← Product Code [FGB](/productcode/FGB) · K221158

# Single-Use Video Flexible Ureterorenoscope System (K221158)

_Guangzhou Red Pine Medical Instrument Co., Ltd. · FGB · Mar 6, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221158

## Device Facts

- **Applicant:** Guangzhou Red Pine Medical Instrument Co., Ltd.
- **Product Code:** [FGB](/productcode/FGB.md)
- **Decision Date:** Mar 6, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The device has been developed to be used with endoscopic accessories such as a biopsy forceps and Endoscopic Video Image Processor for endoscopy and endoscopic surgery within urinary tract and interior of the ureter and kidney.

## Device Story

Single-use flexible ureterorenoscope system; includes sterile ureterorenoscope (handle/insertion portion) and Endoscopic Video Image Processor (touch PC, power adapter, footswitch). CMOS sensor at distal tip captures images; LEDs provide illumination. User controls distal bending via deflection lever on handle. Working channel allows insertion of accessories and irrigation. Video cable transmits signals to processor for display. Used in hospitals by physicians for visualization and surgical procedures in urinary tract/kidney. Output provides real-time video for clinical decision-making; single-use design eliminates reprocessing requirements.

## Clinical Evidence

No clinical data. Bench testing only. Evaluated electrical safety (IEC 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), photobiological safety (IEC 62471), mechanical/optical performance (ISO 8600-1, ISO 12233, ISO 15739), and biocompatibility (ISO 10993).

## Technological Characteristics

Flexible ureterorenoscope with CMOS sensor and LED illumination. Materials compliant with ISO 10993-1. Deflection: 275° up/down. Working channel: 1.2mm. Sterilization: EO (SAL 10^-6). System includes touch PC processor. Bench testing per ISO 8600-1, ISO 12233, ISO 15739. Packaging per ISO 11607, ASTM F1980, ASTM F88, ASTM F1929, ISTA 2A.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Medical Video Endoscope System ([K172098](/device/K172098.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Guangzhou Red Pine Medical Instrument Co., Ltd. % Xiangfei Li Regulatory Affairs Ezisurg Medical Co., Ltd. Rm. 103, Bldg. 2, No.1690 Cailun Rd., China (Shanghai) Pilot Free Trade Zone Shanghai, 201203 China

Re: K221158

Trade/Device Name: Single-Use Video Flexible Ureterorenoscope System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FGB Dated: February 7, 2023 Received: February 7, 2023

Dear Xiangfei Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Bonhye Koo -S

For Mark Antonino Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221158

#### Device Name

Single-Use Video Flexible Ureterorenoscope System

Indications for Use (Describe)

The device has been developed to be used with endoscopic accessories such as a biopsy forceps and Endoscopic Video Image Processor for endoscopy and endoscopic surgery within urinary tract and interior of the ureter and kidney.

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# REDPINE

# Traditional 510(k) Premarket Notification

Guangzhou Red Pine Medical Instrument Co., Ltd.

Single-Use Video Flexible Ureterorenoscope System

Applicant: Guangzhou Red Pine Medical Instrument Co., Ltd. Room 303, 308, No. 12, Luoxuan 3 Road, Guangzhou International Bioisland, 510200 Guangzhou City, Guangdong Province, PEOPLE'S REPUBLIC OF CHINA

> Revision A-3 March 06, 2023

{4}------------------------------------------------

# 510(k) Summary

The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:

# 1. Date of Preparation: 03/06/2023

# 2. Sponsor Identification

# Guangzhou Red Pine Medical Instrument Co., Ltd.

Room 303, 308, No. 12, Luoxuan 3 Road, Guangzhou International Bioisland, 510200 Guangzhou City, Guangdong Province, PEOPLE'S REPUBLIC OF CHINA

Establishment Registration Number: 3021246306

Contact Person: Xiangfei Li Position: Regulatory Affairs Specialist Manager Tel: +86-20-8092-7258 Fax: +86-20-3831-5611 Email: lixiangfei@gzredpine.com

