K220919 · Siemens Medical Solutions USA, Inc. · KPR · May 17, 2022 · Radiology
Device Facts
Record ID
K220919
Device Name
MULTIX Impact E
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
KPR · Radiology
Decision Date
May 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
MULTIX Impact E is a radiographic system used in hospitals, clinics, and medical practices. MULTIX Impact E enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and obese patients. Exposures may be taken with the patient sitting, standing, or in the prone position. MULTIX Impact E uses digital detectors for generating diagnostic images by converting X- rays into image signals. MULTIX Impact E is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes. MULTIX Impact E is not intended for mammography.
Device Story
Modular stationary X-ray system; floor-mounted tube, bucky wall stand, bucky table, generator, wireless digital detector. Inputs: X-ray radiation attenuated by patient anatomy. Transformation: X-rays converted to digital image signals via detector; system manages exposure parameters via generator and collimator. Outputs: Diagnostic radiographic images. Used in hospitals/clinics by healthcare professionals. Tube-side control module (TCM) and touch user interface (TUI) allow operator to manage SID, tube angle, and brakes. System provides visual/audible warnings and blocks functions upon error detection. Benefits: Enables diagnostic imaging for wide range of patient sizes and body parts; supports conventional film/screen or CR cassettes.
Clinical Evidence
Bench testing only. No clinical data presented. Verification and validation testing confirmed performance comparable to predicate device.
Technological Characteristics
Stationary X-ray system; floor-mounted tube, manual collimator, high-frequency generator (40kW/50kW), wireless digital detector (Mars1717VS). Standards: ANSI AAMI ES60601-1, IEC 60601-1-3, IEC 60601-1-2, IEC 62366-1, ISO 14971, IEC 60601-1-6, IEC 62304, IEC 60601-2-28, IEC 60601-2-54, NEMA PS 3.1-3.20, ISO 10993-1. Software: Moderate level of concern.
Indications for Use
Indicated for radiographic imaging of the whole body (skull, chest, abdomen, extremities) in pediatric, adult, and obese patients. Used in hospitals, clinics, and medical practices. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
U.S. FOOD & DRUG
ADMINISTRATION
Siemens Medical Solutions USA, Inc. % Denise Adams Regulatory Affairs Professional 40 Liberty Boulevard MALVERN PA 19355
# Re: K220919
Trade/Device Name: MULTIX Impact E Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: KPR, MQB Dated: March 28, 2022 Received: March 30, 2022
# Dear Denise Adams:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
May 17, 2022
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT 8B: Division of Radiological Imaging and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220919
Device Name MULTIX Impact E
Indications for Use (Describe)
MULTIX Impact E is a radiographic system used in hospitals, clinics, and medical practices.
MULTIX Impact E enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and obese patients. Exposures may be taken with the patient sitting, standing, or in the prone position.
MULTIX Impact E uses digital detectors for generating diagnostic images by converting X- rays into image signals. MULTIX Impact E is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes. MULTIX Impact E is not intended for mammography.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a series of orange dots that are arranged in a pattern.
# 510(k) Summary: MULTIX Impact E
K220919
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355
Date Prepared: May 16, 2022
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
# 1. General Information
Importer/ Distributor Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Establishment Registration Number: 2240869
# Manufacturer
Siemens Shanghai Medical Equipment Ltd. 278 Zhou Zhu Road Shanghai, 201318, China Establishment Registration Number: 3003202425
# Contract Manufacturer
Siemens Healthcare GmbH Siemensstrasse 1 Forchheim, Germany 91301 Establishment Registration Number: 3004977335
# 2. Contact Person
Denise Adams Regulatory Affairs Professional Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 610-448-6545 adams.denise@siemens-healthineers.com
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
# 3. Subject Device Name and Classification
| Trade Name: | MULTIX Impact E |
|----------------------------|-------------------------|
| Classification Name: | Stationary X-Ray System |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1680 |
| Device Class: | Class II |
| Product Code: | KPR |
# 4. Legally Marketed Predicate Device
| Trade Name: | MULTIX Impact (VA20) |
|----------------------------|-------------------------|
| 510(k) #: | K203345 |
| Clearance Date: | March 20, 2020 |
| Classification Name: | Stationary X-Ray System |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1680 |
| Device Class: | Class II |
| Product Code: | KPR |
## 5. Device Description
The MULTIX Impact E Radiography X-ray system is a modular system of x-ray components (floor-mounted x-ray tube, bucky wall stand, bucky table, x-ray generator, portable wireless detector) based on the predicate device, the MULTIX Impact (K203345). The following modifications have been made to the predicate device:
- A new X-ray Tube Assembly 1)
- A new Collimator 2)
- A new Generator 3)
- 4) Fixed patient table
- Tube-side control module (TCM) ર)
- 6) Upgraded software version to VB10 to support hardware modifications
The modified system will be branded as the MULTIX Impact E.
