K221183 · Aeye Health, Inc. · PIB · Nov 10, 2022 · Ophthalmic
Device Facts
Record ID
K221183
Device Name
AEYE-DS
Applicant
Aeye Health, Inc.
Product Code
PIB · Ophthalmic
Decision Date
Nov 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1100
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
More than mild diabetic retinopathy detection
Artificial Intelligence software as a medical device
Sensitivity > 82%, Specificity > 87%
Sensitivity 92.98%, Specificity 91.36% (based on 2 images per eye)
—
—
Pivotal clinical study: 531 subjects
>1 (independent reading center)
Indications for Use
The AEYE-DS device is indicated for use by health care providers to automatically detect more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. The AEYE-DS is indicated for use with the Topcon NW400.
Device Story
AEYE-DS is an AI-based software as a medical device (SaMD) for diagnostic screening of more-than-mild diabetic retinopathy (mtmDR). Input: fundus retinal images (1 or 2 per eye) acquired via Topcon NW400 camera. Operation: Client module on local computer guides image acquisition and transmits images via secure internet connection to a remote server; Analytics module processes images to assess quality and detect mtmDR; results returned to Client module. Used in hospitals/clinics by healthcare providers. Output: diagnostic report (mtmDR detected/not detected or insufficient quality). Benefits: enables automated screening for diabetic retinopathy in asymptomatic diabetic patients, facilitating timely clinical decision-making.
Clinical Evidence
Prospective, multi-center, single-arm, blinded study (n=531). Primary endpoint: sensitivity/specificity for mtmDR detection vs. Reading Center ETDRS-based ground truth. Results (2 images/eye): sensitivity 94.74% (95% CI: 85.63-98.19%), specificity 88.64% (95% CI: 85.18-91.38%). Results (1 image/eye): sensitivity 92.98% (95% CI: 83.30-97.24%), specificity 91.36% (95% CI: 88.22-93.72%). Precision study confirmed repeatability and reproducibility across operators and devices. Human factors study confirmed usability by novice operators.
Technological Characteristics
AI-based SaMD; client-server architecture. Components: Client software (local), Service/Analytics/Security/Reporting modules (remote server). Inputs: 1000x1000 pixel color fundus images (min 45° FOV). Connectivity: secure internet. Standards: IEC 62304, ISO 14971, FDA cybersecurity guidance. Software level of concern: Major.
Indications for Use
Indicated for adults diagnosed with diabetes, without prior diagnosis of diabetic retinopathy, to automatically detect more than mild diabetic retinopathy (mtmDR) using the Topcon NW400 fundus camera.
Regulatory Classification
Identification
A retinal diagnostic software device is a prescription software device that incorporates an adaptive algorithm to evaluate ophthalmic images for diagnostic screening to identify retinal diseases or conditions.
Special Controls
In combination with the general controls of the FD&C Act, the retinal diagnostic software device is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Software verification and validation documentation, based on a comprehensive hazard analysis, must fulfill the following:
(i) Software documentation must provide a full characterization of technical parameters of the software, including algorithm(s).
(ii) Software documentation must describe the expected impact of applicable image acquisition hardware characteristics on performance and associated minimum specifications.
(iii) Software documentation must include a cybersecurity vulnerability and management process to assure software functionality.
(iv) Software documentation must include mitigation measures to manage failure of any subsystem components with respect to incorrect patient reports and operator failures.
(2) Clinical performance data supporting the indications for use must be provided, including the following:
(i) Clinical performance testing must evaluate sensitivity, specificity, positive predictive value, and negative predictive value for each endpoint reported for the indicated disease or condition across the range of available device outcomes.
(ii) Clinical performance testing must evaluate performance under anticipated conditions of use.
(iii) Statistical methods must include the following:
(A) Where multiple samples from the same patient are used, statistical analysis must not assume statistical independence without adequate justification.
(B) Statistical analysis must provide confidence intervals for each performance metric.
(iv) Clinical data must evaluate the variability in output performance due to both the user and the image acquisition device used.
(3) A training program with instructions on how to acquire and process quality images must be provided.
(4) Human factors validation testing that evaluates the effect of the training program on user performance must be provided.
