← Product Code [JOY](/productcode/JOY) · K221309

# AI100 with Shonit (K221309)

_Sigtuple Technologies Pvt. , Ltd. · JOY · Sep 19, 2023 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221309

## Device Facts

- **Applicant:** Sigtuple Technologies Pvt. , Ltd.
- **Product Code:** [JOY](/productcode/JOY.md)
- **Decision Date:** Sep 19, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5260
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** AI/ML

## Indications for Use

AI100 with Shonit™ is a cell locating device intended for in-vitro diagnostic use in clinical laboratories. AI100 with Shonit™ is intended for differential count of White Blood Cells (WBC), characterization of Red Blood Cells (RBC) morphology and Platelet morphology. It automatically locates blood cells on peripheral blood smears and presents images of the blood cells for review. A skilled operator, trained in the use of the device and in the review of blood cells, identifies and classifies each cell according to its characterizations.

## Device Story

AI100 with Shonit is an automated cell-locating device for clinical laboratories. It uses a high-resolution microscope with LED illumination and a motorized XYZ stage to scan Romanowsky-stained peripheral blood smear slides. The device identifies optimal smear areas and captures field-of-view (FOV) images in an outward spiral pattern. An embedded convolutional neural network processes these images to detect and pre-classify WBCs, RBCs, and platelets. The system presents these images to a trained operator for review, verification, and potential re-classification. The operator confirms the final report. This workflow assists in hematological analysis by automating the labor-intensive cell-locating process, potentially improving efficiency and consistency in clinical decision-making regarding blood cell morphology and differential counts.

## Clinical Evidence

Performance validated via method comparison study at four sites with 882 samples (healthy and diseased). Compared AI100 with Shonit to manual microscopy. WBC differential analysis used Passing-Bablok regression (Neutrophil Pearson's r=0.962). WBC morphological abnormality detection showed 91.7% overall agreement (95.3% sensitivity, 90.9% specificity). RBC morphology and platelet morphology assessments also met pre-defined acceptance criteria. Precision studies (repeatability and reproducibility) conducted across multiple sites, runs, and devices confirmed consistent performance across cell types.

## Technological Characteristics

High-resolution microscope with LED illumination, motorized XYZ stage, and camera. Compute components include motherboard, CPU, RAM, SSD, and Wi-Fi. Uses Romanowsky stains. Software utilizes a convolutional neural network for image processing. Operates at 40X magnification. Connectivity includes local storage; no LIS interaction. Safety/EMC standards: IEC 61010-1, IEC 61326-1, IEC 61326-2-6. Risk management per ISO 14971. Usability per IEC 62366-1.

## Regulatory Identification

An automated cell-locating device is a device used to locate blood cells on a peripheral blood smear, allowing the operator to identify and classify each cell according to type. (Peripheral blood is blood circulating in one of the body's extremities, such as the arm.)

## Predicate Devices

- CellaVision® DC-1 ([K200595](/device/K200595.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

INSTRUMENT ONLY

## I Background Information:

A 510(k) Number

K221309

B Applicant

SigTuple Technologies Pvt. Ltd.

C Proprietary and Established Names

AI100 with Shonit

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  JOY | Class II | 21 CFR 864.5260 - Automated Cell-Locating Device | HE - Hematology  |

## II Submission/Device Overview:

A Purpose for Submission:

Clearance of new device

B Type of Test:

White blood cell (WBC) differential, red blood cell (RBC) and platelet morphology evaluation

## III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

K221309 - Page 1 of 12

{1}

B Indication(s) for Use:

AI100 with Shonit is a cell locating device intended for in-vitro diagnostic use in clinical laboratories.

AI100 with Shonit is intended for differential count of White Blood Cells (WBC), characterization of Red Blood Cells (RBC) morphology and Platelet morphology. It automatically locates blood cells on peripheral blood smears and presents images of the blood cells for review.

A skilled operator, trained in the use of the device and in the review of blood cells, identifies and classifies each cell according to type.

C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

IV Device/System Characteristics:

A Device Description:

The AI100 with Shonit device consists of a high-resolution microscope with LED illumination, and compute parts such as the motherboard, CPU, RAM, Wi-Fi dongle, solid state drive (SSD) containing AI100 with Shonit software, motorized XYZ stage, a camera with firmware, Programmable Circuit Board (PCB) and its firmware for driving motor and LED, Switched-mode power supply (SMPS), power supply and a casing. It is capable of handling one Peripheral Blood Smear (PBS) slide at a time. Romanowsky stains are used with the device.

