The QT ECG System is intended to acquire, record and process an electrocardiographic signal so that it can be transmitted digitally via Bluetooth technology to a cell-phone or mobile device, then to a remote location. The QT ECG System is indicated for use on infants, children and adults. It is designed to be operated by adults in the home, or by healthcare workers in non-acute care clinical facilities (such as nursing facilities), to record and transmit a 12lead ECG and rhythm strip in near real-time to enable review at a physician's office, hospital or other medical receiving centers.
Device Story
QT ECG is a hand-held, cordless 12-lead ECG system. Input: electrical cardiac signals captured via disposable, self-adhesive electrode strips. Transformation: recorder digitizes signals and transmits via Bluetooth to a mobile device (smartphone/tablet) running the QT ECG App; data is then uploaded to a cloud-based web service. Output: 12-lead ECG and rhythm strip for remote review by a physician. Used in homes or non-acute clinical settings. The system provides measurement of ECG data but does not perform automated interpretation, diagnostic alarms, or monitoring. Benefits include near real-time transmission of ECG data to medical professionals for clinical decision-making.
Clinical Evidence
Bench testing only. Validation included biocompatibility (cytotoxicity, irritation, sensitization), software validation, cybersecurity evaluation, electromagnetic compatibility (EMC), electrical safety, wireless interference, and human factors/usability testing. Performance testing confirmed the durability and measurement accuracy of the system and electrode strips across all sizes.
Technological Characteristics
Hand-held, cordless 12-lead ECG recorder. Materials: hydrogel and medical-grade tape (electrode strips). Connectivity: Bluetooth to mobile device, then cloud-based web service. Power: battery-operated. Software: non-defibrillator-proof, Class II device. Complies with IEC 60601-2-25, IEC 60601-1-2, and ISO 10993 standards.
Indications for Use
Indicated for infants, children, and adults. Designed for use by adults in home settings or healthcare workers in non-acute clinical facilities (e.g., nursing facilities) to record and transmit 12-lead ECGs and rhythm strips.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 26, 2022
QT Medical, Inc. Jackal Chen Department Director 1360 Valley Vista Drive. Suite 203 Diamond Bar, California 91765
Re: K220795
Trade/Device Name: QT ECG Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter And Receiver Regulatory Class: Class II Product Code: DXH Dated: August 25, 2022 Received: August 29, 2022
Dear Jackal Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220795
Device Name QT ECG
#### Indications for Use (Describe)
The QT ECG System is intended to acquire, record and process an electrocardiographic signal so that it can be transmitted digitally via Bluetooth technology to a cell-phone or mobile device, then to a remote location. The QT ECG System is indicated for use on infants, children and adults. It is designed to be operated by adults in the home, or by healthcare workers in non-acute care clinical facilities (such as nursing facilities), to record and transmit a 12lead ECG and rhythm strip in near real-time to enable review at a physician's office, hospital or other medical receiving centers.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k) 510(k) Summary
# 510(k) SUMMARY
- Type of Submission: Traditional 5.1
- Date of Summary: 03/07/2022 5.2
| 5.3 Submitter: | QT MEDICAL, Inc. |
|-----------------|------------------------------------------------------------------|
| Address: | 1360 Valley Vista Drive, Suite 203, Diamond Bar,<br>CA 91765, US |
| Phone: | 1-909-323-0007 |
| Fax: | 1-310-755-3108 |
| Representative: | Ruey-Kang Chang, CEO<br>(rk.chang@qtmedical.com) |
#### Identification of the Device: 5.4
| Trade name: | QT ECG |
|------------------------------|-------------------------------------------------------|
| Classification Product Code: | DXH |
| Regulation Number: | 870.2920 |
| Regulation Description: | Telephone electrocardiograph transmitter and receiver |
| Review Panel: | Cardiovascular |
| Device Class: | II |
#### ર્ રંડ Identification of the Predicate Device:
| Predicate Device Name: | QT ECG |
|------------------------------|------------------|
| Manufacturer: | QT MEDICAL, Inc. |
| Classification Product Code: | DXH |
| Regulation number: | 870.2920 |
| Device Class: | II |
| 510(k) Number: | K180157 |
## 5.6 Indications for Use of the Device
Indication for Use:
The QT ECG System is intended to acquire, record and process an electrocardiographic signal so that it can be transmitted digitally via Bluetooth technology to a cell-phone or mobile device, then
{4}------------------------------------------------
Traditional 510(k) 510(k) Summary
to a remote location. The OT ECG System is indicated for use on infants, children and adults. It is designed to be operated by adults in the home, or by healthcare workers in non-acute care clinical facilities (such as nursing homes, skilled nursing facilities), to record and transmit a 12-lead ECG and rhythm strip in near real-time to enable review at a physician's office, hospital or other medical receiving centers.
