← Product Code [KWQ](/productcode/KWQ) · K221053

# Magnes-C Anterior Cervical Plate System (K221053)

_Fellowship of Orthopaedic Researchers, Inc. · KWQ · Mar 1, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221053

## Device Facts

- **Applicant:** Fellowship of Orthopaedic Researchers, Inc.
- **Product Code:** [KWQ](/productcode/KWQ.md)
- **Decision Date:** Mar 1, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Magnes-C™ Anterior Cervical Plate System is intended for anterior screw fixation of the cervical spine (C2 to T1). The system is to be used as an adjunct to fusion for the following indications: degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumors, pseudarthrosis or failed previous fusion.

## Device Story

Magnes-C Anterior Cervical Plate System provides anterior screw fixation and stabilization of cervical spine (C2-T1) during interbody spinal fusion. System comprises one- to three-level plates and fixed/variable screws in various sizes to accommodate patient anatomy; includes instruments for unilateral anterior fixation. Used by surgeons in clinical settings to stabilize spine; facilitates fusion process. Benefits patient by providing mechanical support during bone healing. Non-sterile delivery; requires sterilization by user.

## Clinical Evidence

Bench testing only. Mechanical testing of worst-case construct performed per ASTM F1717, including static and dynamic compression and static torsion. MR compatibility testing performed per ASTM F2503.

## Technological Characteristics

Implants manufactured from titanium alloy (ASTM F136). Instruments manufactured from medical grade stainless steel (ASTM F899) with silicone or Radel handles. System includes plates and fixed/variable screws. Non-sterile. MR compatible per ASTM F2503.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Zavation Cervical Plate System ([K130030](/device/K130030.md))
- Cervical Spine Locking Plate (CSLP) ([K945700](/device/K945700.md))
- Biomet MaxAn® Anterior Cervical Plate System ([K133518](/device/K133518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 1, 2023

Fellowship of Orthopaedic Researchers, Inc. % Dr. Karen Warden President BackRoads Consulting Inc. PO Box 566 Chesterland, Ohio 44026

Re: K221053

Trade/Device Name: Magnes-C Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: January 24, 2023 Received: January 26, 2023

Dear Dr. Karen Warden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

542 of the Act); 21 CFR 1000-1050.

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/5 description: The image contains the name "Colin O'neill -S" in a simple, sans-serif font. The text is arranged vertically, with "Colin" on the top line and "O'neill -S" on the bottom line. The text is black and the background is white.

Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

## K221053

Device Name

Magnes-CTM Anterior Cervical Plate System

#### Indications for Use (Describe)

The Magnes-C™ Anterior Cervical Plate System is intended for anterior screw fixation of the cervical spine (C2 to T1). The system is to be used as an adjunct to fusion for the following indications: degenerative disc disease (defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis, deformity (i.e., kyphosis, lordosis or scoliosis), tumors, pseudarthrosis or failed previous fusion.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                              |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="padding-right: 20px;">                      Prescription Use (Part 21 CFR 801 Subpart D)                 </span> <span>                      Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

 PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

# 510(k) Summary

| Date:                                                | 28 February 2023                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:                                             | Fellowship of Orthopaedic Researchers, Inc.<br>320 Metairie Hammond Highway, Suite 406<br>Metairie, LA 70005<br>504-309-9845                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Sponsor Contact:                                     | Stephen D. Cook, PhD, Executive Director                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| 510(k) Contact:                                      | Karen E. Warden, PhD<br>BackRoads Consulting<br>PO Box 566<br>Chesterland, OH 44026<br>Office: 440.729.8457                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Trade Name:                                          | Magnes-C™ Anterior Cervical Plate System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Common Name:                                         | Anterior cervical plate system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Regulatory Class:                                    | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification Name,<br>Regulation, Product<br>Code: | Appliance, fixation, spinal intervertebral body, 888.3060, KWQ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device Description:                                  | The Magnes-C™ Anterior Cervical Plate System components are intended<br>for anterior screw fixation and stabilization of the cervical spine during the<br>development of cervical interbody spinal fusion. The Magnes-C™ Anterior<br>Cervical Plate System consists of one through three-level plates and fixed<br>and variable screws, in a variety of sizes to accommodate individual patient<br>anatomy. The Magnes-C components are provided nonsterile. Instruments<br>to facilitate unilateral anterior fixation are included.                                                   |
| Indications for Use:                                 | The Magnes-C™ Anterior Cervical Plate System is intended for anterior<br>screw fixation of the cervical spine (C2 to T1). The system is to be used as<br>an adjunct to fusion for the treatment of the following indications:<br>degenerative disc disease (defined as neck pain of discogenic origin with<br>the degeneration of the disc confirmed by history and radiographic studies),<br>spondylolisthesis, trauma (i.e., fractures or dislocations), spinal stenosis,<br>deformity (i.e., kyphosis, lordosis or scoliosis), tumors, pseudarthrosis or<br>failed previous fusion. |
| Materials:                                           | The patient contacting portion of the Magnes-C™ Anterior Cervical Plate<br>System implants are manufactured from titanium alloy as described by<br>ASTM F136. All instruments are manufactured from medical grade stainless<br>steels according to ASTM F899, some of which have silicone or Radel<br>handles.                                                                                                                                                                                                                                                                         |
| Primary Predicate:                                   | Zavation Cervical Plate System (Zavation LLC – K130030)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Additional Predicates:                               | Cervical Spine Locking Plate (CSLP) (Synthes Spine – K945700), Biomet<br>MaxAn® Anterior Cervical Plate System (Biomet Spine LLC – K133518)                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Performance Data:                                    | Mechanical testing of the worst case Magnes-C™ Anterior Cervical Plate<br>System construct was performed according to ASTM F1717 and included<br>static and dynamic compression and static torsion.<br>The mechanical test results demonstrate that the Magnes-C™ Anterior<br>Cervical Plate System device performance is substantially equivalent to the<br>predicate devices.<br>Additionally, MR Compatibility testing per ASTM F2503 was performed. The<br>results demonstrate that the Magnes-C™ Anterior Cervical Plate System                                                   |

{4}------------------------------------------------

can be safely scanned in an MR system.

Technological The Magnes-C™ Anterior Cervical Plate System possesses the many of the same technological characteristics as the predicate devices. These include Characteristics: basic design, material, method of stabilization and anatomic location. Differences between the subject and predicate devices did not raise new questions of safety and effectiveness. Therefore the fundamental scientific technology of the Magnes-C™ Anterior Cervical Plate System devices is similar to previously cleared devices. The Magnes-C™ Anterior Cervical Plate System possesses the same Conclusion:

intended use and similar technological characteristics as the predicate devices. Therefore the Magnes-C™ Anterior Cervical Plate System is substantially equivalent for its intended use.

---

**Source:** [https://fda.innolitics.com/device/K221053](https://fda.innolitics.com/device/K221053)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
