K220938 · Aura Wellness, LLC · NGX · Aug 22, 2023 · Physical Medicine
Device Facts
Record ID
K220938
Device Name
Nova HD+
Applicant
Aura Wellness, LLC
Product Code
NGX · Physical Medicine
Decision Date
Aug 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nova HD+ is intended for use by adults for the stimulation of healthy muscles in order to improve or facilitate muscle performance.
Device Story
Nova HD+ is a noninvasive powered muscle stimulator using electrically generated magnetic fields to stimulate healthy muscles. System comprises a main control unit and a large loop coil applicator. Operated by user or clinician in home or clinical settings. User selects treatment duration (1-20 minutes) and intensity (levels 1-10) via control unit interface. Device generates magnetic pulses to facilitate muscle performance. Firmware manages user interface and pulse generator parameters. Benefits include improved muscle performance through noninvasive magnetic field induction.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including signal waveform output measurements, magnetic flux output (mT) across all intensity levels (1-10), and electrical/EMC safety testing per IEC 60601-1, 60601-1-2, 60601-1-11, and 60601-1-6. Biocompatibility testing was performed per ISO 10993-1, 10993-5, and 10993-10.
Technological Characteristics
Powered muscle stimulator; non-invasive magnetic field induction. Main unit dimensions: 9.7" x 10.64" x 3.57". Large loop coil applicator (147.2 in²). Peak frequency: 1.58 kHz; Max flux density: 1.8 mT. Power: 81.6W max. Microprocessor-controlled. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 62366-1, IEC 62304. Biocompatibility: ISO 10993-1, 5, 10.
Indications for Use
Indicated for adults for stimulation of healthy muscles to improve or facilitate muscle performance. No specific disease state or contraindications listed.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
August 22, 2023
Aura Wellness, LLC Scott Blomberg Director of Regulatory Affairs 11530 Electron Drive Louisville, Kentucky 40299
Re: K220938
Trade/Device Name: Nova HD+ Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: August 21, 2023 Received: August 21, 2023
Dear Scott Blomberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Heather L. Dean -S
Heather Dean, Ph.D Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220938
Device Name Nova HD+
Indications for Use (Describe)
The Nova HD+ is intended for use by adults for the stimulation of healthy muscles in order to improve or facilitate muscle performance.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for AURA. The logo consists of a colorful circular design above the word "AURA" in a simple, sans-serif font. The circular design is made up of several small circles in various colors, including shades of blue, green, purple, orange, and red. The circles are arranged in a pattern that radiates outward from the center, creating a visually appealing and modern design.
## 510(k) SUMMARY K220938
## (as required by 21CFR 807.92)
#### I Submitter
Aura Wellness, LLC 11530 Electron Drive Louisville, KY 40299 Phone: 502-714-1993 Fax: 502-369-5226
| Submitter Contact: | Scott Blomberg |
|-----------------------------|--------------------|
| Submission Correspondent: | scott@aurawell.com |
| Establishment Registration: | 10081462 |
| Submission Date: | August 21, 2023 |
#### II Device
| Proprietary or Trade Name: | Nova HD+ |
|----------------------------|------------------------------------------------------|
| Common/Usual Name: | Powered Muscle Stimulator |
| Classification Name: | Stimulator, Muscle, Powered, For Muscle Conditioning |
| Regulation: | 890.5850 |
| Regulatory Class: | 2 |
| Product Code: | NGX |
| Device Panel: | Neurological and Physical Medicine Devices |
| III Predicate Device | Bemer Classic Set and Bemer Pro-Set (K210174) |
#### IV Device Description
The Nova HD+ system is a noninvasive physical medicine device that uses electrically generated magnetic fields to stimulate muscles in order to improve and facilitate muscle performance.
The Nova HD+ is noninvasive, fully reusable (no disposable components such as electrodes) and have configurations allowing both home care and clinical professional use.
In addition to the main control unit, the Nova HD+ includes a large loop coil applicator (coils) which allow the operator to administer optimal treatment to the desired anatomical part of the body.
The device contains firmware that controls the user interface, the operator controls the magnetic pulse generator intensity and duration of the treatment.
#### V Indication for Use
The Nova HD+ is intended for use by adults for the stimulation of healthy muscles in order to improve or facilitate muscle performance.
Environments of use: Over the Counter (OTC), both home care and clinical professional use
Nova HD+ Traditional 510(k)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for a company or organization called "AURA". The logo consists of the word "AURA" in a simple, sans-serif font, positioned below a colorful, circular design. The circular design is made up of many small circles in various colors, including shades of blue, green, purple, and orange. The overall design is modern and visually appealing.
#### VI Comparison of Technological Characteristics and Performance with the Predicate
Table 1 below provides the summarized substantial equivalence comparison of general intended uses/actions, specific indications for use, equivalence of key clinical and technical features between subject and predicate device, along with a full listing of technical and conformance specifications.