The device moves the peripheral whole blood smear slide to the imaging area and performs pre-scan steps which involve identifying the optimal area of the smear for scanning. If the optimal area is not found, the slide is rejected, and the user is notified with appropriate error messages. If an optimal area is identified, scanning (capturing FOV images) proceeds from the center of the optimal area in an outward spiral fashion. Focusing is done at each image location to capture the most optimal image. The image at each field of view (FOV) is checked to ensure appropriate quality for image processing. The scan stops when the required number of FOVs are captured or required number of white blood cells (WBCs) are encountered, depending on the scan mode selected by the user. On each FOV image, image processing is applied to extract and classify WBCs, red blood cells (RBCs), and platelets.

WBCs are classified into Neutrophils, Lymphocytes, Monocytes, Eosinophils, Basophils, Immature Granulocytes (IGs), Atypical Cells/Blasts and Nucleated Red Blood Cell (NRBCs). RBCs are classified according to size (Normocyte, Round Macrocyte, Ovalo Macrocyte, Microcyte) and shape (Normal, Target, Teardrop, Echinocyte, Elliptocyte, Ovalocyte, Fragmented). Platelets are classified into Normal, Large, Giant Platelets and Platelet Clumps. The device then allows the user to review the identified and classified cells, including cells that could not be classified and generate a microscopy report. Users may re-classify cells and add comments before approving the report.

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
K221309 - Page 2 of 12

{2}

B Instrument Description Information:

1. Instrument Name: AI100 with Shonit

2. Specimen Identification: The slide's barcode is captured scanned and assigned to each case. The device also supports typing the barcode manually.

3. Specimen Sampling and Handling: A peripheral blood sample is collected in K2EDTA tubes. A thin blood film is wedged on a clean dry glass slide (a blood smear) and stained with Romanowsky stain. The slide is put on the stage for image capture.

4. Calibration: The device calibration should be performed by the SigTuple support team or by the end user/customer under the guidance of the SigTuple support team with Operational Qualification (OQ) slide every 6 months. Device calibration should be conducted:
- Once every six months
- After the device has been serviced
- When a new OQ slide is provided

5. Quality Control: The AI100 with Shonit device performs self-test during startup after powering up or reset of the system. On startup, all software components and hardware components are checked and confirmed to behave normally before system is allowed to start a peripheral blood smear (PBS) scan. The quality is checked by performing the Operational Qualification (OQ) test and the Performance Qualification (PQ) test. Both tests use the OQ quality control slide. The OQ test should be conducted every three (3) months. In the OQ test, the parameters which determine functional and device performance metrics are tested automatically in a sequential order. The PQ test is for parameters which directly impact the image analysis and they are tested automatically in a sequential order. The PQ test should be conducted daily.

V Substantial Equivalence Information:

A Predicate Device Name(s): CellaVision DC-1

B Predicate 510(k) Number(s): K200595

C Comparison with Predicate(s):

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
K221309 - Page 3 of 12

{3}

|  Device & Predicate Device(s): | K221309 | K200595  |
| --- | --- | --- |
|  Device Trade Name | AI100 with Shonit | CellaVision DC-1  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | AI100 with Shonit is a cell locating device intended for in-vitro diagnostic use in clinical laboratories. AI100 with Shonit is intended for differential count of White Blood Cells (WBC), characterization of Red Blood Cells (RBC) morphology and Platelet morphology. It automatically locates blood cells on peripheral blood smears and presents images of the blood cells for review. A skilled operator, trained in the use of the device and in the review of blood cells, identifies and classifies each cell according to type. | CellaVision DC-1 is an automated cell-locating device intended for in-vitro diagnostic use in clinical laboratories. CellaVision DC-1 is intended to be used by operators trained in the use of the device. Peripheral Blood Application: The CellaVision Peripheral Blood Application is intended for differential count of white blood cells (WBC), characterization of red blood cell (RBC) morphology and platelet estimation. The CellaVision DC-1 with the Peripheral Blood Application automatically locates blood cells on peripheral blood (PB) smears. The application presents images of the blood cells for review. A skilled operator trained in recognition of blood cells, identifies and verifies the suggested classification of each cell according to type.  |
|  Parameters | Automated cell-locating device for cell-location and identification of RBC, WBC or platelets for in-vitro use. Verification of results by human operator. | Same  |
|  Light Source | LED (Light Emitting Diode) | Same  |