### 6. Indications for Use
MULTIX Impact E is a radiographic system used in hospitals, clinics, and medical practices. MULTIX Impact E enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and obese patients. Exposures may be taken with the patient sitting, standing, or in the prone position.
MULTIX Impact E uses digital detectors for generating diagnostic images by converting X- rays into image signals. MULTIX Impact E is also designed to be used with conventional film/screen or Computed Radiography (CR) cassettes.
MULTIX Impact E is not intended for mammography.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the text is a series of orange dots arranged in a circular pattern.
## 7. Substantial Equivalence
The MULTIX Impact E is a modification of the predicate device, the MULTIX Impact (VA20), cleared via K203345. The subject device is within the same classification regulation, has the similar indications for use, and the similar mechanical design as the predicate device. The MULTIX Impact E is substantially equivalent to the predicate device and documentation is provided to support a claim of substantial equivalence.
# 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device
The MULTIX Impact E is substantially equivalent to the commercially available MULTIX Impact (K203345) in terms of the intended use, design, biocompatibility material, functionality, technology, and energy source. The subject device uses the same or similar components cleared in the MULTIX Impact (e.g. detector, BWS and imaging system).
The components of the subject device have many of the same technological characteristics as those in the predicate device. Some technological characteristics that differ slightly are shown in the comparison tables below. Verification and validation testing have been successfully completed and test results show that the subject device, MULTIX Impact E with all its components, is substantially equivalent to the predicate device.
The modifications made to the subject device, MULTIX Impact E, do not affect the intended use of the device nor do they alter its fundamental scientific technology compared to the predicate device, the MULTIX Impact (K203345).
The following tables compare the main performance data of the subject device with the predicate device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots.
| Attribute | MULTIX Impact E<br>VB10<br>(Subject) | MULTIX Impact VA20<br>K203345<br>(Predicate) | Comparison<br>Results |
|------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | MULTIX Impact E is a<br>radiographic system used<br>in hospitals, clinics, and<br>medical practices.<br>MULTIX Impact E<br>enables radiographic<br>exposures of the whole<br>body including: skull,<br>chest, abdomen, and<br>extremities and may be<br>used on pediatric, adult<br>and obese patients.<br>Exposures may be taken<br>with the patient sitting,<br>standing, or in the prone<br>position.<br>MULTIX Impact E uses<br>digital detectors for<br>generating diagnostic<br>images by converting X-<br>rays into image signals.<br>MULTIX Impact E is<br>also designed to be used<br>with conventional<br>film/screen or Computed<br>Radiography (CR)<br>cassettes.<br>MULTIX Impact E is not<br>intended for<br>mammography. | MULTIX Impact is a<br>radiographic system used in<br>hospitals, clinics, and medical<br>practices.<br>MULTIX Impact enables<br>radiographic exposures of the<br>whole body including: skull,<br>chest, abdomen, and<br>extremities and may be used on<br>pediatric, adult and bariatric<br>patients. Exposures may be<br>taken with the patient sitting,<br>standing, or in the prone<br>position. MULTIX Impact is<br>not intended for<br>mammography.<br>MULTIX Impact uses digital<br>detectors for generating<br>diagnostic images by<br>converting X- rays into image<br>signals. MULTIX Impact is<br>also designed to be used with<br>conventional film/screen or<br>Computed Radiography (CR)<br>cassettes. | Same |
| Attribute | MULTIX Impact E<br>VB10<br>(Subject) | MULTIX Impact VA20<br>K203345<br>(Predicate) | Comparison Results |
| Detector | Wireless detector: Mars1717VS | Wireless detector:<br>Mars1717VS | Same |