(5) A protocol must be developed that describes the level of change in device technical specifications that could significantly affect the safety or effectiveness of the device.
(6) Labeling must include:
(i) Instructions for use, including a description of how to obtain quality images and how device performance is affected by user interaction and user training;
(ii) The type of imaging data used, what the device outputs to the user, and whether the output is qualitative or quantitative;
(iii) Warnings regarding image acquisition factors that affect image quality;
(iv) Warnings regarding interpretation of the provided outcomes, including:
(A) A warning that the device is not to be used to screen for the presence of diseases or conditions beyond its indicated uses;
(B) A warning that the device provides a screening diagnosis only and that it is critical that the patient be advised to receive followup care; and
(C) A warning that the device does not treat the screened disease;
(v) A summary of the clinical performance of the device for each output, with confidence intervals; and
(vi) A summary of the clinical performance testing conducted with the device, including a description of the patient population and clinical environment under which it was evaluated.
{1}------------------------------------------------
November 10, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is the Department of Health & Human Services logo, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
AEYE Health, Inc. % John Smith MD, JD Partner Hogan Lovells US LLP 555 13th Street NW Washington, District of Columbia 20004
Re: K221183
Trade/Device Name: Aeye-ds Regulation Number: 21 CFR 886.1100 Regulation Name: Retinal Diagnostic Software Device Regulatory Class: Class II Product Code: PIB Dated: October 7, 2022 Received: October 7, 2022
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Y. Ng -S
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K221183
Device Name
#### AEYE-DS
Indications for Use (Describe)
The AEYE-DS device is indicated for use by health care providers to automatically detect more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. The AEYE-DS is indicated for use with the Topcon NW400.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{4}------------------------------------------------
#### 510(K) SUMMARY
## AEYE-DS DEVICE
#### 510(k) Number K221183
| Applicant Name: | AEYE Health Inc. |
|----------------------|---------------------------------------------------------------------------------------------------------------|
| Contact Person: | Zack Dvey-Aharon, Ph.D. |
| Contact: | AEYE Health Inc.<br>200 Park Ave<br>New York, NY, 10166 USA<br>E-mail: info@aeyehealth.com<br>+1 866 262 7343 |
| Date Prepared: | November 9, 2022 |
| Trade Name: | AEYE-DS |
| Classification Name: | 21 CFR 886.1100; (Product Code PIB)<br>Retinal Diagnostic Software Device |
| Classification: | Class II |
#### Predicate Device:
The AEYE-DS device is substantially equivalent to the following predicate device:
| Predicate | Device | Manufacturer | 510(k) No. |
|-----------|--------|--------------|------------|
| Main | IDx-DR | IDx LLC | DEN180001 |
#### Device Description:
AEYE-DS is a retinal diagnostic software device that incorporates an algorithm to evaluate ophthalmic images for diagnostic screening to identify retinal diseases or conditions. Specifically, the AEYE-DS is designed to perform diagnostic screening for the condition of more-than-mild diabetic retinopathy (mtmDR).
The AEYE-DS is comprised of 5 software components: (1) Client; (2) Service; (3) Analytics; (4) Reporting and Archiving; and (5) System Security. The device configuration of these modules is presented in the figure below, indicating which components are local to the user and which are remotely located.
{5}------------------------------------------------
Image /page/4/Figure/0 description: This image shows a diagram of a system for healthcare providers to analyze fundus images. The diagram is divided into two sections: "Local to User" and "Remote". The "Local to User" section includes a healthcare provider, a user computer, and a client module. The "Remote" section includes secure servers, an analytics module, a service module, a system security module, and a reporting and archiving module. The diagram shows the flow of fundus images from the healthcare provider to the client module, then to the analytics module, and finally to the reporting and archiving module.
Figure 1: Device Configuration
The AEYE-DS device is based on the main technological principle of Artificial Intelligence (AI) software as a medical device. The software as a medical device uses artificial intelligence technology to analyze specific disease features from fundus retinal images for diagnostic screening of diabetic retinopathy.