# 3. Designated Submission Correspondent

Mr. Xiangfei Li (Primary Contact Person)(Primary Contact Person) Mr. Weihua Yang (Alternative Contact Person)(Alternative Contact Person)

# Guangzhou Red Pine Medical Instrument Co., Ltd.

Room 303, 308, No. 12, Luoxuan 3 Road, Guangzhou International Bioisland, 510200 Guangzhou City, Guangdong Province, PEOPLE'S REPUBLIC OF CHINA

Tel: +86-20-8092-7258 Fax: +86-20-3831-5611 Email: lixiangfei@gzredpine.com

# 4. Identification of Subject device

Trade Name: Single-Use Video Flexible Ureterorenoscope System Common Name: Ureteroscope and Accessories, Flexible/rigid Model: RP-U-C12, RP-U-C0304, RP-U-C0305 Classification Name: Endoscope and accessories Classification: II Product Code: FGB Regulation Number: 21 CFR 876.1500 Review Panel: Gastroenterology/Urology

{5}------------------------------------------------

# Indication for Use Statement:

The device has been developed to be used with endoscopic accessories such as a biopsy forceps and Endoscopic Video Image Processor for endoscopy and endoscopic surgery within urinary tract and interior of the ureter and kidney.

# Device Description:

The Single-Use Video Flexible Ureterorenoscope System consists of a sterile Single-Use Video Flexible Ureterorenoscope (consists of handle and insertion portion) and an Endoscopic Video Image Processor (the video system, touch PC ) with its accessories power adapter, footswitch.The Ureterorenoscope is provided sterile (sterilized by EO) and intended to be single-use.The video system is powered by the main line.

The handle includes a deflection lever, a lever lock, and a Luer port for insertion of accessory devices and irrigation to the working channel. The insertion portion contains one working channel and wiring to transmit the image signals to the Video Image Processor. The distal bending section of the insertion portion controlled by the user via the deflection lever on the handle. The distal end of the insertion portion contains a CMOS sensor for capturing image and transmitting it to the Video Image Processor, LEDs for illumination, and the distal opening of the working channel. The video cable connects the endoscope handle to the Video Image Processor, which provides power and processes video signals from the endoscope.

Flexible Ureterorenoscope has the following physical and performance characteristics:

- Flexible insertion portion
- Camera and LED light source at the distal tip

# 5. Identification of Predicate Devices

Trade/Device Name: Medical Video Endoscope System Regulation Number: 21 CFR 876.1500 Regulatory Class: Class II Product Code: FGB 510(k) Number: K172098

# 6. Summary of Technological Characteristics and Comparison

The principles of operation of the subject device is similar to that of the predicate device. The following basic technological elements are the same or similar for the subject and predicate devices:

{6}------------------------------------------------

| Compared                                                       | Subject device                                                                                                                                                                                                                                              | Predicate Device                                                                                                                                                                                                                             | Comparison            |
|----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| items                                                          | Single-Use Video Flexible<br>Ureterorenoscope System                                                                                                                                                                                                        | Medical Video Endoscope<br>system(K172098)                                                                                                                                                                                                   |                       |
| Classification                                                 |                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                              |                       |
| Product Code                                                   | FGB                                                                                                                                                                                                                                                         | FGB                                                                                                                                                                                                                                          | Same                  |
| Regulation<br>Number                                           | 21 CFR 876.1500                                                                                                                                                                                                                                             | 21 CFR 876.1500                                                                                                                                                                                                                              | Same                  |
| Classification                                                 | Class II                                                                                                                                                                                                                                                    | Class II                                                                                                                                                                                                                                     | Same                  |
| Indications For Use                                            |                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                              |                       |
| Indications for<br>Use                                         | The device has been<br>developed to be used with<br>endoscopic accessories such<br>as a biopsy forceps and<br>Endoscopic Video Image<br>Processor for endoscopy and<br>endoscopic surgery within<br>urinary tract and interior of<br>the ureter and kidney. | This instrument has been<br>designed to be used with<br>endo-therapy accessories such<br>as a biopsy forceps and other<br>ancillary equipment for<br>endoscopy and endoscopic<br>surgery within urinary tract<br>and interior of the kidney. | Similar<br>See Note1. |
| anatomical site                                                | Urinary tract and interior of<br>the kidney                                                                                                                                                                                                                 | Urinary tract and interior of<br>the kidney                                                                                                                                                                                                  | Same                  |
| target<br>population                                           | Adults                                                                                                                                                                                                                                                      | Adults                                                                                                                                                                                                                                       | Same                  |
| Environment<br>of use                                          | Hospitals                                                                                                                                                                                                                                                   | Hospitals                                                                                                                                                                                                                                    | Same                  |
| Physical Characteristics                                       |                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                              |                       |
| Type of Scope                                                  | Flexible                                                                                                                                                                                                                                                    | Flexible                                                                                                                                                                                                                                     | Same                  |
| Type of Imager                                                 | CMOS                                                                                                                                                                                                                                                        | CMOS                                                                                                                                                                                                                                         | Same                  |
| Light source                                                   | LED                                                                                                                                                                                                                                                         | LED                                                                                                                                                                                                                                          | Same                  |
| Up/down<br>deflection(°)                                       | Up 275 °<br>Down 275 °                                                                                                                                                                                                                                      | Up 270 °<br>Down 270 °                                                                                                                                                                                                                       | Similar<br>See Note2. |
| Direction of<br>view                                           | 0 °                                                                                                                                                                                                                                                         | 0 °                                                                                                                                                                                                                                          | Same                  |
| Field of view                                                  | 120 °                                                                                                                                                                                                                                                       | 120 °                                                                                                                                                                                                                                        | Same                  |
| Maximum<br>insertion<br>portion width<br>(mm)                  | RP-U-C12: 3.2<br>RP-U-C0304: 3.15<br>RP-U-C0305: 3.18                                                                                                                                                                                                       | 3.2                                                                                                                                                                                                                                          | Similar<br>See Note2. |
| Working<br>length(mm)                                          | 670                                                                                                                                                                                                                                                         | 650                                                                                                                                                                                                                                          | Similar<br>See Note2. |
| Minimum<br>instrument<br>channel width<br>(mm)                 | 1.2                                                                                                                                                                                                                                                         | 1.0                                                                                                                                                                                                                                          | Similar<br>See Note2. |
| Depth of field                                                 | 3mm-50mm                                                                                                                                                                                                                                                    | 3mm-50mm                                                                                                                                                                                                                                     | Same                  |
| System<br>composition                                          | - Endoscopic Video Image<br>Processor<br>- Single-use video flexible<br>ureterorenoscope and                                                                                                                                                                | - Eview (the video system,<br>touch PC with data processing<br>center)<br>- Uscope                                                                                                                                                           | Same                  |
| Patient Contacting Materials                                   |                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                              |                       |
| General<br>material type<br>of main<br>patient-contact<br>part | Compliance with<br>ISO10993-1                                                                                                                                                                                                                               | Compliance with<br>ISO10993-1                                                                                                                                                                                                                | Similar<br>See Note2. |
| Duration and<br>type of<br>contact                             | "Surface –Mucosal<br>Membrane" with a contact<br>duration of "Limited (< 24<br>hours)"                                                                                                                                                                      | "Surface –Mucosal<br>Membrane" with a contact<br>duration of "Limited (< 24<br>hours)"                                                                                                                                                       | Same                  |
| Sterilization Methods                                          |                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                              |                       |
| Number of<br>Users                                             | Single-Use                                                                                                                                                                                                                                                  | Single-Use                                                                                                                                                                                                                                   | Same                  |
| Sterilization                                                  | EO Sterilized,<br>SAL 10-6                                                                                                                                                                                                                                  | EO Sterilized,<br>SAL 10-6                                                                                                                                                                                                                   | Same                  |

{7}------------------------------------------------

# Note.

1. The indications for use statement for the subject device is similar to that of the predicate device. The differences do not alter the intended use of the device nor do they raise different questions of safety and effectiveness of the device relative to the predicate.

