K220990 · Winix, Inc. · FRF · Sep 14, 2022 · General Hospital
Device Facts
Record ID
K220990
Device Name
Qorda QD1
Applicant
Winix, Inc.
Product Code
FRF · General Hospital
Decision Date
Sep 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5045
Device Class
Class 2
Indications for Use
The Qorda QD1 is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles from the air for medical purposes.
Device Story
Freestanding medical recirculating air cleaner; utilizes wheels for mobility. Inputs: ambient room air. Operation: air drawn through front and rear panels; passes through 4-stage purification: Pre-Filter, CD Carbon Filter, True HEPA Filter, and PlasmaWave technology. PlasmaWave field inactivates airborne microorganisms; HEPA filter traps debris; CD Carbon Filter absorbs ozone byproduct. Outputs: purified air. User interface: manual fan speed adjustment (4 stages: Level 1, 3, Turbo, Sleep), child lock, filter replacement indicator, illuminance adjustment, and timer. Used in medical facilities and homes; operated by staff or patients. Benefits: reduction of airborne particles and microorganisms (e.g., MS2 bacteriophage, Bacillus subtilis endospores) to improve air quality.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via chamber testing (579 ft³) demonstrating >4.0 log reduction of MS2 bacteriophage and Bacillus subtilis endospores within 20 minutes at Turbo speed. Ozone emissions confirmed <2.07 ppb, below the 0.050 ppm limit. Air change rates verified across 5 operational modes.
Technological Characteristics
Freestanding air cleaner; 120 VAC power. 4-stage purification: Pre-Filter, CD Carbon Filter, True HEPA Filter (captures 99.99% of 0.003 micron particles), and PlasmaWave technology. Dimensions: 14.6" x 14.6" x 32.6". Standards: IEC 60601-1:2005/A1:2012, IEC 60601-1-2:2014, IEC 60601-1-6:2013, IEC 60601-1-11:2015. Ozone generation <2.07 ppb.
Indications for Use
Indicated for use as a room recirculating air cleaner to filter and inactivate airborne particles for medical purposes in medical facilities and home environments.
Regulatory Classification
Identification
A medical recirculating air cleaner is a device used to remove particles from the air for medical purposes. The device may function by electrostatic precipitation or filtration.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the acronym "FDA" followed by the words "U.S. Food & Drug Administration" in a larger font and "Administration" in a smaller font.
September 14, 2022
Winix Inc. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620
Re: K220990
Trade/Device Name: Qorda QD1 Regulation Number: 21 CFR 880.5045 Regulation Name: Medical Recirculating Air Cleaner Regulatory Class: Class II Product Code: FRF Dated: August 22, 2022 Received: August 23, 2022
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K220990
Device Name Qorda QD1
Indications for Use (Describe)
The Qorda QD is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles from the air for medical purposes.
The Qorda QD1 has been demonstrated to remove the following organisms under the following exposure conditions:
| Organism | Fan Speed | Average Maximum log<br>reduction | Exposure Time |
|----------------------------------------|------------|----------------------------------|---------------|
| MS2 bacteriophage | Turbo MODE | 5.6 log reduction | 20 minutes |
| MS2 bacteriophage | Sleep MODE | 4.9 log reduction | 45 minutes |
| <i>Bacillus subtilis</i><br>endospores | Turbo MODE | 4.4 log reduction | 20 minutes |
| <i>Bacillus subtilis</i><br>endospores | Sleep MODE | 4.3 log reduction | 60 minutes |
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete tine to review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### Submitter
WINIX INC. Daewoon Kang 80-13, Sujeong-ro, Jangan-myeon Hwaseong-si, Gyeonggi-do, 18583 Republic of Korea Email: kdw9478@winix.com Phone: +82-31-5180-7575
#### Device Information
- Trade Name: Qorda QD1
- Classification Name: Medical recirculating air cleaner ●
- Common Name: Air Filtration System, HEPA Air Filtration System
- Product Code: FRF ●
- Panel: General Hospital
- Regulation Number: 21 CFR 880.5045 ●
- Device Class: Class II ●
- Date prepared: 08/22/2022
## Primary Predicate Device
- K200321, Novaerus NV1050 manufactured by Novaerus US Inc. -
#### Indications for use
The Qorda QD1 is intended as a room recirculating air cleaner. The system is used for filtering out and inactivating airborne particles from the air for medical purposes.
The Qorda QD1 has been demonstrated to remove the following organisms under the following exposure conditions:
| Organism | Fan Speed | Average Maximum log<br>reduction | Exposure Time |
|---------------------------------|------------|----------------------------------|---------------|
| MS2 bacteriophage | Turbo MODE | 5.6 log reduction | 20 minutes |
| MS2 bacteriophage | Sleep MODE | 4.9 log reduction | 45 minutes |
| Bacillus subtilis<br>endospores | Turbo MODE | 4.4 log reduction | 20 minutes |
| Bacillus subtilis<br>endospores | Sleep MODE | 4.3 log reduction | 60 minutes |
## Device Description
Oorda OD1 is a freestanding medical recirculation air cleaner that facilitates movement with wheels. The device includes two pre-filters, two deodorizing filters, two HEPA filters - 3 filters in the front panel and 3 filters in the rear panel), and a manual for users.
