← Product Code [QAS](/productcode/QAS) · K221248

# Rapid LVO (K221248)

_Ischemaview, Inc. · QAS · May 31, 2022 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221248

## Device Facts

- **Applicant:** Ischemaview, Inc.
- **Product Code:** [QAS](/productcode/QAS.md)
- **Decision Date:** May 31, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2080
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Rapid LVO is a radiological computer aided triage and notification software indicated for use in the analysis of CTA head images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive ICA or MCA-M1 Large Vessel Occlusion (LVO) findings in head CTA images. Rapid LVO uses a software algorithm to analyze images and highlight cases with suspected LVO on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO findings. Notifications include compressed preview images. These are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of Rapid LVO are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage /prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

## Device Story

Rapid LVO is a radiological computer-aided triage and notification (CADt) software. It processes DICOM-compliant head CTA images to detect high-grade stenosis or occlusion of the intracranial ICA or horizontal segment of the MCA-M1. The device operates on a server or standalone desktop application in parallel to standard-of-care image interpretation. It uses traditional programming algorithms to analyze images; upon detecting suspected LVO, it sends priority notifications to clinicians via email or mobile applications. Notifications include compressed preview images for informational purposes. The device does not alter original images. Clinicians use these notifications to prioritize workflow and triage patients; they must review full, non-compressed images on a diagnostic viewer before making clinical decisions. The system aims to reduce time-to-notification for potential LVO cases, facilitating faster clinical intervention.

## Clinical Evidence

Performance validation used 217 scans (135 positive, 82 negative) from 8 sites, truthed by three expert neuroradiologists (2:3 concurrence). Results: Sensitivity 0.96 (95% CI: 0.91-0.97), Specificity 0.98 (95% CI: 0.93-0.99), PPV 0.98, NPV 0.96, and ROC AUC 0.99. Time-to-notification was 3.18 minutes (95% CI: 3.11-3.25), meeting the ≤3.5-minute goal. Subgroup analyses performed across geography, age, gender, scanner manufacturer, and slice thickness.

## Technological Characteristics

Software-based CADt device; operates on standard off-the-shelf PC/server or virtual platforms (e.g., VMware). DICOM-compliant (NEMA PS 3.1-3.20). Uses traditional programming algorithms for image analysis. Connectivity via PACS, email, and mobile. Software lifecycle follows IEC 62304; risk management per ISO 14971; usability per IEC 62366.

## Regulatory Identification

Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.

## Special Controls

Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

## Predicate Devices

- Rapid LVO ([K200941](/device/K200941.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

iSchemaView Inc. James Rosa SVP Regulatory and Quality 1120 Washington Ave., Ste 200 Golden. Colorado 80401

May 31 ,2022

### Re: K221248

Trade/Device Name: Rapid LVO Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: April 28, 2022 Received: May 2, 2022

Dear James Rosa:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Jessica Lamb, Ph.D. Imaging Software Division of Radiological Imaging Device and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221248

Device Name

Rapid LVO

#### Indications for Use (Describe)

Rapid LVO is a radiological computer aided triage and notification software indicated for use in the analysis of CTA head images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive ICA or MCA-M1 Large Vessel Occlusion (LVO) findings in head CTA images.

Rapid LVO uses a software algorithm to analyze images and highlight cases with suspected LVO on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO findings. Notifications include compressed preview images. These are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of Rapid LVO are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage /prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Type of Use (*Select one or both, as applicable*)

| <div style="display:flex; align-items:center;"><div>☑</div><div>Prescription Use (Part 21 CFR 801 Subpart D)</div></div> |
|--------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><div>☐</div><div>Over-The-Counter Use (21 CFR 801 Subpart C)</div></div>  |

CONTINUE ON A SEPARATE PAGE IF NEEDED.#### This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

### iSchemaView, Inc.'s Rapid LVO

This document contains the 510(k) summary for the iSchemaView Rapid. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).