The AEYE-DS device is based on the principle of operation, whereby a fundus camera is used to obtain retinal images. The fundus camera is attached to a computer, where the Client module/software is installed. The Client module/software guides the user to acquire the images and enables the user to interact with the server-based analysis software over a secure internet connection. Using the Client module/software, users identify the fundus images per eye to be dispatched to the Service module/software. The Service module/software is installed on a server hosted at a secure datacenter, receives the fundus images and transfers them to the Analytics module/software. The Analytics module/software, which runs alongside the Service module/software, processes the fundus images and returns information on the image quality and the presence or absence of mtmDR to the Service module/software. The Service module/software then returns the results to the Client module/software.
{6}------------------------------------------------
#### Intended Use/Indication for Use:
The AEYE-DS device is indicated for use by health care providers to automatically detect more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. The AEYE-DS is indicated for use with the Topcon NW400.
Prescription Use only: Federal law restricts this device for sale by or on the order of a physician.
### Performance Standards:
The AEYE-DS device complies with the following FDA recognized consensus standards:
- Software Verification and Validation Testing Software verification and validation . testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Major" level of concern, since a failure or latent flaw in the software could result in serious injury to the patient through incorrect or delayed information or through the action of a care provider.
- IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device . software – Software life cycle processes
- ISO 14971 Medical devices Application of risk management to medical devices .
### Non-Clinical (Bench) Performance Data:
Software validation testing in compliance with FDA guidelines for software validation and IEC 62304 standard requirements was also conducted.
The software hazard analysis was performed as part of the system hazard analysis. The hazards of the software influencing the operations of the system, the hardware problems impairing the software's integrity, and the incorrect operations of the system by the user that could affect the software's correct functioning, were handled as part of the system hazard analysis. The risks and the risk reductions are found in the Risk Analysis for the AEYE-DS device.
The cybersecurity requirements for the AEYE-DS device were identified according to the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices. A threat analysis was performed and documented in the Cybersecurity Report. Updates to the software are also detailed in the Cybersecurity Report.
{7}------------------------------------------------
The results of the performance tests, including software validation, cybersecurity and hazard analysis demonstrated that the AEYE-DS device (version 5.00) is substantially equivalent to the predicate devices.
#### Animal Performance Data / Histology Data:
Not Applicable
#### Clinical Performance Data:
The AEYE-DS device performance for automated more than mild Diabetic Retinopathy (mtmDR) detection from digital funduscopic images was demonstrated in a pivotal clinical study. The study was a prospective, multi-center, single-arm, blinded study. The study was conducted at 8 study sites in the United States (7 sites) and Israel (1 site), with active enrollment from October 2020 through November 2021. A total of 531 subjects were screened and enrolled in the study. The study population represented the target population for the use of this device and consisted of stable, visually asymptomatic subjects who were previously diagnosed with diabetes and had no prior diagnosis of mtmDR. Subjects participated in a routine retinal screening test for diabetic retinopathy (DR) in hospitals, primary care clinics or medical research centers. Patients of both genders, all ethnicities and ≥22 years of age were recruited to the study. General patient demographics, medical history, concomitant medications, fundoscopy system used, OCT system used, etc., were obtained for each study subject.
Novice operators, who had not previously performed ocular imaging, obtained fundoscopy images from each eye of the patient, using the Topcon Model NW400 funduscopic camera. Upon submission of the fundoscopy images to the AEYE-DS client software, a diagnostic result (and PDF diagnostics report) of more than mild DR (mtmDR) detected or more than mild DR not detected was produced. A result of 'insufficient quality' was determined if the novice operator reached a maximum of 6 image submission attempts and one or more of the images was still of insufficient image quality. After the novice operator generated an AEYE-DS diagnostic output, each participant underwent additional retinal imaging captured by a professional ophthalmic photographer, to obtain dilated four widefield stereo color fundus images, lens photography for media opacity assessment and macular optical coherence tomography (OCT) imaging. The professional images were sent to an independent reading center where the severity of retinopathy and diabetic macular edema (DME) were determined according to the Early Treatment for Diabetic Retinopathy Study severity (ETDRS) scale. The Reading Center diagnostic results formed the reference standard (ground truth) for the study. As part of the final clinical assessment, each participant was categorized as mtmDR+ or mtmDR-, based on the worst of two eyes. The final clinical assessment based on the worst of two eyes was compared with the AEYE- DS output, at the participant level.
