K221302 · Serpex Medical, Inc. · EOQ · Aug 15, 2022 · Ear, Nose, Throat
Device Facts
Record ID
K221302
Device Name
Recon Steerable Sheath
Applicant
Serpex Medical, Inc.
Product Code
EOQ · Ear, Nose, Throat
Decision Date
Aug 15, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Indications for Use
The Recon Steerable Sheath is an extended working channel intended to be used with a compatible bronchoscope to guide endotherapy accessories to the target area within the respiratory system.
Device Story
Recon Steerable Sheath is a sterile, single-use transbronchial sheath; provides access to intrapulmonary regions. Device includes steerable sheath, stylet, and bronchoscope adapters. Used with compatible 2.8mm working channel bronchoscopes. Clinician inserts device through bronchoscope; advances into lung. Handle controls rotation, extension, retraction, and distal tip articulation (minimum 90° via internal pull wire). Once positioned, stylet is removed, allowing insertion of endotherapy tools (up to 1.9mm diameter). Used in bronchoscopy suites by trained clinicians. Provides a path for endotherapy tools to reach target tissue; facilitates diagnostic or therapeutic procedures.
Clinical Evidence
No clinical data. Bench testing included visual, dimensional, insertion force, articulation, tensile, pneumostasis, simulated use, and lumen patency testing. Usability validation (formative and summative) conducted in simulated bronchoscopy suites per ANSI/AAMI/IEC 62366-1:2015. Biocompatibility testing performed per ISO 10993-1. Sterilization validated per ISO 11135.
Technological Characteristics
Sterile, single-use, transbronchial sheath. Features manual unidirectional distal tip articulation (min 90°). Materials biocompatible per ISO 10993-1. Sterilized via ethylene oxide per ISO 11135. Compatible with 2.8mm working channel bronchoscopes. Dimensions: ~1095mm length, ≤2.73mm outer diameter, ≥1.93mm inner diameter. Radiopaque. Mechanical actuation via internal pull wire.
Indications for Use
Indicated for use as an extended working channel with a compatible bronchoscope to guide endotherapy accessories to target areas within the respiratory system in patients undergoing bronchoscopy.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 15, 2022
Serpex Medical, Inc. % Laurie Lewandowski Vice President Honkanen Consulting, Inc. 738 Saddle Wood Drive Eagan, Minnesota 55123
Re: K221302
Trade/Device Name: Recon Steerable Sheath Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ Dated: July 14, 2022 Received: July 15, 2022
Dear Laurie Lewandowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221302
Device Name Recon Steerable Sheath
Indications for Use (Describe)
The Recon Steerable Sheath is an extended working channel intended to be used with a compatible bronchoscope to guide endotherapy accessories to the target area within the respiratory system.
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
### 1. SUBMITTER INFORMATION
Submitter: Serpex Medical, Inc. Sasha Schrode 3350 Scott Blvd, Suite 37B Santa Clara, CA 95054 sschrode(@serpexmedical.com Email:
Primary Contact: Laurie Lewandowski Consultant, Honkanen Consulting, Inc. 738 Saddle Wood Drive Eagan, MN 55123 Telephone: 612-770-4038 (cell) llewandowski(@serpexmedical.com Email:
DATE PREPARED:
May 4, 2022
### 2. DEVICE INFORMATION
### Recon Steerable Sheath
Proprietary Name: Classification Name: Common/Usual Name: Regulatory Class: Product Code: Regulation Number:
Bronchoscope (flexible or rigid) and accessories Guide Catheter Class II EOQ 21 CFR 874.4680
### 3. PREDICATE DEVICE INFORMATION
Proprietary Name: Common/Usual Name: Classification Name: Regulatory Class: Product Code: Regulation Number: 510K Number:
Lung Vision Tool Guide Catheter Bronchoscope (flexible or rigid) and accessories Class II EOQ 21 CFR 874.4680 K172955
### 4. DEVICE DESCRIPTION
The Recon Steerable Sheath is a sterile, single use transbronchial sheath with a unidirectional, steerable distal tip that provides access to the intrapulmonary regions. It is
{4}------------------------------------------------
designed to guide endotherapy tools with an outer diameter of up to 1.9 mm and minimum length of 1100 mm to the target tissue.
The Recon Steerable Sheath (Model RSS1000) includes the Steerable Sheath, Stylet, Bronchoscope Adapter and Pentax Adapter.
The Recon Steerable Sheath is compatible with Olympus® 190 or Pentax® bronchoscopes that have a 2.8mm working channel and 600 mm working length. It is coupled to the selected bronchoscope using an adapter(s).
