Accuryn Monitoring System, Accuryn SmartFoley IAP UO Temp Tray System 12 F, Accuryn SmartFoley IAP UO Temp Tray System 14 F
K221020 · Potrero Medical, Inc. · EZL · Jan 13, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K221020
Device Name
Accuryn Monitoring System, Accuryn SmartFoley IAP UO Temp Tray System 12 F, Accuryn SmartFoley IAP UO Temp Tray System 14 F
Applicant
Potrero Medical, Inc.
Product Code
EZL · Gastroenterology, Urology
Decision Date
Jan 13, 2023
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Indications for Use
The Accuryn Monitoring System is intended for use in the drainage and/or collection of urine, and in the monitoring of urine output and core body temperature, in degrees Fahrenheit and degrees Celsius. The Accuryn Monitoring System with the Accuryn Sensing Urinary Catheter (SmartFoley®) - IAP UO Temp is also intended for use in the monitoring of intraabdominal pressure. The measured pressures can be used as an aid in the diagnosis of intra-abdominal hypertension (IAH) and the associated clinical syndrome of abdominal compartment syndrome (ACS). The Accuryn Sensing Urinary Catheter is a single use device intended for short term use (less than 30 days).
Device Story
Accuryn Monitoring System comprises a sensing urinary catheter, urine collection set, and electronic monitor. Catheter features four lumens: urine drainage, retention balloon, thermistor for temperature, and pressure lumen for IAP. Monitor processes signals from catheter to display urine output, core temperature, and IAP. Used in hospitals (OR/ICU) by clinicians (surgeons, anesthesiologists, nurses). Provides real-time monitoring and trending data to assist in clinical decision-making for IAH/ACS diagnosis. Benefits include automated, continuous monitoring of critical physiological parameters, reducing manual measurement burden and improving diagnostic accuracy for abdominal complications.
Clinical Evidence
No clinical data was required; substantial equivalence was demonstrated through bench testing, design verification, and validation.
Technological Characteristics
Silicone 4-lumen urinary catheter (12F, 14F, 16F) with thermistor and pressure sensing. Electronic monitor for data processing and display. Standards: ASTM F623, ISO 20696, IEC 60601-1, IEC 60601-1-2, IEC 80601-2-49, ISO 80601-2-56, ISO 10993. MR conditional. Software-based monitoring.
Indications for Use
Indicated for patients requiring urine drainage/collection, urine output monitoring, core body temperature monitoring, and intra-abdominal pressure (IAP) monitoring. Used as an aid in diagnosing intra-abdominal hypertension (IAH) and abdominal compartment syndrome (ACS). Intended for short-term use (<30 days) in hospital settings including operative or intensive care environments.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 13, 2023
Potrero Medical, Inc. Priscila Tapia Director of Regulatory and Quality Assurance 26142 Eden Landing Road Hayward, CA 94545
K221020 Re: Trade/Device Name: Accuryn Monitoring System Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL, EXY, PHU Dated: December 12, 2022 Received: December 13, 2022
Dear Priscila Tapia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jessica K. Nguyen -S
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K221020
Device Name Accuryn Monitoring System
#### Indications for Use (Describe)
The Accuryn Monitoring System is intended for use in the drainage and/or collection of urine, and in the monitoring of urine output and core body temperature, in degrees Fahrenheit and degrees Celsius. The Accurvn Monitoring System with the Accuryn Sensing Urinary Catheter (SmartFoley®) - IAP UO Temp is also intended for use in the monitoring of intraabdominal pressure. The measured pressures can be used as an aid in the diagnosis of intra-abdominal hypertension (IAH) and the associated clinical syndrome of abdominal compartment syndrome (ACS). The Accuryn Sensing Urinary Catheter is a single use device intended for short term use (less than 30 days).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