The baseline demographic data and characteristics analysis showed that the mean age was 55 years (range 21-88), 47% were male and 53% were African-American, 39% White
{8}------------------------------------------------
and 29% Hispanic or Latino, the remainder were of other racial/ethnic origins. Approximately 95% of the subjects in the study were diagnosed with type 2 diabetes while approximately 5% percent were diagnosed with type 1 diabetes. Duration of diabetes since diagnosis ranged from 1 day to approximately 44 years, with a mean of approximately 10 years (SD=7.9). Most recent HbA1c levels were reported for 446 (86.3%) subjects. HbA1c levels ranged from 5% to 15.6% with a mean of 8.26% (SD=2.15).
The primary efficacy objective of this study was the sensitivity and specificity of the AEYE-DS device to detect mtmDR on digital funduscopic images, acquired by the Topcon NW400 fundoscopy device, based on two macula-centered images (one image from each eye of the patient).
AEYE-DS correctly identified 53 of the 57 fully analyzable subjects with mtmDR+, thus the sensitivity was 92.98% [CI: 83.30%; 97.24%] with a lower one-sided 97.5% confidence bound of 83.3%, which is higher than the pre-defined performance goal of 82%. Sensitivity was identical when calculated for fundus-based mtmDR+ and for multi-modal mtmDR+. Of the 405 fully analyzable subjects who were mtmDR-, according to the Reading Center diagnosis, 370 subjects were correctly identified by AEYE-DS as mtmDR-, based on fundus images alone. Thus, the specificity is 91.36% [CI: 88.22%; 93.72%] with a lower one-sided 97.5% confidence bound of 88.22%, which is higher than the pre-defined performance goal of 87%. Specificity was almost identical when calculated for multi-modal mtmDR-. Therefore, the null hypotheses for sensitivity and specificity were rejected and it is concluded that the study is deemed successful. Positive Predictive Value (PPV) was 60.23% [CI: 49.78%; 69.82%] and the Negative Predictive Value (NPV) was 98.93% [CI: 97.28%; 99.58%]. The PPV was influenced by the actual prevalence of mtmDR+ patients in the pivotal study diabetic population (i.e., 12.3%), as this was not an enriched study.
Further sub-analyses showed that there were no significant effects of sex, race/ethnicity, HbA1c (<10% vs >10%), age (<55 years vs >55 years), diabetic duration since diagnosis (<10 years vs
>10 years, if available) and lens status on sensitivity and specificity. As sequential enrollment provided a sufficient number of mtmDR+ subjects, enrichment was not performed. Therefore, subset analysis by cohort (not enriched, enriched) was not performed.
The powered secondary endpoint of this study was the sensitivity and specificity of the AEYE-DS device to detect mtmDR from digital funduscopic images, acquired by the Topcon NW400 fundoscopy device, based on four images (one macula centered image and one optic disc centered image per eye). Diagnostic results and AEYE-DS device diagnostic results and showed a sensitivity and specificity of 94.74% [CI: 85.63%; 98.19%] and 88.64% [CI: 85.18%; 91.38%], respectively. The sensitivity is 94.74% with a lower one-sided 97.5% confidence bound of 85.63%, which is higher than the performance goal of 82%, therefore the null hypothesis is rejected in favor of the alternative hypothesis for the sensitivity. The specificity is 88.64% with a lower one-sided 97.5% confidence bound of 85.18%, which is slightly lower than the
{9}------------------------------------------------
performance goal of 87%, therefore the null hypothesis is not rejected in favor of the alternative hypothesis for the specificity. The Positive Predictive Value (PPV) was 54% [CI: 44.26%; 63.44%] and the Negative Predictive Value (NPV) was 99.17% [CI: 97.59%; 99.72%]. Sensitivity and specificity (and PPV and NPV) results were the same when calculated for fundus-based mtmDR+ and for multi-modal mtmDR+. The sensitivity of the AEYE-DS device based on 4 images passed the study performance goals, but the specificity was slightly lower. The implications of the slightly lower specificity mean that more subjects may be determined to be mtmDR+ and sent for follow-up ophthalmic examinations than necessary. As the sensitivity results were successful and even higher than the sensitivity with 2 images, and true-mtmDR+ subjects will not be missed by the AEYE-DS device, the price of a slightly lower specificity based on 4 images is not too high and does not involve any risks. In any case, as the use of two images is less complicated and time consuming than four images, it is more probable that AEYE-DS device users will rely on the use of 2 (macula-centered) images per eye. For this reason, the Instructions for Use will allow the use of 2 or 4 images per eye to obtain an EYE-DS diagnostic output.