| Tube Stand<br>(TS) | Integrated fully manual TS<br>- Manual tube tilting<br>- Manual longitudinal<br>movement<br>- Manual tube lifting | Independent semi-motorized TS<br>- Manual tube tilting<br>- Manual longitudinal<br>movement<br>- Motorized tube lifting | Different.<br>New tube modified from<br>motorized to manual for<br>lower cost. Verification<br>and Validation testing<br>concluded no impact on<br>safety and effectiveness. |
| X-ray Tube | RAY-12 3S tube with<br>following key performance:<br>-Heat capacity of anode: 170KJ<br>(230kHU)<br>-Input Power with Ref.300W:<br>54KW<br>-Anode rotary frequency: 50/60<br>Hz (3000/3600 rpm) | RAY-14S 3F tube with<br>following key performance:<br>-Heat capacity of anode: 260KJ<br>(350kHU)<br>-Input Power with Ref.300W:<br>78KW<br>-Anode rotary frequency:<br>150/180 Hz(~ 8500 to 10800<br>rpm) | Different.<br>Performance reduced<br>for lower cost and low-<br>end market. Verification<br>and Validation testing<br>concluded no impact on<br>safety and effectiveness. |
| Collimator | Manual collimator with<br>feedback following information<br>to system:<br>-Blade positions<br>-Cu filter status | Manual collimator without any<br>feedback information to system | Different.<br>Function improved to<br>support more<br>operational possibilities.<br>Verification and<br>Validation testing<br>concluded no impact on<br>safety and effectiveness. |
| Generator | 50KW high frequency X-ray<br>Generator with line voltage 3-<br>phase, 380V / 400V / 440V<br>(50/60Hz), 480 V (60Hz) | 55KW/65KW/80KW high<br>frequency X-ray Generator with<br>line voltage 3-phase, 380V /<br>400V / 440V (50/60Hz), 480 V<br>(60Hz) | Different.<br>Configuration reduced<br>for lower cost and low-<br>end market.<br>Verification and<br>Validation testing<br>concluded no impact on<br>safety and effectiveness. |
| | 40KW high frequency X-ray<br>Generator with line voltage 1-<br>phase, 208-230V<br>/(50/60Hz) | N.A. | Different.<br>Modified to support<br>more operational<br>possibilities. |
| Attribute | MULTIX Impact E<br>VB10<br>(Subject) | MULTIX Impact VA20<br>K203345<br>(Predicate) | Comparison Results |
| | | | Verification and<br>Validation testing<br>concluded no impact on<br>safety and effectiveness. |
| Automatic<br>Exposure<br>Control<br>(AEC) | 3 fields AEC chamber with<br>analog interface to system | 5 fields AEC module with CAN<br>interface to system | Different.<br>Modified for lower cost<br>and low-end market.<br>Verification and<br>Validation testing<br>concluded no impact on<br>safety and effectiveness. |
| Patient Table | Fixed table with integrated rail<br>mounting tube stand | Fixed table without rail | Different.<br>Modified for lower cost<br>and low-end market.<br>Verification and<br>Validation testing<br>concluded no impact on<br>safety and effectiveness. |
| Human<br>Machine<br>Interface<br>(HMI) | Tube-side control module<br>(TCM) with following<br>functions:<br>-SID display<br>-Tube angle display<br>-Release brakes | N.A. | Different.<br>Modified to support<br>more operational<br>possibilities.<br>Verification and<br>Validation testing<br>concluded no impact on<br>safety and effectiveness. |
| | Touch User Interface module<br>(TUI) | Touch User Interface module<br>(TUI) | Same |
| | N.A. | Remote Interface supported by<br>Siemens provided tablet that<br>meets minimum requirements. | Different.<br>Function reduced for<br>lower cost and low-end<br>market. Verification and<br>Validation testing<br>concluded no impact on<br>safety and effectiveness. |
| UI (User<br>Interface) on<br>Imaging<br>System | Siemens UI concept | Siemens UI concept | Same |
| Attribute | MULTIX Impact E<br>VB10<br>(Subject) | MULTIX Impact VA20<br>K203345<br>(Predicate) | Comparison Results |
| Software<br>version | VB10 | VA20 | Different<br>Updated to support all<br>hardware modifications.<br>Verification and<br>Validation testing<br>concluded no impact on<br>safety and effectiveness. |
# Table 1: Indications for Use Comparison:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the words is a graphic of orange dots.
