SPY Laparoscope

K221217 · Stryker · GCJ · Jun 17, 2022 · Gastroenterology, Urology

Device Facts

Record IDK221217
Device NameSPY Laparoscope
ApplicantStryker
Product CodeGCJ · Gastroenterology, Urology
Decision DateJun 17, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesPediatric

Indications for Use

The SPY Laparoscope is intended to be used for gynecological and general procedures that clinicians deem appropriate for the adult or pediatric patient aged one month or older, when the dimensions of the SPY Laparoscope are appropriate for the patient size and anatomy.

Device Story

SPY Laparoscope is a rigid endoscope for visualization during minimally invasive procedures; transmits visible and near-infrared light to illuminate anatomy; forms and relays images of surgical site to external camera system for processing and display; used by clinicians in clinical settings; provides visual feedback to assist in examination, diagnosis, or therapy; benefits patient through minimally invasive access.

Clinical Evidence

Bench testing only. Performance verified via optical and mechanical testing per ISO 8600-1:2015, electrical safety (IEC 60601-1, IEC 60601-2-18), biocompatibility (ISO 10993 series), and validation of cleaning, disinfection, and sterilization protocols (AAMI/ISO standards). Design validation performed per IEC 62366-1:2020.

Technological Characteristics

Rigid endoscope; rod lens optical system; visible and near-infrared illumination. Materials: Stainless steel, epoxy, optical glass (sapphire), glass fibers. Dimensions: 5.4mm/10mm OD; 30cm/33cm/45cm length; 0°/30°/45° DOV. Reusable; manual/automated cleaning/disinfection; sterilization via autoclave, Steris VPRO, or Sterrad. Connectivity: Connects to external camera system.

Indications for Use

Indicated for gynecological and general endoscopic procedures in adult and pediatric patients aged one month or older, provided device dimensions are appropriate for patient size and anatomy.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 17, 2022 Stryker Christie Samsa Manager, Regulatory Affairs 5900 Optical Court San Jose, California 95138 Re: K221217 Trade/Device Name: SPY Laparoscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ, FCW Dated: April 25, 2022 Received: April 27, 2022 Dear Christie Samsa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221217 Device Name SPY Laparoscope Indications for Use (Describe) The SPY Laparoscope is intended to be used for gynecological and general procedures that clinicians deem appropriate for the adult or pediatric patient aged one month or older, when the dimensions of the SPY Laparoscope are appropriate for the patient size and anatomy. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------|-----------------------------------------------------------| | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "Stryker" in a bold, sans-serif font. The letters are all capitalized and black. The word is centered and takes up most of the frame. ## 510(k) Summary This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R Part 807.92(c) ### Submitter: | Applicant | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138 | |----------------|-------------------------------------------------------------------------------------------------------------| | Contact Person | Christie Samsa<br>Manager, Regulatory Affairs<br>Phone: (978) 500-1303<br>Email: christie.samsa@stryker.com | | Date Prepared | June 9, 2022 | ### Subject Device: | Name of Device | SPY Laparoscope | |----------------------|------------------------------------------------| | Common or Usual Name | Laparoscope | | Classification Name | Endoscope And Accessories, 21 C.F.R. §876.1500 | | Regulatory Class | Class II | | Product Code | GCJ | ### Predicate Device: | Name of Device | Stryker AIM (Advanced Imaging Modality) System, K210088 | |----------------|---------------------------------------------------------| |----------------|---------------------------------------------------------| NOTE: The predicate device has not been subject to a design-related recall. ### Device Description: Stryker's SPY Laparoscope is part of Stryker's rigid endoscope product portfolio. Stryker's SPY Laparoscopes are tubular optical instruments used to visualize or image a patient's anatomy during minimally invasive, endoscopic procedures for examination, diagnosis or therapy. The SPY Laparoscope transmits light in both the visible and near infrared spectrum to illuminate and image the anatomy, then forms and relays the image of the surgical site to a camera system for processing and display. {4}------------------------------------------------ ## Indications for Use: The SPY Laparoscope is intended to be used for gynecological and general procedures that clinicians deem appropriate for the adult or pediatric patient aged one month or older, when the dimensions of the SPY Laparoscope are appropriate for the patient size and anatomy. | Feature | Subject Device | Predicate Device | |-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------| | | SPY Laparoscope<br>(Current Submission) | Stryker AIM (Advanced Imaging<br>Modality) System<br>(K210088) | | Manufacturer | Stryker Endoscopy | Same as subject device | | Submission Reference | Current Submission | K210088 | | Intended Use | Endoscopic white light and near-infrared<br>illumination and imaging during<br>endoscopic procedures. | Same as subject device | | Indications for Use Statement | The SPY Laparoscope is intended to be<br>used for gynecological and general<br>procedures that clinicians deem<br>appropriate for the adult or pediatric<br>patient aged one month or older, when<br>the dimensions of the SPY Laparoscope<br>are appropriate for the patient size and<br>anatomy. | Same as subject device | | Image Transmission | Rigid rod lenses | Same as subject device | | Outer diameter | 5.4mm, 10mm | Same as subject device | | Working Length | 30cm, 33cm & 45cm | 30cm, 33cm | | DOV | 0°, 30°, 45° | 0°, 30° | | FOV | 75° | Same as subject device | | Patient-Contacting Materials | Stainless Steel, Epoxy, Optical Glass<br>(sapphire), Glass Fibers | Same as subject device | | Single Use or Reusable | Reusable | Same as subject device | | Cleaning | Manual and Automated | Same as subject device | | Disinfection | Manual and Automated | Same as subject device | | Sterilization<br>Methods | Autoclave, Steris VPRO,<br>Sterrad | Same as subject device | | Sterility Assurance Level | $10^{-6}$ | $10^{-6}$ | Comparison of Technological Characteristics with the Predicate Device: {5}------------------------------------------------ # Performance Testing: The following performance data were provided in support of the substantial equivalence determination. | Test | Method | Result | |----------------------------|----------------------------------------------------------------------------------------------------------------------------|--------| | Electrical Safety | ● IEC 60601-1:2005 + A1:2012 | PASS | | | ● IEC 60601-2-18:2009 | | | Packaging | ● ASTM D4169:2016 | PASS | | Biocompatibility | ● ISO 10993-1:2018 | PASS | | | ● ISO 10993-5:2009 | | | | ● ISO 10993-10:2010 | | | | ● ISO 10993-11:2017 | | | | ● FDA Guidance: Use of International Standard ISO 10993-1 | | | Cleaning &<br>Disinfection | ● AAMI TIR 12:2020 | PASS | | | ● AAMI TIR 30:2011 | | | | ● ISO 15883-5:2021 | | | | ● ISO 15883-2:2009 | | | | ● ISO 15883-1:2009 | | | Sterilization | ● AAMI ST79:2017 | PASS | | | ● AAMI ST58:2013 | | | | ● AAMI TIR 12:2020 | | | | ● ISO 17665-1:2021 | | | | ● ISO 14937:2009 | | | | ● FDA Guidance: Reprocessing Medical Devices in Health Care Settings:<br>Validation Methods and Labeling | | | Performance -<br>Bench | ● Optical and mechanical verification in accordance with device input<br>specifications and comparison to predicate device | PASS | | | ● ISO 8600-1:2015 | | | | | | | Design Validation | ● IEC 62366-1:2020 | PASS | | | ● In accordance with device user needs | | ### Conclusions: The SPY Laparoscope is the same or similar in design, intended use, principles of operation, technological characteristics and safety features to the predicate device. There are no new issues of safety and/or effectiveness introduced by the SPY Laparoscope when used as instructed.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...