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

K221309 - Page 4 of 12

{4}

|  Sample Source | Stained blood film on glass slides of peripheral whole blood. | Same  |
| --- | --- | --- |
|  Sample Staining | Romanowsky stain | Same  |
|  Image interpretation requirements | A skilled operator is required to differentiate and finally modify and/or confirm the pre-classification/characterization of the located blood cells. | Same  |
|  Result format for WBC, RBC | The differential proportional count is normally based on 100 white blood cells. The number of WBCs can be modified if required. The result can be presented as a Count (count of located cells for the WBC cell type) or as % of total number of WBCs. The result of RBC characterization is presented as a grading for each morphology. | Same  |
|  User Interface | The User Interface is primarily designed to allow the user to view the images of the WBCs, RBCs and Platelets and review the classification. The user will be able to make corrections to the results and generate a report. | Same  |
|  Loading capacity | 1 slide | Same  |
|  Immersion oil application | Manual Application | Same  |
|  Neural network | Neural network of convolutional type | Same  |
|  General Device Characteristic Differences | AI100 with Shonit | CellaVision DC-1  |
|  Analysis technique: WBC | White blood cells are located and counted by moving according to the outwardly increasing spiral path. | White blood cells are located/counted by moving according to the battlement track pattern  |

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

K221309 - Page 5 of 12

{5}

|  Analysis technique: Platelet | Platelets are pre-classified based on morphology and images are displayed to the user. The operator verifies the suggested classification and confirms the qualitative output of ‘Detected’ vs ‘Not Detected’ for each platelet type. The user is presented with platelet images, classified based on morphology. The user can review the classification and confirm the qualitative output of Detected vs Not Detected against each platelet morphology category. | The operator manually counts and estimates the platelet concentration from the overview image according to a standardized procedure. From an overview image corresponding to eight high power fields the platelet level is estimated. The concentration of platelets is estimated by the user.  |
| --- | --- | --- |
|  Image Magnification | The device has one objective lens at 40X magnification. | The device has two objective lenses, one at 10X and one at 100X magnification.  |
|  Information transfer from instrument to Printer or network | The current system does not interact with a laboratory information system (LIS) | The system can interact with a laboratory information system (LIS) The system will retrieve order data from the LIS and send results back to the LIS.  |

VI Standards/Guidance Documents Referenced:

- EVS-EN ISO 14971: 2019 Medical devices - Application of risk management to medical devices, (Recognition Number: 5-125)
- ISO 15223-1, Fourth Edition 2021-07 Medical devices - Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements
- IEC 61010-1, Edition 3.1: 2017-01 - Standard for Safety for Electrical Equipment for Measurement, Control and Laboratory Use; Part 1: General Requirements
- IEC 62304, Edition 1.1: 2015-06 - Medical device software - Software life cycle processes
- IEC 62366-1, Edition 1.1:2020-06 - Medical devices - Part 1: Application of usability engineering to medical devices
- CLSI EP05-A3 - Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
K221309 - Page 6 of 12

{6}

- CLSI EP12-A2 - User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition
- CLSI H20-A2 - Reference Leukocyte (WBC) Differential Count (Proportional) an Evaluation of Instrumental Methods; Approved Standard - Second Edition
- ISTA 3B: 2013 - Packaged-Products for Less-Than-Truckload (LTL) Shipment
- IEC 61326-1 Edition 3.0 2020-10 - Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
- IEC 61326-2-6 Edition 3.0 2020-10 - Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment

## VII Performance Characteristics (if/when applicable):

## A Analytical Performance:

### 1. Precision:

a) Repeatability was conducted at one site using 12 blood samples. The slides were either prepared manually or using an automated smearer and stained with Romanowsky stain. The repeatability study was conducted over 20 days x 2 runs per day x 2 replicates per run. The WBC repeatability results with the highest recorded standard deviations are provided. The Mean and Standard Deviation with the maximum 95% CI are provided for each cell type. The RBC and Platelet repeatability results are provided as the mean overall agreement for each subgroup. All results met the pre-defined acceptance criteria.