Intended Use:
The OT ECG System is intended to condition an electrocardiographic signal so that it can be transmitted digitally via Bluetooth technology and cell-phone or communication device to a remote location. The OT ECG System is designed to be used by an adult or a healthcare worker to transmit a 12-lead ECG and rhythm strip in near real-time to enable review at a physician's office, hospital or other medical receiving center.
# 5.7 Device Description
The OT ECG system is a non-defibrillator-proof, hand-held, cordless 12-lead electrocardiograph (ECG) system with Bluetooth connectivity. The OT ECG system consists of 5 major components:
- ·The QT ECG Recorder—Compact device that records 12-lead, resting electrocardiograms, then transmits the recorded data to a mobile device (smartphone, tablet, etc.) paired via Bluetooth. A Bluetooth-enabled mobile device (not included) is needed to operate the QT ECG Recorder, and to send the recoded rhythm strip to a cardiologist or licensed physician for review.
- . The QT ECG Electrode Strip-Disposable, patented electrodes that are prepositioned on a self-adhesive strip.
- The QT ECG App - Software that lets the user uses their mobile device to operate the QT ECG recorder, then send the recorded data via cloud to a certified medical professional for review.
- . Analysis - The analysis module provides ECG measurement from the collected data. It does not make any interpretation of the intervals provided based on factors such as heart rate, ORS duration, etc.
- . Web Service — The web service provides an interface for communication.
The recorded ECG data is saved temporarily on the mobile device until it is transferred via the Internet to the cloud server. The QT ECG System does not have monitoring capabilities and does
{5}------------------------------------------------
Traditional 510(k) 510(k) Summary
not have diagnostic alarm function. The QT ECG System is intended for use on infants, children and adults patients to acquire ECG signals to be transmitted wirelessly via Bluetooth to a mobile device, and then over the Internet to a web service. The QT ECG System is designed to be used by a patient to record and transmit ECG data to a physician's office, hospital or other medical receiving center for review.
## 5.8 Non-clinical Testing
A series of validation activities were conducted on the subject device, QT ECG.
- . Biocompatibility
- in vitro cytotoxicity test -
- white rabbit skin irritation test -
- skin sensitization study (maximization test)
- . Software validation and cybersecurity evaluation
- . Electromagnetic compatibility and electrical safety
- electrical safety test -
- external defibrillation assessment -
- The system is not defibrillation-proof. The defibrillation safety assessment manifests that the patient's safety is assured.
- -EMC test
- cable performance test
- electrical safety in home healthcare environment
- battery safety test
- . Performance
- performance test of electrode and the recorder -
- durability and measurement test of the system -
- wireless interference test -
- Human Factor (Usability) .
All the test results demonstrate OT ECG meets the requirements of its pre-defined acceptance criteria and indication for use, and is substantially equivalent to the predicate device.
