The DT-703 is a portable dental X-ray system that captures radiographic images for dental diagnosis using intraoral imaging sensors. Only trained and qualified dental practitioner or radiologist shall use DT-703 to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children.
Device Story
Portable handheld dental X-ray system; generates X-rays for dental diagnosis. Input: manual operator settings (exposure time). Transformation: X-ray tube (70 kV, 3 mA) generates radiation directed at intraoral image receptors (not included). Output: radiographic images captured by external sensors. Used in dental clinics by dentists/radiologists. Features: rechargeable Li-ion polymer battery, backscatter shielding glass, user interface, optional remote exposure switch. Benefits: enables portable dental imaging for diagnosis of teeth/jaw diseases.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing and adherence to international standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, and IEC 60601-2-65.
Indicated for use by trained dental practitioners or radiologists for diagnostic dental X-ray imaging of teeth, jaws, and oral structures in adult and pediatric patients using intraoral image receptors.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
June 27, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
ECOTRON Co., Ltd % Mr. Dave Kim Regulatory Affairs Mtech Group 7505 Fannin St, Suite 610 HOUSTON TX 77054
Re: K221233
Trade/Device Name: DT-703 Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: EHD Dated: April 20, 2022 Received: April 29, 2022
## Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221233
Device Name DT-703
### Indications for Use (Describe)
The DT-703 is a portable dental X-ray system that captures radiographic images for dental diagnosis using intraoral imaging sensors. Only trained and qualified dental practitioner or radiologist shall use DT-703 to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;"> <span style="font-family:Wingdings;">ü</span> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt;">□ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for eco & electronics eddTRON. The logo is composed of a blue square with the letters "edd" in white, followed by the text "eco & electronics" in black. Below this is the word "TRON" in gray. The logo is simple and modern, and it is likely used to represent a company that specializes in electronics.
# 510(k) Summary K221233
This summary of 510(k) is being submitted in accordance with requirements of SMDA 1990 and 21 CFR Part 807.92.
Date 510(k)submitted: 4/25/2022
| 1. | Submitter Information: | | | | | | | | | | | | | | | | | | | | | | | | |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|------------------|----------|-------------------------------------------|--|---------------------------------|--|--------------------------------|-----------------|----------------------------|-------------------|----------------|--------|-------------------|-------------------------|---------------|--|-------------|----------|----------------------------------------------|------|--------------|--------|-----------------------|
| | <table><tr><td>Sumitter's Name:</td><td>ECOTRON Co., Ltd</td></tr><tr><td>Address:</td><td>404, 504, 505Ho, Hanshin IT Tower II, 47,</td></tr><tr><td></td><td>Digital-ro 9-gil, Geumcheon-Gu,</td></tr><tr><td></td><td>Seoul, Republic of Korea 08511</td></tr><tr><td>Contact Person:</td><td>Sang Bong Lee, R&D Manager</td></tr><tr><td>Telephone Number:</td><td>(02)-2025-3760</td></tr><tr><td>Email:</td><td>lsb@ecotron.co.kr</td></tr><tr><td>Official Correspondent:</td><td>Dave Kim, MBA</td></tr><tr><td></td><td>Mtech Group</td></tr><tr><td>Address:</td><td>7505 Fannin St. Suite 610, Houston, TX 77054</td></tr><tr><td>Tel:</td><td>713-467-2607</td></tr><tr><td>Email:</td><td>davekim@mtech-inc.net</td></tr></table> | Sumitter's Name: | ECOTRON Co., Ltd | Address: | 404, 504, 505Ho, Hanshin IT Tower II, 47, | | Digital-ro 9-gil, Geumcheon-Gu, | | Seoul, Republic of Korea 08511 | Contact Person: | Sang Bong Lee, R&D Manager | Telephone Number: | (02)-2025-3760 | Email: | lsb@ecotron.co.kr | Official Correspondent: | Dave Kim, MBA | | Mtech Group | Address: | 7505 Fannin St. Suite 610, Houston, TX 77054 | Tel: | 713-467-2607 | Email: | davekim@mtech-inc.net |