|  Cell Type | Repeatability |   | Between Run |   | Between Day |   | Within Lab  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Mean (%) | SD (95%CI) | Mean (%) | SD (95%CI) | Mean (%) | SD (95%CI) | Mean (%) | SD (95%CI)  |
|  Neutrophil | 65.4 | 2.605 (2.139, 3.333) | 32.4 | 1.176 (0.994, 1.442) | 65.4 | 2.487 (2.084, 3.086) | 65.4 | 3.67 (3.049, 4.612)  |
|  Lymphocyte | 24.4 | 3.212 (2.637, 4.109) | 60.2 | 1.419 (1.198, 1.742) | 60.2 | 1.42 (1.173, 1.798) | 24.4 | 3.324 (2.871, 3.947)  |
|  Eosinophil | 17.5 | 1.408 (1.156, 1.801) | 2.1 | 0.352 (0.298, 0.429) | 5.2 | 0.838 (0.711, 1.021) | 17.5 | 1.496 (1.293, 1.777)  |
|  Monocyte | 3.8 | 1.156 (0.949, 1.479) | 3.8 | 0.832 (0.704, 1.019) | 3.1 | 0.653 (0.541, 0.825) | 3.8 | 1.507 (1.287, 1.818)  |
|  Basophil | 0.5 | 0.417 (0.342, 0.533) | 0.4 | 0.121 (0.102, 0.148) | 0.5 | 0.232 (0.195, 0.286) | 0.5 | 0.477 (0.408, 0.574)  |
|  Immature Granulocyte | 9.7 | 1.608 (1.320, 2.058) | 3 | 0.625 (0.530, 0.763) | 9.7 | 0.801 (0.679, 0.975) | 9.7 | 1.796 (1.542, 2.153)  |
|  Atypical Cell/Blast | 49.9 | 2.614 (2.146, 3.345) | 2.7 | 0.58 (0.486, 0.719) | 49.9 | 1.448 (1.223, 1.777) | 49.9 | 2.989 (2.557, 3.598)  |
|  NRBC | 1.3 | 0.533 (0.438, 0.682) | 1.3 | 0.199 (0.168, 0.245) | 1.3 | 0.361 (0.302, 0.451) | 1.3 | 0.674 (0.571, 0.822)  |

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
K221309 - Page 7 of 12

{7}

Mean Agreement for RBC Size

|  RBC Size | Overall Agreement Mean Values  |
| --- | --- |
|  Round Normocytes | 100%  |
|  Round Macrocytes | 92%  |
|  Oval Macrocytes | 93%  |
|  Anisocytosis | 95%  |
|  Microcytosis | 99%  |
|  Macrocytosis | 95%  |

Mean Agreement for RBC Shape

|  RBC Shape | Overall Agreement Mean Values  |
| --- | --- |
|  Normal | 100%  |
|  Elliptocyte | 99%  |
|  Teardrop | 94%  |
|  Acanthocyte | 96%  |
|  Target | 91%  |
|  Fragmented | 92%  |
|  Ovalocyte | 99%  |
|  Poikilocytosis | 99%  |

Mean Agreement for Platelets

|   | Overall Agreement Mean Value  |
| --- | --- |
|  Platelet | 100%  |
|  Giant Platelet | 96%  |
|  Platelet Clump | 83%  |

b) Reproducibility was conducted using 13 blood samples. The study was conducted over five (5) days x five (5) runs x three (3) devices. Each slide was scanned 5 times per day per device. Each scan was considered one replicate. The WBC reproducibility results with the highest standard deviations are provided. The Mean and Standard Deviation with the maximum 95% CI is provided for each cell type. All results met the pre-defined acceptance criteria.