The device can adjust the air volume in four stages through manual operation (Level 1, Level3, Turbo) and Sleep mode. When operating in sleep mode, the equipment LED and PlasmaWave are turned off and operated with low noise.
## Official Correspondent Withus Group Inc April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: +1-909-274-9971
{5}------------------------------------------------
The device contains the following additional functions.
(1)Child lock: Button lock function that prevents children or pets from changing their current motion mode even if they accidentally operate device.
(2)Filter Replacement Guide: The remaining filter life is displayed on the front display of the device and tells you when to replace the filter.
(3)Adjustment of illuminance: You can adjust the illuminance in four stages (100%, 50%, 20%, OFF).
(4)You can turn on or off the PlasmaWave.
(5)The timer allows the equipment to turn off automatically after a certain period of time.
Qorda QD1 inhales air in both directions on the front and rear of the device and purifies it through the four cleaning steps below:
[Pre-Filter]
The air is drawn in through the cover front, passes through the Pre-Filter then moves to two filters stages(CD Carbon Filter, True HEPA Filter). The Pre-Filter protects CD Carbon Filter, True HEPA Filter inside the Pre-Filter.
[CD Carbon Filter]
Reduces any ozone generated as a byproduct of the plasma field. It also prolongs the True HEPA Filter lifespan.
[True HEPA Filter]
True HEPA Filter captures 99.99%* of airborne ultrafine particles as small as 0.003 microns. *Based on independent laboratory test conducted on particles as small as 0.003 microns in size.
[PlasmaWave]
PlasmaWave Technology has been tested by an independent third-party to remove particles from the air for medical purposes.
{6}------------------------------------------------
# Summary of Technological Characteristics
The subject and primary predicate device (K20032 ) are similar in indications, design, fundamental technology, functions, and principle of operation.
A table to compare the technical characteristics is below.
| | Subject Device | Primary Predicate | Discussion |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| K number | K220990 | K200321 | |
| Product Name | Qorda QD1 | Novaerus NV 1050 | |
| Common Name | Air Filtration System, HEPA Air Filtration System | Air Filtration System, HEPA Air Filtration<br>System | Identical |
| Product Code | FRF | FRF | Identical |
| Regulation Number | 21 CFR 880.5045 | 21 CFR 880.5045 | Identical |
| Class | Class II | Class II | Identical |
| Device Illustration | Image: Qorda QD1 device | Image: Novaerus NV 1050 device | Identical |
| Intended<br>Use/Indications for<br>Use | The Qorda QD1 is intended as a room recirculating air cleaner.<br>The system is used for filtering out and inactivating airborne<br>particles from the air for medical purposes. | The Novaerus NV1050 is intended as a room<br>recirculating air cleaner. The system is used for<br>filtering out and inactivating airborne particles<br>from the air for medical purposes. | Identical |
| Number of Filters | 6 | 3 | Difference |
| Air purifying<br>Mechanism | 4 steps (Pre-Filter, CD Carbon Filter, True HEPA Filter, Plasma<br>technology) | 4 steps (Pre-Filter, CD Carbon Filter, HEPA<br>Filter, Plasma technology) | Identical |
| Use Location | Medical Facilities and Home | Medical Facilities | Difference |
| Technology | Air from the room is passed through a plasma field to inactivate<br>airborne micro-organisms. A True HEPA Filter traps the<br>resulting debris and a CD Carbon Filter absorbs any ozone<br>generated as a byproduct of the plasma field. | Air from the room is passed through a plasma<br>field to inactivate airborne micro-organisms. A<br>HEPA filter traps the resulting debris and an<br>activated carbon filter absorbs any ozone<br>generated as a byproduct of the plasma field. | Identical |
| Device size (inches) | 14.6 in(W) x 14.6 in(D) x 32.6 in(H) | 19.0 (w) x 19.1 (D) x 36.5 (h) | Difference |
| Device weight | 14.1 kg (30.9 lbs) | 112lb (51 kg) | Difference |
| Power source | 120 VAC | 110V AC | Similar |
| Air change rates | 6,000 to 27,900 ft3/h (170 to 790 m3/h) in 5 steps | 6,400 to 31,925 ft3/h (180 to 904 m3/h) in 5 steps | Difference |
| Reduction of biological agents | The QD1 unit, set to 'Turbo', was effective at reducing both organisms(the MS2 bacteriophage and <i>Bacillus subtilis</i> endospores) by the FDA required net log reduction of 4.0 or greater (equivalent to 99.99% or greater reduction) within 20 minutes when operating in a room of 579 ft3 (16.4 m3). | <i>Bacillus Globigii</i> endospores and MS2 phage reduced by 3 log reduction in 10 minutes and 4 log reduction in 15 minutes when operating at full fan speed in a room of 580ft3 (16.4m3) | Difference |