#### Applicant Name and Address:

| Name:             | iSchemaView, Inc.                                                |
|-------------------|------------------------------------------------------------------|
| Address:          | 1120 Washington Ave<br>Ste. 200<br>Golden, CO 80401              |
| Official Contact: | Jim Rosa<br>Phone: (303) 704-3374<br>Email: rosa@ischemaview.com |

Summary Preparation Date: April 29, 2022

#### Device Name and Classification:

| Trade Name:           | iSchemaView Rapid LVO                                           |
|-----------------------|-----------------------------------------------------------------|
| Common Name:          | Radiological computer aided triage and<br>notification software |
| Classification:       | II                                                              |
| Product Code:         | QAS                                                             |
| Regulation No:        | 21 C.F.R. §892.2080                                             |
| Classification Panel: | Radiology Devices                                               |

#### Predicate Devices:

The iSchemaView Rapid LVO is claimed to be substantially equivalent to iSchemaView's Rapid LVO (K200941)

#### Device Description:

Rapid LVO is a radiological computer-assisted triage and notification software device. The Rapid LVO module is a contrast enhanced CTA module which operates within the integrated Rapid Platform to provide triage and notification of suspected ICA and MCA-M1 Large Vessel Occlusion (LVO) based on the following definitions:

ICA Occlusion: A high-grade stenosis or occlusion of the intracranial portion of the ICA.

MCA-M1 Occlusion: A high-grade stenosis or occlusion of the horizontal segment of the MCA-M1, defined as the segment which extends from the ICA terminus until the vessel has turned upward into the Sylvian fissure. This includes post-bifurcation M1 segments in some patients.

{4}------------------------------------------------

The LVO module uses traditional programming algorithms. The output of the module is a priority notification to clinicians indicating the suspicion of LVO based on positive findings. The Rapid LVO module uses the basic services supplied by the Rapid Platform including DICOM processing, job management, imaging module execution and imaging output including the notification and compressed image.

## Indications for Use:

Rapid LVO is a radiological computer aided triage and notification software indicated for use in the analysis of CTA head images. The device is intended to assist hospital networks and trained radiologists in workflow triage by flagging and communication of suspected positive ICA or MCA-M1 Large Vessel Occlusion (LVO) findings in head CTA images.

Rapid LVO uses a software algorithm to analyze images and highlight cases with suspected LVO on a server or standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO findings. Notifications include compressed preview images. These are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device.

The results of Rapid LVO are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage /prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Contraindications/Exclusions/Cautions:

- Rapid LVO is a triage and notification (CADt) device indicated for workflow . prioritization only, not for diagnostic decision making.
- Excessive patient motion may lead to artifacts that make the scan technically inadequate.
- Images previewed through email and the mobile/web applications are compressed ● and are for informational purposes only and not intended for diagnostic use beyond notification.
- Notified clinicians are responsible for viewing non-compressed images on a . diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests.

# Technological Characteristics:

Rapid LVO does not raise new questions of safety or effectiveness compared to the previously cleared Rapid LVO (K200941). Both devices are radiological computer-aided triage and notification software applications for use with CTA input. There are minor differences in intended use to clarify the definition of the vasculature of interest; however, with the minor change the clinical use for Rapid LVO is the same the predicate with no additional risk. Thus, the Rapid LVOdeviceis substantially equivalent.

