Immunoglobulin G (IgG)

K221114 · Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd. · CFN · Aug 2, 2023 · Immunology

Device Facts

Record IDK221114
Device NameImmunoglobulin G (IgG)
ApplicantBeckman Coulter Laboratory Systems (Suzhou) Co., Ltd.
Product CodeCFN · Immunology
Decision DateAug 2, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5510
Device ClassClass 2

Intended Use

System reagent for the quantitative determination of IgG immunoglobulins in human serum, plasma and cerebrospinal fluid on Beckman Coulter AU/DxC AU analyzers. The measurement of IgG aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Device Story

Quantitative immunoturbidimetric assay for IgG measurement in human serum, plasma, and CSF. Reagents (R1 buffer, R2 goat anti-IgG antiserum) mixed with patient sample; human IgG reacts with anti-human IgG to form insoluble aggregates. Aggregates scatter light; turbidimeter measures reduction in transmitted light intensity (absorbance increase) proportional to IgG concentration. Used in clinical laboratory settings on Beckman Coulter AU/DxC AU analyzers (AU480, AU680, AU5800, DxC 700 AU, DxC 500 AU). Operated by laboratory technicians. Output is quantitative IgG concentration (mg/dL). Results assist clinicians in diagnosing protein metabolism disorders and assessing immune function.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP05-A3 (N=80 for within-run/day/run; N=75 for instrument-to-instrument). Linearity verified per CLSI EP06 (R²=0.998 for serum, 1.000 for CSF). Method comparison against predicate (K162208) using 147 serum and 114 CSF samples showed high correlation (R=0.9981 for serum, 0.9995 for CSF). No prozone effect observed up to 36,785 mg/dL (serum) and 7,419 mg/dL (CSF).

Technological Characteristics

Immunoturbidimetric assay. Reagents: R1 (Tris buffer, PEG 6000), R2 (goat anti-IgG). Preservative: sodium azide. Analyzers: Beckman Coulter AU/DxC AU series. Measurement: light scatter/absorbance. Storage: liquid, on-board. Calibration: Serum Protein Multi-Calibrator.

Indications for Use

Indicated for quantitative determination of IgG in human serum, plasma, and cerebrospinal fluid to aid in diagnosis of abnormal protein metabolism and immune resistance. No specific age or gender restrictions provided.