{3}------------------------------------------------
K221020 Page 1 of 7
# 510(k) Summary
This 510(k) summary was prepared in accordance with 21 CFR 807.92 Prepared on January 13, 2023
| 510(k) Submitter / Holder: | Potrero Medical, Inc.<br>26142 Eden Landing Road<br>Hayward, CA 94545<br>Establishment Registration No: 3011383638 |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Priscila Tapia<br>Director of Regulatory and Quality Assurance<br>Phone: 510-826-2033<br>Email: ptapia@potreromed.com |
| Subject Device | |
| Device Trade Name: | Accuryn Monitoring System<br>Accuryn SmartFoley IAP UO Temp Tray System 12 F<br>Accuryn SmartFoley IAP UO Temp Tray System 14 F<br>Accuryn SmartFoley IAP UO Temp Tray System 16 F |
| Device Common Name: | Temperature-sensing foley catheter,<br>Intra-abdominal pressure monitoring device |
| Device Class: | II |
| Classification Regulation: | 21 CFR §876.5130, Urological Catheter and Accessories |
| Regulation Description: | Urological Catheter and Accessories |
| Product Codes: | EZL, EXY, PHU |
| 510(k) Type: | Traditional |
| Catalog Numbers: | FGN-06-2687, FGS-06-IAP.12FST, FGS-06-IAP.14FST |
# Predicate Device
| | The Accuryn Monitoring System with the Accuryn SmartFoley IAP UO Temp Tray System 12F<br>and 14F is being compared to the following legally marketed predicate device: |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number: | K153655 |
| Manufacturer: | Potrero Medical, Inc. |
| Trade Name: | Accuryn Monitoring System<br>Accuryn SmartFoley UO Temp Tray System<br>Accuryn SmartFoley IAP UO Temp Tray System |
| Device Common Name: | Temperature-sensing foley catheter,<br>Intra-abdominal pressure monitoring device |
| Catalog Numbers: | FGN-06-2687, FGS-06-IAP.16FST, FGS-06-UOT.16FST |
The predicate has not been subject to a design-related recall
{4}------------------------------------------------
# Device Description
The Accuryn Monitoring System consists of three components, the Accuryn Sensing Urinary Catheter, Accuryn Urine Collection Set or Accuryn SmartFoley Adapter, and the Accuryn Monitor.
The Accuryn Sensing Urinary Catheter is a single use, sterile, MR conditional, disposable (12F, 14F, and 16 F), two-way silicone urinary bladder catheter with two opposing drainage eyes and four lumens. The first lumen is for urine drainage, the second lumen for the retention balloon, the third lumen embedded with a thermistor, and the fourth lumen to measure intra-abdominal pressure (IAP).
The Accuryn SmartFoley® comes in three configurations:
- IAP, U/O, Temp (intra-abdominal pressure, urine output, and core temperature)
- U/O. Temp (urine output and core temperature) ●
- Adapter (that is compatible with commercially available Foley catheters between 6-24Fr)
The Accuryn Urine Collection Set is provided permanently pre-connected to the Accuryn Sensing Urinary Catheter's drainage lumen conveys urine to a urine measurement cassette. which connects to the Accuryn Monitor. The cassette feeds the urine collection bag. Potrero Medical also provides a configuration of the Accuryn Urine Collection Set that is not provided pre-connected to the catheter/foley. It is known as Accuryn SmartFoley Adapter and enables the user to connect a Foley catheter of their choice to the Accuryn Urine Collection Set. The Accuryn Urine Collection Set and Accuryn SmartFoley Adapter function in conjunction with the Accuryn Monitor to measure and display urine output is considered a 510(k) exempt device per 21 CFR 876.1800 (urine flow or volume measuring system, product code EXY).
The Accuryn Monitor is a portable electronic device which maintains urine flow from the foley through the Accuryn Urine Collection Set or Accuryn SmartFoley Adapter, measures the urine output volume in the measurement cassette, and displays the urine output. The Accuryn Monitor also measures and displays the temperature from the thermistor and intra-abdominal pressure (IAP) from the pressure lumen. The monitor contains software.