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
K221309 - Page 8 of 12

{8}

|  Cell Type | Repeatability |   | Between Day |   | Between Site |   | Reproducibility  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  Mean (%) | SD (95%CI) | Mean (%) | SD (95%CI) | Mean (%) | SD (95%CI) | Mean (%) | SD (95%CI)  |
|  Neutrophil | 25.6 | 2.158 (1.832, 2.627) | 67.2 | 1.777 (1.002, 6.773) | 62.1 | 2.024 (1.634, 2.659) | 62.1 | 2.318 (1.330, 8.102)  |
|  Lymphocyte | 46.5 | 2.806 (2.382, 3.417) | 46.5 | 2.986 (1.644, 13.029) | 46.5 | 2.606 (2.073, 3.510) | 46.5 | 4.856 (3.485, 7.999)  |
|  Eosinophil | 10.8 | 1.501 (1.274, 1.828) | 10.8 | 2.095 (1.667, 2.822) | 9.9 | 1.147 (0.969, 1.406) | 10.8 | 2.578 (2.026, 3.545)  |
|  Monocyte | 9.4 | 1.033 (0.877, 1.258) | 5.9 | 1.094 (0.674, 2.814) | 9.4 | 0.844 (0.715, 1.032) | 9.4 | 1.424 (1.019, 2.362)  |
|  Basophil | 0.2 | 0.292 (0.248, 0.356) | 0.1 | 0.13 (0.070, 0.647) | 0.2 | 0.205 (0.171, 0.255) | 0.2 | 0.275 (0.175, 0.628)  |
|  Immature Granulocyte | 5.3 | 1.231 (1.044, 1.498) | 2.6 | 0.703 (0.394, 2.763) | 6.1 | 1.03 (0.820, 1.383) | 6.1 | 1.219 (0.714, 3.874)  |
|  Atypical Cell/Blast | 31.8 | 2.096 (1.779, 2.552) | 31.8 | 1.754 (0.937, 9.201) | 31.8 | 3.222 (2.627, 4.169) | 31.8 | 4.225 (2.639, 10.333)  |
|  NRBC | 13 | 1.044 (0.886, 1.271) | 13 | 1.466 (0.878, 4.214) | 2 | 1.038 (0.878, 1.269) | 2 | 1.27 (0.766, 3.541)  |

## Mean Agreement for RBC Size

|  RBC Size | Overall Agreement Mean Value  |
| --- | --- |
|  Round Normocytes | 100%  |
|  Round Macrocytes | 97%  |
|  Oval Macrocytes | 89%  |
|  Anisocytosis | 99%  |
|  Microcytosis | 100%  |
|  Macrocytosis | 96%  |

## Mean Agreement for RBC Shape

|  RBC Shape | Overall Agreement Mean Values  |
| --- | --- |
|  Normal | 97%  |
|  Elliptocyte | 96%  |
|  Teardrop | 96%  |
|  Acanthocyte | 91%  |
|  Target | 97%  |
|  Fragmented | 91%  |
|  Ovalocyte | 99%  |
|  Poikilocytosis | 95%  |

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
K221309 - Page 9 of 12

{9}

Mean Agreement for platelet:

|   | Overall Agreement Mean Values  |
| --- | --- |
|  Platelet | 100%  |
|  Giant Platelet | 92%  |
|  Platelet Clumps | 93%  |

2. Linearity:

Not applicable.

3. Analytical Specificity/Interference:

Not applicable.

4. Accuracy (Instrument):

A method comparison study was conducted at four (4) sites comparing the AI100 with Shonit to manual microscopy. A total of 882 samples were collected encompassing healthy subjects and patients with specific disease conditions. Blood samples (K₂EDTA) were collected, and one peripheral blood smear (PBS) slide was prepared for each sample (Romanowsky stain). The slides were randomly selected, blinded and evaluated by two examiners at each site. Passing-Bablok Regression analysis was performed for WBC differential and overall agreement, sensitivity and specificity were measured for WBC Distributional and Morphological abnormalities. Overall agreement, sensitivity and specificity were evaluated for RBC and PLT morphology. Calculations of 95% confidence intervals are also provided. The study met the pre-defined acceptance criteria.

