The AHI System is intended for use by healthcare professionals managing patients 18 years or older who are receiving continuous physiological monitoring with electrocardiography (ECG) in hospitals. AHI provides a frequently updated binary output over time based on pattern analysis of a lead-II ECG waveform intended to describe a patient's hemodynamic status and indicate if a patient is showing stability or instability or instability. Signs of hemodynamic instability (HI) are defined as hypotension (systolic blood pressure <90 mmHq or mean arterial pressure (MAP) <70 mmHg) combined with tachycardia (heart rate ≥ 100 bpm). AHI-Pl provides the clinician with physiological insight into a patient's likelihood of a future episode of HI is defined as 10 continuous minutes or more where HI is present. The goal of this adjunctive monitoring method is to enable identification of patients who are showing Hl or are likely to experience a future episode of H, and to allow clinicians an opportunity to increase vigilance. This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.
Device Story
AHI System processes single-lead ECG waveform data to monitor hemodynamic status; utilizes pattern analysis of heart rate variability (HRV) and ECG morphology. System provides two outputs: AHI (binary stable/unstable status) and AHI-PI (predictive indicator for future hemodynamic instability episodes). Used in hospital settings by healthcare professionals; provides adjunctive insight alongside standard vital signs. Output displayed to clinicians to increase vigilance; not intended to independently direct therapy. Benefits include early identification of patients at risk for hemodynamic instability.
Clinical Evidence
Software verification and validation testing conducted per FDA guidance for moderate level of concern software. No clinical trial data presented; device relies on bench testing to ensure system performance as designed.
Technological Characteristics
Software-based system; processes single-lead ECG input. Features include HRV and ECG morphology analysis. User interface modified for contraindication workflow. Software level of concern: moderate.
Indications for Use
Indicated for patients 18+ years old receiving continuous ECG monitoring in hospital settings. Used as an adjunctive tool to identify current hemodynamic instability (hypotension + tachycardia) or predict future episodes of hemodynamic instability.
Regulatory Classification
Identification
An adjunctive hemodynamic indicator with decision point is a device that identifies and monitors hemodynamic condition(s) of interest and provides notifications at a clinically meaningful decision point. This device is intended to be used adjunctively along with other monitoring and patient information.
Special Controls
In combination with the general controls of the FD&C Act, the adjunctive hemodynamic indicator with decision point is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Software description, verification, and validation based on comprehensive hazard analysis and risk assessment must be provided, including:
(i) Full characterization of technical parameters of the software, including algorithm(s);
(ii) Description of the expected impact of all applicable sensor acquisition hardware characteristics on performance and any associated hardware specifications;
(iii) Specification of acceptable incoming sensor data quality control measures;
(iv) Mitigation of impact of user error or failure of any subsystem components (signal detection and analysis, data display, and storage) on output accuracy; and
(v) The sensitivity, specificity, positive predictive value, and negative predictive value in both percentage and number form for clinically meaningful pre-specified time windows consistent with the device output.
(2) Scientific justification for the validity of the hemodynamic indicator algorithm(s) must be provided. Verification of algorithm calculations and validation testing of the algorithm must use an independent data set.
(3) Usability assessment must be provided to demonstrate that risk of misinterpretation of the status indicator is appropriately mitigated.
(4) Clinical data must support the intended use and include the following:
(i) The assessment must include a summary of the clinical data used, including source, patient demographics, and any techniques used for annotating and separating the data;
(ii) Output measure(s) must be compared to an acceptable reference method to demonstrate that the output represents the measure(s) that the device provides in an accurate and reproducible manner;
(iii) The data set must be representative of the intended use population for the device. Any selection criteria or limitations of the samples must be fully described and justified;
(iv) Where continuous measurement variables are displayed, agreement of the output with the reference measure(s) must be assessed across the full measurement range; and
(v) Data must be provided within the clinical validation study or using equivalent datasets to demonstrate the consistency of the output and be representative of the range of data sources and data quality likely to be encountered in the intended use population and relevant use conditions in the intended use environment.
(5) Labeling must include the following:
(i) The type of sensor data used, including specification of compatible sensors for data acquisition, and a clear description of what the device measures and outputs to the user;
(ii) Warnings identifying factors that may impact output results;
(iii) Guidance for interpretation of the outputs, including warning(s) specifying adjunctive use of the measurements;
(iv) Key assumptions made in the calculation and determination of measurements; and
(v) A summary of the clinical validation data, including details of the patient population studied (
*e.g.,* age, gender, race/ethnicity), clinical study protocols, and device performance with confidence intervals for all intended use populations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
July 14, 2022
Fifth Eye Inc. % Donna-Bea Tillman Senior Consultant Biologics Consulting 1555 King Street. Suite 300 Alexandria, Virginia 22314
Re: K221203
Trade/Device Name: AHI System Regulation Number: 21 CFR 870.2220 Regulation Name: Adjunctive hemodynamic indicator with decision point Regulatory Class: Class II Product Code: QNV,QNL Dated: June 17, 2022 Received: June 21, 2022
Dear Donna-Bea Tillman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
Submission Number (if known)
K221203
Device Name
AHI System (v 2.2.0)
#### Indications for Use (Describe)
The AHI System is intended for use by healthcare professionals managing patients 18 years or older who are receiving continuous physiological monitoring with electrocardiography (ECG) in hospitals.
