← Product Code [FET](/productcode/FET) · K221174

# Telepack + (K221174)

_Karl Storz Endoscopy America, Inc. · FET · May 13, 2022 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221174

## Device Facts

- **Applicant:** Karl Storz Endoscopy America, Inc.
- **Product Code:** [FET](/productcode/FET.md)
- **Decision Date:** May 13, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The TELE PACK + is an all-in-one Imaging System, which comprises a light source for illumination, Camera Control Unit (CCU) for use with compatible camera heads or video endoscopes for image processing, as well as a monitor for image display, intended for the visualization of endoscopic and microscopic procedures.

## Device Story

TELEPACK+ is a portable, all-in-one endoscopic imaging system; includes 18.5-inch monitor, camera control unit (CCU), and internal LED light source. Device connects to compatible camera heads or video endoscopes to capture, process, and display images for endoscopic/microscopic procedures and stroboscopy. Operated by clinicians in clinical settings. System features include image capture, zoom, brightness/light source control, and white balance. Users can customize button functions. Output is visual display of endoscopic/microscopic field; assists clinicians in real-time visualization and documentation. Modifications from predicate include touchscreen camera swapping (X-link/office link), added capacitors, mechanical isolation/protective earth changes, video playback/image review functionality, and additional cybersecurity features.

## Clinical Evidence

No clinical data. Substantial equivalence established via non-clinical bench testing, including electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62471), software verification/validation, and performance testing (illumination, spatial resolution, color performance, latency, white balance, AE step response, button functionality).

## Technological Characteristics

Portable all-in-one imaging system (450 x 350 x 150 mm, 9kg). Includes 18.5-inch monitor, CCU, LED light source. Electrical safety: Class I protection, Type BF (stroboscopy/camera), Type CF (light). IPX1 drip-water protection. Software level of concern: Moderate. Connectivity: Compatible with camera heads/video endoscopes.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Telepack + ([K193235](/device/K193235.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

May 13, 2022

KARL STORZ Endoscopy America, Inc. Mario Trujillo Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245

Re: K221174 Trade/Device Name: TELEPACK+ Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FET Dated: April 22, 2022 Received: April 25, 2022

Dear Mario Trujillo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K221174

Device Name TELEPACK+

Indications for Use (Describe)

The TELE PACK + is an all-in-one Imaging System, which comprises a light source for illumination, Camera Control Unit (CCU) for use with compatible camera heads or video endoscopes for image processing, as well as a monitor for image display, intended for the visualization of endoscopic and microscopic procedures.

| Type of Use (Select one or both, as applicable)                                                |                                                                       |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| <div> <span> <b> </b> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is in blue and features the word "STORZ" in large, bold letters with a ring around the "O". Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller, sans-serif font. The logo is simple and professional, and it is likely used on the company's products and marketing materials.

## 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

| Applicant:                 | KARL STORZ Endoscopy-America, Inc<br>2151 E. Grand Avenue<br>El Segundo, CA 90245                                                                                                                                                                                                                                                                                                                   |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                   | Mario Trujillo<br>Regulatory Affairs Specialist<br>Tel.: (424) 218-8184<br>Email: <a href="mailto:Mario.Trujillo@karlstorz.com">Mario.Trujillo@karlstorz.com</a>                                                                                                                                                                                                                                    |
| Date of Preparation:       | April 22, 2022                                                                                                                                                                                                                                                                                                                                                                                      |
| Type of 510(k) Submission: | Special                                                                                                                                                                                                                                                                                                                                                                                             |
| Device Identification:     | Trade Name:TELEPACK +<br>Common Name: Endoscopic Video Imaging System/Component<br>Classification Name: Endoscope and Accessories                                                                                                                                                                                                                                                                   |
| Product Code:              | FET                                                                                                                                                                                                                                                                                                                                                                                                 |
| Regulation:                | 21 CFR 876.1500 (Endoscope and Accessories)                                                                                                                                                                                                                                                                                                                                                         |
| Predicate Device(s):       | Trade Name: Telepack + (K193235)                                                                                                                                                                                                                                                                                                                                                                    |
| Device Description:        | The Telepack + is a portable and compact all-in-one<br>imaging system that includes a 18.5 inch screen<br>display, a camera control unit and internal LED light<br>source, that is intended to be connected to a<br>compatible device (camera head or videonendoscope)<br>for the purpose of visualization and documentation of<br>endoscopic and microscopic procedures as well as<br>stroboscopy. |