| | AEYE-DS Device | |
|---------------|-----------------------------------|------------------------------------|
| | 1 image per eye<br>(Topcon NW400) | 2 images per eye<br>(Topcon NW400) |
| Sensitivity: | 93% [CI: 83.3%; 97.2%] | 94.7% [CI: 85.6%; 98.2%] |
| Specificity: | 91.4% [CI: 88.2%; 93.7%] | 88.6% [CI: 85.2%; 91.4%] |
| Imageability: | 99.1% [CI: 97.8%; 99.7%] | 99.1% [CI: 97.8%; 99.7%] |
| PPV: | 60.2% [CI: 49.8%; 69.8%] | 54% [CI: 44.3%; 63.4%] |
| NPV: | 99% [CI: 97.3%; 99.6%] | 99.2% [CI: 97.6%; 99.7%] |
Key results from the pivotal study are summarized in the table below.
The imageability results from the pivotal study. reflecting data on the usability in the hands of the study novice operators, reported >99% imageability for the AEYE-DS device using the Topcon NW400 fundoscopy camera. In obtaining 2 sufficient quality images per eye, the vast majority of the study subjects (413 (88%)) did not require pupil dilation in the study, resulting in 91.67% sensitivity and 91.78% specificity, based on 1 image per eye. The endpoint analyses results show that the sensitivity and specificity for non-mydriatic (non-dilated subjects) are almost identical to the pivotal study efficacy results presented for the device. Moreover, it is apparent that only a small percent (~12%) of subjects required pupil dilation in order to obtain a diagnostic output. In this small percent of mydriatic (dilated) subjects, the results showed 100% sensitivity and 87.50% specificity, based on 1 image per eye.
#### Precision Study
Overall, twenty two (22) participants were included in the final statistical analysis. All 22 participants completed the entire AEYE-DS device imaging protocol and diagnostic output
{10}------------------------------------------------
twelve consecutive times, imaged by three different novice operators, using two different Topcon NW400 fundoscopy devices. Overall, 12 image sets/diagnoses were obtained for each participant, for a total of 264 image sets/diagnoses in total. Per protocol, operators allowed subjects at least 8 minutes between successive imaging. The results are presented in the following tables.
Precision Tables based on 1 image per eye:
| Intra-Operator Repeatability | Repeat 2 - mtmDR | | |
|------------------------------|-------------------------------------------------------|---------|----------------------|
| Repeat 1 - mtmDR | mtmDR + | mtmDR - | Insufficient quality |
| mtmDR + | 49 | 0 | 0 |
| mtmDR - | 2 | 81 | 0 |
| Insufficient quality | 0 | 0 | 0 |
| OA | 98.48% (130/132) [94.64% ;99.58% ] | | |
| APA | 98.00% [93.00% ;99.45% ] | | |
| ANA | 98.78% [95.66% ;99.66% ] | | |
| AUA | Not presented as all cases were of sufficient quality | | |
| Between<br>Operator<br>Reproducibility | Operator 3 | | | | | | | | |
|----------------------------------------|-------------------------------------------------------|---------|-----|-----------------------|---------|-----|------------------------------------|---------|-----|
| Operator 1 | mtmDR +<br>Operator 2 | | | mtmDR -<br>Operator 2 | | | Insufficient quality<br>Operator 2 | | |
| | mtmDR + | mtmDR - | I Q | mtmDR + | mtmDR - | I Q | mtmDR + | mtmDR - | I Q |
| mtmDR + | 16 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| mtmDR - | 0 | 1 | 0 | 0 | 27 | 0 | 0 | 0 | 0 |
| Insufficient<br>quality | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| OA | 97.73% (43/44) [88.19% ;99.60% ] | | | | | | | | |
| APA | 97.96% [89.31% ;99.64% ] | | | | | | | | |
| ANA | 98.80% [93.49% ;99.79% ] | | | | | | | | |
