K220916 · Beta Bionics, Inc. · QJI · May 19, 2023 · Clinical Chemistry
Device Facts
Record ID
K220916
Device Name
iLet® Dosing Decision Software
Applicant
Beta Bionics, Inc.
Product Code
QJI · Clinical Chemistry
Decision Date
May 19, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1356
Device Class
Class 2
Attributes
Software as a Medical Device, Therapeutic, Pediatric
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Glycemic control (HbA1c)
Adaptive control algorithms
—
—
—
—
Prospective, multi-center, randomized controlled trial (RCT) with 440 adult and child participants
—
Indications for Use
The iLet Dosing Decision Software is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps. A self-monitoring of blood glucose (SMBG) meter may also be used for manual input of blood glucose values to continue insulin dosing for a limited period of time when input from the iCGM is temporarily not available. The iLet Dosing Decision Software autonomously determines and commands an increase, decrease, maintenance, or suspension of all basal doses of insulin and autonomously determines and commands correction doses of insulin based on input from an iCGM, and it autonomously determines and commands meal doses of insulin based on meal announcements. iLet Dosing Decision Software is intended for the management of type 1 diabetes mellitus in people 6 years of age or older. iLet Dosing Decision Software is intended for single patient use and requires a prescription.
Device Story
iLet Dosing Decision Software, part of iLet Bionic Pancreas, manages type 1 diabetes by automating insulin delivery. Inputs include real-time glucose data from compatible iCGM, user-provided body mass, and qualitative meal announcements (breakfast, lunch, dinner). Device autonomously calculates and commands basal rate adjustments, correction doses, and meal doses via compatible ACE pump. Features 'BG-Run mode' for temporary manual blood glucose input if iCGM data is unavailable. Used by patients in home/daily settings; requires prescription. Healthcare providers use output to monitor glycemic control. Benefits include automated glucose management, reduced burden of manual dosing, and improved time-in-range. System supports iOS-based 'iLet Go App' for data sharing with Beta Bionics Cloud and over-the-air updates.
Clinical Evidence
Randomized, prospective, multicenter pivotal trial (N=440; 275 adults, 165 pediatrics) compared iLet Bionic Pancreas to Usual Care over 13 weeks. Primary endpoints: HbA1c and time < 54 mg/dL. Results showed significant improvement in time in range (70-180 mg/dL) for iLet groups (p<0.001) and reduction in mean CGM glucose compared to control. Ancillary study (N=54) validated 'BG-Run mode' safety using manual fingerstick inputs for ~49.5 hours with no severe hypoglycemia or DKA events. Human factors study (N=60) confirmed safe use by intended users.
Technological Characteristics
Software-only device (iAGC). Operates as an adaptive control algorithm system. Integrates with iCGM and ACE pumps via wireless communication. Features include autonomous basal, bolus, and meal dose controllers. Cybersecurity mitigations include software bill of materials and penetration testing. Compliant with 21 CFR 862.1356 (OJI).
Indications for Use
Indicated for management of type 1 diabetes mellitus in patients 6 years of age or older. Requires prescription. Intended for single-patient use.
Regulatory Classification
Identification
An interoperable automated glycemic controller is a device intended to automatically calculate drug doses based on inputs such as glucose and other relevant physiological parameters, and to command the delivery of such drug doses from a connected infusion pump. Interoperable automated glycemic controllers are designed to reliably and securely communicate with digitally connected devices to allow drug delivery commands to be sent, received, executed, and confirmed. Interoperable automated glycemic controllers are intended to be used in conjunction with digitally connected devices for the purpose of maintaining glycemic control.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) An appropriate, as determined by FDA, clinical implementation strategy, including data demonstrating appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use.
(A) The clinical data must be representative of the performance of the device in the intended use population and in clinically relevant use scenarios and sufficient to demonstrate appropriate, as determined by FDA, clinical performance of the device for its intended use, including all of its indications for use.
(B) For devices indicated for use with multiple therapeutic agents for the same therapeutic effect (
*e.g.,* more than one type of insulin), data demonstrating performance with each product or, alternatively, an appropriate, as determined by FDA, clinical justification for why such data are not needed.(C) When determined to be necessary by FDA, the strategy must include postmarket data collection to confirm safe real-world use and monitor for rare adverse events.