Regulatory Classification

Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. August 2, 2023 Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd. Tracy Jin Senior Analyst, Regulatov Affairs No. 181 West Su Hong Road, Suzhou Industrial Park Suzhou, Jiangsu 215021 People's Republic of China Re: K221114 Trade/Device Name: Immunoglobulin G (IgG) Regulation Number: 21 CFR 866.5510 Regulation Name: Immunoglobulins A. G. M. D. and E immunological test system Regulatory Class: Class II Product Code: CFN Dated: February 14, 2023 Received: February 16, 2023 Dear Tracy Jin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Ying Mao -S Ying Mao, Ph.D. Branch Chief Division of Immunology and Hematology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221114 Device Name Immunoglobulin G (IgG) ### Indications for Use (Describe) System reagent for the quantitative determination of IgG immunoglobulins in human serum, plasma and cerebrospinal fluid on Beckman Coulter AU/DxC AU analyzers. The measurement of IgG aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> | |----------------------------------------------|-----------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | <span style="font-size: 20px;">☐</span> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red oval shape with two white curved lines inside, placed to the left of the company name. The text "BECKMAN" is on the top line and "COULTER" is on the bottom line, both in a bold, sans-serif font. #### 1.0 Submitted By Tracy Jin Sr. Regulatory Affairs Beckman Coulter Laboratory Systems (Suzhou) Co., Ltd. 181 West Su Hong Road, Suzhou Industrial Park, Suzhou, Jiangsu, P.R. China, 215021 Telephone: 86-512- 6895 5129 Fax: 86-512- 6742 1069 #### 2.0 Date of Preparation Jul 25,2023 #### 3.0 Device Name(s) | Proprietary Name: | IgG | |----------------------|---------------------------------------------------------| | Common Name: | IgG | | Classification: | Class II | | Classification Name: | Immunoglobulins A, G, M, D, E Immunological Test System | | Product Codes: | CFN | | Regulation Number: | 21 CFR §866.5510 | #### 4.0 Predicate Device | Candidate | Predicate | Manufacturer | |-------------|-----------------------|-----------------------| | IgG Reagent | IgG Reagent (K162208) | Beckman Coulter, Inc. | #### 5.0 Device Description The device consists of two reagents: R1 buffer (Tris buffer pH 7.2, polyethylene glycol 6000) and R2 (goat anti-IgG antiserum). The reagents contain sodium azide as preservative. When a sample is mixed with R1 buffer and R2 antiserum solution, human IqG reacts specifically with anti-human IgG antibodies to yield insoluble aggregates. Immune complexes formed in solution scatter light in proportion to their size, shape, and concentration. Turbidimeters then measure the reduction of incidence light due to reflection, absorption, or scatter. The decrease in intensity of light transmitted (increase in absorbance) through particles suspended in solution is a result of complexes formed during the antigen-antibody reaction. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red symbol on the left and the company name "BECKMAN COULTER" on the right. The symbol is a stylized, abstract design with curved lines, resembling a stylized letter "B". The company name is written in a bold, sans-serif font, with "BECKMAN" stacked above "COULTER". #### 6.0 Indications for Use System reagent for the quantitative determination of IgG immunoglobulins in human serum, plasma, and cerebrospinal fluid on Beckman Coulter AU/DxC AU analyzers. The measurement of IgG aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents. For in vitro diagnostic use. #### 7.0 Comparison to the Predicate Table 7.0 shows the similarities and differences between the candidate and predicate devices. | Feature | Predicate Device:<br>IgG Reagent<br>(DxC 700 AU Clinical<br>Chemistry Analyzer)<br>(K162208) | Candidate Device:<br>IgG Reagent<br>(DxC 500 AU Clinical<br>Chemistry Analyzer) | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | System reagent for the<br>quantitative determination of IgG<br>immunoglobulins in human<br>serum, plasma, and<br>cerebrospinal fluid on Beckman<br>Coulter AU analyzers | Same; updated<br>instrument family<br>branding only.