| AUA | Not presented as all cases were of sufficient quality | | | | | | | | |
| Between-Device Reproducibility | Device 2 mtmDR | | |
|--------------------------------|-------------------------------------------------------|---------|----------------------|
| Device 1 - mtmDR | mtmDR + | mtmDR - | Insufficient quality |
| mtmDR + | 24 | 1 | 0 |
| mtmDR - | 0 | 41 | 0 |
| Insufficient quality | 0 | 0 | 0 |
| OA | 98.48% (65/66) [91.90% ;99.73% ] | | |
| APA | 97.96% [89.31% ;99.64% ] | | |
| ANA | 98.80% [93.49% ;99.79% ] | | |
| AUA | Not presented as all cases were of sufficient quality | | |
{11}------------------------------------------------
Precision Tables based on 2 images per eye:
| Intra-Operator Repeatability | Repeat 2 - mtmDR | | |
|------------------------------|-------------------------------------------------------|---------|----------------------|
| Repeat 1 - mtmDR | mtmDR + | mtmDR - | Insufficient quality |
| mtmDR + | 56 | 1 | 0 |
| mtmDR - | 3 | 72 | 0 |
| Insufficient quality | 0 | 0 | 0 |
| OA | 96.97% (128/132) [92.47% ;98.82%] | | |
| APA | 96.55% [91.47% ;98.65%] | | |
| ANA | 97.30% [93.26% ;98.94%] | | |
| AUA | Not presented as all cases were of sufficient quality | | |
| Between<br>Operator<br>Reproducibility<br>Operator 1 | Operator 3 | | | | | | | | |
|------------------------------------------------------|-------------------------------------------------------|------------|--------|-----------------------|------------|--------|------------------------------------|------------|--------|
| | mtmDR +<br>Operator 2 | | | mtmDR -<br>Operator 2 | | | Insufficient quality<br>Operator 2 | | |
| | mtmD<br>R+ | mtmDR<br>- | I<br>Q | mtmDR<br>+ | mtmDR<br>- | I<br>Q | mtmDR<br>+ | mtmDR<br>- | I<br>Q |
| mtmDR + | 18 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| mtmDR - | 0 | 1 | 0 | 0 | 24 | 0 | 0 | 0 | 0 |
| Insufficient<br>quality | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| OA | 95.45% (42/44) [84.87% ;98.74% ] | | | | | | | | |
| APA | 96.49% [88.08% ;99.03% ] | | | | | | | | |
| ANA | 97.33% [90.79% ;99.27% ] | | | | | | | | |
| AUA | Not presented as all cases were of sufficient quality | | | | | | | | |
| Between-Device Reproducibility | | Device 2 mtmDR | | |
|--------------------------------|-------------------------------------------------------|----------------|----------------------|--|
| Device 1 - mtmDR | mtmDR + | mtmDR - | Insufficient quality | |
| mtmDR + | 27 | 1 | 0 | |
| mtmDR - | 2 | 36 | 0 | |
| Insufficient quality | 0 | 0 | 0 | |
| OA | 95.45% (63/66) [87.47% ;98.44% ] | | | |
| APA | 94.74% [85.63% ;98.19% ] | | | |
| ANA | 96.00% [88.89% ;98.63% ] | | | |
| AUA | Not presented as all cases were of sufficient quality | | | |
### Human Factors Validation Study
Usability of the AEYE-DS device was also assessed in a human factors validation study, including User Manual comprehension and usability of the device in the hands of potential users. Once the users underwent initial, one-time training and practice, all users stated that the device was easy and straightforward and all were successful in submitting images for diagnosis and obtaining a diagnosis output result and PDF report. All users stated that the user manual was clear and easy to use.
In summary, the AEYE-DS device using 2 or 4 images acquired from the Topcon NW400 fundoscopy device demonstrates successful performance, in terms of sensitivity, specificity, PPV and NPV, as well as imageability, usability, and precision.
