K221024 · Thorasys Thoracic Medical Systems, Inc. · PNV · Jun 20, 2023 · Anesthesiology
Device Facts
Record ID
K221024
Device Name
tremoflo C2 Airwave Oscillometry System
Applicant
Thorasys Thoracic Medical Systems, Inc.
Product Code
PNV · Anesthesiology
Decision Date
Jun 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The tremoflo C2 Airwave Oscillometry System is intended to measure respiratory system impedance using the Forced Oscillation Technique (FOT). The tremoflo C2 Airwave Oscillometry System is intended for use with pediativ and adult patients 4 years of age or older. The device is designed to be used by pulmonologists, general practitioners, nurses, respiratory therapists, laboratory technologists, medical researchers and similarly trained personnel in hospitals, clinics, and private physician offices.
Device Story
Portable lung function testing device using Forced Oscillation Technique (FOT). Device superimposes small sinusoidal pressure waves (1-3 cmH2O) onto patient's normal breathing. Input: patient mouth pressure and airflow response. Output: mechanical impedance of respiratory system (complex ratio of pressure to airflow) and breathing pattern parameters. Used in hospitals, clinics, and private offices by clinicians (pulmonologists, nurses, respiratory therapists). Handheld unit contains electronics, sensors, and actuator. Patient breathes through device with nose clip. Data transferred to computer via USB, Wi-Fi, or Bluetooth for analysis and presentation. Output assists clinicians in assessing lung mechanics and respiratory function.
Clinical Evidence
Bench testing only. Comparative performance testing demonstrated equivalent oscillometry performance between subject and predicate. Parameters tested: resistance, reactance, and impedance accuracy; intra-device repeatability for resistance, reactance, impedance, and tidal volume. Reproducibility and repeatability variations < 3.0%. Compliance with electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), biocompatibility (ISO 10993), and gas pathway emission standards confirmed.
Technological Characteristics
Handheld device; Forced Oscillation Technique (FOT) using sinusoidal pressure waves (5-41 Hz). Components: pneumotach, static mesh, piston-based actuator. Connectivity: USB, Wi-Fi, Bluetooth. Power: 3.7V rechargeable battery or AC/DC adapter. Biocompatibility: ISO 10993 compliant. Software: standalone package for data analysis/management. Sterilization: non-sterile, validated for low-level disinfection.
Indications for Use
Indicated for pediatric and adult patients 4 years of age or older requiring respiratory system impedance measurement via Forced Oscillation Technique (FOT).
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
Predicate Devices
tremoFlo C100 Airwave Oscillometry System (K170185)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
June 20, 2023
Thorasys Thoracic Medical Systems, Inc. Christopher McLean RA/QA Representative 6560 Avenue de l'Esplanade Suite #103 Montreal, Quebec H2V 4L5 Canada
Re: K221024
Trade/Device Name: tremoflo C2 Airwave Oscillometry System Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: PNV Dated: May 18, 2023 Received: May 18, 2023
Dear Christopher McLean:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221024
Device Name tremoflo C2 Airwave Oscillometry System
#### Indications for Use (Describe)
The tremoflo C2 Airwave Oscillometry System is intended to measure respiratory system impedance using the Forced Oscillation Technique (FOT). The tremoflo C2 Airwave Oscillometry System is intended for use with pediativ and adult patients 4 years of age or older. The device is designed to be used by pulmonologists, general practitioners, nurses, respiratory therapists, laboratory technologists, medical researchers and similarly trained personnel in hospitals, clinics, and private physician offices.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "THORASYS" in a bold, sans-serif font. The letters are a golden yellow color, and the word is slightly angled upwards from left to right. There is a trademark symbol in the upper right corner of the image. The background is white.
# 510(k) Summary
#### Submitter Information:
| Company Name: | THORASYS Thoracic Medical Systems, Inc. |
|------------------------|-----------------------------------------------------------------|
| Company Address: | 6560 Avenue de l'Esplanade Suite #103<br>Quebec, Canada H2V 4L5 |
| Contact Person: | Christopher McLean |
| Date Summary Prepared: | June 14, 2023 |
| Trade Name: | tremoflo C2 Airwave Oscillometry System |
| Common/Usual Name: | Respiratory Impedance Measurement Device |
| Classification Name: | Impedance Measuring Device Utilizing Oscillation Techniques |
| Product Code: | PNV |
| Device Class: | Class II |
| Regulation Number: | 21 CFR 868.1840 |
#### Predicate Device:
| Manufacturer's Name | Brand Name | 510(k) Number |
|-----------------------------------------|----------------------------------------------|---------------|
| THORASYS Thoracic Medical Systems, Inc. | tremoFlo C100 Airwave<br>Oscillometry System | K170185 |
### Device Description:
The tremoflo C2 Airwave Oscillometry System (tremoflo C2) is a portable lung function testing device that implements methods known per the Forced Oscillation Technique (FOT), or oscillometry, to assess lung function in humans. The Forced Oscillation Technique (FOT) is a non-invasive test that provides a full report of lung mechanics per the FOT. In general, the FOT usually consists of superimposing given external multi-frequency sinusoidal excitation small pressure waves (1-3 cmH₂O peak) onto the normal breathing of the patient through the device and then deriving the mechanical properties from the patient's mouth pressure and airflow response while breathing. The main outcome reported by FOT is the mechanical impedance of the respiratory system which is the complex ratio between pressure and airflow at the given excitation frequencies.