<br>System reagent for the<br>quantitative determination<br>of IgG immunoglobulins in<br>human serum, plasma,<br>and cerebrospinal fluid on<br>Beckman Coulter AU/DxC<br>AU analyzers | | Measurement | Quantitative | Same | | Instrument Required | AU400/400e/480, AU640/640e<br>/680, AU2700/5400/AU5800 and<br>DxC 700 AU Beckman Coulter<br>Analyzers. | AU480, AU680, AU5800,<br>DxC 500 AU and DxC<br>700 AU Beckman Coulter<br>Analyzers. | | Methodology | Immunoturbidimetric | Same | | Antibody | Goat anti-IgG | Same | | Reagent form and storage | Liquid, on-board storage | Same | | Feature | Predicate Device:<br>IgG Reagent<br>(DxC 700 AU Clinical<br>Chemistry Analyzer)<br>(K162208) | Candidate Device:<br>IgG Reagent<br>(DxC 500 AU Clinical<br>Chemistry Analyzer) | | Specimen Type | Serum, EDTA or Lithium heparin<br>plasma, and cerebrospinal fluid | Same | | Calibration | Serum Protein Multi-Calibrator<br>(Cat # ODR3021) | Same | | Onboard Stability | 90 Days | Same | | Calibration<br>Stability | Serum &<br>Plasma<br>90 Days | Same | | | CSF<br>2 Days | Same | | Analytical<br>Range | Serum &<br>Plasma<br>75 - 3000 mg/dL | Same | | | CSF<br>2- 50 mg/dL | Same | | Precision | Serum &<br>Plasma<br>Within-run Precision:<br>≤ 3.5% CV<br>Total Precision:<br>< 6% CV | Same | | | CSF<br>Within-run Precision:<br>≤ 6% CV or ≤0.4 mg/dL<br>Total Precision:<br>< 7.5% CV or <0.5mg/dL | Same | | Sensitivity | Serum &<br>Plasma<br>LoQ: <75 mg/dL using a within<br>laboratory CV of 20% | Same | | | CSF<br>LoQ: <2 mg/dL using a within<br>laboratory CV of 20% | Same | | Interference | Serum &<br>Plasma<br>Bilirubin: NSI* up to 40 mg/dL<br>Hemolysis: NSI up to 500 mg/dL<br>Lipemia: NSI up to 1000 mg/dL<br>RF: NSI up to 1200 IU/mL | Same | | | CSF<br>Bilirubin: NSI up to 36 mg/dL<br>Hemolysis: NSI up to 500 mg/dL | Same | # Table 7.0 - IgG Predicate Device Comparison Table {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red oval shape with two curved white lines inside, followed by the company name "BECKMAN COULTER" in bold, black font. The word "BECKMAN" is on the top line and "COULTER" is on the bottom line. *NSI – No significant interference is recovery within 10% of the initial value {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red oval shape with two white curved lines inside, placed to the left of the company name. The text "BECKMAN COULTER" is written in bold, black letters, with "BECKMAN" on the top line and "COULTER" on the bottom line. #### 8.0 Comparison testing Substantial equivalence is demonstrated through non-clinical (bench) studies as shown below. In accordance with FDA's Guidance for Industry and FDA Staff -Replacement Reagent and Instrument Family Policy, CLSI study protocols were used to verify the performance claims stated in the reagent IFU and ensure that the technological differences between the candidate and predicate analyzer models did not adversely affect safety and effectiveness. - Method Comparison Linearity Sensitivity Reference Interval Interference Precision #### 9.0 Summary of Performance Data The data contained in the Premarket Notification supports a finding of substantial equivalence to the measurand test systems already in commercial distribution. #### 9.1 Method Comparison with Predicate Device Method comparison and bias estimation experiments were designed using CLSI Guideline EP09C-ED3 "Measurement Procedure Comparison and Bias Estimation Patient Samples; Third Edition". These patient correlation studies Using demonstrate equivalence between Beckman Coulter's IgG assay on the candidate DxC 500 AU analyzer and the predicate DxC 700 AU analyzer. The results summary in Table 9.1. is based on Weighted Deming regression analysis. | Sample Type | Units | N | Slope | Intercept | R | |-------------|-------|-----|-------|-----------|--------| | Serum | mg/dL | 147 | 1.015 | -25.422 | 0.9981 | | CSF | mg/dL | 114 | 0.998 | 0.1141 | 0.9995 | | | Table 9.1 IqG Method Comparison Data Summary | | | |--|----------------------------------------------|--|--| {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, and the words "BECKMAN COULTER" in black, with "BECKMAN" on top of "COULTER". The font is bold