{12}------------------------------------------------
## Substantial Equivalence:
The subject AEYE-DS device, manufactured by AEYE Health Inc., is substantially equivalent to the cleared IDx-DR device (manufactured by IDx LLC and the subject of De Novo DEN180001).
| Technological<br>Characteristic | AEYE-DS Device | IDx-DR Device<br>(DEN180001) |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code,<br>Class | PIB<br>Class II | PIB<br>Class II |
| Indications for<br>Use | The AEYE-DS is indicated for use by<br>health care providers to automatically<br>detect more than mild diabetic<br>retinopathy (mtmDR) in adults<br>diagnosed with diabetes who have not<br>been previously diagnosed with<br>diabetic retinopathy. The AEYE-DS is<br>indicated for use with the Topcon<br>NW400. | IDx-DR is indicated for use by health<br>care providers to automatically detect<br>more than mild diabetic retinopathy<br>(mtmDR) in adults diagnosed with<br>diabetes who have not been previously<br>diagnosed with diabetic retinopathy.<br>IDx-DR is indicated for use with the<br>Topcon NW400. |
| Target Population | Adult subjects diagnosed with Diabetes | Adult subjects diagnosed with Diabetes |
| Anatomical Sites | Eye examination | Eye examination |
| Environment<br>Used | Hospitals, Clinics | Hospitals, Clinics |
| Energy Used /<br>Delivered | Not Applicable | Not Applicable |
| Design: | A fundus camera is attached to a<br>computer, where the AEYE-DS Client<br>module is installed. The Client module<br>allows the user to interact with the<br>server-based analysis software over a<br>secure internet connection. Using the<br>Client module, users identify one<br>(macular) or two (macular and disc)<br>fundus images per eye to be dispatched<br>to the Service module. The Service is<br>installed on a server hosted at a secure<br>datacenter. The Analytics module,<br>which runs alongside the Service<br>module, processes the fundus images<br>and returns information on the image<br>quality and the presence or absence of<br>mtmDR to the Service module. The<br>Service then returns the results to the<br>Client module. | A fundus camera is attached to a<br>computer, where the Idx-DR Client<br>software is installed. The Client<br>software allows the user to interact with<br>the server-based analysis software over<br>a secure internet connection. Using the<br>Client software, users identify two<br>fundus images per eye to be dispatched<br>to Idx-Service. Idx-Service is installed<br>on a server hosted at a secure<br>datacenter. Idx-DR Analysis, which<br>runs inside Idx-Service, processes the<br>fundus images and returns information<br>on the image quality and the presence or<br>absence of mtmDR to Idx-Service. Idx-<br>Service then returns the results to the<br>Idx-DR Client. |
| Technological<br>Characteristic | AEYE-DS Device | IDx-DR Device<br>(DEN180001) |
| -Mechanism of<br>Action | Artificial Intelligence software as a<br>medical device | Artificial Intelligence software as a<br>medical device |
| - Components | The AEYE-DS device consists of the<br>following components:<br>- Client software on computer<br>connected to fundoscopy camera<br>- Server including Service software and<br>Analytics software | The Idx-DR device consists of the<br>following components:<br>- Client software on computer<br>connected to fundoscopy camera<br>- Server including Service software and<br>Analytics software |
| - Inputs | Macula and disc centered color fundus<br>images with at least 45° field of view, 2<br>images per eye; Or<br>Macula centered color fundus images<br>with at least 45° field of view, 1 image<br>per eye. | Macula and disc centered color fundus<br>images with 45° field of view, 2 images<br>per eye |
| - Outputs | More than mild diabetic retinopathy<br>(mtmDR) detected, not detected or<br>insufficient quality | More than mild diabetic retinopathy<br>(mtmDR) detected, not detected or<br>insufficient quality |
| - Indicated<br>Cameras | Topcon NW400 camera | Topcon NW400 camera |
| - Minimum image<br>size | 1000 x 1000 pixels per image | 1000 x 1000 pixels per image |