{5}------------------------------------------------

| Parameter                            | Rapid LVO (K200941)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Rapid LVO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code                         | QAS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | QAS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Regulation                           | 21 CFR §892.2080                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 21 CFR §892.2080                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Intended Use/<br>Indications for Use | Rapid LVO is a radiological<br>computer aided triage and<br>notification software indicated for<br>use in the analysis of CTA head<br>images. The device is intended to<br>assist hospital networks and trained<br>radiologists in workflow triage by<br>flagging and communication of<br>suspected positive Large Vessel<br>Occlusion (LVO) findings in head<br>CTA images.<br><br>Rapid LVO uses a software<br>algorithm to analyze images and<br>highlight cases with suspected<br>LVO on a server or standalone<br>desktop application in parallel to<br>the ongoing standard of care image<br>interpretation. The user is<br>presented with notifications for<br>cases with suspected LVO findings.<br>Notifications include compressed<br>preview images, that are meant for<br>informational purposes only and<br>not intended for diagnostic use<br>beyond notification. The device<br>does not alter the original medical<br>image and is not intended to be<br>used as a diagnostic device. | Rapid LVO is a radiological<br>computer aided triage and<br>notification software indicated for<br>use in the analysis of CTA head<br>images. The device is intended to<br>assist hospital networks and trained<br>radiologists in workflow triage by<br>flagging and communication of<br>suspected positive ICA or MCA-M1<br>Large Vessel Occlusion (LVO)<br>findings in head CTA images.<br><br>Rapid LVO uses a software<br>algorithm to analyze images and<br>highlight cases with suspected LVO<br>on a server or standalone desktop<br>application in parallel to the ongoing<br>standard of care image<br>interpretation. The user is presented<br>with notifications for cases with<br>suspected LVO findings.<br>Notifications include compressed<br>preview images. These are meant<br>for informational purposes only and<br>are not intended for diagnostic use<br>beyond notification. The device<br>does not alter the original medical<br>image and is not intended to be used<br>as a diagnostic device. |
|                                      | The results of Rapid LVO are<br>intended to be used in conjunction<br>with other patient information and<br>based on professional judgment, to<br>assist with triage /prioritization of<br>medical images. Notified clinicians<br>are responsible for viewing full<br>images per the standard of care.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | The results of Rapid LVO are<br>intended to be used in conjunction<br>with other patient information and<br>based on professional judgment, to<br>assist with triage /prioritization of<br>medical images. Notified clinicians<br>are responsible for viewing full<br>images per the standard of care.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| PACS Functionality                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Stroke/Head                          | Intracranial Stroke/Head                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Region of Interest                   | ICA and MCA-M1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Parameter                            | Rapid LVO (K200941)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Rapid LVO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Code                         | QAS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | QAS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Regulation                           | 21 CFR §892.2080                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 21 CFR §892.2080                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Computer Platform                    | Standard off-the-shelf PC<br>workstation/server<br>Virtual platform such as VMware                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| DICOM<br>Compliance                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Imaging Type                         | CT Angiography                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Data Acquisition                     | Acquires medical image data<br>from DICOM compliant imaging<br>devices and modalities                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Technical Implementation             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| SaMD                                 | Traditional Algorithms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Notification/Workflow                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Pathways                             | PACS, email, mobile                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Preview/Prioritizati<br>on           | Notification Message of Suspected<br>LVO.<br>Presentation of a compressed preview<br>of the study for initial assessment not<br>meant for diagnostic purposes.<br>The device operates in parallel with<br>the standard of care, which remains.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| SoC Workflow                         | In parallel to the SoC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Original Image                       | No Alteration                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Primary Users                        | Clinician                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

The following table summarizes and compares data on the Rapid LVO (K200941) to the Rapid LVO that is the subject of this Special 510(k) submission.

{6}------------------------------------------------

# Performance Standards:

Rapid has been developed in conformance with the following standards, as applicable:

| ISO 14971:2019     | Application of Risk Management to Medical Devices       |
|--------------------|---------------------------------------------------------|
| IEC 62304:2015     | Medical device software – Software lifecycle processes  |
| IEC 62366:2015     | Application of Usability Engineering to Medical Devices |
| NEMA PS 3.1 - 3.20 | Digital Imaging and Communications in Medicine (DICOM)  |

# Performance Data:

Rapid complies with DICOM (Digital Imaging and Communications in Medicine) - Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20.

{7}------------------------------------------------

Additionally, iSchemaView conducted extensive performance validation testing and software verification and validation testing of the Rapid system. This performance validation testing demonstrated that the Rapid system provides accurate representation of key processing parameters under a range of clinically relevant parameters and perturbations associated with the intended use of the software. Software performance, validation and verification testing demonstrated that the Rapid system met all design requirements and specifications.

iSchemaView performed standalone performance in accordance with the 892.2080 special controls to show acceptance of the clinical performance of the Rapid LVO module. The Standalone Performance exceeded the 80% Goal using the lower bound of the 95% Confidence Interval for Sensitivity (Se) and Specificity (Sp). The observed results are Se: 0.96 (95% CI: 0.91 - 0.97) and Sp: 0.98 (95% CI: 0.93-0.99). Additionally, PPV = 0.98 and NPV = 0.96. The RoC AUC is 0.99:

Image /page/7/Figure/2 description: The image is a plot of sensitivity versus 1-specificity, also known as a receiver operating characteristic (ROC) curve. The x-axis represents 1-specificity or the false positive rate, ranging from 0.00 to 1.00. The y-axis represents sensitivity or the true positive rate, also ranging from 0.00 to 1.00. The ROC curve starts at approximately (0, 0.85), rises sharply to nearly 1.00, and then plateaus, indicating high performance of the model being evaluated.

In addition, an analysis on time to notification using Rapid LVO to notify of suspicion/nonsuspicion as 3.18 min (95% CI: 3.11 - 3.25) was achieved meeting the goal of ≤ 3.5min established by the predicate. The Rapid LVO time-to-notification includes the time to get the DICOM exam, de-identify it (if required), analyze and send a notification to the attending physician(s) email and mobile. The time to notification consists primarily of the processing time, as the notification time via electronic transmission is considered in seconds via email and mobile.