2. The subject and predicate device have similar principles of operation, type of scope, type of imager, light source, sterilization method, number of users and environment. The subject device differs from the predicate in deflection, maximum insertion portion width, working length, minimum instrument channel width, and patient-contacting materials, most of them are better than the predicate device in clinic use. These differences do not raise different questions of safety and effectiveness as compared to the predicate, and can be evaluated through performance testing.

# 7. Non clinical testing summary

# Electrical Safety and Electromagnetic Compatibility Summary

The electrical safety and EMC data included in the submission are in compliance with the following FDA recognized standards:

{8}------------------------------------------------

•IEC 60601-1:2012, and AAMI/ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 •IEC 60601-2-18:2009: Part 2-18

- •IEC 60601-1-2:2014 and A1:2020
# Photobiological safety

The subject device were tested according to the following FDA recognized standards:

• IEC 62471:2006 Medical electrical equipment, Photobiological safety of lamps and lamp systems

# Bench Testing Summary

# Mechanical and Optical Performance

The subject device was designed to comply with applicable parts of ISO 8600-1:2015. Optical measurements were performed according to applicable part of ISO 8600-1:2015, ISO 12233:2017 and ISO 15739:2017 standard.

The following mechanical characteristics were evaluated: water delivery, fatigue stress of rocker, stress of the bending part, tensile strength of the insertion portion, maximum insertion width, minimum instrument channel width deflection.

Optical performance characteristics relevant to functions as intended were tested include field of view, direction of view, resolution, depth of field, distortion, SNR, illumination, dynamic range,color performance.

# Biocompatibility Summary

Biocompatibility evaluation for the Flexible Ureterorenoscope was conducted in accordance with the FDA Guidance document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process".

The following tests were conducted based contact category of "Surface - Mucosal Membrane" with a contact duration of "Limited (< 24 hours):

- Cytotoxicity: ISO 10993-5:2009/(R) 2014
- Sensitization, Intracutaneous reactivity/irritation: ISO 10993-10:2010
- · Material-mediated pyrogenicity: ISO 10993-11:2017
- · Acute systemic toxicity: ISO 10993-11:2017

# Sterilization and shelf life testing

Sterile barrier systems were evaluated in accordance with ISO 11607:2006. Sterilization Process has been validated accordance with ISO 11135:2014. EO/ECH residual test was performed according to ISO 10993-7:2008. The shelf life of the Flexible Ureterorenoscope is determined based on stability study which includes ageing test according to ASTM F1980-16 and ASTM F 1980

{9}------------------------------------------------

07 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices

# Package Validation

Package validation was conducted according to ISO 11607-1:2019 and ISO 11607-2:2019, and F88/F88M-15, ASTM F 1929-15 Transport and shipping testing as per ISTA 2A :2011.

The collective results of Non-clinical testing demonstrate that the materials chosen, the manufacturing processes and design of Ureterorenoscope System meet the established specifications necessary for consistent performance during its intended use. In addition, the Bench Test results demonstrate that compared with the predicate , Ureterorenoscope System has the same or similar image quality, optical performance and mechanical performance.

# 8. Clinical Test Conclusion

No clinical study is included in this submission.

# 9. Conclusion

Non clinical tests demonstrated that the Single-Use Video Flexible Ureterorenoscope System is as safe and effective as the predicate. Therefore, the Single-Use Video Flexible Ureterorenoscope System is substantially equivalent to the predicate devices."

---

**Source:** [https://fda.innolitics.com/device/K221158](https://fda.innolitics.com/device/K221158)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
