← Product Code [KLE](/productcode/KLE) · K221033

# UniPrime A, UniPrime E, UniPrime E/C (K221033)

_S&C Polymer Silicon- Und Composite Spezialitäten GmbH · KLE · Jul 6, 2022 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K221033

## Device Facts

- **Applicant:** S&C Polymer Silicon- Und Composite Spezialitäten GmbH
- **Product Code:** [KLE](/productcode/KLE.md)
- **Decision Date:** Jul 6, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3200
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

UniPrime A: UniPrime A is a MDP containing universal primer 1) for the bonding of methacrylate based dental materials to zirconia and metal/alloy frameworks, crowns and bridges and 2) for the intraoral restorations made of zirconia and metal/alloy. UniPrime E: UniPrime E is a MDP containing universal primer 1) for the bonding of methacrylate based dental materials to zirconia, composite and metal/alloy frameworks, crowns and bridges and 2) for the intraoral repair of fractured restorations made of zirconia, composite and metal/allov. UniPrime E/C: UniPrime E/C is a MDP containing univer 1) for the bonding of methacrylate based dental materials to ceramic, zirconia, composite and metal/alloy frameworks, crowns and bridges and 2) for the intraoral restorations made of ceramic, zirconia, composite and metal/alloy.

## Device Story

UniPrime A, E, and E/C are MDP-containing universal primers; used by dental professionals in clinical settings to enhance adhesion between methacrylate-based dental materials and prosthetic surfaces (ceramic, zirconia, composite, metal/alloy). Primers are applied to restoration surfaces to facilitate bonding during crown/bridge placement or intraoral repair of fractured restorations. The chemical composition (solvent, MDP, silane) acts as an adhesive interface; improves bond strength; ensures structural integrity of dental restorations. Benefits include reliable adhesion for various restorative materials; simplified clinical workflow for repairs and cementation.

## Clinical Evidence

Bench testing only. Evaluations included adhesive strength to indicated surfaces, pH, and biocompatibility testing.

## Technological Characteristics

Universal dental primers containing MDP, silane, and solvent. Delivered in bottles. Chemical-based adhesive mechanism. No electronic components, software, or energy sources.

## Regulatory Identification

A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.

## Predicate Devices

- Clearfil Ceramic Primer Plus ([K150703](/device/K150703.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 6, 2022

S&C Polymer Silicon- und Composite Spezialitäten GmbH Christian Böttcher Reg. Compliance Officer, Official Correspondent to FDA Robert-Bosch-Str. 2 Elmshorn, 25335 Germany

Re: K221033

Trade/Device Name: UniPrime A, UniPrime E, UniPrime E/C Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: March 31, 2022 Received: April 7, 2022

Dear Christian Böttcher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

# K221033

Device Name UniPrime A UniPrime E UniPrime E/C

### Indications for Use (Describe)

UniPrime A

UniPrime A is a MDP containing universal primer 1) for the bonding of methacrylate based dental materials to zirconia and metal/alloy frameworks, crowns and bridges and 2) for the intraoral restorations made of zirconia and metal/alloy.

UniPrime E

UniPrime E is a MDP containing universal primer 1) for the bonding of methacrylate based dental materials to zirconia, composite and metal/alloy frameworks, crowns and bridges and 2) for the intraoral repair of fractured restorations made of zirconia, composite and metal/allov.

UniPrime E/C

UniPrime E/C is a MDP containing univer 1) for the bonding of methacrylate based dental materials to ceramic, zirconia, composite and metal/alloy frameworks, crowns and bridges and 2) for the intraoral restorations made of ceramic, zirconia, composite and metal/alloy.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

PSC Publishing Sensces (301) 443-6740

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

# 510(k) Summary - K221033

#### 1. Submitter

| Name of company: | S&C Polymer Silicon- und Composite Spezialitäten GmbH |
|------------------|-------------------------------------------------------|
| Address:         | Robert-Bosch-Str. 2, 25335 Elmshorn, Germany          |
| Phone:           | 0049 4121 483 0                                       |
| Fax:             | 0049 4121 483 184                                     |
| Contact person:  | Dr. Christian Böttcher                                |
| Date prepared:   | March 31st 2022                                       |

#### 2. Device name

| Trade name:                 | UniPrime A<br>UniPrime E<br>UniPrime E/C |
|-----------------------------|------------------------------------------|
| Common name:                | Universal primer                         |
| Device classification name: | Agent, tooth bonding resin               |
| Regulatory number:          | 872.3200                                 |
| Product code:               | KLE                                      |

#### 3. Device description

UniPrime A is a MDP containing universal primer for the bonding of methacrylate based dental materials to zirconia and metal/alloy frameworks, crowns and bridges (indications from labeling).

