006, ed 2.1. RFID testing was successfully completed in accordance with AIM 7351731.
### Software Verification, Validation Testing, and Hazard Analysis
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device is considered as a "moderate" level of concern. Hazard Analysis/Risk Management activities for the device comply with ISO 14971:2019 Medical Devices – Application of risk management to medical devices.
### Bench / Performance Testing -
Comparative performance testing was performed to demonstrate substantial equivalence and to meet the Special Controls requirements of 21 CFR 868.5454 High flow humidified oxygen delivery device, and included:
- . Precision Flow and HVT 2.0 Comparison Testing to demonstrate the same High Velocity Nasal Insufflation (HVNI) technological characteristics and performance
- Oxygen accuracy/Blender Performance ●
- Temperature accuracy/Thermal Stability ●
- Flow rate accuracy
- . Humidification output ISO 80601-2-74
- Patient contacting surface temperature/Thermal Safety
- Volume of liquid exiting the humidification chamber outlet
- Continuous use (Use Lifespan)
- Operating environment
- . Nurse call compatibility
- Usability/Human Factors ●
- Cleaning Validation (for hardware) .
#### Substantial Equivalence Conclusion
HVT 2.0 is substantially equivalent to the predicate device based on the intended use and comparison of the technological characteristics and performance testing.
Any differences in technological characteristics of the devices do not raise different questions of safety and effectiveness based on the performance testing, further supported by the prior clearance of the Reference Device HVT 2.0, cleared under K203357. The devices are substantially equivalent.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.