(ii) Results obtained through a human factors study that demonstrates that an intended user can safely use the device for its intended use.
(iii) A detailed and appropriate, as determined by FDA, strategy to ensure secure and reliable means of data transmission with other intended connected devices.
(iv) Specifications that are appropriate, as determined by FDA, for connected devices that shall be eligible to provide input to (
*e.g.,* specification of glucose sensor performance) or accept commands from (*e.g.,* specifications for drug infusion pump performance) the controller, and a detailed strategy for ensuring that connected devices meet these specifications.(v) Specifications for devices responsible for hosting the controller, and a detailed and appropriate, as determined by FDA, strategy for ensuring that the specifications are met by the hosting devices.
(vi) Documentation demonstrating that appropriate, as determined by FDA, measures are in place (
*e.g.,* validated device design features) to ensure that safe therapy is maintained when communication with digitally connected devices is interrupted, lost, or re-established after an interruption. Validation testing results must demonstrate that critical events that occur during a loss of communications (*e.g.,* commands, device malfunctions, occlusions, etc.) are handled and logged appropriately during and after the interruption to maintain patient safety.(vii) A detailed plan and procedure for assigning postmarket responsibilities including adverse event reporting, complaint handling, and investigations with the manufacturers of devices that are digitally connected to the controller.
(2) Design verification and validation documentation must include appropriate design inputs and design outputs that are essential for the proper functioning of the device that have been documented and include the following:
(i) Risk control measures to address device system hazards;
(ii) Design decisions related to how the risk control measures impact essential performance; and
(iii) A traceability analysis demonstrating that all hazards are adequately controlled and that all controls have been validated in the final device design.
(3) The device shall include appropriate, as determined by FDA, and validated interface specifications for digitally connected devices. These interface specifications shall, at a minimum, provide for the following:
(i) Secure authentication (pairing) to connected devices;
(ii) Secure, accurate, and reliable means of data transmission between the controller and connected devices;
(iii) Sharing of necessary state information between the controller and any connected devices (
*e.g.,* battery level, reservoir level, sensor use life, pump status, error conditions);(iv) Ensuring that the controller continues to operate safely when data is received in a manner outside the bounds of the parameters specified;
(v) A detailed process and procedures for sharing the controller's interface specification with connected devices and for validating the correct implementation of that protocol; and
(vi) A mechanism for updating the controller software, including any software that is required for operation of the controller in a manner that ensures its safety and performance.
(4) The device design must ensure that a record of critical events is stored and accessible for an adequate period to allow for auditing of communications between digitally connected devices, and to facilitate the sharing of pertinent information with the responsible parties for those connected devices. Critical events to be stored by the controller must, at a minimum, include:
(i) Commands issued by the controller, and associated confirmations the controller receives from digitally connected devices;
(ii) Malfunctions of the controller and malfunctions reported to the controller by digitally connected devices (
*e.g.,* infusion pump occlusion, glucose sensor shut down);(iii) Alarms and alerts and associated acknowledgements from the controller as well as those reported to the controller by digitally connected devices; and
(iv) Connectivity events (
*e.g.,* establishment or loss of communications).(5) The device must only receive glucose input from devices cleared under § 862.1355 (integrated continuous glucose monitoring system), unless FDA determines an alternate type of glucose input device is designed appropriately to allow the controller to meet the special controls contained within this section.
(6) The device must only command drug delivery from devices cleared under § 880.5730 of this chapter (alternate controller enabled infusion pump), unless FDA determines an alternate type of drug infusion pump device is designed appropriately to allow the controller to meet the special controls contained within this section.
(7) An appropriate, as determined by FDA, training plan must be established for users and healthcare providers to assure the safety and performance of the device when used. This may include, but not be limited to, training on device contraindications, situations in which the device should not be used, notable differences in device functionality or features compared to similar alternative therapies, and information to help prescribers identify suitable candidate patients, as applicable.
(8) The labeling required under § 809.10(b) of this chapter must include:
(i) A contraindication for use in pediatric populations except to the extent clinical performance data or other available information demonstrates that it can be safely used in pediatric populations in whole or in part.