| Filtration of particles | The QD1 produces a 3 log reduction in 0.5 to 4.0 μm sized particles in 10 minutes when operating at Turbo speed in a room of 579 ft3 (16.4 m3). | NV1050 produces a 4 log reduction in 0.5 to 2.0 μm sized particles in 10 minutes in a 580ft3 (16.4m3) sealed room | Difference |
| Ozone emitted | Ozone is being generated at a level of less than 2.07 ppb during the operation of the device. | Ozone emissions below 10 ppb (1/5th FDA limit for medical devices) | Similar |
| Operational range | Operating temperature: 5°C~40°C<br>Relative humidity: 15% ~ 90% | Temperature: 50°F to 95°F (10°C to 35 °C)<br>Relative humidity: 10 to 75%RH | Similar |
| Storage range | Operating temperature: 5°C~40°C<br>Relative humidity: 10% ~ 80% | Temperature: 13°F to 160°F (-10°C to + 71°C)<br>Relative humidity: 10 to 93%RH | Similar |
| Standards used | IEC 60601-1:2005/A1:2012<br>IEC 60601-1-2:2014 & IEC 60601-1-6:2013<br>IEC 60601-1-11:2015 | IEC 60601-1:2005/A1:2012<br>IEC 60601-1-2:2014 | Similar |
{7}------------------------------------------------
{8}------------------------------------------------
## Non-clinical Testing
1) Electrical Safety and Electromagnetic Compatibility
Electrical, mechanical, environmental safety and performance testing were conducted according to standard IEC 60601-1:2012, IEC 60601-1-11:2015 and EMC testing was conducted according to IEC 60601-1- 2:2014. All test results were satisfying with the standards.
## 2) Performance data
The following performance data has been provided to demonstrate that the subject device meets the acceptance criteria of the standards listed below:
| Test name/<br>Methodology/<br>Standard name | Purpose | Acceptance Criteria | Result |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ozone emissions | Confirm ozone<br>emissions are<br>below<br>the maximum<br>permitted levels<br><0.050 ppm | Ozone emitted to be<br><0.050 ppm | Pass with the device operating in normal<br>operating conditions under both TURBO<br>and SLEEP Modes. The QD1 units<br>averaged under 1.0 ppb of ozone emission<br>throughout the 24-hour testing, with a<br>maximum generation level less than 2.07<br>ppb. Testing was conducted in a chamber<br>(internal dimensions are 9.1 ft x 9.1 ft x 7<br>ft, with a displacement volume of 579 ft³,<br>or 16.4 m³.) |
| Filtration of<br>particles | To demonstrate<br>that the device can<br>produce a 4 log<br>reduction in<br>particles | The device<br>produces a 4 log<br>reduction in the<br>concentration of µm<br>sized the two organism<br>types(the MS2<br>bacteriophage and Bacillus<br>subtilis endospores) and<br>poly-styrene latex (PSL)<br>microspheres | The QD1 produces a 3 log reduction in<br>0.5 to 4.0 µm sized particles in 10<br>minutes when operating at Turbo speed in<br>a chamber (internal dimensions are 9.1 ft<br>x 9.1 ft x 7 ft, with a displacement<br>volume of 579 ft³, or 16.4 m³ |
| Combined<br>Operation | To demonstrate the<br>performance of the<br>Qorda QD1 at<br>maximum speed to<br>inactivate and filter<br>out specified micro-<br>organisms | To produce a 4 log<br>reduction in the<br>specified microorganisms | QD1 showed the following results in the<br>the MS2 bacteriophage and Bacillus<br>subtilis endospores removal performance<br>test.<br>The QD1 unit, set to 'Turbo', was<br>effective at reduction of both<br>organisms by the required net log<br>reduction of 4.0 or greater<br>(equivalent to 99.99% or greater<br>reduction) within 20 minutes when<br>operating in a chamber (internal<br>dimensions are 9.1 ft x 9.1 ft x 7 ft,<br>with a displacement volume of 579<br>ft³, or 16.4 m³.) |
{9}------------------------------------------------
| Air Change Rate<br>Test | To demonstrate the<br>performance from<br>6,000 to 27,900<br>ft3/h (170 to 790<br>m3/h) in 5 steps.<br>(Level 1, Level 2,<br>Level 3, Turbo,<br>Sleep) | The criteria of air change<br>rate in each mode is higher<br>than the following:<br>Sleep mode is 6,000 ft3/h.<br>Level 1 mode is 11,400 ft3/h.<br>Level 2 mode is 13,800 ft3/h.<br>Level 3 mode is 16,200 ft3/h.<br>Turbo mode is 27,900 ft3/h. | Air change rate of the QD1 showed that<br>all results were higher than the standard<br>air change range required for each stage. |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Software<br>Validation<br>Report | Software verification and validation testing was conducted, and documentation was provided<br>as recommended by FDA's Guidance for Industry and FDA Staff,<br>"Guidance for the Content of Premarket Submissions for Software Contained in<br>Medical Devices." The software for this device was considered as a "moderate(Class B)"<br>level of concern. | | |
# Clinical Data
Not applicable
# Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(k) submission K220990, Qorda QD1, is as safe, as effective, and performs as well as or better than the legally marketed predicate device cleared under K200321.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.