IPL Hair Removal (Model: SL-B080, SL-B126, SL-B136)

K221246 · Shenzhen Semlamp Intelligent Technology Co., Ltd. · OHT · Jul 7, 2022 · General, Plastic Surgery

Device Facts

Record IDK221246
Device NameIPL Hair Removal (Model: SL-B080, SL-B126, SL-B136)
ApplicantShenzhen Semlamp Intelligent Technology Co., Ltd.
Product CodeOHT · General, Plastic Surgery
Decision DateJul 7, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The IPL Hair Removal is an over the-counter device intended for removal of unwanted body and/or facial hair.

Device Story

Personal light-based hair reduction device; uses Intense Pulsed Light (IPL) technology via Xenon flashlamp; intended for OTC home use by consumers. Device requires external power adapter; features skin contact sensor to prevent light pulse emission unless fully applied to skin. User operates device via buttons on main unit (Power, Mode/Flash, or Auto/Manual buttons). Output is optical radiation (510-1100nm) delivered directly to tissue to reduce hair growth. Benefits include hair reduction without cutting or pulling, with minimal pain. Safety features include skin contact sensor and manual finger switch activation. User controls treatment intensity via gear selection. Device is for single-person use.

Clinical Evidence

Bench testing only. No clinical data provided. Safety and performance validated via biocompatibility (ISO 10993-5, ISO 10993-10), electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and usability (IEC 60601-1-6).

Technological Characteristics

Xenon arc flashlamp source; 510-1100nm wavelength; 9-12ms pulse duration; 2.2-4.0 J/cm² energy density; 3.0cm² spot size. Powered by external adapter. Includes skin contact sensor. Biocompatibility per ISO 10993. Software moderate level of concern. Standards: IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-57, 60601-1-6, IEC 62471.