| Performance | 1 image per eye (Topcon NW400)<br>Sensitivity: 93% [CI: 83.3%; 97.2%]<br>Specificity: 91.4% [CI: 88.2%;<br>93.7%]<br>Imageability: 99.1% [CI: 97.8%;<br>99.7%]<br>PPV: 60.2% [CI: 49.8%; 69.8%]<br>NPV: 99% [CI: 97.3%; 99.6%] | 1 image per eye (Topcon NW400)<br>NOT APPLICABLE |
| | 2 images per eye (Topcon NW400)<br>Sensitivity: 94.7% [CI: 85.6%;<br>98.2%]<br>Specificity: 88.6% [CI: 85.2%;<br>91.4%]<br>Imageability: 99.1% [CI: 97.8%;<br>99.7%]<br>PPV: 54% [CI: 44.3%; 63.4%]<br>NPV: 99.2% [CI: 97.6%; 99.7%] | 2 images per eye (Topcon NW400)<br>Sensitivity: 87.4% (95%CI, 81.9% -<br>92.9%).<br>Specificity: 89.5% (95% CI, 86.9%-<br>93.1%)<br>Imageability: 96%<br>PPV (Positive Predictive Value): 73%<br>NPV (Negative Predictive Value): 96% |
| Technological<br>Characteristic | AEYE-DS Device | IDx-DR Device<br>(DEN180001) |
| Human Factors | The AEYE-DS device uses the Client<br>module as the user interface. The safe<br>and efficient use of the device was<br>established in a Usability study with<br>novice operators. | The IDx-DR device uses the Client<br>software as the user interface. The safe<br>and efficient use of the device was<br>established in a Usability study with<br>novice operators. |
| Standards Met | IEC 62304 and<br>FDA Guidance for the Content of<br>Premarket Submissions for Software<br>Contained in Medical Devices<br>ISO 14971<br>FDA Guidance - Content of Premarket<br>Submissions for Management of<br>Cybersecurity in Medical Devices | FDA Guidance for the Content of<br>Premarket Submissions for Software<br>Contained in Medical Devices<br>ISO 14971<br>FDA Guidance - Content of Premarket<br>Submissions for Management of<br>Cybersecurity in Medical Devices |
| Materials | No patient contacting materials | No patient contacting materials |
| Biocompatibility | Not Applicable | Not Applicable |
| Compatibility<br>With the<br>Environment and<br>Other Devices | The AEYE-DS device is compatible for<br>use with the Topcon NW400 device.<br>Compatibility with the environment is<br>not applicable. | The IDx-DR device is compatible for<br>use with the Topcon NW400 device.<br>Compatibility with the environment is<br>not applicable. |
| Sterility | Not Applicable | Not Applicable |
| Electrical Safety | Not Applicable | Not Applicable |
| Mechanical Safety | Not Applicable | Not Applicable |
| Chemical Safety | Not Applicable | Not Applicable |
| Thermal Safety | Not Applicable | Not Applicable |
| Radiation Safety | Not Applicable | Not Applicable |
# Table 1: Comparison of the AEYE-DS Device to the predicate IDx-DR Device (DEN180001)
{13}------------------------------------------------
{14}------------------------------------------------
### Conclusions:
The subject AEYE-DS device, manufactured by AEYE Health Inc., is substantially equivalent to the cleared IDx-DR device (manufactured by IDx LLC and the subject of De Novo DEN180001).
The subject AEYE-DS device has the same intended use and indications for use as the cleared IDx-DR device. The subject device and the cleared IDx-DR device are similar in terms of their intended prescription use only, suitable for the adult population diagnosed with diabetes, indicated for use in the same anatomical site (i.e., for eye examinations) and to be used in hospital or clinic settings..
The subject AEYE-DS and cleared IDx-DR devices are composed of the same components,
{15}------------------------------------------------
including a Client module, and a Server including the Service and Analytics modules. The AEYE- DS and the IDx-DR devices are both compatible for use with the Topcon NW400 device. The subject AEYE-DS device has the same mechanism of operation and uses the same underlying technology as the predicate IDx-DR device. That is, a user interface Client module communicates with the Server, where the fundus images (one macular and one disc centered image per eye) are received and analyzed in the Analytics module to provide information regarding the presence or absence of mtmDR, which is returned to the Client module. The only difference in the mechanism of action is the additional ability of the AEYE-DS device to receive only one (m…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.