The main tremoflo C2 Unit is a lightweight handheld device. It contains electronics, pressure and flow sensors, and the actuator providing the forced oscillations. The tremoflo software is a complete standalone software package for patient management, testing, result analysis, and presentation.
{4}------------------------------------------------
During use the operator holds the handheld device using the ergonomic handle while the patient is seated, wearing a standard nose clip and with hands on cheeks. To perform the test, the patient then breathes quietly through the device into a standard single use Pulmonary Function Testing (PFT) filter connected at the front of the C2 Unit via the PFT filter interface. The tremoflo C2 is not intended to be used as a stand-alone diagnostic device.
As compared to the tremoFlo C100 Airwave Oscillometry System predicate (tremoflo C100), the modifications are summarized as follows:
- . Changes to the enclosures and the internal hardware and related firmware components such to integrate the two main units of the prior tremoflo C100 version (i.e., the Handheld Unit and the Cradle Unit in the C100) into a single main handheld unit with a new display screen to show the breathing traces and help the user connect their computer to the tremoflo.
- Changes to include Wi-Fi and Bluetooth wireless interfaces and also a USB interface on the main unit for communication to the user's computer instead of the ethernet cable connection in the tremoflo C100
- . Changes to the software to accommodate the new communication modes and the hardware/firmware changes.
- Changes to include a rechargeable battery module to allow operation without connections to the mains supply as was required in the C100.
## Predicate Product Comparisons:
### Indications for Use - No Change
| Predicate: | Subject: |
|------------------------------------------------------------|-----------------------------------------------------------|
| tremoFlo C-100 Airwave Oscillometry System | tremoflo C2 Airwave Oscillometry System |
| The tremoFlo C-100 Airwave Oscillometry System is | The tremoflo C2 Airwave Oscillometry System is |
| intended to measure respiratory system impedance | intended to measure respiratory system impedance |
| using the Forced Oscillation Technique (FOT). tremoFlo | using the Forced Oscillation Technique (FOT). The |
| C-100 Airwave Oscillometry System is intended for use | tremoflo C2 Airwave Oscillometry System is intended |
| with pediatric and adult patients 4 years of age or older. | for use with pediatric and adult patients 4 years of age |
| The device is designed to be used by pulmonologists, | or older. The device is designed to be used by |
| general practitioners, nurses, respiratory therapists, | pulmonologists, general practitioners, nurses, |
| laboratory technologists, medical researchers and | respiratory therapists, laboratory technologists, medical |
| similarly trained personnel in hospitals, clinics, and | researchers and similarly trained personnel in hospitals, |
| private physician offices. | clinics, and private physician offices. |
{5}------------------------------------------------
# Predicate Product Comparison
| Technical Feature/<br>Specification | C100 (Predicate) | C2 (Subject) |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Fundamental Scientific<br>Technology | Forced Oscillation Technique and Pneumotach per ERS<br>FOT recommendations [1]. | Identical |
| Pneumotach Flow Range | ± 2.5 L/s | Identical |
| Flow Resolution | ±1.4 ml/s | Identical |
| Flow Linearity | ± 2% up to 1.0 L/s | Identical |
| Common Mode Rejection<br>Ratio (CMRR) | Dynamic software compensation signal processing | Identical |
| Device Load to Patient | 1.0 ± 5% cmH2O.s/L at 1 L/s | Modified: 1.1 ± 5% cmH2O.s/L at 1 L/s<br>Discussion: The difference in resistance to flow is small and within<br>the potential variation of resistance for bacterial/viral filters that<br>may be used with the tremoflo C100 system and does not raise<br>different questions of safety and effectiveness. |
| Volume Range | +3 Liters | Identical |
| Volume Accuracy | <3.0% or 0.050 L (whichever is greater) | Identical |
| Mouth Pressure (PM) | Piezo Resistive | Identical |
| Mouth Pressure Linearity | 2% of full Scale up to 5 cm H2O | Identical |
| Mouth Pressure Resolution | 0.0053 cmH2O (0.0039 mmHg) | 0.00265 cmH2O (0.00195 mmHg)<br>Discussion: improved resolution does not raise different questions<br>of safety and effectiveness. |
| Mouth Pressure Range | ± 10 cmH2O | Modified: ± 5 cmH20<br>Discussion: By design the pressure range in FOT measurements<br>never exceeds +/- 5 cm H2O. This is sufficient for human<br>measurements, does not change the performance specifications,<br>and does not raise different questions of safety and effectiveness. |
| Effective Device Dead<br>Space | 35 ml | Identical |