(ii) A prominent statement identifying any populations for which use of this device has been determined to be unsafe.
(iii) A prominent statement identifying by name the therapeutic agents that are compatible with the controller, including their identity and concentration, as appropriate.
(iv) The identity of those digitally connected devices with which the controller can be used, including descriptions of the specific system configurations that can be used, per the detailed strategy submitted under paragraph (b)(1)(iii) of this section.
(v) A comprehensive description of representative clinical performance in the hands of the intended user, including information specific to use in the pediatric use population, as appropriate.
(vi) A comprehensive description of safety of the device, including, for example, the incidence of severe hypoglycemia, diabetic ketoacidosis, and other relevant adverse events observed in a study conducted to satisfy paragraph (b)(1)(i) of this section.
(vii) For wireless connection enabled devices, a description of the wireless quality of service required for proper use of the device.
(viii) For any controller with hardware components intended for multiple patient reuse, instructions for safely reprocessing the hardware components between uses.
Predicate Devices
Control-IQ Technology (Tandem Diabetes Care, Inc., K200467)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
## I Background Information:
A 510(k) Number
K220916
B Applicant
Beta Bionics, Inc.
C Proprietary and Established Names
iLet® Dosing Decision Software
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| QJI | Class II | 21 CFR 862.1356 – Interoperable Automatic Glycemic Controller | CH – Clinical Chemistry |
E Purpose for Submission:
New device
## II Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
The iLet Dosing Decision Software is intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps. A self-monitoring of blood glucose (SMBG) meter may also be used for manual input of blood glucose values to continue insulin dosing for a limited period of time when input from the iCGM is temporarily not available.
The iLet Dosing Decision Software autonomously determines and commands an increase, decrease, maintenance, or suspension of all basal doses of insulin and autonomously determines
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
and commands correction doses of insulin based on input from an iCGM, and it autonomously determines and commands meal doses of insulin based on meal announcements.
iLet Dosing Decision Software is intended for the management of type 1 diabetes mellitus in people 6 years of age or older. iLet Dosing Decision Software is intended for single patient use and requires a prescription.
## C Special Conditions for Use Statement(s):
Rx – For prescription use only.
Do not use the iLet Dosing Decision Software if you are unable or unwilling to test blood glucose (BG) levels with an SMBG meter when input from the iCGM is not available.
Do not use the iLet Dosing Decision Software if you are unable or unwilling to recognize and respond to iLet safety alerts.
Do not use the iLet System if you are taking hydroxyurea, also known as Hydrea. This medication is sometimes used in the treatment of blood disorders and some kinds of cancer. The use of hydroxyurea can result in falsely elevated sensor glucose readings. The iLet System relies on sensor glucose readings to adjust insulin, provide insulin doses, and provide high and low glucose alerts. If the iLet System receives sensor readings that are higher than actual glucose levels, it could result in missed hypoglycemia alerts and potential errors in diabetes management, such as too much insulin being delivered. Hydroxyurea can also result in errors when reviewing, analyzing, and interpreting historical patterns for assessing glucose control.
Do not use the iLet ACE Pump and Dosing Decision Software in people under 6 years of age. The iLet ACE Pump and Dosing Decision Software have not been studied in these populations.
Do not use the iLet Dosing Decision Software in people who are pregnant, on dialysis or critically ill. The iLet Dosing Decision Software has not been studied in these populations.
The system should NOT be used in hospitalized people as the safety of the technology has not been evaluated in this population.
The iLet Dosing Decision Software is only for use with insulin U-100 lispro (Humalog), insulin U-100 aspart (Novolog), or insulin U-100 aspart (Fiasp).
Do not use the iLet Dosing Decision Software with the U-100 Fiasp in individuals under 18 years of age. The iLet Dosing Decision Software has not been studied with the U-100 Fiasp in these populations.
The iLet Dosing Decision Software is only for use with the Dexcom G6 iCGM. When using the iLet device, wear an iCGM.
K220916 - Page 2 of 13
{2}
Remove the iLet device, steel infusion set, CGM sensor, and CGM transmitter before undergoing radiation therapy, Magnetic Resonance Imaging (MRI), Computed Tomography (CT) scan, or diathermy treatment procedures. Exposure of the iLet device, steel infusion set, CGM sensor, or CGM transmitter to any of these may damage them.