## Table 1: Comparison of Subject vs. Predicate
| | | Nova HD+ | BEMER Pro Set |
|----------------------------------------|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Subject Device<br>K220938 | Predicate Device<br>K210174 |
| Classification Code | Identical | Primary: Powered Muscle<br>Stimulator NGX 890.5850 | Same |
| Indication for Use | Similar | The Nova HD+ is intended for<br>the stimulation of healthy muscles<br>in order to improve or facilitate<br>muscle performance. | The BEMER therapy is<br>indicated:<br>• To temporarily increase local<br>blood circulation in healthy leg<br>muscles.<br>• To stimulate healthy muscles in<br>order to improve and facilitate<br>muscle performance |
| Primary Mode of<br>Action | Identical | Non-invasive tissue stimulation<br>via magnetic field induction | Same |
| Therapy Timer | Similar | 1-20 minutes | 8-20 minutes |
| Model (System) | Similar | Nova HD+ | Pro Set |
| Weight | Similar | System:<br>5.6 pounds (2.5 kg) | System:<br>2.9 pounds (1.3 kg)<br>[B.BOX Classic]<br>System:<br>3.1 pounds (1.4 kg)<br>[B.BOX Professional] |
| Dimensions (Main<br>Unit): | Similar | 9.7" x 10.64" x 3.57" | Public information not available |
| Size of Therapy Area | Similar | 147.2 in² (Large Loop) | Public information not available |
| Magnetic Field Peak<br>Frequency | Different but<br>does not raise<br>any safety<br>concerns | 1.58 kHz | Public information not available. |
| Average Flux<br>Density (Applicator) | Different but<br>does not raise<br>any safety<br>concerns | 1.8 mT (max level) | 150 µT (max level) |
| Power Consumption | Different but<br>does not raise<br>safety or<br>effectiveness<br>concerns | 81.6 Watts maximum | 30 Watts maximum |
| Input Power | Similar | 100-130 Vac, 60 Hz, 0.68 A | 100-240 Vac, 50-60 Hz, 0.6 A; |
| | | Nova HD+ | BEMER Pro Set |
| | | Subject Device<br>K220938 | Predicate Device<br>K210174 |
| Number of Output<br>Channels | Similar | 1 | 2 |
| Microprocessor<br>controlled | Identical | Yes | Same |
| Therapy Amplitude<br>Control | Identical | Power level 1-10 controlled by<br>operator | Power level 1-10 controlled by<br>operator |
| User-controlled<br>therapy start/stop | Identical | Yes | Yes |
| Compliance with<br>voluntary standards | Identical | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 62366-1<br>IEC 60601-1-11 | Same |
| Firmware developed<br>per IEC 62304 | Identical | Yes | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo for a company called AURA. The logo consists of the word "AURA" in a simple, sans-serif font, with the letters in a light gray color. Above the word "AURA" is a circular design made up of many small, colored dots. The dots are arranged in concentric circles, with the colors ranging from green and blue on the outside to purple, orange, and yellow in the center.
#### VII Equivalence Rationale
## Principle of Operation
The principle of operation for the Nova HD+ is identical to the predicate, Bemer Classic Set and Bemer Pro-Set (K210174). The Nova HD+ and the predicate systems deliver therapeutic massage to the patient through the use of low-level magnetic fields administered directly to the body.
## Use Mode
The use of the Nova HD+ is essentially the same as the predicate device. Both units have main control units have interchangeable treatment applicators (coils) for administering the therapy to desired area of the controls on the control unit allow the user to select the time and intensity of the therapy and allow for adjustment of the intensity during the session.
## Non-clinical Testing
Performance testing demonstrated that the subject treatment applicator performed equivalent to the predicate applicator. Testing included:
Performance testing involved multiple measurements of:
- Signal waveform output (volts) generated from Nova HD+ with its applicators. ●
- Magnetic flux output (mT) generated from the applicator at all signal intensity input levels 1-10.
## Software
- All software features documented and tested in accordance with IEC 62304 for 'Moderate' level of concern.
Electrical / EMC evaluated per
- IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for Basic ● Safety And Essential Performance
- IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements For ● Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for AURA. The logo consists of a colorful circular design made up of many small circles in shades of green, blue, purple, and orange. Below the circular design, the word "AURA" is written in a simple, sans-serif font.
- . IEC 60601-1-11 - Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety And Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in The Home Healthcare Environment
- IEC 60601-1-6 - Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability
- . IEC 62366-1 - Medical devices - Part 1: Application of usability engineering to medical devices
## Biocompatibility
- ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management ● process
. ISO 10993-5 - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests for Irritation And . Skin Sensitization
## Clinical Testing
Not applicable.
## CONCLUSIONS
The analysis of the differences between Nova HD+ and the predicate device does not raise new questions of safety and effectiveness for the subject device performance test results, Aura Wellness determines that the Nova HD+ system performs within its design specifications and is equivalent to the predicate device.
The information in this 510(k) submission demonstrates that the Nova HD+ system is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.