K220929 · Merit Cro, Inc. · NFJ · Oct 31, 2022 · Radiology
Device Facts
Record ID
K220929
Device Name
Excelsior
Applicant
Merit Cro, Inc.
Product Code
NFJ · Radiology
Decision Date
Oct 31, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The EXCELSIOR Software is intended for use in importing, processing, measurement, and analysis of ophthalmic clinical images as well as in management of clinical data, through a computerized network for use in analysis of images and data obtained in clinical trials. Excelsior is a software solution intended to be used for viewing, manipulation, annotation, and ysis, and comparison of medical images from multiple imaging modalities and/or multiple time points. The application supports images, functional data such as PET as well as anatomical datasets, such as CT or MR. Excelsior is a software only medical device to be deployed through a cloud-based computerized network via web applications and customized user interfaces for use in the analysis of images and data obtained in clinical trials. Excelsior enables visualization of information that would otherwise have to be visually compared disjointedly. Excelsior provides and workflow automation tools to help the user assess and document the extent of a disease and/or the response to therapy in accordance with user selected standards and assess changes in imaging findings over multiple time-points. Excelsior supports the interpretation and evaluations and follow-up documentation of findings for radiologic oncology imaging and data obtained in clinical trials. The product is intended to be used as a workflow automation tool by trained medical professionals. It is intended to provide image and related information that is interpreted by a trained professional but does not directly generate any diagnosis or potential findings. Note: The medical professional retains the ultimate responsibility for making the pertinent observations based on their standard practices and established practices related to clinical trial outcomes. Excelsior is a complement to these standard procedures. Excelsior is not to be used in mammography.
Device Story
Cloud-based software platform for clinical trial data management; integrates remote data collection, quantitative analysis, and storage of ophthalmic and radiological DICOM images. Operates via web applications and customized user interfaces; used by trained medical professionals in clinical trial settings. Transforms raw imaging data into visualized, annotated, and measured outputs; enables side-by-side comparison of images from multiple modalities and time points. Provides workflow automation tools for documenting disease extent and therapy response. Does not use AI/ML. Healthcare providers use output to assist in clinical trial evaluations; device acts as a complement to standard clinical procedures; does not provide automated diagnosis. Benefits include streamlined data management and improved visualization of longitudinal imaging findings.
Clinical Evidence
Bench testing only. Software verification and validation testing performed to confirm performance of additional tools and features. Conformance to DICOM standards demonstrated.
Technological Characteristics
Cloud-based software; DICOM compliant; web-based deployment. Features include image processing, measurement, annotation, and video playback. No AI/ML algorithms. Software-only device.
Indications for Use
Indicated for trained medical professionals to import, process, measure, analyze, and manage ophthalmic and radiological clinical trial images and data. Supports viewing, manipulation, annotation, and comparison of multi-modality images (CT, MR, PET) across time points. Not for mammography. Does not generate diagnoses.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
October 31, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of a human figure, while the right side features the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Merit CRO, Inc. % Maureen O'Connell President O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, Massachusetts 02180
Re: K220929
Trade/Device Name: Excelsior Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: NFJ, LLZ Dated: September 23, 2022 Received: September 23, 2022
Dear Maureen O'Connell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220929
Device Name EXCELSIOR
# Indications for Use (Describe)
#### Ophthalmology
The EXCELSIOR Software is intended for use in importing, processing, measurement, and analysis of ophthalmic clinical images as well as in management of clinical data, through a computerized network for use in analysis of images and data obtained in clinical trials.
#### Radiology
Excelsior is a software solution intended to be used for viewing, manipulation, annotation, and ysis, and comparison of medical images from multiple imaging modalities and/or multiple time points. The application supports images, functional data such as PET as well as anatomical datasets, such as CT or MR. Excelsior is a software only medical device to be deployed through a cloud-based computerized network via web applications and customized user interfaces for use in the analysis of images and data obtained in clinical trials. Excelsior enables visualization of information that would otherwise have to be visually compared disjointedly. Excelsior provides and workflow automation tools to help the user assess and document the extent of a disease and/or the response to therapy in accordance with user selected standards and assess changes in imaging findings over multiple time-points. Excelsior supports the interpretation and evaluations and follow-up documentation of findings for radiologic oncology imaging and data obtained in clinical trials.
The product is intended to be used as a workflow automation tool by trained medical professionals. It is intended to provide image and related information that is interpreted by a trained professional but does not directly generate any diagnosis or potential findings.
Note: The medical professional retains the ultimate responsibility for making the pertinent observations based on their standard practices and established practices related to clinical trial outcomes. Excelsior is a complement to these standard procedures. Excelsior is not to be used in mammography.