Do not expose your iLet device, steel infusion set, CGM transmitter, or CGM sensor to equipment used in procedures for Pacemaker/Automatic Implantable Cardioverter Defibrillator (AICD) placement or reprogramming, Cardiac Catheterization, or Nuclear Stress Test.
The iLet Go App is compatible with the iOS platform. The iLet Go App provides the ability to perform over-the-air updates and / or pull data from an iLet device to share with the Beta Bionics Cloud. The iLet Go App is not currently compatible with Android or other platforms.
If your CGM is offline for an extended period of time, dosing will stop and you should switch to alternative therapy until you are able to reconnect to a CGM sensor. A countdown timer will appear before dosing would stop.
## III Device Description
The iLet Dosing Decision software is part of the iLet Bionic Pancreas and is intended for use by people with diabetes. The iLet Dosing Decision software works in conjunction with a compatible alternate controller enabled (ACE) pump. The iLet Dosing Decision Software requires initialization with the user's body mass (body weight), as well as meal announcements. When initiating a meal announcement with the iLet Dosing Decision Software, the user qualitatively approximates carbohydrate content (meal size) relative to the usual carbohydrate content for each of the three meal types (breakfast, lunch, or dinner). The iLet Dosing Decision Software autonomously determine the size of the insulin dose in response to a meal announcement by the user.
The iLet Dosing Decision software works to control glucose to a user-set glucose target of "lower" (110 mg/dL), "usual" (120 mg/dL), or "higher" (130 mg/dL) within the device settings. Users can also set glucose targets specific for sleep.
The iLet Dosing Decision software also includes a feature, called "BG-Run mode", which enables the device to continue insulin delivery in the event CGM-data is unavailable. Use of the feature, however, should be temporary and always for the shortest duration possible, with the goal to resume CGM-guided insulin dosing as soon as possible.
## IV Substantial Equivalence Information:
A Predicate Device Name(s):
Control IQ-Technology
B Predicate 510(k) Number(s):
K200467
K220916 - Page 3 of 13
{3}
K220916 - Page 4 of 13
C Comparison with Predicate(s):
| Device & Predicate Device(s): | K220916 | K200467 |
| --- | --- | --- |
| Device Trade Name | iLet® Dosing Decision Software | Control-IQ Technology |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Intended for use with compatible integrated continuous glucose monitors (iCGM) and alternate controller enabled (ACE) pumps to automatically increase, decrease, and suspend delivery of basal insulin, as well as command correction doses, based on glucose values. | Same |
| Age Range of Intended Users | 6 years and older | Same |
| General Device Characteristic Differences | | |
| Specific Drug/Biologic Use | U-100 insulins: NovoLog, Humalog, and Fiasp | U-100 insulins: NovoLog and Humalog |
V Standards/Guidance Documents Referenced:
Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices: Guidance for Industry and FDA Staff
General Principles of Software Validation: Guidance for Industry and FDA Staff
Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices: Guidance for Industry and Food and Drug Administration Staff
AAMI TIR36:2007 - Validation of software for regulated processes
AAMI/AAMI HE75:2009(R2018)- Human factors engineering - Design of medical devices
AAMI/IEC TIR62366-2:2016 -Medical Devices Part 2: Guidance on the Application of Usability Engineering to Medical Devices
{4}
IEC 62304:2006+A1:2015 - Medical device software - Software life-cycle processes
AAMI/IEC 62366-1:2015 - Medical devices – Application of usability engineering to medical devices
ISO 14155:2020 - Clinical investigation of medical devices for human subjects - Good clinical practice
VI Performance Characteristics:
A. Analytical Performance
For the purposes of analytical and clinical validation testing, the iLet Dosing Decision Software was installed on the iLet ACE pump (K223846), which was paired with the Dexcom G6 continuous glucose monitoring system (K223931). Details on the performance characteristics of these devices can be found in the public decision summaries for each device.
