← Product Code [LZA](/productcode/LZA) · K221269

# Powder Free Nitrile Examination Gloves (Blue),Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (K221269)

_Better Care Plastic Technology Co., Ltd. · LZA · Jul 27, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221269

## Device Facts

- **Applicant:** Better Care Plastic Technology Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Jul 27, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05(2019)

## Device Story

Disposable, non-sterile, ambidextrous nitrile examination gloves; worn by clinicians to prevent cross-contamination between patient and examiner. Device provides a physical barrier; tested for permeation resistance against specific chemotherapy agents and Fentanyl Citrate per ASTM D6978-05(2019). Output is the physical barrier itself; used in clinical settings. Benefits include protection of the examiner from exposure to hazardous drugs and patient contaminants. Warning: Not for use with Carmustine or Thio Tepa due to low permeation breakthrough times.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM D6319-19 (physical dimensions, properties), ASTM D5151-19 (water leak), ASTM D6124-06 (powder residue), ASTM D6978-05 (chemotherapy permeation), and ISO 10993 series (biocompatibility: irritation, sensitization, cytotoxicity, systemic toxicity).

## Technological Characteristics

Nitrile material; powder-free; blue color; ambidextrous; non-sterile. Dimensions: 230mm min length, 0.05mm min thickness. Tensile strength 14MPa min (before/after aging). Connectivity: N/A. Standards: ASTM D6319-19, ASTM D6978-05(2019), ASTM D5151-19, ASTM D6124-06, ASTM D412-16, ISO 10993-5, 10, 11.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Blue Colored, Power Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate ([K192954](/device/K192954.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.

July 27, 2022

Better Care Plastic Technology Co., Ltd. % Kathy Liu Project Manager Hongray USA Medical Products Inc. 3973 Schaefer Avenue, Chino, CA 91710, USA Chino, California 91710

Re: K221269

Trade/Device Name: Powder Free Nitrile Examination Gloves (Blue),Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, ODO Dated: April 18, 2022 Received: May 2, 2022

Dear Kathy Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K221269

Device Name

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentany) Citrate

Indications for Use (Describe)

The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05(2019)

| Chemotherapy Drug                                                         | Minimum Breakthrough Detection Time (BDT) in Minutes |
|---------------------------------------------------------------------------|------------------------------------------------------|
| Bleomycin Sulfate 15mg/ml (15000 ppm)                                     | >240                                                 |
| Busulfan 6mg/ml (6,000 ppm)                                               | >240                                                 |
| Carboplatin 10mg/ml (10,000 ppm)                                          | >240                                                 |
| Carmustine (BCNU)3.3 mg/ml (3,300 ppm)                                    | 11.1                                                 |
| Chloroquine 50mg/ml (50,000ppm)                                           | >240                                                 |
| Cisplatin 1mg/ml (1,000 ppm)                                              | >240                                                 |
| Cyclophosphamide 20mg/ml (20,000 ppm)                                     | >240                                                 |
| Cyclosporin 100 mg/ml (100,000 ppm)                                       | >240                                                 |
| Cytarabine HCL, 100 mg/ml (100,000 ppm)                                   | >240                                                 |
| Dacarbazine 10 mg/ml (10,000 ppm)                                         | >240                                                 |
| Daunorubicin HCL, 5 mg/ml (5,000 ppm)                                     | >240                                                 |
| Docetaxel, 10 mg/ml (10,000 ppm)                                          | >240                                                 |
| Doxorubicin HCL, 2 mg/ml (2,000 ppm)                                      | >240                                                 |
| Epirubicin HCL, 2 mg/ml (2,000 ppm)                                       | >240                                                 |
| Etoposide, 20 mg/ml (20,000 ppm)                                          | >240                                                 |
| Fludarabine, 25 mg/ml (25,000 ppm)                                        | >240                                                 |
| Fluorouracil, 50mg/ml (50,000ppm)                                         | >240                                                 |
| Gemcitabine, 38mg/ml (38,000ppm)                                          | >240                                                 |
| Idarubicin HCL, 1mg/ml (1,000ppm)                                         | >240                                                 |
| Ifosfamide, 50mg/ml (50,000ppm)                                           | >240                                                 |
| Irinotecan, 20mg/ml (20,000ppm)                                           | >240                                                 |
| Mechlorethamine HCI, 1mg/ml (1,000ppm)                                    | >240                                                 |
| Melphalan, 5mg/ml (5,000ppm)                                              | >240                                                 |
| Methotrexate, 25mg/ml (25,000ppm)                                         | >240                                                 |
| Mitomycin C, 0.5mg/ml (500ppm)                                            | >240                                                 |
| Mitoxantrone HCL, 2mg/ml (2,000ppm)                                       | >240                                                 |
| Oxaliplatin, 5mg/ml (5,000ppm)                                            | >240                                                 |
| Paclitaxel, 6mg/ml (6,000ppm)                                             | >240                                                 |
| Paraplatin, 10mg/ml (10,000ppm)                                           | >240                                                 |
| Retrovir, 10mg/ml (10,000ppm)                                             | >240                                                 |
| Rituximab, 10mg/ml (10,000ppm)                                            | >240                                                 |
| Thio Tepa, 10mg/ml (10,000ppm)                                            | 21.6                                                 |
| Topotecan, 1mg/ml (1,000ppm)                                              | >240                                                 |
| Trisenox, 1mg/ml (1,000ppm)                                               | >240                                                 |
| Velcade, 1mg/ml (1,000ppm)                                                | >240                                                 |
| Vincristine Sulfate, 1mg/ml (1,000ppm)                                    | >240                                                 |
| Fentanyl Citrate Injection (100 mcg/2ml)                                  | >240                                                 |
| Please note that the following drugs have extremely low permeation times: |                                                      |