| Technical Feature/<br>Specification | C100 (Predicate) | C2 (Subject) |
| Test Signal / Frequency<br>range | Sinusoidal signal at specific frequencies, between 5-<br>41Hz | Identical |
| Pseudo-Random Noise<br>(PSRN) Stimulus | 5-37 Hz, 7-41 Hz | Identical |
| Single Frequency Stimuli<br>for a within-breath analysis<br>of respiratory impedance | Can offer 5 to 41 Hz | Identical |
| Multi-Frequency Stimulus<br>for a with-in breath<br>analysis of respiratory<br>impedance and an<br>estimation of the<br>frequency-dependence of<br>respiratory impedance | 5-11-19, 5 to 37 and 7 to 41 Hz | Identical |
| Energy Type | AC/DC Power adapter 110-240 V / 47-63Hz to 24 V DC | Modified: Integrated rechargeable battery pack, 3.7 volts<br>5000mAh 18.5Wh, charged via AC/DC Power adaptor 110-240 V to<br>5 V DC<br>Discussion: In addition to the option of being powered through a<br>medical grade power supply as in the C100, the C2 can also be<br>used in battery operation without the need to connect the device<br>to the mains via a power supply. No effect to functional<br>performance and does not raise different questions of safety and<br>effectiveness. |
| Compatible Bacterial/Viral<br>Filter and nose clip | Single use,<br>510(k) cleared PFT Filter (K111587) and nose clip | Identical |
| Patient Contact/<br>Biocompatibility | Externally communicating (Indirect), Tissue, limited<br>duration<br>Surface, Skin, limited duration for nose clip and inlet of<br>filter. | Identical.<br>The C2 was tested identically as in the C100 to ensure that the<br>indirect gas contact from their breathing flow pathways present<br>acceptable emissions of VOCs, CO, CO2, Ozone and particulate |
| Technical Feature/<br>Specification | C100 (Predicate) | C2 (Subject) |
| Calculated Impedance<br>Parameters | Total Resistance (Rtot)<br>Inspiratory Resistance (Rinsp)<br>Expiratory Resistance (Rexp)<br>Total Reactance (Xtot)<br>Inspiratory Reactance (Xinsp)<br>Expiratory Reactance (Xexp)<br>R5-R19, R5-R20<br>AX<br>Fres | Identical |
| Calculated Breathing<br>Pattern Parameters | Tidal Volume (Vt)<br>Inspiratory Time (Ti)<br>Expiratory Time (Te)<br>Respiratory Duty Cycle (Ti/Ttot)<br>Respiratory Rate (RR)<br>Mean Inspiratory Flow (Vt/Ti)<br>Mean Expiratory Flow (Vt/Te)<br>Ventilation (Ve) | Identical |
| Breathing Circuit | Includes a pneumotach with a mesh for the flow<br>measurement. A breath-through low resistance<br>Vibrating Mesh (patented) is placed in series with the<br>pneumotach and used to generate the oscillation<br>stimulus during the test. | Modified: includes a pneumotach with a mesh for the flow<br>measurement. A breath-through low resistance mesh (static) is<br>placed in series with the pneumotach at the exhaust. This circuit<br>connects in parallel to a space with a piston which provides the<br>oscillation stimulus during the test.<br>Discussion: Both include a pneumotach and mesh in the breathing<br>circuit, except that the mesh in the C2 is static with the piston<br>being separate in a closed line connected and providing the<br>oscillations to the breathing circuit as opposed to being combined<br>with the mesh in the C100 (vibrating mesh). The flow path and<br>component designs are sufficiently close to that in the predicate<br>device to provide the same input oscillatory excitation profiles and<br>the same respiratory response measurements in both two devices.<br>This does not raise different questions of safety and effectiveness. |
| Technical Feature/<br>Specification | C100 (Predicate) | C2 (Subject) |
| Test duration | A minimum of 3 measurements with a minimum<br>duration of 20 seconds each and 2 valid breaths<br>minimum per measurements for a total of 60 seconds<br>and 6 breaths minimum per test. | Identical |
| Electrical Safety<br>Electromagnetic<br>Compatibility | Compliant with<br>• IEC60601-1, Ed. 3.1<br>• IEC 60601-1-2, Ed 4 | Identical for:<br>• IEC60601-1, Ed. 3.1<br>• IEC 60601-1-2, Ed 4<br>Additionally tested for immunity to EM energy in RF range per:<br>• AIM 7351731 Rev 2.0<br>Added compliance to wireless communication standards:<br>• FCC Part 15, Subpart B<br>• coexistence test (EEE/ANSI C63.27-2017)<br>Added compliance to lithium battery standard:<br>• IEC/UL 62133 2012 and IEC 62133-2 :2017<br>Discussion: these are to ensure compliance to the standards for<br>the added wireless communication and rechargeable battery<br>functions. |
| Cleaning/<br>Disinfection | Non-sterile and Non-Critical device with validated and<br>labelled cleaning and low-level disinfection method per<br>FDA Guidance ("Reprocessing Medical Devices in Health<br>Care Settings: Validation Methods and Labeling", 2015) | Identical - Validated per FDA guidance as done for the C100 |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
[1]: Oostveen E, MacLeod D, Lorino H, et al. The forced oscillation technique in clinical practice: methodology, recommendations and future developments. Eur Respir J 2003.