B. Other Supportive Instrument Performance Characteristics Data
1. Summary of Clinical Testing
The sponsor conducted a randomized-controlled, prospective, multicenter pivotal clinical trial consisting of 440 participants (275 subjects ≥ 18 years old, and 165 subjects 6-17 years old) with type 1 diabetes. Subjects are randomized into the Bionic Pancreas (BP) Group or the Usual Care (UC) Group to complete the 13-week study period. The sponsor also conducted an ancillary study in which the iLet system’s functioning was based only on SMBG manual entry, and no CGM data was available to the system.
K220916 - Page 5 of 13
{5}
| Study Feature | Description |
| --- | --- |
| Title | The Insulin-Only Bionic Pancreas Pivotal Trial: Testing the iLet in Adults and Children with Type 1 Diabetes |
| Study Design | Randomized, prospective, multicenter clinical trial |
| Number of Sites | 16 |
| Population | Adults and children who were diagnosed with type 1 diabetes and were using insulin for at least a year were eligible for the study. |
| Sample Size | 440 participants (275 adults, 165 pediatrics) |
| Treatment Groups | **Pediatric (6-17 years old) Study:** Random assignment in a 2:1 ratio to the BP Group (with Novolog or Humalog, hereafter referred to as the iLet-N/H group, ~ 110 participants) or the UC Group (~55 participants), such that a minimum of 100 in the BP group will complete the 13-week Randomized Controlled Trial (RCT).
**Adult (≥18 years) Study:** Random assignment in a 2:2:1 ratio to the BP Group (with Novolog or Humalog, iLet-N/H, ~ 110 participants), the BP Fiasp Group (iLet-F, ~ 110 participants), or the UC Group (~55 participants), such that a minimum of 100 in each BP Group complete the 13-week RCT. |
| Study Duration | 3 months for primary study |
| Protocol Overview/Synopsis | The 13-week, parallel-group, multi-center RCT Period is designed to compare the insulin-only iLet BP Group using insulin lispro, insulin aspart, or Fiasp (adults only); and a control group following standard care (SC Group).
**Key Eligibility Criteria**
**Inclusion**
1. Clinical diagnosis of T1D for at least one year and using insulin for at least 1 year
2. Diabetes managed using the same regimen (either pump or MDI, with or without CGM) for ≥ 3 months
3. Age ≥ 6 years old
4. If a GLP-1 agonist or pramlintide is being used, participant must be |
K220916 - Page 6 of 13
{6}
| Study Feature | Description |
| --- | --- |
| | willing to discontinue use while the iLet BP system is being used, including the randomized trial and extension study.
**Exclusion**
1. Plan to change usual diabetes regimen in the next 3 months
2. Use of Apidra as the pre-study rapid-acting insulin analog and unwilling to switch to lispro or aspart for the duration of the study
3. Known hemoglobinopathy (sickle cell trait is not an exclusion)
4. History of cystic fibrosis, pancreatitis, or other pancreatic disease, including pancreatic tumor or insulinoma, or history of complete pancreatectomy
5. Current use of SGLT2 inhibitors or a sulfonylurea drug (use more than 3 months prior to enrollment is acceptable)
6. Pregnant, breast feeding, plan to become pregnant in the next 3 months, or sexually active without use of contraception
7. Renal failure on dialysis or with an eGFR <30mL/min
**Primary Endpoints**
1. HbA1c at 13 weeks
2. CGM time < 54 mg/dl
**Key Safety Outcomes**
• severe hypoglycemia
• diabetic ketoacidosis
• other serious adverse events |
K220916 - Page 7 of 13
{7}
Demographics
| Adult Study | Standard Care (N = 54) | iLet-F (N = 114) | iLet-N/H (N = 107) |