The list of claimed standards and regulations for compliance:
{6}------------------------------------------------
| Testing Item | FDA Recognition No. | Standards and Regulations Applied |
|--------------------------------------------------------------|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility | - | Good Laboratory Practice for Nonclinical Laboratory Studies. Title 21 of the U.S. Code of Federal Regulations,<br>Part 58 United States Food and Drug Administration. |
| | - | Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and<br>testing within a risk management process". |
| | 2-222 | ISO 10993-2:2006, Biological evaluation of medical devices -- Part 2: Animal welfare requirements. |
| | 2-245 | ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity. |
| | 2-245 | ANSI/AAMI/ISO 10993-5:2009/(R)2014, Biological Evaluation Of Medical Devices - Part 5: Tests For In<br>Vitro Cytotoxicity. |
| | 2-174 | ISO 10993-10:2010, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin<br>sensitization. |
| | 2-191 | ISO 10993-12:2012, Biological evaluation of medical devices -- Part 12: Sample preparation and reference<br>materials. |
| | 2-191 | ANSI/AAMI/ISO 10993-12:2012, Biological Evaluation Of Medical Devices - Part 12: Sample Preparation<br>And Reference Materials. |
| Software<br>Validation and<br>Cybersecurity<br>Evaluation | - | Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices: 2005. |
| | - | Guidance on Off-The-Shelf Software Use in Medical Devices: 2019. |
| | - | Guidance for content of Premarket Submissions for Management of Cybersecurity in Medical Devices: 2014. |
| | 13-79 | IEC 62304:2006+AMD1:2015, Medical device software - Software life cycle processes. |
| | 5-125 | ISO 14971:2019, Medical devices - Application of risk management to medical devices. |
| | 3-105 | ANSI/AAMI/IEC 60601-2-25:2011/(R)2016, Medical electrical equipment - Part 2-25: Particular requirements<br>for the basic safety and essential performance of electrocardiographs. |
| | | Guidance on Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered<br>Medical Devices: 2016. |
| Electromagnetic<br>Compatibility<br>and Electrical<br>Safety | 19-4 | ANSI/AAMI ES60601-1: 2005 / A2:2010, Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance. |
| | 19-8 | IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and<br>essential performance - Collateral standard: Electromagnetic compatibility. |
| | 19-14 | IEC 60601-1-11:2015, Medical electrical equipment - Part 1-11: General requirements for basic safety and<br>essential performance - Collateral standard: Requirements for medical electrical equipment and medical<br>electrical systems used in the home healthcare environment. |
| | 3-105 | IEC 60601-2-25:2011, Medical electrical equipment - Part 2-25: Particular requirements for the basic safety<br>and essential performance of electrocardiographs. |
| | 19-13 | IEC 62133:2012, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety<br>requirements for portable sealed secondary cells, and for batteries made from them, for use in portable<br>applications. |
| | Performance | 3-52 |
| 19-13 | | IEC 62133:2012, Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety<br>requirements for portable sealed secondary cells, and for batteries made from them, for use in portable<br>applications. |
| Human Factor | | |
{7}------------------------------------------------
#### Traditional 510(k) 510(k) Summary
#### Substantial Equivalence Determination 5.9
The QT ECG submitted in this 510(k) file is compared with the cleared device, former QT ECG (K180157). Differences between the devices are cited as below, and other technological specifications are all the same as that of K180157. The patient population is extended,
{8}------------------------------------------------
Traditional 510(k) 510(k) Summary
| Item | Subject device | Predicate device | Substantial equivalence<br>determination |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Brand Name | QT ECG | QT ECG | |
| Model Name | PCA 500 | - | |
| 510(k) No. | (to be assigned) | K180157 | |
| Indications for<br>Use | The QT ECG System is intended to acquire,<br>record and process an electrocardiographic<br>signal so that it can be transmitted digitally<br>via Bluetooth technology to a cell-phone or<br>mobile device, then to a remote location. The<br>QT ECG System is indicated for use on<br>infants, children and adults. It is designed to<br>be operated by adults in the home, or by<br>healthcare workers in non-acute care clinical<br>facilities (such as nursing homes, skilled<br>nursing facilities), to record and transmit a<br>12-lead ECG and rhythm strip in near<br>real-time to enable review at a physician's<br>office, hospital or other medical receiving<br>centers. | The QT ECG System is intended to acquire,<br>record and process an electrocardiographic<br>signal so that it can be transmitted digitally via<br>Bluetooth technology to a cell-phone or mobile<br>device, then to a remote location. The QT ECG<br>System is indicated for use on adult patients and<br>pediatric patients age 18 – 22 years. It is<br>designed to be used by a patient or other<br>layperson in the home, or by healthcare workers<br>in non-acute care clinical facilities (such as<br>nursing homes, skilled nursing facilities), to<br>record and transmit a 12-lead ECG and rhythm<br>strip in near real-time to enable review at a<br>physician's office, hospital or other medical<br>receiving centers. | Equivalent<br>Both devices are designed to<br>acquire, record and process an<br>electrocardiographic signal.<br>The difference in patient population<br>has been evaluated, verified &<br>validated, and the difference in<br>wording is to specify the user more<br>clearly.<br>Above all, it does not raise new<br>issues of SE. |
so the risk management and verification & validation activities in our design and development control process are all completed.