AHI provides a frequently updated binary output over time based on pattern analysis of a lead-II ECG waveform intended to describe a patient's hemodynamic status and indicate if a patient is showing signs of hemodynamic stability or instability. Signs of hemodynamic instability (HI) are defined as hypotension (systolic blood pressure <90 mmHg or mean arterial pressure (MAP) <70 mmHq) combined with tachycardia (heart rate ≥ 100 bpm).
AHI-PI provides the clinician with physiological insight into a patient's likelihood of a future episode of HI. An episode of HI is defined as 10 continuous minutes or more where HI is present.
The goal of this adjunctive monitoring method is to enable identification of patients who are showing HI or are likely to experience a future episode of HI, and to allow clinicians an opportunity to increase vigilance. This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.
Type of Use (Select one or both, as applicable)
< Prescription Use (Part 21 CFR 801 Subpart D)
The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Prepared on: 2022-06-14
### Contact Details
| Applicant Name | Fifth Eye Inc. |
|---------------------------------|---------------------------------------------------------------|
| Applicant Address | 110 Miller Avenue, Suite 300 Ann Arbor MI 48104 United States |
| Applicant Contact Telephone | 734-260-4800 |
| Applicant Contact | Ms. Jennifer Baird |
| Applicant Contact Email | jbaird@fiftheye.com |
| Correspondent Name | Biologics Consulting |
| Correspondent Address | 1555 King Street, Suite 300 Alexandria VA 22314 United States |
| Correspondent Contact Telephone | (410) 531- 6542 |
| Correspondent Contact | Dr. Donna-Bea Tillman |
| Correspondent Contact Email | dtillman@biologicsconsulting.com |
21 CFR 807.92(a)(2)
| Device Trade Name | AHI System (v 2.2.0) |
|---------------------|------------------------------------------------------|
| Common Name | Adjunctive hemodynamic indicator with decision point |
| Classification Name | Adjunctive Hemodynamic Indicator With Decision Point |
| Regulation Number | 870.2220 |
| Product Code | QNV |
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K212219 | AHI System | QNV |
21 CFR 807.92(a)(4)
The AHI System is a multiparameter system designed to meet clinicians' need to identify patient hemodynamic status and predict patient hemodynamic instability episodes using two analytics:1. Analytic for Hemodynamic Instability (AHI): Utilizing data from a single existing lead of a non-invasive electrocardiograph (ECG), AH analyzes heart rate variability (HRV) and ECG morphology features to rapidly detect signs of hemodynamic stability and categorize each window of data as either "AHI Stable" or "AHI Unstable." Time trending of AHI outputs is also provided.
2. Analytic for Hemodynamic Instability Predictive Indicator (AH-P); Utilizing AH outputs from up to the most recent 30 minutes of EG data, AH-PI indicates the likelihood of a future episode of hemodynamic instability, defined as ten continuous minutes or more where signs of hemodynamic instability are present.
{4}------------------------------------------------
## Intended Use/Indications for Use
The AHI System is intended for use by healthcare professionals managing patients 18 years or older who are receiving continuous physiological monitoring with electrocardiography (ECG) in hospitals.
AHI provides a frequently updated binary output over time based on pattern analysis of a lead-ll ECG waveform intended to describe a patient's hemodynamic status and indicate if a patient is showing stability or instability or instability. Signs of hemodynamic instability (HI) are defined as hypotension (systolic blood pressure <90 mmHq or mean arterial pressure (MAP) <70 mmHg) combined with tachycardia (heart rate ≥ 100 bpm).
AHI-Pl provides the clinician with physiological insight into a patient's likelihood of a future episode of HI is defined as 10 continuous minutes or more where HI is present.
The goal of this adjunctive monitoring method is to enable identification of patients who are showing Hl or are likely to experience a future episode of H, and to allow clinicians an opportunity to increase vigilance. This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.
## Indications for Use Comparison
The subject device and predicate device have identical indications for use statements.
## Technological Comparison
Technological Characteristics: The subject device have very similar technological characteristics. The only difference is a minor modification to the user interface to allow users to contraindicate only the whom contraindications apply, rather than having to proactively assert "not contraindicated" on all patients.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Software: Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a moderate level of concern. Testing was conducted to ensure the AH System works as designed.
### 21 CFR 807.92(a)(5)
### 21 CFR 807.92(a)(5)
# 21 CFR 807.92(a)(6)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.