Demographic. Scanner and Performance data:

| LVO Performance by Geography |         |          |              |              |
|------------------------------|---------|----------|--------------|--------------|
| Location                     | Measure | Estimate | Lower 95% CI | Upper 95% CI |
| US                           | Se      | 0.963    | 0.897        | 0.987        |
| US                           | Sp      | 0.963    | 0.875        | 0.990        |
| OUS                          | Se      | 0.964    | 0.823        | 0.994        |
| OUS                          | Sp      | 1.000    | 0.934        | 1.000        |

Demographics:

{8}------------------------------------------------

| Location | Measure | Estimate | Lower 95% CI | Upper 95% CI |
|----------|---------|----------|--------------|--------------|
| Female   | Se      | 0.981    | 0.902        | 0.997        |
|          | Sp      | 1.000    | 0.923        | 1.000        |
| Male     | Se      | 0.963    | 0.875        | 0.990        |
|          | Sp      | 0.968    | 0.890        | 0.991        |

| Age Group Performance |         |          |              |              |
|-----------------------|---------|----------|--------------|--------------|
| Age Groups            | Measure | Estimate | Lower 95% CI | Upper 95% CI |
| 20-39                 | Se      | 1.000    | 0.646        | 1.000        |
|                       | Sp      | 1.000    | 0.722        | 1.000        |
| 40-59                 | Se      | 1.000    | 0.883        | 1.000        |
|                       | Sp      | 0.974    | 0.868        | 0.995        |
| 60+                   | Se      | 0.959    | 0.886        | 0.986        |
|                       | Sp      | 0.983    | 0.909        | 0.997        |

Scanner/Manufacturer:

| Performance by Scanner Manufacturer |         |          |              |              |
|-------------------------------------|---------|----------|--------------|--------------|
| Brand                               | Measure | Estimate | Lower 95% CI | Upper 95% CI |
| GE                                  | Se      | 0.969    | 0.893        | 0.991        |
|                                     | Sp      | 0.970    | 0.847        | 0.995        |
| SIEMENS                             | Se      | 1.000    | 0.785        | 1.000        |
|                                     | Sp      | 0.978    | 0.887        | 0.996        |
| TOSHIBA                             | Se      | 0.929    | 0.774        | 0.980        |
|                                     | Sp      | 1.000    | 0.879        | 1.000        |

| LVO Performance by Slice Thickness |         |          |              |              |
|------------------------------------|---------|----------|--------------|--------------|
| Thickness(mm)                      | Measure | Estimate | Lower 95% CI | Upper 95% CI |
| ≤ 0.65                             | Se      | 0.981    | 0.899        | 0.997        |
| ≤ 0.65                             | Sp      | 0.959    | 0.863        | 0.989        |
| > 0.65                             | Se      | 0.965    | 0.881        | 0.990        |
| > 0.65                             | Sp      | 1.000    | 0.939        | 1.000        |

### Data:

The validation of Rapid LVO included 217 (Pos:135, Neg: 82) scans from 8 sites/studies. The data was truthed using three expert neuroradiologists with a 2:3 concurrence.

{9}------------------------------------------------

| Reference LVO by Site |              |    |    |     |
|-----------------------|--------------|----|----|-----|
| Location              | Study/Site   | 1  | 0  | All |
| US                    | Study/Site 1 | 65 | 3  | 68  |
|                       | Study/Site 2 | 0  | 29 | 29  |
|                       | Study/Site 3 | 16 | 0  | 16  |
|                       | Study/Site 4 | 0  | 15 | 15  |
|                       | Study/Site 5 | 0  | 7  | 7   |
| OUS                   | Study/Site 6 | 1  | 27 | 28  |
|                       | Study/Site 7 | 11 | 26 | 37  |
|                       | Study/Site 8 | 16 | 1  | 17  |

### Prescriptive Statement:

Caution: Federal law restricts this device to sale by or on the order of a physician.

### Safety & Effectiveness:

Rapid has been designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with EN ISO 14971:2019 (risk management). The Rapid LVO (1.0) device performance has been validated through the use of clinical data.

### Conclusion:

In conclusion, the iSchemaView Rapid LVO is substantially equivalent in intended use, technological characteristics, safety, and performance characteristics to the legally marketed predicate device, Rapid LVO (K200941).

---

**Source:** [https://fda.innolitics.com/device/K221248](https://fda.innolitics.com/device/K221248)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