| --- | --- | --- | --- |
| Age (years) | | | |
| Mean (SD) | 44 (16) | 42 (16) | 44 (15) |
| 18 to < 25 | 7 (13%) | 21 (18%) | 16 (15%) |
| 25 to < 45 | 21 (39%) | 46 (40%) | 38 (36%) |
| 45 to < 60 | 13 (24%) | 29 (25%) | 33 (31%) |
| ≥ 60 | 13 (24%) | 18 (16%) | 20 (19%) |
| Range | 18 to 79 | 18 to 83 | 18 to 73 |
| Diabetes Duration (years) | | | |
| Mean (SD) | 29 (14) | 24 (14) | 26 (14) |
| HbA1c Level at Randomization (%)* | | | |
| Mean (SD) | 7.6 (1.2) | 7.8 (1.2) | 7.6 (1.2) |
| <7.0 | 18 (34%) | 31 (27%) | 37 (35%) |
| 7.1-7.4 | 6 (11%) | 17 (15%) | 17 (16%) |
| 7.5%-<9.5 | 26 (49%) | 56 (50%) | 46 (43%) |
| ≥9.5 | 3 (6%) | 9 (8%) | 7 (7%) |
| Range | 5.5 to 11.3 | 5.3 to 14.9 | 5.5 to 13.1 |
| Sex – Female n (%) | 26 (48%) | 62 (54%) | 52 (49%) |
| Race/Ethnicity Group n (%) | | | |
| White non-Hispanic | 47 (87%) | 98 (86%) | 85 (79%) |
| Black non-Hispanic | 2 (4%) | 10 (9%) | 14 (13%) |
| Hispanic or Latino | 3 (6%) | 6 (5%) | 7 (7%) |
| Asian | 1 (2%) | 0 (0%) | 0 (0%) |
| American Indian/Alaskan Native | 1 (2%) | 0 (0%) | 0 (0%) |
| More than one race | 0 (0%) | 0 (0%) | 1 (<1%) |
| Unknown | 0 (0%) | 0 (0%) | 0 (0%) |
| Pediatric Study | Standard Care (N = 53) | iLet-N/H (N = 114) |
| --- | --- | --- |
| Age (years) | | |
| Mean (SD) | 12 (3) | 12 (3) |
| 6 to <12 | 47 (42%) | 47 (42%) |
| 12 to <18 | 65 (58%) | 65 (58%) |
| Range | 6 to17 | 6 to17 |
| Diabetes Duration (years) | | |
| Mean (SD) | 7 (4) | 6 (4) |
K220916 - Page 8 of 13
{8}
| Range | 1 to 15 | 1 to 16 |
| --- | --- | --- |
| HbA1c Level at Randomization (%) | | |
| Mean (SD) | 7.8 (1.1) | 8.1 (1.2) |
| <7.0 | 12 (23%) | 18 (16%) |
| 7.1-7.4 | 19 (36%) | 34 (30%) |
| 7.5%-<9.5 | 15 (28%) | 33 (29%) |
| ≥9.5 | 7 (13%) | 27 (24%) |
| Range | 5.8 to 10.6 | 6.1 to 12.2 |
| Sex – Female n (%) | 15 (28%) | 55 (49%) |
| Race/Ethnicity Group n (%) | | |
| White non-Hispanic | 36 (68%) | 72 (64%) |
| Black non-Hispanic | 3 (6%) | 13 (12%) |
| Hispanic or Latino | 8 (15%) | 16 (14%) |
| Asian | 2 (4%) | 2 (2%) |
| American Indian/Alaskan Native | 0 (0%) | 1 (<1%) |
| More than one race | 4 (8%) | 6 (5%) |
| Unknown | 0 (0%) | 2 (2%) |
Safety Results
| | Adult (Age ≥ 18 yr) | | | Pediatrics (Age < 18 yr) | |
| --- | --- | --- | --- | --- | --- |
| | iLet – N/H (N=107) | iLet – F (N=113) | Standard of Care (N=54) | iLet – N/H (N=112) | Standard of Care (N=53) |
| Total Number of Adverse Events (AEs) | 63 | 83 | 6 | 181 | 4 |
| Severe Hypoglycemic (SH) Events | | | | | |
| Number of SH Events per Participant n (%) | | | | | |
| 0 | 100 (93%) | 111 (97%) | 53 (98%) | 109 (97%) | 52 (98%) |
| 1 | 7 (7%) | 3 (3%) | 0 (0%) | 3 (3%) | 1 (2%) |
| 2 | 0 (0%) | 0 (0%) | 1 (2%) | 0 (0%) | 0 (0%) |
| Incidence Rate per 100 Person-Years | 25.5 | 10.2 | 14.2 | 10.4 | 7.3 |
| Diabetic Ketoacidosis (DKA) Events | | | | | |
| Number of OKA Events | | | | | |
| 0 | 100 (93%) | 108 (96%) | 53 (98%) | 108 (96%) | 52 (98%) |
| 1 | 7 (7%) | 3 (3%) | 0 (0%) | 3 (3%) | 1 (2%) |
| 2 | 0 (0%) | 0 (0%) | 1 (2%) | 0 (0%) | 0 (0%) |
K220916 - Page 9 of 13
{9}
K220916 - Page 10 of 13
| | Adult (Age ≥ 18 yr) | | | Pediatrics (Age < 18 yr) | |
| --- | --- | --- | --- | --- | --- |
| | iLet – N/H (N=107) | iLet – F (N=113) | Standard of Care (N=54) | iLet – N/H (N=112) | Standard of Care (N=53) |
| per Participant n (%) | | | | | |
| 0 | 107 (100%) | 112 (98%) | 54 (100%) | 112 (100%) | 53 (100%) |
| 1 | 0 (0%) | 2 (2%) | 0 (0%) | 0 (0%) | 0 (0%) |
| Incidence Rate per 100 Person-Years | 0 | 6.8 | 0 | 0 | 0 |