| Sumitter's Name: | ECOTRON Co., Ltd | | | | | | | | | | | | | | | | | | | | | | | | |
| Address: | 404, 504, 505Ho, Hanshin IT Tower II, 47, | | | | | | | | | | | | | | | | | | | | | | | | |
| | Digital-ro 9-gil, Geumcheon-Gu, | | | | | | | | | | | | | | | | | | | | | | | | |
| | Seoul, Republic of Korea 08511 | | | | | | | | | | | | | | | | | | | | | | | | |
| Contact Person: | Sang Bong Lee, R&D Manager | | | | | | | | | | | | | | | | | | | | | | | | |
| Telephone Number: | (02)-2025-3760 | | | | | | | | | | | | | | | | | | | | | | | | |
| Email: | lsb@ecotron.co.kr | | | | | | | | | | | | | | | | | | | | | | | | |
| Official Correspondent: | Dave Kim, MBA | | | | | | | | | | | | | | | | | | | | | | | | |
| | Mtech Group | | | | | | | | | | | | | | | | | | | | | | | | |
| Address: | 7505 Fannin St. Suite 610, Houston, TX 77054 | | | | | | | | | | | | | | | | | | | | | | | | |
| Tel: | 713-467-2607 | | | | | | | | | | | | | | | | | | | | | | | | |
| Email: | davekim@mtech-inc.net | | | | | | | | | | | | | | | | | | | | | | | | |
| 2. | Device Name: | | | | | | | | | | | | | | | | | | | | | | | | |
| Device Name: |
|-------------------|
| Proprietary Name: |
| Manufacturer Name: | |
|----------------------|--|
| Common Name: | |
| Classification Name: | |
| CFR Number: | |
| Device Class: | |
| Product Code: | |
DT-703 ECOTRON Handheld X-ray System Extraoral Source X-ray System 872.1800 II EHD
#### 3. Predicate Device:
Proprietary Name: Manufacturer Name: Common Name: Classification Name: CFR Number: Device Class: Product Code:
KaVo NOMAD Pro 2 Handheld X-ray System (K173319) Aribex Handheld X-ray System Extraoral Source X-ray System 872.1800 II EHD
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "eco & electronics edotron". The logo has a blue square design on the left side, followed by the text "eco & electronics" in black font. Below this text is the word "edoTRON" in a larger, gray font. The logo appears to be for an electronics company with an eco-friendly focus.
#### 4. Description of Device:
DT-703 is a handheld X-ray system powered by a rechargeable Li-ion polymer battery pack. DT-703 is an extra source X-ray generating device that is mainly designed for dental examination (on teeth, etc.). DT-703 is composed of X-ray generator with X-ray tube including device controller, power controller, user interface, beam limiting part, backscatter shielding glass, and optional remote exposure switch. DT-703 is designed to diagnose teeth and jaw through X-ray irradiation using an intra-oral image receptor. DT-703 is controlled by software (firmware) and the software level of concern is Moderate. The x-ray image detector (an integral part of a complete dental diagnostic system) is not part of the DT-703 X-ray device.
Optional parts available:
- Remote Exposure Switch
- . Backscatter Shield
- . Rectangular Cover (FOV 2x3)
- . Rectangular Cover (FOV 3x4)
#### 5. Indications for Use:
DT-703 is a portable dental X-ray system that captures radiographic images for dental diagnosis using intraoral imaging sensors. Only trained and qualified dental practitioner or radiologist shall use DT-703 to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children.
### 6. Description of Substantial Equivalence: Technological Characteristics:
DT-703 has similar design components andoperating features as the predicate device, KaVo NOMAD Pro 2 Handheld X-ray System (K173319).
The handheld device features a main unit (tube head), rechargeable battery (handset) and charger. The main components of the tube head including X-ray tube (70 kV, 3 mA), internal shielding and external backscatter shielding are similar to the predicate device. The functionality of the user interface is also similar to the predicate device.
The power is supplied by a rechargeable Lithium Ion battery core pack built into a handset. The design for DT-703 is equipped with a 22.2 VDC, 1.0 Ahr battery core pack compared to the predicate device (KaVo NOMAD Pro 2 Handheld X-ray System - K173319) design of 21.6 VDC, 1.7 Ahr. The DT-703 battery core pack is compliant with IEC 62133-2.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "eco & electronics eddotron". The logo has a blue square design on the left side of the logo. The text "eco & electronics" is in black and is above the word "eddotron", which is in gray.