{9}------------------------------------------------
## Discussion of Technological Differences
The main technology of the device is unchanged in the proposed to the C100 predicate. lt involves the same Forced Oscillation Technique (FOT) using the same methods, and the user selected multi-frequency sinusoidal excitation pressure profile specifications (i.e. the waveforms) are the same as in the C100.
The corresponding output and display of pulmonary function characteristics, the patient's respiratory system resistance and reactance at each sinusoidal frequency, the related or other derived outcome parameters (e.g., breathing rate, inspiratory/expiratory volume, etc.) are identical as in the C100 per the same corresponding unchanged Host Software functionalities and algorithms and as was described in the original C100 510(k) aside from the communication protocols and interface changes as described below.
While many of the subcomponents and the breathing circuit were redesigned or re-organized, the C2's performance specifications were re-verified and found to be the same or to be negligibly impacted as compared to the C100. The added wireless communication modes in addition to the predicate's wired communication are for the same functions as the predicate wired mode, which is solely for data and information transfer between the user's computer and the main device unit and do not involve any time critical or time sensitive transfer requirements. These were implemented and tested following telecommunications standards and the FDA guidance Radio Frequency Wireless Technology in Medical Devices. The added battery pack powers the C2 with DC power equivalently as in the C100 via its AC/DC power supply. It was also implemented and tested in compliance with safety standards. The software updates were implemented and tested similarly as in the C100 predicate.
### Performance Data:
- . Device performance measurement - comparative testing and results as noted above demonstrating equivalent oscillometry performance between the predicate and the proposed C2 device and compliance with the ERS FOT recommendations ([1] as noted above) for the impedance accuracy. The tested parameters included:
- -Resistance Accuracy
- -Reactance Accuracy
- ー Impedance Accuracy
- -Intra-Device Total Resistance Repeatability
- Intra-Device Total Reactance Repeatability -
- -Intra-Device Total Impedance Repeatability
- -Tidal Volume Accuracy
- -Intra-Device Tidal Volume Repeatability
- Reproducibility and Repeatability testing demonstrated consistency between the predicate and proposed C2 device with measurement variations < 3.0%.
- . Electrical safety and Electromagnetic Compatibility - as per current standards IEC 60601-1:2005+A1:2012 and IEC 60601-1-2:2014+A1:2020 and additionally per AIM 7351731:2017 as noted above.
- . Biocompatibility assessment and gas pathway assessment of volatile organic compounds (VOCs), particulates, ozone, CO, and CO2 emission levels per ISO 10993-1:2018, 10993-5:2009, ISO 10993-10:2010, ISO 10993-12:2013, ISO-18562-1/2/3:2017, 21 CFR 801.415, and National Air Quality and
{10}------------------------------------------------
OSHA limits. Contact Type and Duration: Externally communicating (Indirect), Tissue, limited duration (<24 hours); Surface, Intact Skin and Oral Mucosa, limited duration (<24 hours)
- Cleaning/Disinfection Validation Testing - validated as per the FDA guidance; Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (2015) and AAMI TIR30:2011.
- . Hardware performance validation tests – bench tests conducted to validate that hardware requirements were met.
- . Software verification and validation testing – per IEC 62304:2006+A1:2015 and FDA Guidances: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document, May, 2005; Off-The-Shelf Software Use in Medical Devices, September 1999; Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, October 2014).
## Conclusion:
Based on the above, the tremoflo C2 Airwave Oscillometry System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.