UniPrime E is a MDP containing universal primer for the bonding of methacrylate based. dental materials to zirconia, composite and metallalloy frameworks, crowns and bridges (indications from labeling).

UniPrime E/C is a MDP containing universal primer for the bonding of methacrylate based dental materials to ceramic, zirconia, composite and metal/alloy frameworks, crowns and bridges (indications from labeling).

#### 4. Intended use of the devices

UniPrime A improves the bonding of methacrylate based dental materials to zirconia and metal/alloy frameworks, crowns and bridges.

UniPrime E improves the bonding of methacrylate based dental materials to zirconia, composite and metal/alloy frameworks, crowns and bridges.

UniPrime E/C improves the bonding of methacrylate based dental materials to ceramic, zirconia, composite and metal/alloy frameworks, crowns and bridges.

{4}------------------------------------------------

#### 5. Indication for use of the devices

UniPrime A is a MDP containing universal primer 1) for the bonding of methacrylate based dental materials to zirconia and metal/alloy frameworks, crowns and bridges and 2) for the intraoral repair of fractured restorations made of zirconia and metal/alloy.

UniPrime E is a MDP containing universal primer 1) for the bonding of methacrylate based dental materials to zirconia, composite and metal/alloy frameworks, crowns and bridges and 2) for the intraoral repair of fractured restorations made of zirconia, composite and metal/alloy.

UniPrime E/C is a MDP containing universal primer 1) for the bonding of methacrylate based dental materials to ceramic, zirconia, composite and metal/alloy frameworks, crowns and bridges and 2) for the intraoral repair of fractured restorations made of ceramic, zirconia, composite and metal/alloy.

#### 6. Device for which substantial equivalence is claimed

| Subject device:   | UniPrime A                                   |
|-------------------|----------------------------------------------|
|                   | UniPrime E                                   |
|                   | UniPrime E/C                                 |
| Predicate device: | Clearfil Ceramic Primer Plus (Kuraray)       |
| 510(k) Number:    | K150703 (introduced into the US-Market 2015) |

#### 6.1. Device comparison with the predicate device

The predicate device has been found to be substantially equivalent under the 510(k) premarket notification as class II dental device under CFR 872.3200 product code KLE.

|                    | Subject devices<br>UniPrime A<br>UniPrime E<br>UniPrime E/C                                                                                                                               | Predicate device<br>Clearfil Ceramic Primer<br>Plus (Kuraray)                                                                                                                          |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device description | MDP containing universal<br>primer for the bonding of<br>methacrylate based dental<br>materials to ceramic,<br>zirconia, composite and<br>metal/alloy frameworks,<br>crowns and bridges.* | Clearfil Ceramic Primer Plus                                                                                                                                                           |
| Intended use       | UniPrime improves the<br>bonding of methacrylate<br>based dental materials<br>to ceramic, zirconia,<br>composite and metal/alloy<br>frameworks, crowns and<br>bridges.*                   | Clearfil Ceramic Primer Plus is<br>a dental universal prosthetic<br>primer that provides an<br>enhanced adhesive surface to<br>ceramic, hybrid ceramics,<br>composite resin and metal. |
| Prescription use   | yes                                                                                                                                                                                       | yes                                                                                                                                                                                    |

{5}------------------------------------------------

| Mechanism of action | equal                                                                                                                                                        | equal                                                                                                                                                                                                                                                                                                          |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Form of delivery    | bottle                                                                                                                                                       | bottle                                                                                                                                                                                                                                                                                                         |
| Indications for use | UniPrime II improves the<br>bonding of methacrylate<br>based dental materials<br>to ceramic, zirconia and<br>metal/alloy frameworks,<br>crowns and bridges.* | Clearfil Ceramic Primer Plus<br>is indicated for the following<br>uses:<br>-Surface treatment of<br>prosthetic restorations made of<br>ceramic, hybrid ceramics,<br>composite resin or metal<br>Intraoral repairs of fractured<br>Restorations made of ceramic<br>hybrid ceramics, composite<br>resin or metal |
| Ingredients         | Solvent<br>MDP<br>Silane *                                                                                                                                   | Solvent<br>MDP<br>Silane*                                                                                                                                                                                                                                                                                      |

* product specific information

#### 7. 1 Conclusion

The comparison worked out above and the further elaboration of information within this 510(k) submission demonstrate that the subject devices UniPrime A, UniPrime E and UniPrime E/C are substantially equivalent to the predicate device Clearfil Ceramic Primer Plus (Kuraray) in terms of description, intended use, indications for use, chemical composition and physical properties. The subject device has been evaluated for adhesive strength to indicated surfaces, pH, and biocompatiblity. The information given above do not raise different questions of safety and effectiveness. The devices are as safe and effective as the predicate device.

---

**Source:** [https://fda.innolitics.com/device/K221033](https://fda.innolitics.com/device/K221033)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