# Table 2: Subject Device Compared to Predicate
Tel.: +1-888-826-9702
usa.siemens.com/healthineers
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
40 Liberty Boulevard
Malvern, PA 19355
USA
Tel.: +1-888-826-9702
usa.siemens.com/healthineers
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots.
# 9. Nonclinical Performance Testing
Non-clinical tests were conducted for the MULTIX Impact E during product development. The modifications described in this Premarket Notification are supported with verification and validation testing.
MULTIX Impact E conforms to the following standards: ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012; IEC 60601-1-3:2013; IEC 60601-1-2:2014; IEC 62366-1:2015; ISO 14971:2019; IEC 60601-1-6:2013; IEC 62304:2015; IEC 60601-2-28:2017; IEC 60601-2-54:2018; NEMA PS 3.1-3.20 (2016) and ISO 10993-1:2018.
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005, is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests (integration and functional) were conducted on the MULTIX Impact E during product development.
The risk analysis was completed and risk controls were implemented to mitigate identified hazards. The test results support that all the software specifications have met the acceptance criteria. Verification and validation testing were found acceptable to support the claim of substantial equivalence.
#### 10. General Safety and Effectiveness Concerns
Instructions for use are included within the device labeling and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features, including visual and audible warnings, are incorporated into the system design. In addition, the MULTIX Impact E Radiography X-ray system is continually monitored and if an error occurs the system functions will be blocked and an error message will be displayed.
40 Liberty Boulevard
Malvern, PA 19355
USA
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
Risk management is ensured via a hazard analysis which is used to identify potential hazards. These potential hazards are controlled via software development, verification, and validation testing. To minimize electrical, and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore, the operators are healthcare professionals familiar with and responsible for the x-ray examinations to be performed.
#### 11. Conclusion as to Substantial Equivalence
The MULTIX Impact E has the similar indications for use as the predicate device, MULTIX Impact (K203345). The operating environment is same and mechanical design is similar.
Verification and validation testing demonstrate that the MULTIX Impact E performs as intended. The non-clinical test data demonstrate that the MULTIX Impact E device performance is comparable to the predicate device that is currently marketed for the same intended use.
In summary, Siemens concludes that the MULTIX Impact E does not introduce any new potential safety risks and is substantially equivalent to and performs as well as the predicate device, MULTIX Impact (K203345).
#### 12. Guidance documents
The following FDA guidance documents were utilized in the documentation of this Premarket Notification:
- Content of Premarket Submissions for Management of Cybersecurity in Medical ● Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: October 2, 2014
- Information to Support a Claim of Electromagnetic Compatibility (EMC) of . Electrically-Powered Medical Devices Guidance for Industrv and Food and Drug Administration Staff Document issued on July 11, 2016.
- Pediatric Information for X-ray Imaging Device Premarket Notifications Guidance . for Industry and Food and Drug Administration Staff Document issued on November 28, 2017.
- Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: September 1, 2016
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots.
- Guidance for Industry and FDA Staff Guidance for the Content of Premarket . Submissions for Software contained in Medical Devices Document issued on: May 11, 2005
- Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: September 14, 2018
- The 510(k) Program: Evaluating Substantial Equivalence in Premarket . Notifications [510(k)] Guidance for Industry and Food and Drug Administration Staff Document issued on: July 28, 2014
- • Radio Frequency Wireless Technology in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document issued on: August 14, 2013
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.