| **Other Serious Adverse Events (SAEs)** | | | | | |
| Number of SAEs per Participant n (%) | | | | | |
| 0 | 106 (>99%) | 114 (100%) | 53 (98%) | 110 (98%) | 52 (98%) |
| 1 | 1 (1%) | 0 (0%) | 1 (2%) | 2 (2%) | 1 (2%) |
| Incidence Rate per 100 Person-Years | 3.6 | 0 | 7.1 | 6.9 | 7.3 |
| **Participants with Worsening of HbA1c from baseline to 13 weeks by >0.5%** | | | | | |
| n (%) | 4 (4%) | 7 (6%) | 4 (8%) | 13 (12%) | 4 (8%) |
| **Other Adverse Events (N Events / N Participants)** | | | | | |
| Hyperglycemia with or without Ketosis Related to Study Device | 34 / 27 | 52 / 40 | 0 / 0 | 126 / 68 | 0 / 0 |
| Hyperglycemia with or without Ketosis Not Related to Study Device | 13 / 12 | 14 / 12 | 0 / 0 | 41 / 32 | 2 / 1 |
| Non-Severe Hypoglycemia | 1 / 1 | 0 / 0 | 0 / 0 | 1 / 1 | 0 / 0 |
| Other Reportable Adverse Events | 7 / 7 | 12 / 10 | 3 / 3 | 8 / 7 | 0 / 0 |
{10}
Observed Results
| | Adult (Age ≥ 18 years) | | | Pediatric (Age < 18 years) | |
| --- | --- | --- | --- | --- | --- |
| | iLet – N/H (N=107) | iLet – F (N=113) | Standard of Care (N=54) | iLet – N/H (N=112) | Standard of Care (N=53) |
| % Time<54 mg/dL | | | | | |
| Baseline median (IQR) | 0.21% (0.02%, 0.57%) | 0.12% (0.02%, 0.68%) | 0.11% (0.00%, 0.37%) | 0.20% (0.03%, 0.59%) | 0.22% (0.03%, 0.46%) |
| Week 13 median (IQR) | 0.33% (0.14%, 0.52%) | 0.00% (-0.07%, 0.05%) | -* | -0.04% (-0.13%, 0.03%) [0.24] | -* |
| % Time 70-180 mg/dL | | | | | |
| Baseline mean (SD) | 56% (19%) | 54% (18%) | 53% (21%) | 47% (17%) | 48% (19%) |
| Week 13 mean (SD) | 69% (8%) | 71% (8%) | 58% (17%) | 60% (8%) | 50% (16%) |
| % Time 70-180 mg/dL | | | | | |
| Adjusted Group Difference over 13 weeks (95% CI) [p-value] | 11% (8%, 13%) [<0.001] | 14% (11%,17%) [<0.001] | -* | 10% (7%, 13%) [<0.001] | -* |
| Mean CGM Glucose (mg/dL) | | | | | |
| Baseline mean (SD) | 179 (41) | 181 (34) | 186 (42) | 195 (39) | 195 (42) |
| Week 13 mean (SD) | 157 (12) | 155 (11) | 174 (30) | 172 (14) | 187 (34) |
| Adjusted Group Difference over 13 weeks (95% CI) [p-value] | -16 (-20, -11) [<0.001] | -18 (-22, -14) [<0.001] | -* | -15 (-21, -9) [<0.001] | -* |
| Change in HbA1c after 13 Weeks | | | | | |
| Baseline | 7.6 | 7.8 | 7.6 | 8.1 | 7.8 |
| Week 13 | 7.1 | 7.1 | 7.5 | 7.5 | 7.8 |
* no data available
K220916 - Page 11 of 13
{11}
BG-Run Mode
The sponsor conducted an ancillary study after the conclusion of the RTC to evaluate the safety of utilizing BG values obtained by fingerstick instead of CGM as input into the iLet System. Fifty-four subjects (43 adults and 11 pediatrics) participated in the ancillary-phase of the study. Median time using BG measurements as input for the iLet System was 49.5 hours, with all participants having at least 46.5 hours of continued BG-run mode use. Median number of BG measurements entered during the first 24 hours was 9 and during the second 24 hours was 10. There were no incidences of severe hypoglycemia, DKA, or other serious adverse events (SAEs). Results of the ancillary study indicated that the frequency of low-sensor glucose events during the BG-run period appeared similar to the pre-randomization period and the RCT period, and the frequency of high-sensor glucose events during the BG-run period appeared higher than in the RCT-period.
2. Human Factors Testing