Indications for Use

Indicated for removal of unwanted body and/or facial hair in over-the-counter (OTC) settings.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below. July 7, 2022 Shenzhen Semlamp Intelligent Technology Co., Ltd % Cassie Lee Manger Guangzhou GLOMED Biological Technology Co., Ltd. 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District Guangzhou, Guangdong 510530 China Re: K221246 Trade/Device Name: IPL Hair Removal (Model: SL-B080, SL-B126, SL-B136) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHT Dated: April 20, 2022 Received: May 2, 2022 Dear Cassie Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221246 Device Name IPL Hair Removal (Model: SL-B080, SL-B126, SL-B136) Indications for Use (Describe) The IPL Hair Removal is an over the-counter device intended for removal of unwanted body and/or facial hair. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary for K221246 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 878.4810. #### 1. Submitter's Information Company Name: Shenzhen Semlamp Intelligent Technology Co., Ltd Address: Room 915, Lankun Building, Minkang Road 213, Zhangkeng Community, Minzhi Avenue, Longhua District, Shenzhen, Guangdong, China Contact Person (including title): Qi Wenjun (CEO) Tel: +86 13428733261 Fax: +86 0755-23490305 E-mail: 2320792967@qq.com ## Application Correspondent: Contact Person: Ms. Cassie Lee Guangzhou GLOMED Biological Technology Co., Ltd. Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China Tel: +86 20 8266 2446 Email: regulatory@glomed-info.com ## 2. Date of the summary prepared: July 6, 2022 #### 3. Subject Device Information Classification Name: Light Based Over-The-Counter Hair Removal Trade Name: IPL Hair Removal Model Name: SL-B080, SL-B126, SL-B136 Review Panel: General & Plastic Surgery Product Code: OHT Regulation Number: 21 CFR 878.4810 Regulatory Class: II #### 4. Predicate Device Information Predicate Device 1 (Primary Predicate Device) Sponsor: Shenzhen Bosidin Technology Co., Ltd. Trade Name: IPL Home Use Hair Removal Device, model: D-1128, D-1103, D-1119, D-1129, D-1130 {4}------------------------------------------------ Classification Name: Light Based Over-The-Counter Hair Removal 510(K) Number: K192432 Review Panel: General & Plastic Surgery Product Code: OHT Regulation Number: 21 CFR 878.4810 Regulation Class: 2 ## Predicate Device 2 (Secondary Predicate Device) Sponsor: Shenzhen Mismon Technology Co., Ltd. Trade Name: Home Use IPL Beauty Device Model(s): MS-208B, MS-212B, MS-216B, MS-218B, MELSYA-M6, MELSYA-M7 Classification Name: Laser surgical instrument for use in general and plastic surgery and in dermatology 510(K) Number: K210311 Review Panel: General & Plastic Surgery Product Code: OHT Regulation Number: 21 CFR 878.4810 Regulation Class: 2 ## 5. Device Description The IPL Hair Removal is a personal, light-based, hair reduction device intended to be sold over-thecounter directly to the end user. The device provides hair reduction using IPL technology. The device provides hair reduction using Intense Pulsed Light (IPL) technology. It works below the skin's surface and does not involve any cutting or pulling, reducing hair growth with minimal pain. The device is only powered by the external power adapter and its IPL emission activation is by finger switch. The device mainly contains a Xenon lamp and a skin sensor to detect appropriate skin contact. If the device is not properly applied to the treatment area (in full contact with the skin), the device cannot emit the treatment light pulses. The device is for single-person use only. The IPL Hair Removal includes SL-B080, SL-B126, and SL-B136 three models. The difference of all models is mainly the exception of appearance and number of buttons which do not affect the intended use. For model SL-B080, it consists of an IPL beauty device main body and power adapter two parts, and one non-removable lamp head (light-emitting treatment window) located in the main body which is the source of optical radiation, namely a Xenon flashlamp. In addition, the device is equipped with a manual shaver, sunglasses and replacement lampshade in addition to the IPL beauty device main body and adapter. The user can control the device effectively by the buttons on the main unit. There are 2 operation buttons: Power button (turning on/off the device and selecting the level of gears) and Mode/Flash button. {5}------------------------------------------------ For model SL-B126 and SL-B136, consisting of IPL beauty device main body and power adapter two parts, and one non-removable lamp head (light-emitting treatment window) located in the main body which is the source of optical radiation, namely a Xenon flashlamp. In addition, the device is equipped with a manual shaver and sunglasses in addition to the IPL beauty device main body and adapter. The user can control the device effectively by the buttons on the main unit. There are 3 operation buttons: Power button (turning on/off the device), Auto/Manual button and Flash button. #### 6. Intended Use / Indications for Use The IPL Hair Removal is an over the-counter device intended for removal of unwanted body and/or facial hair. #### 7. Test Summary IPL Hair Removal (Model: SL-B080, SL-B126, SL-B136) has been evaluated the safety and performance by lab bench testing as following: ## 1) Biocompatibility Testing The biocompatibility evaluation for the body-contacting components of the IPL Home Use Hair Removal Device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices –Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on June 16, 2016", as recognized by FDA. The battery of testing was performed to, and passed, including: > ISO 10993-5:2009/(R)2014, Biological Evaluation of Medical Devices –Part 5: Tests for In Vitro Cytotoxicity > ISO 10993-10:2010/(R)2014, Biological Evaluation of Medical Devices –Part 10: Tests for Irritation and Skin Sensitization ## 2) Electrical Safety and Eye Safety Electrical safety and Eye safety testing was performed to, and passed, the following standards: > IEC 60601-1-2 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance -Collateral standard: electromagnetic compatibility > IEC 60601-1 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance > IEC 60601-1-11 Medical Electrical Equipment –Part 1: General Requirements for Basic Safety and Essential Performance -Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment > IEC 60601-2-57 Medical electrical equipment -Part 2-57: Particular requirements for the basic safety and essential performance of non-laser source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use {6}------------------------------------------------ > IEC 60601-1-6 Edition 3.1 2013-10, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability ## 3) Eye Safety > IEC 62471 Photobiological safety of lamps and lamp systems #### 4) Software Verification and Validation Software documentation consistent with moderate level of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels. #### 8. Comparison to predicate device and conclusion Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness. | | | Predicate Device 1 | Predicate Device 2 | | |--------------------------------------------|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------| | Elements of<br>Comparison | Subject Device | (Primary Predicate<br>Device) | (Secondary Predicate<br>Device) | Remark | | Company | Shenzhen Semlamp<br>Intelligent Technology<br>Co., Ltd | Shenzhen Bosidin<br>Technology Co.,Ltd. | Shenzhen Mismon<br>Technology Co., Ltd. | -- | | Trade Name | IPL Hair Removal, Model:<br>SL-B080, SL-B126, SL-<br>B136 | IPL Home Use Hair<br>Removal Device | Home Use IPL Beauty<br>Device, model: MS-<br>208B, MS-212B, MS-<br>216B, MS-218B,<br>MELSYA-M6,<br>MELSYA-M7 | -- | | Classification<br>Name | Laser surgical instrument<br>for use in general and<br>plastic surgery and in<br>dermatology | Laser surgical instrument<br>for use in general and<br>plastic surgery and in<br>dermatology | Laser surgical<br>instrument for use in<br>general and plastic<br>surgery and in<br>dermatology | -- | | 510(k) Number | K221246 | K192432 | K210311 | -- | | Product Code | OHT | OHT | OHT | Same | | Intended Use / | The IPL Hair Removal is | IPL Home Use Hair | The Home Use IPL | Same | | Elements of<br>Comparison | Subject Device | Predicate Device 1<br>(Primary Predicate<br>Device) | Predicate Device 2<br>(Secondary Predicate<br>Device) | Remark | | Indications for<br>Use | an over the-counter<br>device intended for<br>removal of unwanted<br>body and/or facial hair. | Removal Device is an<br>over the-counter device<br>intended for removal of<br>unwanted body and/or<br>facial hair. | Beauty Device is an<br>over the-counter device<br>intended for removal of<br>unwanted hair such as<br>but not limited to small<br>areas such as<br>underarm and facial<br>hair below the chin line<br>and large areas such<br>as legs. | | | Energy medium | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Xenon Arc Flashlamp | Same | | Wavelength<br>Range | 510-1100nm | Regular window: 510nm-<br>1100nm<br>Filter window: 600-<br>1100nm | Not publicly available | Same | | Pulse duration | 9-12 milliseconds | 7.5~14ms | 9-12 milliseconds | Same | | Energy density | 2.2-4.0J/cm² | 2.0~4.0J/cm² (applicable<br>for model D-1128,<br>D1119, D-1129, D1130)<br>2.5~4.5J/cm² (applicable<br>for model D-1103) | 2.6~5.1J/cm² | Similar<br>Note | | Spot size (size<br>of treatment<br>window) | 3.0cm² | Regular window: 4.5cm²,<br>2.0cm², 3.0cm²<br>Filter window: 2.5cm² | MS208B/MS212B/MEL<br>SYAM6: 3.6cm² MS-<br>216B/MS218B/MELSY<br>A-M7: 3.0cm² | Same | | Delivery device | Direct illumination to<br>tissue | Direct illumination to<br>tissue | Direct illumination to<br>tissue | Same | | Pulsing control | Finger switch | Finger switch | Finger switch | Same | | Location for use | OTC | OTC | OTC | Same | | Elements of<br>Comparison | Subject Device | Predicate Device 1<br>(Primary Predicate<br>Device) | Predicate Device 2<br>(Secondary Predicate<br>Device) | Remark | | Safety and<br>EMC | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57<br>IEC 62471<br>IEC 60601-1-6 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 60601-2-57<br>IEC 62471 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 60601-1-11<br>IEC 60601-2-57<br>IEC 62471 | Same | | Biocompatibility | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | Same | {7}------------------------------------------------ {8}------------------------------------------------ ## Note: Although the "Energy density" is a similar from the predicate devices, but the value of the predicate device can override the value of the subject device, also they all complied with the IEC 60601-1-2, IEC 60601-2-57 and IEC 62471 safety standards' requirements. So, these slight differences will not raise any safety or effectiveness issue. ## Final Conclusion: The subject device is as safe, as effective, and performs as well as or better than the legally marketed predicated device K192432 and K210311.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%