and sans-serif. #### Linearity/Reportable Range: 9.2 Analytical range (linearity) studies were designed to meet the requirements of CLSI quideline EP06-A "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline". High and low pools were prepared and inter-diluted to achieve concentrations spanning the required linear range. Each dilution was assayed in quadruplicate on a DxC 500 AU analyzer. The performance data for the study demonstrates linearity throughout the claimed dynamic range of the IgG candidate assay, as represented in Table 9.2 below. | | Acceptance Criteria | | | Results | | | |----------------|---------------------|--------------------------------|------------------------------------|----------------|--------------|---------------| | Sample<br>Type | Linear<br>Range | Allowable<br>Difference<br>± % | Allowable<br>Difference<br>± units | Linear<br>From | Linear<br>To | Pass/<br>Fail | | Serum | 75-3,000 | 8% between<br>375-3,000 | 30 mg/dL<br>between<br>75 - 375 | 73.2868 | 3261.9190 | Pass | | CSF | 2-50 | 10% between<br>2-50 | 0.5 mg/dL<br>between<br>2.0 - 5 | 1.9 | 53.0 | Pass | #### 9.3 Sensitivity (Detection Limits): LOB, LOD and LOQ studies were designed in accordance with CLSI quideline EP17-A2 "Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures Approved Guideline - Second Edition". For the LOB, LOD, and LOQ determinations, the experimental design consisted of replicate measurements on blank and low-level sample pools using 2 lots of reagent across multiple days. A total of 72 blank replicates were generated per reagent lot (4 blank native serum samples 'analyte-depleted' ran n=6 for 3 days) to generate the LOB value that serves as a component of the LOD determination. For the LOD and LOQ determinations, 500 replicates were generated for each reagent lot for each application; this was comprised of 10 low-level samples for IgG which were run 10fold for 5 days. The LOB, LOD, and LOQ sample preparation information, including the matrices of the spiking solutions, are describe in Table 9.3 below | Sample<br>Type | Units | Low End of<br>Measuring<br>Range | LOQ Spec | LOB | LOD | LOQ | |----------------|-------|----------------------------------|---------------|------|------|----------------| | Serum | mg/dL | 75 | <75 at ≤20%CV | 5.6 | 8.6 | 18.5 at 10% CV | | CSF | mg/dL | 2 | <2 at ≤20%CV | 0.11 | 0.31 | 0.63 at 20% CV | Table 9.3 IgG Sensitivity Data Summary {8}------------------------------------------------ Image /page/8/Picture/0 description: The image contains the logo for Beckman Coulter. The logo consists of a red oval shape with two white curved lines inside, placed to the left of the company name. The text "BECKMAN" is stacked on top of "COULTER" in a bold, sans-serif font. #### 9.4 Precision: Repeatability (within-run) and within- laboratory (total) precision studies were performed in accordance with the CLSI guideline EP05-A3 "Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline - Third Edition" using a 20-day study design. A separate 5-day study design was used to evaluate reproducibility using three DxC 500 AU analyzers. The candidate IgG assay met the performance specifications as stated in the IgG reagent IFU. The results are summarized in the Tables 9.4.1 and 9.4.2 below: | Sample<br>(Units) | Sample<br>Levels | Mean<br>(n=80) | Repeatability<br>(Within Run) | | Within Laboratory<br>(Total) | | |-------------------|------------------|----------------|-------------------------------|-----|------------------------------|-----| | Serum<br>(mg/dL) | 1 | 503.99 | 5.8 | 1.2 | 6.3 | 1.2 | | Serum<br>(mg/dL) | 2 | 1797.53 | 43.1 | 2.4 | 43.2 | 2.4 | | Serum<br>(mg/dL) | 3 | 2576.22 | 86.9 | 3.4 | 94.9 | 3.7 | | Serum<br>(mg/dL) | 4 | 88.31 | 1.6 | 1.8 | 4.3 | 4.9 | | Serum<br>(mg/dL) | 5 | 835.56 | 7.2 | 0.9 | 14.8 | 1.8 | | CSF<br>(mg/dL) | 1 | 4.14 | 0.2 | 4.7 | 0.2 | 5.6 | | CSF<br>(mg/dL) | 2 | 11.20 | 0.2 | 2.1 | 0.4 | 3.1 | | CSF<br>(mg/dL) | 3 | 36.32 | 0.6 | 1.6 | 0.9 | 2.6 | Table 9.4.1 IqG 20-day Precision Data Summary Table 9.4.2 IgG 5-Day Precision Data Summary | Sample<br>(Units) | Sample<br>Levels | Mean<br>(n=75) | Repeatability<br>(Within Run) | Within<br>Laboratory<br>(Total) | Reproducibility | |-------------------|------------------|----------------|-------------------------------|---------------------------------|-----------------| | | 1 | 481.20 | 0.8% CV | 1.0% CV | 1.0% CV | | | | | 3.67 SD | 4.63 SD | 4.63 SD | | Serum<br>(mg/dL) | 2 | 1669.57 | 1.5% CV | 1.7% CV | 1.7% | | | | | 24.45 SD | 29.04 SD | 29.07 SD | | | 3 | 2404.48 | 2.8% CV | 3.6 % CV | 3.6 % CV | | | | | 66.69 SD | 86.08 SD | 86.08 SD | | | 1 | 3.53 | 1.7% CV, | 4.4% CV | 4.7% CV | | | | | 0.06 SD | 0.16 SD | 0.16 SD | | CSF<br>(mg/dL) | 2 | 11.31 | 0.9% CV, | 1.9% CV | 1.9% CV | | | | | 0.10 SD | 0.22 SD | 0.22 SD | | | 3 | 37.18 | 0.9% CV. | 1.4% CV | 1.4% CV | | | | | 0.33 SD | 0.50 SD | 0.50 SD | {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, and the words "BECKMAN COULTER" in black, with "BECKMAN" on top of "COULTER". The logo is simple and modern, and the colors are eye-catching. #### 9.5 Interferences (Analytical Specificity): Interference studies were designed based on the CLSI Guideline EP07, 3d Edition "Interference Testing in Clinical Chemistry; Approved Guideline". All test samples were assayed n=5 at two analyte levels. The sample pools tested were at different levels of interferents to determine the magnitude of their effect. The data analysis involved calculating the difference in recovery of the samples with and without the potential interfering substances. The results are summarized in Table 9.5 below. The candidate IgG assay met the performance specifications as stated in the IgG reagent IFU. | Sample | Interference Threshold | | | | | |--------|-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|---------------| | | Lipemic<br>(Intralipid)1 | Icteric<br>Unconjugated2 | Hemolytic3 | RF | Pass/<br>Fail | | Serum | Intralipid<br>(1000 mg/dL)<br>intf. ≤10%<br>at IgG conc. of<br>1000 mg/dL &<br>2000 mg/dL | Bilirubin<br>(40 mg/dL)<br>intf. ≤10%<br>at IgG conc. of<br>1000 mg/dL &<br>2000 mg/dL | Hemolysate<br>(500 mg/dL)<br>intf. ≤10%<br>at IgG conc. of<br>1000 mg/dL &<br>2000 mg/dL | RF<br>(1200 IU/mL)<br>intf. ≤10%<br>at IgG conc. of<br>1000 mg/dL &<br>2000 mg/dL | Pass | | CSF | N/A | Bilirubin<br>(36 mg/dL)<br>intf. ≤10%<br>at IgG conc. of<br>5 mg/dL & 20<br>mg/dL | Hemolysate<br>(500 mg/dL)<br>intf. ≤10%<br>at IgG conc. of<br>5 mg/dL & 20<br>mg/dL | N/A | Pass | # Table 9.5 IqG Interferences Data Summary 1 Intralipid is a 20% fat emulsion used to emulate extremely turbid samples. 2 Unconjugated bilirubin (porcine source) 3 Lysed human red blood cells {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the logo for Beckman Coulter. The logo consists of a red circle with two white curved lines inside, followed by the company name in black, bold letters. The word "BECKMAN" is on top of the word "COULTER". #### 9.6 Reference Interval The Reference Interval study utilized a transference approach in accordance with the CLSI guidelines EP028-A3c "Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition". The purpose of the study was to validate the acceptability of the original IgG serum reference interval for use on the candidate DxC 500 AU analyzer. Transference evaluations were not performed for the CSF sample type. The results are summarized in Table 9.6 below. | Sample Type | Reference Range (mg/dL) | Result | |------------------|----------------------------------------------------------------------------------------------------------------------|--------| | Serum<br>(Adult) | 635 - 1,741 | Pass | | CSF | 15 – 20 years: 3.5 mg/dL ± 2.0 mg/dL<br>21 - 40 years: 4.2 mg/dL ± 1.4 mg/dL<br>41 - 60 years: 4.7 mg/dL ± 1.0 mg/dL | N/A* | # Table 9.6 IgG Reference Interval Summary *Literature reference used #### 10.0 Conclusion The IgG Reagent on the DxC 500 AU Clinical Chemistry Analyzer is identical in design and composition as the IgG Reagent on the DxC 700 AU Clinical Chemistry Analyzer cleared under K162208. Method comparison, linearity, sensitivity, reference interval, interference, and precision testing demonstrate that the assay performance is substantially equivalent between the candidate predicate test systems. This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing requlation 21 CFR 807.92.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...