Regression results for WBC cell types

|  WBC cell type | Parameter | Result (n = 882) with 95% CI  |
| --- | --- | --- |
|  Neutrophil % | Slope | 1.024 (1.016, 1.032)  |
|   |  Intercept | -1.78 (-2.249, -1.346)  |
|   |  Pearson's Correlation Coefficient | 0.962  |
|  Lymphocyte % | Slope | 1.025 (1.016, 1.034)  |
|   |  Intercept | -0.587 (-0.881, -0.306)  |
|   |  Pearson's Correlation Coefficient | 0.96  |
|  Eosinophil % | Slope | 1.029 (1.012, 1.05)  |
|   |  Intercept | -0.039 (-0.07, -0.01)  |
|   |  Pearson's Correlation Coefficient | 0.907  |
|  Monocyte % | Slope | 1.083 (1.051, 1.117)  |
|   |  Intercept | -0.462 (-0.66, -0.304)  |
|   |  Pearson's Correlation Coefficient | 0.789  |

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
K221309 - Page 10 of 12

{10}

Distributional and Morphological abnormalities for WBCs

|  WBC abnormality | Parameter | Results (n = 882) with 95% CI  |
| --- | --- | --- |
|  Morphological abnormality | Overall Agreement | 91.7% (90.4%, 92.8%)  |
|   |  Sensitivity | 95.3% (92.8%, 96.7%)  |
|   |  Specificity | 90.9% (89.4%, 92.2%)  |
|  Distributional abnormality | Overall Agreement | 96.4% (95.5%, 97.2%)  |
|   |  Sensitivity | 91.0% (86.8%, 93.9%)  |
|   |  Specificity | 97.2% (96.3%, 97.9%)  |
|  Overall abnormality | Overall Agreement | 95.0% (94.0%, 95.9%)  |
|   |  Sensitivity | 92.7% (89.2%, 95.0%)  |
|   |  Specificity | 95.4% (94.35%, 96.25%)  |

RBC Morphology

|  RBC subgroup | Parameter | Results (n = 882) with 95% CI  |
| --- | --- | --- |
|  Anisocytosis | Overall agreement | 94.7% (93.6%, 95.7%)  |
|   |  Sensitivity | 91.1% (88.1%, 93.4%)  |
|   |  Specificity | 95.9% (94.7, 96.9%)  |
|  Poikilocytosis | Overall agreement | 92.1% (90.7, 93.2%)  |
|   |  Sensitivity | 96.3% (94.8%, 97.3%)  |
|   |  Specificity | 88.1% (85.8%, 90.0%)  |
|  Macrocytosis | Overall agreement | 95.5% (94.5%, 96.4%)  |
|   |  Sensitivity | 90.7% (87.0, 93.5%)  |
|   |  Specificity | 96.6% (95.5%, 97.4%)  |

Platelet Morphology

|  PLT Cell type | Parameter | Results (n = 882) with 95% CI  |
| --- | --- | --- |
|  Normal platelets | Overall agreement | 100% (99.8%, 100%)  |
|   |  Sensitivity | 100% (99.8%, 100%)  |
|   |  Specificity | 100% (34.2%, 100%)  |
|  Giant platelets | Overall agreement | 96.4% (95.4%, 97.1%)  |
|   |  Sensitivity | 99.1% (98.4%, 99.5%)  |
|   |  Specificity | 92.4% (90.3%, 94.1%)  |
|  Platelet clumps | Overall agreement | 94.2% (93.0%, 95.2%)  |
|   |  Sensitivity | 91.6% (89.5%, 93.4%)  |
|   |  Specificity | 96.3% (94.9%, 97.3%)  |
|  Overall platelets | Overall agreement | 96.8% (96.0%, 97.4%)  |
|   |  Sensitivity | 97.9% (97.1%, 98.4%)  |
|   |  Specificity | 94.6% (92.8%, 95.9%)  |

5. Carry-Over:

Not applicable.

B Other Supportive Instrument Performance Characteristics Data:

1. Scan Mode Equivalence:

The scan mode equivalency study was conducted at one site with 200 samples collected and slides prepared using Romanowsky stain. The slides were scanned in each mode (i.e.,

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
K221309 - Page 11 of 12

{11}

100WBC, 200WBC, and 120 FOVs). There were 90 healthy individuals and 100 abnormal samples included in the study. Regression analysis with slope, intercept and 95% CI are calculated for the WBC differential. Overall agreement, positive percent agreement (PPA), Negative percent agreement (NPA) with 95% CI are evaluated for WBC Distributional and Morphological abnormalities, RBC morphology, and PLT morphology. The study met the pre-defined acceptance criteria.

## VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
K221309 - Page 12 of 12

---

**Source:** [https://fda.innolitics.com/device/K221309](https://fda.innolitics.com/device/K221309)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