{3}------------------------------------------------

| Carmustine: | 11.1 minutes |
|-------------|--------------|
| Thio Tepa:  | 21.6 minutes |

*Warning: Do not use with Carmustine and Thio Tepa.

Type of Use (Select one or both, as applicable)

_ Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China

### 510K Summary

The assigned 510(K) numbers: K221269 Date Prepared: July 22, 2022

### 1. Owner's Identification:

Better Care Plastic Technology Co., Ltd. Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China Tel:86-311-66179668 Contact: Ms. Kathy Liu, Project Manager Address: 3973 Schaefer Avenue, Chino, CA 91710, USA Tel:909-590-1611 Email: kathyliu@hongrayusa.com or fdareg@hongray.com.cn

#### 2. Name of the Device:

Trade / Product Name: Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Common Name: Exam Gloves Classification Name: Patient Examination Glove Specialty Classification Regulation: 21 CFR 880.6250 Product Code: LZA, LZC, QDO Classification Panel: General Hospital Device Class: Class I

### 3. Predicate Device Information:

Comfort Rubber Gloves Industries Sdn. Bhd. Blue Colored, Power Free Nitrile Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (K192954)

### 4. Device Description:

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate Are Class I Patient Examination Gloves and Specialty Chemotherapy Gloves. They are ambidextrous and come in different sizes = Extra Small, Medium, Large, Extra Large and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05(2019). The gloves are single use, disposable, and provided non-sterile.

### 5. Indications for Use:

Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

Gloves have been tested for use with chemotherapy drugs and Fentanyl Citrate using ASTM D6978-05(2019)

{5}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China

# 510K Summary

The following chemicals have been tested with these gloves:

| Chemotherapy Drug                        | Minimum Breakthrough Detection Time (BDT) | in Minutes |
|------------------------------------------|-------------------------------------------|------------|
| Bleomycin Sulfate 15mg/ml (15000 ppm)    | >240                                      |            |
| Busulfan 6mg/ml (6,000 ppm)              | >240                                      |            |
| Carboplatin 10mg/ml (10,000 ppm)         | >240                                      |            |
| Carmustine (BCNU)3.3 mg/ml (3,300 ppm)   | 11.1                                      |            |
| Chloroquine 50mg/ml (50,000ppm)          | >240                                      |            |
| Cisplatin 1mg/ml (1,000 ppm)             | >240                                      |            |
| Cyclophosphamide 20mg/ml (20,000 ppm)    | >240                                      |            |
| Cyclosporin 100 mg/ml (100,000 ppm)      | >240                                      |            |
| Cytarabine HCL, 100 mg/ml (100,000 ppm)  | >240                                      |            |
| Dacarbazine 10 mg/ml (10,000 ppm)        | >240                                      |            |
| Daunorubicin HCL, 5 mg/ml (5,000 ppm)    | >240                                      |            |
| Docetaxel, 10 mg/ml (10,000 ppm)         | >240                                      |            |
| Doxorubicin HCL, 2 mg/ml (2,000 ppm)     | >240                                      |            |
| Epirubicin HCL, 2 mg/ml (2,000 ppm)      | >240                                      |            |
| Etoposide, 20 mg/ml (20,000 ppm)         | >240                                      |            |
| Fludarabine, 25 mg/ml (25,000 ppm)       | >240                                      |            |
| Fluorouracil, 50mg/ml (50,000ppm)        | >240                                      |            |
| Gemcitabine, 38mg/ml (38,000ppm)         | >240                                      |            |
| Idarubicin HCL, 1mg/ml (1,000ppm)        | >240                                      |            |
| Ifosfamide, 50mg/ml (50,000ppm)          | >240                                      |            |
| Irinotecan, 20mg/ml (20,000ppm)          | >240                                      |            |
| Mechlorethamine HCI, 1mg/ml (1,000ppm)   | >240                                      |            |
| Melphalan, 5mg/ml (5,000ppm)             | >240                                      |            |
| Methotrexate, 25mg/ml (25,000ppm)        | >240                                      |            |
| Mitomycin C, 0.5mg/ml (500ppm)           | >240                                      |            |
| Mitoxantrone HCL, 2mg/ml (2,000ppm)      | >240                                      |            |
| Oxaliplatin, 5mg/ml (5,000ppm)           | >240                                      |            |
| Paclitaxel, 6mg/ml (6,000ppm)            | >240                                      |            |
| Paraplatin, 10mg/ml (10,000ppm)          | >240                                      |            |
| Retrovir, 10mg/ml (10,000ppm)            | >240                                      |            |
| Rituximab, 10mg/ml (10,000ppm)           | >240                                      |            |
| Thio Tepa, 10mg/ml (10,000ppm)           | 21.6                                      |            |
| Topotecan, 1mg/ml (1,000ppm)             | >240                                      |            |
| Trisenox, 1mg/ml (1,000ppm)              | >240                                      |            |
| Velcade, 1mg/ml (1,000ppm)               | >240                                      |            |
| Vincristine Sulfate, 1mg/ml (1,000ppm)   | >240                                      |            |
| Fentanyl Citrate Injection (100 mcg/2ml) | >240                                      |            |

{6}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base,

Shenze County, Hebei, 050000, China

### 510K Summary

*Please note that the following drugs have extremely low permeation times: Carmustine: 11.1 minutes, Thio Tepa: 21.6 minutes Warning: Do not use with Carmustine and Thio Tepa.