{4}------------------------------------------------

|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The Telepack + includes a LED illumination light<br>source to illuminate the intended area and a 18.5 inch<br>monitor for display. It also allows the users to redefine<br>the functions that take place when a button is pressed.<br>The Telepack + is a non-patient contacting and require<br>only wipe down as needed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use and<br>Indications for use:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The TELE PACK + is an all-in-one Imaging System,<br>which comprises a light source for illumination,<br>Camera Control Unit (CCU) for use with compatible<br>camera heads or video endoscopes for image<br>processing, as well as a monitor for image display,<br>intended for the visualization of endoscopic and<br>microscopic procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Technological<br>Characteristics:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | The Telepack + is a portable and compact all-in-one<br>imaging system that includes a 18.5 inch screen<br>display, a camera control unit and internal LED light<br>source, that is intended to be connected to a<br>compatible device (camera head or videoendoscope)<br>for the purpose of visualization and documentation of<br>endoscopic and microscopic procedures as well as<br>stroboscopy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The Telepack + includes, but not limited to, the<br>following features:<br>● Image capture<br>● Zoom<br>● Brightness control<br>● Light source control<br>● White Balance                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | The dimension of the subject device is 450 x 350 x<br>150 mm and weighs 9kg. It is not intended to be<br>soiled and is non-patient contacting. It includes<br>moderate level of concern software. The device has<br>been tested and passed the electrical safety and EMC<br>testing, which is certified to be Class I protection<br>against electrical shock, Type BF protection against<br>electrical shock from stroboscopy and camera applied                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | parts, Type CF protection against electrical shock from<br>light and lastly drip-water protection against moisture<br>per IPX1.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Non-Clinical<br>Performance Data:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | There are no performance standards or special<br>controls developed under Section 514 of the FD&C Act<br>for endoscopes. However, the Telepack + follows the<br>FDA recognized consensus standards and is tested<br>according to the following standards and FDA<br>Guidance:<br>Electrical Safety and EMC  IEC 60601-1 IEC 60601-1-2 IEC 60601-2-18 IEC 62471   Software Verification and Validation Testing  Guidance for the Content of Premarket<br>Submissions for Software Contained in<br>Medical Device Level of concern: Moderate   Performance Testing  Minimum Illumination Spatial Resolution Color Performance Latency White Balance AE Step Response Head Button Functionality                Additional bench testing was performed to ensure the<br>device met its design specifications. The bench<br>testing performed verified and validated that the<br>Telepack + has met all its design specification and is<br>substantially equivalent to its predicate devices. |
| Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The subject device, Telepack +, marketed by KARL<br>STORZ Endoscopy America, Inc., is a modification of<br>the previous clearance of the same device in<br>K193235.<br>Predicate Comparison Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| The intended use, operating principles, technological<br>characteristics and features are identical to those<br>originally cleared in K193235. The minor differences<br>between the originally cleared device and the modified<br>device that do not raise new or different questions of<br>safety and effectiveness are:<br>Swapping of two cameras types (X-link and<br>office link) via touchscreen. Capacitors were added to the office link front<br>end Mechanical changes:  Isolation is added to microphone Additional cover for power supply Mechanical metal part added as a<br>protective earth   Addition of video playback and image review<br>functionality. Additional cybersecurity features Additional camera and videoscope compatibility          As proven by the comparisons, the above differences<br>do not raise different questions of safety and<br>effectiveness because the intended use, operating<br>principles, technological characteristics, and features<br>are similar, if not identical. The modified system also<br>complies with identical standards and safety testing as<br>the originally cleared device. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Clinical Performance Data:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Clinical performance is not required to demonstrate<br>substantial equivalence to the predicate devices. Non-<br>clinical bench testing was sufficient to establish<br>substantial equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Conclusion:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | The modified Telepack + is substantially equivalent to<br>the originally cleared device. The non-clinical bench<br>and comparative testing demonstrate that the device is<br>as safe and effective as the legally marketed devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

{5}------------------------------------------------

{6}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K221174](https://fda.innolitics.com/device/K221174)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