## Intended and Indications for Use
The Accuryn Monitoring System is intended for use in the drainage and/or collection of urine, and in the monitoring of urine output and core body temperature, in degrees Fahrenheit and degrees Celsius. The Accuryn Monitoring System with the Accuryn Sensing Urinary Catheter (SmartFoley®) – IAP UO Temp is also intended for use in the monitoring of intra-abdominal pressure. The measured pressures can be used as an aid in the diagnosis of intra-abdominal hypertension (IAH) and the associated clinical syndrome of abdominal compartment syndrome (ACS). The Accuryn Sensing Urinary Catheter is a single use device intended for
{5}------------------------------------------------
short term use (less than 30 days).
## Technological Characteristics
The Accuryn SmartFoley IAP UO Temp Tray System 12 F and 14 F introduce two smaller size catheters which are similar in design characteristics and performance to the predicate device. The Accuryn Monitoring System displays the trending IAP value in the trending IAP graph screen, which differs from the predicate which only displays the trending IAP graph. However, the fundamental scientific technology used in the Accuryn Monitoring System, including the SmartFoley devices, is not affected by the addition of these two foley's sizes or the trending IAP display, and the mechanism of action, general principle of operation, and intended use, remain unchanged from the predicate. Performance testing data has demonstrated that the Accuryn Monitoring System is safe and effective for its intended use.
| Comparison<br>Feature | Subject Device | Predicate Device | Comments |
|-----------------------|----------------|------------------|----------|
| Product Class | II | II | Same |
| Product Code | EZL, PHU, EXY | EZL, PHU, EXY | Same |
#### Comparison of Technological Characteristics and Device Use with Predicate Device
{6}------------------------------------------------
| Indications for<br>Use | The Accuryn Monitoring<br>System is intended for<br>use in the drainage<br>and/or collection of<br>urine, and in the<br>monitoring of urine<br>output and core body<br>temperature, in degrees<br>Fahrenheit and degrees<br>Celsius. The Accuryn<br>Monitoring System with<br>the Accuryn Sensing<br>Urinary Catheter<br>(SmartFoley®) – IAP UO<br>Temp is also intended<br>for use in the monitoring<br>of intra- abdominal<br>pressure. The measured<br>pressures can be used<br>as an aid in the<br>diagnosis of intra-<br>abdominal hypertension<br>(IAH) and the associated<br>clinical syndrome of<br>abdominal compartment<br>syndrome (ACS). The<br>Accuryn Sensing Urinary<br>Catheter is a single use<br>device intended for short | The Accuryn Monitoring<br>System is intended for<br>use in the drainage<br>and/or collection of<br>urine, and in the<br>monitoring of urine<br>output and core body<br>temperature, in degrees<br>Fahrenheit and degrees<br>Celsius. The Accuryn<br>Monitoring System is<br>also intended for use in<br>the monitoring of intra-<br>abdominal pressure.<br>The measured<br>pressures can be used<br>as an aid in the<br>diagnosis of intra-<br>abdominal hypertension<br>(IAH) and the<br>associated clinical<br>syndrome of abdominal<br>compartment syndrome<br>(ACS). The Accuryn<br>Sensing Urinary | Subject device<br>specifies the<br>Accuryn Sensing<br>Urinary Catheter<br>(SmartFoley®) –<br>IAP UO Temp is<br>the catheter with<br>intra-abdominal<br>pressure<br>measurement<br>ability. The<br>Accuryn<br>SmartFoley UO<br>Temp is an<br>alternative option<br>for users who do<br>not need the IAP<br>functionality. |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{7}------------------------------------------------
| | term use (less than 30<br>days). | Catheter is a single use<br>device intended for<br>short term use (less<br>than 30 days). | |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|
| Use environment | The device is intended<br>for use in the hospital,<br>including the operative<br>or intensive care<br>environments. | The device is intended<br>for use in the hospital,<br>including the operative<br>or intensive care<br>environments. | Same |
| Users | The device is intended<br>to be used by surgeons,<br>anesthesiologists,<br>nurses, and other clinical<br>care personnel. | The device is intended<br>to be used by surgeons,<br>anesthesiologists,<br>nurses, and other<br>clinical care personnel. | Same |
| IAP UO Temp<br>Catheter Size | 12Fr<br>14Fr<br>16Fr | 16Fr | Two smaller<br>French sizes are<br>being introduced. |
| Lumens | 4 (IAP, UO, Temp,<br>Retention) | 4 (IAP, UO, Temp,<br>Retention) | Same |
| Performance<br>Standard | ASTM F623 | ASTM F623 | Same |
| Use Duration | Foley: <30 days<br>Monitor: Re-usable | Foley: <30 days<br>Monitor: Re-usable | Same |
| Sterile | Disposable: Yes<br>Monitor: No | Disposable: Yes<br>Monitor: No | Same |
# Summary of Non-Clinical Performance Data
The following testing was conducted to validate and verify that the subject device met all acceptance criteria and was substantially equivalent to the predicate device:
# Design Verification and Validation Testing
- · Urinary Catheter Tests per ASTM F623-19 and ISO 20696
- Performance Testing, including compliance with ASTM D3078-02, ASTM F88/F88M-15, . ASTM D4332-14, ASTM D4169-16, ASTM F1980-16
- MR Compatibility per FDA Guidance "Testing and Labeling Medical Devices for Safety in .