Type of Use (*Select one or both, as applicable*)
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------|
| <div> <span> <b> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) SUMMARY
Merit CRO, Inc. EXCELSIOR K220929
#### 510(k) Owner
Merit CRO, Inc. 6527 Normandy Lane Suite 100 Madison, WI 53719
#### Submission Correspondent
Maureen O'Connell O'Connell Regulatory Consultants, Inc. 44 Oak Street Stoneham, MA 02180 Phone: 978-207-1245
Date Prepared: September 20, 2022
Trade Name of Device EXCELSIOR
Common or Usual Name Image Processing Software
#### Classification Name
Image Processing Software, Ophthalmology, 21 C.F.R. §892.2050 Image Processing Software, Radiology, 21 C.F.R. §892.2050 Class II Product Codes: NFJ and LLZ
# Predicate Devices
Primary: Merit CRO, Inc. (formerly Eyekor) EXCELSIOR software cleared in K130453
Secondary: Mint Lesion cleared in K142647
#### Device Description
The EXCELSIOR software is a cloud-based software that provides a central reading platform integrating remote data collection, quantitative analysis and measurement, storage, and management of ophthalmic and radiological data from DICOM images for clinical trials. The software does not use artificial intelligence or machine learning algorithms.
{5}------------------------------------------------
### Indications for Use
### Ophthalmology
The EXCELSIOR software is intended for use in importing, processing, measurement, and analysis of ophthalmic clinical images as well as in management of clinical data, through a computerized network for use in analysis of images and data obtained in clinical trials.
# Radiology
EXCELSIOR is a software solution intended to be used for viewing, manipulation, communication, annotation, analysis, and comparison of medical images from multiple imaging modalities and/or multiple time points. The application supports images, functional data such as PET as well as anatomical datasets, such as CT or MR. EXCELSIOR is a software only medical device to be deployed through a cloud-based computerized network yia web applications and customized user interfaces for use in the analysis of images and data obtained in clinical trials. EXCELSIOR enables visualization of information that would otherwise have to be visually compared disjointedly. EXCELSIOR provides analytical and workflow automation tools to help the user assess and document the extent of a disease and/or the response to therapy in accordance with user selected standards and assess changes in imaging findings over multiple time-points. EXCELSIOR supports the interpretation and evaluation of examinations and follow-up documentation of findings for radiologic oncology imaging and data obtained in clinical trials.
The product is intended to be used as a workflow automation tool by trained medical professionals. It is intended to provide image and related information that is interpreted by a trained professional but does not directly generate any diagnosis or potential findings.
Note: The medical professional retains the ultimate responsibility for making the pertinent observations based on their standard practices and established procedures related to clinical trial outcomes. EXCELSIOR is a complement to these standard procedures. EXCELSIOR is not to be used in mammography.
# Substantial Equivalence
Merit CRO, Inc. believes that the EXCELSIOR software described in this notification and for use under the conditions of the proposed labeling is substantially equivalent to legally marketed predicate devices that are Class II medical devices. For Ophthalmic indications for use the primary predicate device is the Merit CRO (formerly Eyekor) EXCELSIOR software cleared in K130453. For radiology indications for use. EXCELSIOR is substantially equivalent to the mint Lesion software cleared in K142647, which is the secondary predicate device.
Regarding Ophthalmology features, both the new version of EXCELSIOR and the version cleared in K130453 are ophthalmic management software devices for use in clinical trials. Additionally, the indications for use are identical. The new EXCELSIOR software is a modification of the cleared EXCELSIOR software. The majority of technological characteristics are identical, however, the new version of EXCELSIOR includes additional image overlay tools, and a video storage and playback capability has been added to the software. These tools and features do not change the intended use of the product and performance data shows that each of these features performs as intended.
{6}------------------------------------------------
The additional tools and features do not raise different questions of safety and effectiveness but represent minor modifications to the software features. The methodology used to validate and verify that the software performs as intended was used to confirm performance of the additional tools and features. The performance data provided supports substantial equivalence with the EXCELSIOR predicate device.
Regarding the Radiology features, the both the new version of EXCELSIOR and the mint Lesion software cleared in K142647 are image management software devices for use in radiology. EXCELSIOR has similar technological characteristics relating to radiology image management. Minor differences with the technological characteristics of the two devices do not change the intended use of the product and performance data is provided which shows that each of these features performs as intended. The additional tools and features do not raise different questions of safety and effectiveness compared to the predicate device. These represent minor modifications to the software features. The methodology used to validate and verify that the software performs as intended was used to confirm performance of the additional tools and features. The performance data provided supports substantial equivalence with the EXCELSIOR predicate device by demonstrating that the software performs as intended.
### Performance Data
Software validation and verification testing was performed which showed that the software performs as intended supporting substantial equivalence. Additionally, EXCELSIOR conforms with the DICOM standards which supports substantial equivalence as the primary predicate device conformed with the DICOM standard.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.