Human factors validation tests were conducted with the iLet Dosing Decision Software installed on the iLet ACE insulin pump (K223846). The summative human factors validation study was performed with 60 total users (45 primary users and 15 secondary users). All study participants received training that was consistent with the training that patients would receive with the commercial product. Usability evaluations assessed comprehension and usability of the device for critical device tasks. Results of the study demonstrated that the device could be used safely by intended users in the intended use environment when used in combination with a digitally connected device.
3. Insulin Compatibility
The iLet Dosing Decision Software used with the iLet Bionic Pancreas System is designed to use rapid-acting U-100 insulins including: NovoLog® (insulin aspart), Humalog® (insulin lispro), and Fiasp® (insulin aspart). These insulins were used in the pivotal clinical study for this device and no other insulins have been tested for use with the device.
4. Data Logging
Software verification testing has demonstrated the device records timestamped critical events, including information related to its state, user inputs, and device settings, as required by the special controls.
5. Interoperability
A plan and approach for interoperability were provided according to the FDA Guidance "Design Considerations and Pre-market Submission Recommendations for Interoperable Medical Devices - Guidance for Industry and Food and Drug Administration Staff" (issued on September 6, 2017) and determined to be adequate to support and clearly specify expectations, requirements, and interface specifications to potential interoperable devices. In addition, the plans provided by the sponsor covered their approach to working with connected device companies regarding contractual issues, interfaces for data communication and exchange, and post-market reporting procedures and responsibilities (e.g., who is responsible for investigating and reporting complaints, malfunctions, and adverse events).
K220916 - Page 12 of 13
{12}
The sponsor additionally provided validated software protocols intended to ensure secure, accurate, and reliable communication with digital interfacing devices, as well as failsafe design features to mitigate the risks associated with interruption of communication with digitally connected devices. These protocols were reviewed and found to be adequate.
## 6. Cybersecurity
Detailed information on cybersecurity of the device was reviewed and found to be acceptable. The sponsor also provided a software bill of materials, which provided details on all software used in the device and the hardware platform that the device was installed on. This included all manufacturer-developed, commercially licensed, open source, and off-the-shelf software components (including firmware as relevant), along with an identification of the hardware runtime environment in which each resides, with relevant version and/or model information, as well as details on whether each component was actively supported by its manufacturer or legacy licensed.
## 7. Postmarket Surveillance Study
There is uncertainty remaining regarding the risk/benefit profile of the device when used in the broader intended use population. While the premarket clinical study provided to support the 510(k) showed some benefits, it was not adequately powered to assess differences in the rates of safety events (e.g., diabetic ketoacidosis and severe hypoglycemia). Accordingly, a postmarket surveillance study will be ordered by FDA to confirm understanding of safety.
## VII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
## VIII Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K220916 - Page 13 of 13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.