{8}------------------------------------------------
the Magnetic Resonance (MR) Environment", ASTM F2052-15, ASTM F2119-07, ASTM F2182-19e2, ASTM F2213-17, and ASTM F2503-20
- IAP Functionality Testing ●
- Usability per ANSI AMMI IEC 62366-1 ●
- System and Software Verification and Validation per IEC 62304 ●
## Electromagnetic Compatibility and Electrical Safety
- Testing and evaluation per IEC 60601-1, IEC 60601-1-2, IEC 80601-2-49, and ISO 80601-2-56
## Cybersecurity
- Testing and risk assessment per FDA quidance document "Content of Premarket ● Submissions for Management of Cybersecurity in Medical Devices" and AAMI TIR57
## Sterilization
- Product adoption equivalency and evaluation per AAMI TIR28:2016, ISO 11607-1, ISO ● 11135, and FDA Guidance "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile"
## Biocompatibility
- Biological Risk Assessment per ISO 10993, Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing within a Risk Management Process (2018) and the FDA Guidance "Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" (2020), considering:
- Cytotoxicity O
- Sensitization O
- Irritation: Intracutaneous Injection Test O
- Acute Systemic Toxicity O
- Material-Mediated Pyrogenicity O
- o Subacute system toxicity
- O Genotoxicity
- Implantation O
## Clinical Data
No clinical data was required to demonstrate safe use of the Accuryn Monitoring System with the Accuryn SmartFolev IAP UO Temp Trav System 12 F and 14 F.
{9}------------------------------------------------
# Conclusion
Based on the similarities in design between the subject and predicate devices currently in use, and the performance data, the addition of the Accuryn SmartFoley IAP UO Temp Tray System 12 F and 14 F to the Accuryn Monitoring System for the proposed indication does not raise new questions related to safety and effectiveness compared with the predicate.
The Indications for Use statement for the subject device is not identical to the predicate device; however, the differences do not alter the intended therapeutic use of the device, nor do they affect the safety and effectiveness of the device relative to the predicate.
Both the subject and the predicate devices have intended use for the drainage and/or collection of urine, and in the monitoring of urine output and core body temperature, in degrees Fahrenheit and degrees Celsius. Both the subject and predicate devices are also intended for use in the monitoring of intra-abdominal pressure. The Accuryn SmartFoley UO Temp is an alternative option for users who do not need the IAP functionality.
The fundamental scientific technology used in the Accuryn Monitoring System, including the SmartFoley devices, is not affected and the mechanism of action and general principle of operation, remain unchanged from the predicate. Performance testing and analysis of results provide assurance that the device meets its specifications, is acceptable for device functionality, and is safe and effective for its intended use without raising any new issues of safety and effectiveness as compared to the predicate. Therefore, the Accuryn Monitoring System with the Accuryn SmartFoley IAP UO Temp Tray System 12 F and 14 F is substantially equivalent to the Accuryn Monitoring System with the Accuryn SmartFoley IAP UO Temp Tray System 16 F.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.