The handle provides the user with control of device rotation, extension, retraction, and distal tip articulation of 90° minimum within a plane with the stylet inserted. A Luer connector on the proximal end of the device provides the connection for the stylet.
The Recon Steerable Sheath with stylet is inserted and coupled to a bronchoscope via a clip on the handle. The telescope advances the device beyond the bronchoscope into the lung. Depressing the plunger articulates the distal end of the sheath by tensioning an internal pull wire. Once in the desired location, the Stylet is removed from the Steerable Sheath and the tool of choice can be inserted for access.
## 5. INTENDED USE/INDICATION FOR USE
The Recon Steerable Sheath is an extended working channel intended to be used with a compatible bronchoscope to guide endotherapy accessories to the target area within the respiratory system.
# 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The Recon Steerable Sheaths have the same technological characteristics as the predicate, Lung Vision Tool, cleared under K172955. The subject and predicate device both provide a path for endotherapy tools to reach target tissue.
A comparison of the Recon Steerable Sheath and the Lung Vision Tool is in the following table.
| Attribute | Proposed Device<br>Recon Steerable Sheath | Body Vision<br>Lung Vision Tool<br>K172955<br>Predicate | Difference and Impact<br>on Substantial<br>Equivalence |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | To guide endoscopic tools<br>to target tissue | Identical | Identical to predicate |
| Attribute | Proposed Device<br>Recon Steerable Sheath | Body Vision<br>Lung Vision Tool<br>K172955<br>Predicate | Difference and Impact<br>on Substantial<br>Equivalence |
| Indications for<br>use | The Recon Steerable<br>Sheath is an extended<br>working channel intended<br>to be used with a<br>compatible bronchoscope<br>to guide endotherapy<br>accessories to the target<br>area within the respiratory<br>system. | LungVision Tool is an<br>instrument designed as<br>a working channel<br>intended to be used<br>with standard<br>bronchoscopes,<br>endotherapy<br>accessories and<br>ultrasound probe to<br>guide the endotherapy<br>accessories or<br>ultrasound probe to<br>the target area,<br>specifically within the<br>respiratory system. | Similar to predicate;<br>minor wording<br>differences including the<br>use of an ultrasound<br>probe for guiding for the<br>predicate |
| Users | Clinicians trained in<br>bronchoscopy devices,<br>accessories, and<br>procedures | Identical | Identical to predicate |
| Anatomic Site | Lung | Identical | Identical to predicate |
| Method of<br>Introduction | Delivered through a<br>flexible bronchoscope | Identical | Identical to predicate |
| Recommended<br>Channel Size | 2.8 mm | Identical | Identical to predicate |
| Components | Handle,<br>Shaft,<br>Stylet,<br>Adapters | Handle,<br>Sheath,<br>Stylet,<br>Adapters | Identical to predicate |
| Compatible<br>Bronchoscope<br>Size | 2.8 mm | Identical | Identical to predicate |
| Articulation /<br>Curve | Manual articulation<br>(depress plunger) to<br>articulate the distal tip a<br>minimum of 90° | Preformed curve<br>shapes of 90° and 210° | Similar<br>Subject device<br>articulates to 90°, the<br>same articulation as one<br>of the preformed curves. |
| Extension | Yes<br>140mm | Yes<br>Not listed | Identical to predicate |
| Extension<br>beyond scope<br>with device<br>retracted | ≤ 10 mm | 17 mm | Similar |
| Attribute | Proposed Device<br>Recon Steerable Sheath | Body Vision<br>Lung Vision Tool<br>K172955<br>Predicate | Difference and Impact<br>on Substantial<br>Equivalence |
| Outer<br>Diameter | $ \u2264 $ 2.73mm | $ \u2264 $ 2.72mm | Similar<br>Both are intended to fit<br>within a 2.8 mm<br>bronchoscope |
| Inner<br>Diameter | $ \u2265 $ 1.93mm | $ \u2265 $ 2.08mm | Similar – accommodates<br>endotherapy tools up to<br>1.90mm |
| Total Length | 1095 mm ± 5 mm | 1000mm | Similar |
| Stylet OD | 1.83 mm ± 0.05 mm | Not listed | Sized to fit sheath |
| Radiopaque | Yes | Yes | Identical to predicate |
| Life/Sterility | Single Use, EO sterilized | Identical | Identical to predicate |
| Biocompatible | Yes | Yes | Identical to predicate |
{5}------------------------------------------------
{6}------------------------------------------------
The Recon Steerable Sheath has the same intended use, similar indications for use with the ability to articulate distal tip 90° minimum unidirectionally to reach target tissue. The angle of articulation is identical to one of the preformed curve shapes of the predicate device. The clinician manually articulates the subject device and manually manipulates the predicate device to reach target tissue.