### 6. Comparison of Subject Device and Predicate Device:

| General Comparison Table:                           |                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                  |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
|                                                     | Proposed Device<br>K221269                                                                                                                                                                                                                                                                                                                                                                      | Predicate Device<br>K192954                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comparison                       |
| Trade Name                                          | Powder Free Nitrile<br>Examination Gloves (Blue),<br>Tested for Use with<br>Chemotherapy Drugs and<br>Fentanyl Citrate                                                                                                                                                                                                                                                                          | Blue Colored, Power Free Nitrile<br>Examination Gloves Tested for<br>Use with Chemotherapy Drugs<br>and Fentanyl Citrate                                                                                                                                                                                                                                                                                                                                      | Similar                          |
| Product Code                                        | LZA, LZC, QDO                                                                                                                                                                                                                                                                                                                                                                                   | LZA, LZC,QDO                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                             |
| Regulation Number                                   | 21 CFR 880.6250                                                                                                                                                                                                                                                                                                                                                                                 | 21 CFR 880.6250                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                             |
| Class                                               | I                                                                                                                                                                                                                                                                                                                                                                                               | I                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                             |
| Indications for Use                                 | Powder Free Nitrile<br>Examination Gloves (Blue),<br>Tested for Use with<br>Chemotherapy Drugs and<br>Fentanyl Citrate is a disposable<br>device intended for medical<br>purposes that is worn on the<br>examiner's hand to prevent<br>contamination between patient<br>and examiner. Gloves have<br>been tested for use with<br>chemotherapy drugs and<br>Fentanyl Citrate using ASTM<br>D6978 | Blue Colored, Powder Free Nitrile<br>Examination Gloves, Non-sterile,<br>and Tested for Use with<br>Chemotherapy Drugs and Fentanyl<br>Citrate is a patient medical exam<br>glove which is a disposable device<br>intended for medical purpose that<br>is worn on the examiner's hand or<br>finger to prevent contamination<br>between examiner and patient.<br>Glove was tested for use with<br>Chemotherapy Drugs and Fentanyl<br>Citrate as per ASTM D6978 | Similar                          |
| Material                                            | Nitrile                                                                                                                                                                                                                                                                                                                                                                                         | Nitrile                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                             |
| Powder or Powder Free                               | Powder Free                                                                                                                                                                                                                                                                                                                                                                                     | Powder Free                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                             |
| Color                                               | Blue                                                                                                                                                                                                                                                                                                                                                                                            | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same                             |
| Single use                                          | Single use                                                                                                                                                                                                                                                                                                                                                                                      | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                             |
| Chemotherapy Drugs<br>and Fentanyl Citrate<br>Claim | See below comparison table                                                                                                                                                                                                                                                                                                                                                                      | See below comparison table                                                                                                                                                                                                                                                                                                                                                                                                                                    | See below<br>comparison<br>table |
| Technological Characteristic Comparison Table:      |                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                  |
| Technological<br>Characteristics                    | Proposed Device<br>K221269                                                                                                                                                                                                                                                                                                                                                                      | Predicate Device<br>K192954                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comparison                       |
| Length                                              | Minimum 230mm                                                                                                                                                                                                                                                                                                                                                                                   | Minimum 240mm                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Similar                          |
| Palm Width (size) (mm)                              |                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                  |
| XS                                                  | 70±10                                                                                                                                                                                                                                                                                                                                                                                           | 70±10                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                             |
| S                                                   | 80±10                                                                                                                                                                                                                                                                                                                                                                                           | 80±10                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                             |
| M                                                   | 95±10                                                                                                                                                                                                                                                                                                                                                                                           | 95±10                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                             |
| L                                                   | 110±10                                                                                                                                                                                                                                                                                                                                                                                          | 110±10                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                             |
| XL                                                  | 120±10                                                                                                                                                                                                                                                                                                                                                                                          | 120±10                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                             |
| XXL                                                 | 130±10                                                                                                                                                                                                                                                                                                                                                                                          | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Different                        |
| Thickness(mm)                                       |                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                  |
| Finger                                              | Minimum 0.05                                                                                                                                                                                                                                                                                                                                                                                    | Minimum 0.05                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                             |
| Palm                                                | Minimum 0.05                                                                                                                                                                                                                                                                                                                                                                                    | Minimum 0.05                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                             |
| Tensile Strength, Before<br>Aging                   | 14MPa, min                                                                                                                                                                                                                                                                                                                                                                                      | 14MPa, min                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                             |
| Ultimate Elongation,<br>Before Aging                | 500%, min                                                                                                                                                                                                                                                                                                                                                                                       | 500%, min                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                             |
| Tensile Strength, After<br>Accelerated Aging        | 14MPa, min                                                                                                                                                                                                                                                                                                                                                                                      | 14MPa, min                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                             |
| Ultimate Elongation, After<br>Accelerated Aging     | 400%, min                                                                                                                                                                                                                                                                                                                                                                                       | 400%, min                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                             |
| Watertight (1000ml)                                 | 21 CFR 800.20<br>ASTM D5151<br>AQL 2.5                                                                                                                                                                                                                                                                                                                                                          | 21 CFR 800.20<br>ASTM D5151<br>AQL 2.5                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                             |
| Powder-Content                                      | ≤ 2 mg per glove                                                                                                                                                                                                                                                                                                                                                                                | ≤ 2 mg per glove                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                             |
| 10993-10:2010 Skin<br>Irritation Study              | Under the conditions of<br>the study, not an irritant                                                                                                                                                                                                                                                                                                                                           | Under the conditions of the<br>study, the subject device is<br>non-irritating                                                                                                                                                                                                                                                                                                                                                                                 | Same                             |
| 10993-10:2010 Maximization<br>Sensitization Study   | Under the conditions of the<br>study, not a sensitizer                                                                                                                                                                                                                                                                                                                                          | Under the conditions of the<br>study, the subject device is<br>non sensitization                                                                                                                                                                                                                                                                                                                                                                              | Same                             |
| 10993-5:2009 Cytotoxicity<br>Test                   | Under the conditions of this<br>study, the test article extract<br>showed potential toxicity to<br>L929 cells.<br>Cytotoxicity concern was<br>addressed by acute systematic<br>toxicity testing.                                                                                                                                                                                                | Exhibits severe cytotoxicity<br>reactivity at 100%, and 66%<br>extract concentrations and<br>no cytotoxicity reactivity at<br>44%, 30%, 20% and 15%<br>extract concentrations under<br>the condition of this test.<br>Cytotoxicity concern was                                                                                                                                                                                                                | Same                             |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                 | addressed by acute<br>systematic toxicity testing.                                                                                                                                                                                                                                                                                                                                                                                                            |                                  |
| ISO 10993-11:2017<br>Acute Systemic toxicity study  | Under the conditions of this<br>study, there was no evidence<br>of systemic toxicity.                                                                                                                                                                                                                                                                                                           | Under the conditions of the<br>study, the subject showed no<br>adverse biological reaction.                                                                                                                                                                                                                                                                                                                                                                   | Same                             |