Testing has been completed on basic safety and essential performance and the device complies with IEC 60601-1; IEC 60601-1-2 (Ed. 4.1); IEC 60601-1-3, and IEC 60601-2-65.
| | DT-703 Handheld<br>Dental X-ray System<br>(Subject Device) | KaVo NOMAD Pro2<br>Handheld X-ray System<br>(K173319) |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| INDICATIONS<br>FOR USE: | DT-703 is a portable dental X-ray system that captures radiographic images for dental diagnosis using intraoral imaging sensors. Only trained and qualified dental practitioner or radiologist shall use DT-703 to diagnose and treat diseases related to the teeth, jaws, and/or other oral structures in adults and children. | The KaVo NOMAD Pro 2 Handheld X-ray System is indicated for use only by a trained and qualified dentist or dental technician for both adult and pediatric subjects as an extraoral diagnostic dental X-ray source to produce X-ray images using intraoral image receptors. |
| MECHANICAL: | | |
| Size: Body (mm/inch) | 259 x 277 x 165 mm<br>(10"L x 11"H x 6.5"W) | 11"L x 10.5"H x 5.5"W |
| Weight | 1.5 kg (3.3 lbs) | 6.0 lbs. |
| Source to skin distance | 20 cm | 21 cm |
| Cone diameter | 6 cm | 6 cm |
| User Interface | Same, plus several user-selectable preset times. | |
| Backscatter radiation protection | 165mm (6.5") dia. Pb-filled acrylic plastic, backscatter shield | 6.75" dia. Pb-filled acrylic plastic scatter shield |
| Exposure switch | Trigger located on handset / optional remote switch | Trigger located on handset |
| Tubehead mounting | Handheld | Handheld |
| ELECTRICAL: | | |
| Tubehead | CEI OX/70-3 | Cyprus PSOC CY8C29866 |
| Energy Source | Rechargeable 22.2 V Li-ion polymer battery pack | Rechargeable 21.6 V DC Li-ion battery core pack |
| Capacity | 1.0 A-hr | 1.7 A-hr |
| Exposure Time | 0.02 – 0.5 seconds in 0.01 increments | 0.02 – 1.0 seconds in 0.01 increments |
| mA | 3 mA fixed | 2.5 mA fixed |
| kVp | 70 kVp fixed | 60 kVp fixed |
{6}------------------------------------------------
| Image: edotron logo | | |
|--------------------------------|-----------------------------|--------------------------------------------|
| Electrical Safety<br>Standards | IEC 60601-1:2005, AMD1:2012 | AAMI ES60601-1:2005/(R)2012 And<br>A1:2012 |
| EMI Standards | IEC60601-1-2: Ed. 4.1 | IEC60601-1-2 Ed. 4 |
| X-RAY<br>PERFORMANCE: | | |
Similar technological characteristics for the proposed device DT-703 and the predicate device KaVo Nomad Pro2 seem to indicate substantial equivalence.
### Non-Clinical Test Data:
Testing was performed in accordance with the following international standards:
IEC 60601-1:2005, AMD 1: 2012
IEC 60601-2-65 Edition 1.0 2012-09 Medical Electrical Equipment - Part 2-65: Particular Requirements for the Basic Safety And Essential Performance of Dental Intra-Oral X-Ray Equipment
IEC 60601-1-2 Edition 4.1 2014-02 Medical Electrical Equipment – Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
IEC 60601-1-3 Edition 2.1 2013-04 Medical Electrical Equipment - Part 1-3: General Requirements for Basic Safety And Essential Performance – Collateral Standard: Radiation Protection in Diagnostic X-ray Equipment
ISO 14971 Third Edition 3 2019 Medical Devices – Application Of Risk ManagementTo Medical Devices
IEC 62133 Edition 2.0 2012-12 Secondary Cells And Batteries Containing Alkaline Or Other Non-Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From them, For Use In Portable Applications [Including: Corrigendum 1 (2013)]
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for "eco & electronics edotron". The logo has a blue icon on the left, followed by the text "eco & electronics" in a smaller font. Below the text is the word "edoTRON" in a larger, gray font. The logo is simple and modern, with a focus on the company's name and its association with eco-friendly electronics.
### FDA Guidance Documents Utilized:
Radiation Safety Considerations for X-ray Equipment Designed for Hand-Held Use, issued December 24, 2008
Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued May 11, 2005
Pediatric Information for X-ray Imaging Device Premarket Notifications, issued May 10, 2012
### Clinical Performance Data:
Clinical data is not needed to characterize performance and establish substantial equivalence. The non-clinical test data characterize all performance aspects of the device based on wellestablished scientific and engineering principles. Clinical testing has not been conducted on this product.
### Conclusion as to Substantial Equivalence:
Based on a comparison of intended use, indications, technological characteristics, principleof operation, features and performance data, the sponsor believes that the subject device, DT-703 is deemed to be substantially equivalent to the predicate device, KaVo NOMAD Pro 2.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.