The different technological characteristics of the new device do not raise different questions of safety and effectiveness.
### 7. PERFORMANCE DATA
## Bench Testing
All testing was performed pre and post aging.
- Visual ●
- Dimensional ●
- Insertion force ●
- Articulation angle, length, curve profile, curve stability, and planarity ●
- Coupling Tensile ●
- Tensile for all bonds/joints ●
- Bronchoscope Adapter Pneumostasis / Adapter Vacuum,
- Simulated Use / Device robustness
- Plunger force
- Bending, Buckling
- Torque Transmission and Integrity ●
- Lumen Patency ●
### Packaging
Packaging was subjected to Environmental, Distribution Simulation and Aging with testing performed pre and post aging
Serpex Medical, Inc. Recon Steerable Sheath
{7}------------------------------------------------
Distribution, Environmental and Aging of Packages
- Distribution per ASTM D4169:2016 ●
- Environmental Conditioning per ISTA 3A:2011 ●
- Aging per ASTM F1980:2016,
- . Packaging per ISO 11607-1:2006
- Visual per ASTM F1886-16/F1886M-16 o
- Bubble Leak per ASTM F2096-11:2019 o
- Seal Strength per ASTM F88/F88M-15 o
Packages met the specifications pre and post aging.
All tests met the predefined acceptance criteria. The test results demonstrated that differences in device characteristics between the subject device and predicate device do not raise any new questions of safety or effectiveness.
## Validation testing
Validation testing was performed demonstrating the Recon Steerable Sheath user needs and intended use were met. Testing under simulated use conditions was performed in accordance with the Instructions for Use. The Recon Steerable Sheath met the performance requirements and was found to be clinically acceptable by all evaluators.
## Usability
A human factors (HF) engineering process was followed in accordance with the following:
- ANSI/AAMI/IEC 62366-1:2015, Medical devices Application of usability . engineering to medical devices
- . FDA Guidance: "Applying Human Factors and Usability Engineering to Medical Devices - Guidance for Industry and Food and Drug Administration Staff." 2016
Formative and summative validation studies were conducted to identify and minimize use errors related to the use of the Recon Steerable Sheath. Studies were conducted by intended user groups in a simulated bronchoscopy suite and involved preoperative preparation and simulated procedures. The Instructions for Use and Reference Guide were assessed in the usability study. No new use errors were identified. The Recon Steerable Sheath has been assessed and found to be safe and effective for its intended use, by the intended users, in its intended use environment.
# Biocompatibility
Biocompatibility testing was performed in accordance with:
- ISO 10993-1:2018 for a limited (<24 hour) tissue contacting device and .
- FDA Guidance: "Use of International Standard ISO-10993, "Biological Evaluation of . Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process," June 2016
The following testing was performed:
{8}------------------------------------------------
- Cytotoxicity: MEM Extraction Cytotoxicity Assay per ISO 10993-5:2009/ (R) 2014 ●
- Sensitization: Guinea Pig Maximization Test per ISO 10993-10:2010 / (R) 2014 .
- Irritation: Intracutaneous Reactivity Test, per ISO10993- 23:2021 .
- . Toxicity: Acute Systemic Toxicity per ISO 10993-11:2017
- Toxicity: Materials Medicated Rabbit Pyrogen Test ISO 10993-11:2017 ●
- Pyrogen testing per USP 39 <151> .
- . Bacterial Endotoxin Testing per USP 39 <85>
## Sterilization:
Sterilization (ethylene oxide) and packaging of the Recon Steerable Sheaths were validated using the following standards:
- ANSI/AAMI/ISO 11607-1:2019 Packaging for terminally sterilized medical . devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems
- ANSI/AAMI/ISO 11135-1:2014 Sterilization of health-care products ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices
- FDA Guidance: Submission and Review of Sterility Information in Pre-Market ● Notification 510(k) Submissions for Device Labeled as Sterile (January 21, 2016)
- AAMI ST72:2011/(R)2016 Bacterial Endotoxins Test Methods, Routine ● Monitoring,
- ISO 10993-7:2008 Amendment 1:2019 Biological evaluation of medical devices -● Part 7: Ethylene oxide sterilization residuals
Testing demonstrated that all endpoints were met.
## 8. CONCLUSION
The Recon Steerable Sheath is substantially equivalent in terms of the indications for use, technological characteristics, performance testing and comparison to the cited predicate, and does not present any new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.