{7}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China

# 510K Summary

Technological Characteristic Comparison Table:

{8}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base,

Shenze County, Hebei, 050000, China

# 510K Summary

Chemotherapy Permeation and Fentanyl Citrate Comparison Claim:

| Tested Chemotherapy Drug and<br>Concentration | Minimum BDT (Minutes)      |                             | Comparison |
|-----------------------------------------------|----------------------------|-----------------------------|------------|
|                                               | Proposed Device<br>K221269 | Predicate Device<br>K192954 |            |
| Bleomycin Sulfate 15mg/ml (15000 ppm)         | >240                       | N/A                         | Different  |
| Busulfan 6mg/ml (6,000 ppm)                   | >240                       | N/A                         | Different  |
| Carboplatin 10mg/ml (10,000 ppm)              | >240                       | N/A                         | Different  |
| Carmustine (BCNU)3.3 mg/ml (3,300 ppm)        | 11.1                       | 18.2                        | Similar    |
| Chloroquine 50mg/ml (50,000ppm)               | >240                       | N/A                         | Different  |
| Cisplatin 1mg/ml (1,000 ppm)                  | >240                       | >240                        | Same       |
| Cyclophosphamide 20mg/ml (20,000 ppm)         | >240                       | >240                        | Same       |
| Cyclosporin 100 mg/ml (100,000 ppm)           | >240                       | N/A                         | Different  |
| Cytarabine HCL, 100 mg/ml (100,000 ppm)       | >240                       | N/A                         | Different  |
| Dacarbazine 10 mg/ml (10,000 ppm)             | >240                       | >240                        | Same       |
| Daunorubicin HCL, 5 mg/ml (5,000 ppm)         | >240                       | N/A                         | Different  |
| Docetaxel , 10 mg/ml (10,000 ppm)             | >240                       | N/A                         | Different  |
| Doxorubicin HCL, 2 mg/ml (2,000 ppm)          | >240                       | >240                        | Same       |
| Epirubicin HCL, 2 mg/ml (2,000 ppm)           | >240                       | N/A                         | Different  |
| Etoposide, 20 mg/ml (20,000 ppm)              | >240                       | >240                        | Same       |
| Fludarabine, 25 mg/ml (25,000 ppm)            | >240                       | N/A                         | Different  |
| Fluorouracil, 50mg/ml (50,000ppm)             | >240                       | >240                        | Same       |
| Gemcitabine, 38mg/ml (38,000ppm)              | >240                       | N/A                         | Different  |
| Idarubicin HCL, 1mg/ml (1,000ppm)             | >240                       | N/A                         | Different  |
| Ifosfamide, 50mg/ml (50,000ppm)               | >240                       | N/A                         | Different  |
| Irinotecan, 20mg/ml (20,000ppm)               | >240                       | N/A                         | Different  |
| Mechlorethamine HCI, 1mg/ml (1,000ppm)        | >240                       | N/A                         | Different  |
| Melphalan, 5mg/ml (5,000ppm)                  | >240                       | N/A                         | Different  |
| Methotrexate, 25mg/ml (25,000ppm)             | >240                       | N/A                         | Different  |
| Mitomycin C, 0.5mg/ml (500ppm)                | >240                       | N/A                         | Different  |
| Mitoxantrone HCL, 2mg/ml (2,000ppm)           | >240                       | N/A                         | Different  |
| Oxaliplatin, 5mg/ml (5,000ppm)                | >240                       | N/A                         | Different  |
| Paclitaxel, 6mg/ml (6,000ppm)                 | >240                       | >240                        | Same       |
| Paraplatin, 10mg/ml (10,000ppm)               | >240                       | N/A                         | Different  |
| Retrovir, 10mg/ml (10,000ppm)                 | >240                       | N/A                         | Different  |