{9}------------------------------------------------
| QT MEDICAL, Inc. |
|------------------|
| QT ECG |
# Traditional 510(k)
510(k) Summary
| Intended Use | The QT ECG System is intended to condition an electrocardiographic signal so that it can be transmitted digitally via Bluetooth technology and cell-phone or communication device to a remote location. The QT ECG System is designed to be used by an adult or a healthcare worker to transmit a 12-lead ECG and rhythm strip in near real-time to enable review at a physician's office, hospital or other medical receiving center. | The QT ECG System is intended to condition an electrocardiographic signal so that it can be transmitted digitally via Bluetooth technology and cell-phone or communication device to a remote location. The QT ECG System is designed to be used by a patient or another layperson or a healthcare worker to transmit a 12-lead ECG and rhythm strip in near real-time to enable review at a physician's office, hospital or other medical receiving center. | Equivalent<br>The intended use for both devices is the same.<br>The difference in wording is to specify the user more clearly, and it does not raise new issues of SE. |
|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of protection<br>against electrical<br>shock | Type: CF; non-defibrillator proof | | Identical<br>Both devices do not have the defibrillation-proof specification and both devices have stated warning on labeling to control the relevant risk. |
| Recorder storage<br>and transport<br>temperature | -25°C~70°C | -20°C~40°C | Equivalent<br>The storage and transport condition has been tested and met its pre-defined criteria, and it does not raise new issues of SE. |
| Electrode Strip<br>sizes | 7 sizes: S, M, L, XL for adults,<br>1, 2 ,3 for pediatrics | 4 sizes: S, M, L ,XL for adults | Equivalent<br>The performance of new strips for pediatrics has been tested and met |
| | | | its pre-defined criteria, and it does<br>not raise new issues of SE. |
| Electrode Strip<br>materials in<br>contact with the<br>human body | Hydrogel, and<br>Medical grade tape | Hydrogel, and<br>Medical grade tape | Identical<br>Both devices use the same materials<br>on the electrode strip in contact<br>with the human body. |
| Electrode Strip<br>shelf life | 3 year | 1 year | Equivalent<br>The extended shelf life has been<br>tested and met its pre-defined<br>criteria, and it does not raise new<br>issues of SE. |
| Electrode Strip<br>storage and<br>transportation<br>temperature | 5°C~30°C | 5°C~27°C | Equivalent<br>The storage condition has been<br>tested and met its pre-defined<br>criteria, and it does not raise new<br>issues of SE. |
| Electrode Strip<br>operation<br>temperature | 5°C~30°C | < 32°C | Equivalent<br>The operation condition has been<br>tested and met its pre-defined<br>criteria, and it does not raise new<br>issues of SE. |
{10}------------------------------------------------
### Traditional 510(k) 510(k) Summary
{11}------------------------------------------------
# Traditional 510(k)
510(k) Summary
| Item | Subject device | Predicate device | |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Brand Name | QT ECG | QT ECG | Substantial equivalence |
| Model Name | PCA 500 | | determination |
| 510(k) No. | (to be assigned) | K180157 | |
| App modules | add Configuration module | • Recorder control module<br>• ECG data module<br>• User/Patient identification module | Equivalent<br>The added or modified modules<br>have been tested and met its<br>pre-defined criteria, and it does not<br>raise new issues of SE. |