{9}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China

### 510K Summary

| Rituximab, 10mg/ml (10,000ppm)           | >240 | N/A  | Different |
|------------------------------------------|------|------|-----------|
| Thio Tepa, 10mg/ml (10,000ppm)           | 21.6 | 57.3 | Similar   |
| Topotecan, 1mg/ml (1,000ppm)             | >240 | N/A  | Different |
| Trisenox, 1mg/ml (1,000ppm)              | >240 | N/A  | Different |
| Velcade, 1mg/ml (1,000ppm)               | >240 | N/A  | Different |
| Vincristine Sulfate, 1mg/ml (1,000ppm)   | >240 | N/A  | Different |
| Fentanyl Citrate Injection (100 mcg/2ml) | >240 | >240 | Same      |

### 7. Summary of Non-Clinical Performance Data

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device met the performance criteria with the following standards:

| Methodology                               | Test Performed                       | Acceptance Criteria                                                                                                       | Results                                        |
|-------------------------------------------|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| ASTM D6319- 19                            | Physical Dimensions<br>Length        | Minimum 230mm for all<br>sizes                                                                                            | Pass                                           |
| ASTM D6319- 19                            | Physical Dimensions<br>Palm Width    | XS: 70±10mm<br>S: 80±10mm<br>M: 95±10mm<br>L:110±10mm<br>XL: 120±10mm<br>XXL: 130±10mm                                    | Pass                                           |
| ASTM D6319- 19                            | Physical Dimensions<br>Thickness     | Finger: 0.05mm (min)<br>Palm: 0.05mm (min)                                                                                | Pass                                           |
| ASTM D6319- 19<br>ASTM D412-16(2021)      | Physical Properties                  | Tensile Strength (Min14<br>MPa) and Elongation<br>(Before Aging 500% and<br>after aging 400%) Min<br>AQL 2.5 (ISO 2859-1) | Pass                                           |
| ASTM D6319- 19<br>ASTM D5151-19           | Water leak test                      | AQL 2.5 (ISO 2859-1)                                                                                                      | Pass                                           |
| ASTM D6319- 19<br>ASTM D6124-06<br>(2017) | Powder Residue                       | Max 2mg/glove                                                                                                             | Pass                                           |
| ASTM D6978-05<br>(2019)                   | Permeation by<br>Chemotherapy Drugs  | Refer the above table 1                                                                                                   | Pass                                           |
| ISO 10993-10:2010                         | Irritation and Skin<br>Sensitization | Skin sensitization and<br>Skin irritation                                                                                 | Is non-sensitization<br>and Non-irritation     |
| ISO 10993-5:2009                          | Cytotoxicity                         | Cytotoxicity reactivity                                                                                                   | showed potential<br>toxicity to L929<br>cells. |
| ISO 10993-11:2017                         | Acute systemic toxicity<br>study     | Subject showed no adverse<br>biological reaction                                                                          | no evidence of<br>systemic toxicity.           |

● ASTM D6319-19, Standard Specification for Nitrile Examination Gloves for Medical Application.

{10}------------------------------------------------

Fuqian Xi Road, West district of Shenze, Industrial Base, Shenze County, Hebei, 050000, China

## 510K Summary

- . ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.
- . ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves
- . ASTM D412-16 (2021) Standard Test Methods for Vulcanized Rubber and Thermoplastic Elastomers-Tension
- . ASTM D6978-05 (Reapproved 2019), Assessment of Reissuance of Medical Gloves to Permeation by Chemotherapy Drugs.
- ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests For Irritation And ● Skin Sensitization.
- . ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
- ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic ● toxicity

### 8. Clinical Performance Data

N/A

### 9. Conclusion:

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

---

**Source:** [https://fda.innolitics.com/device/K221269](https://fda.innolitics.com/device/K221269)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