| Web Service<br>modules | change User management module to<br>Authentication module<br>add RBAC module, Organization<br>module, and Configuration module | • User management module<br>• ECG data storage module<br>• Report module | |
| Analysis modules | add Pre-processing module and Noise<br>detection module | Measurement module | |
| Mobile | on Phone: | iOS 10 or later; | |
| application | iPhone 11 Pro Max (iOS 13.5) | Android 5.0 or later. | |
| Mobile device | iPhone 11 Pro (iOS 13.5) | on Phone: | Equivalent<br>The mobile device compatibility has been tested and met its pre-defined criteria, and it does not raise new issues of SE. |
| | iPhone 11 (iOS 13.5) | iPhone 6, 6 Plus, 6s, 6s Plus | |
| | iPhone XS Max (iOS 13.5) | iPhone 7, 7 Plus | |
| | iPhone XS (iOS 13.5) | Google Pixel | |
| | iPhone XR (iOS 13.5) | Samsung Galaxy A7 | |
| | iPhone X (iOS 13.5) | LG G5 | |
| | iPhone 8/ 8 Plus (iOS 13.5) | | |
| | iPhone 7/ 7 Plus (iOS 13.5) | on Tablet: | |
| iPhone 6S/ 6S Plus (iOS 13.5) | iPad Air (or later) | | |
| iPhone 6/ 6 Plus (iOS 12.4.7) | iPad Air 2 | | |
| iPhone SE (iOS 13.5) | iPad Mini 2 (or later) | | |
| iPhone SE 2 (iOS 13.5) | iPad Mini 3 | | |
| iPod touch (6th) (iOS 12.4.7) | iPad Mini 4 | | |
| Samsung Galaxy S9 (Android 9) | HTC nexus 9 | | |
| Samsung Galaxy S9+ (Android 9) | Samsung Galaxy Tab S2 | | |
| Samsung Galaxy S8 (Android 9) | | | |
| Samsung Galaxy S8+ (Android 9) | | | |
| LG G7 (Android 8.0.0) | | | |
| LG G6 (Android 8.0.0) | | | |
| LG G5 (Android 8.0.0) | | | |
| Google Pixel 2 (Android 10) | | | |
| Google Pixel 2 XL (Android 10) | | | |
| Google Pixel (Android 10) | | | |
| Google Pixel XL (Android 10) | | | |
| on Tablet: | | | |
| iPad 7th/2019 (iOS 13.5) | | | |
| iPad 6th/2018 (iOS 13.5) | | | |
| iPad 5th/2017 (iOS 13.5) | | | |
| iPad mini 5 (iOS 13.5) | | | |
| iPad mini 4 (iOS 13.5) | | | |
| iPad mini 3 (iOS 12.4.7) | | | |
| QT MEDICAL, Inc.<br>QT ECG | Traditional 510(k)<br>510(k) Summary | | |
| iPad mini 2 (iOS 12.4.7) | | | |
| iPad Air 2 (iOS 13.5) | | | |
| iPad Air 1 (iOS 12.4.7) | | | |
| Samsung Tab S4 (Android 9) | | | |
| Samsung Tab S3 (Android 9) | | | |
| Samsung Tab S2 (Android 7.0) | | | |
| Samsung Tab A 10.1" 2019 (Android 9) | | | |
| Lenovo Tab 4 8 Plus (Android 8.1.0) | | | |
| Asus Zenpad 3s 10 (Android 7.0) | | | |
| Asus Zenpad 3 8.0 (Android 7.0) | | | |
{12}------------------------------------------------
#### Traditional 510(k) 510(k) Summary
{13}------------------------------------------------
{14}------------------------------------------------
## 5.10 Similarity and Difference
The QT ECG has been compared with former "QT ECG". The subject device has similar indications for use, same technology/mechanism of action, and similar safety and performance as the predicate device. The patient population is extended to infant and children in the subject device, which is operated through 3 new sizes Electrode Strip.
Although there are some different specifications between two devices, the software validation, performance test and usability test have been completed to demonstrate that the differences between these parameters would not impact the safety and effectiveness of the subject device. The subject device has also undergone all safety and performance tests, and the results complied with the test requests.
Therefore, the difference between the subject device and the predicate device did not raise any problem of substantial equivalence. The subject device is substantially equivalent to the predicate device in intended use. safety and performance claims.
## 5.11 Conclusion
After analyzing all testing data and comparing with predicated device, it can be concluded that the QT ECG is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.