Straumann BLX Temporary Abutment, VITA CAD-Temp, PMMA
K220751 · Institut Straumann AG · NHA · Jun 2, 2022 · Dental
Device Facts
Record ID
K220751
Device Name
Straumann BLX Temporary Abutment, VITA CAD-Temp, PMMA
Applicant
Institut Straumann AG
Product Code
NHA · Dental
Decision Date
Jun 2, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Straumann® temporary prosthetic components are indicated to be used prior to the final components to maintain, stabilize, and shape the soft tissue during the healing phase. They must not be placed in occlusion. Strauman® temporary abutments are indicated to be placed into Straumann® dental implants to provide a support structure for the functional and esthetic oral rehabilitation of partially edentulous patients with temporary crowns and bridges.
Device Story
Temporary abutments for Straumann BLX dental implants; used to support temporary crowns/bridges during healing phase; maintains/shapes soft tissue. Device consists of PMMA body with Titanium Alloy (TAN) inlay. Used in dental clinics by clinicians. Not for use in occlusion. Provides support structure for esthetic/functional rehabilitation. Benefits patient by facilitating soft tissue healing and temporary restoration before final prosthetic placement.
Clinical Evidence
Bench testing only. Dynamic fatigue and static strength testing performed per FDA guidance for root-form endosseous dental implants and abutments. Biocompatibility and sterilization validation leveraged from existing cleared devices (K122192, K162890). No clinical data presented.
Technological Characteristics
Materials: PMMA and Titanium Alloy (Ti-6Al-7Nb). Connection: TorcFit. Form factor: Temporary abutment for RB/WB and WB platforms. Sterilization: Non-sterile, requires user reprocessing. Mechanical testing per FDA guidance for endosseous dental abutments.
Indications for Use
Indicated for partially edentulous patients requiring temporary crowns and bridges for oral rehabilitation; used to maintain, stabilize, and shape soft tissue during healing phase; not for use in occlusion.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Institut Straumann AG % Jennifer Jackson Director, Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K220751
Trade/Device Name: Straumann® BLX Temporary Abutment, VITA CAD-Temp, PMMA Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: March 11, 2022 Received: March 15, 2022
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K220751
### Device Name:
Straumann® BLX Temporary Abutment, VITA CAD-Temp, PMMA
### Indications for Use (Describe)
Straumann® temporary prosthetic components are indicated to be used prior to the final components to maintain, stabilize, and shape the soft tissue during the healing phase. They must not be placed in occlusion. Strauman® temporary abutments are indicated to be placed into Straumann® dental implants to provide a support structure for the functional and esthetic oral rehabilitation of partially edentulous patients with temporary crowns and bridges.
Type of Use (Select one or both, as applicable) 図Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{4}------------------------------------------------
## Straumann® BLX Temporary Abutment, VITA CAD-Temp, PMMA
510(k) Summary
### ട് 510(k) Summary
### Submitter's Contact Information 5.1
| Submitter: | Straumann USA, LLC<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315 Owner/Operator No.: 9005052<br>On the behalf of:<br>Institut Straumann AG<br>Peter Merian-Weg 12<br>4052 Basel, Switzerland. |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs and Quality<br>Phone Number: +1-978-747-2509<br>Fax Number: +1-978-747-0023 |
| Prepared By & Alternate Contact: | Viviana Horhoiu<br>Regulatory Affairs and Compliance manager<br>Institut Straumann AG<br>Phone number: +41 61 965 1512 |
| Date of Submission: | May 30, 2022 |
#### 5.2 Name of the Device
| Trade Names: | Straumann® BLX Temporary Abutment, VITA CAD-Temp,<br>PMMA |
|------------------------|-----------------------------------------------------------|
| Common Name: | Endosseous dental implant abutment |
| Classification Name: | Endosseous dental implant abutment |
| Regulation Number: | §872.3630 |
| Device Classification: | Class II |
| Product Code(s): | NHA |
| Classification Panel: | Dental |
{5}------------------------------------------------
## Straumann® BLX Temporary Abutment, VITA CAD-Temp, PMMA
510(k) Summary
#### 5.3 Predicate Device(s)
Primary Predicate:
- K173961 Straumann® BLX Implant System Institut Straumann AG .
Reference Device:
- . K122192 – Straumann Temporary Abutments, PMMA – Institut Straumann AG
#### Device Description 5.4 -
The subject temporary abutments will be marketed as Straumann® BLX Temporary Abutment, VITA CAD-Temp, PMMA, however, throughout this 510(k) Summary the subject devices will be referred to as BLX Temporary Abutment, PMMA.
Straumann® BLX Temporary Abutments PMMA are temporary abutments intended for placement on Straumann BLX Dental Implants with RB/WB and WB connection platforms. The temporary abutments are made of polymethyl methacrylate (PMMA), with a Titanium Alloy (TAN) inlay.
### 5.5 Indications for Use
Straumann® temporary prosthetic components are indicated to be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase. They must not be placed in occlusion. Straumann® temporary abutments are indicated to be placed into Straumann® dental implants to provide a support structure for the functional and esthetic oral rehabilitation of partially edentulous patients with temporary crowns and bridges.
{6}------------------------------------------------
# Straumann® BLX Temporary Abutment, VITA CAD-Temp, PMMA
510(k) Summary
# 5.6 Technological Characteristics
| FEATURE | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | | K173961 | K122192 | |
| Indications<br>for Use | Straumann® temporary<br>prosthetic components are<br>indicated to be used prior to the<br>insertion of the final components<br>to maintain, stabilize and shape<br>the soft tissue during the healing<br>phase. They must not be placed<br>in occlusion. Straumann®<br>temporary abutments are<br>indicated to be placed into<br>Straumann® dental implants to<br>provide a support structure for<br>the functional and esthetic oral<br>rehabilitation of partially<br>edentulous patients with<br>temporary crowns and bridges. | Straumann® BLX Basal Screws<br>and Temporary Abutments<br>Prosthetic components directly or<br>indirectly connected to the<br>endosseous dental implant are<br>intended for use as an aid in<br>prosthetic rehabilitations.<br>Temporary components can be<br>used prior to the insertion of the<br>final components to maintain,<br>stabilize, and shape the soft<br>tissue during the healing phase;<br>they may not be placed into<br>occlusion. Final abutments may<br>be placed into occlusion when the<br>implant is fully osseointegrated.<br>BLX Temporary Abutments have<br>a maximum duration of usage of<br>180 days. | Straumann® temporary<br>abutments are indicated to be<br>placed into Straumann® dental<br>implants to provide a support<br>structure for the functional and<br>esthetic oral rehabilitation of<br>partially edentulous patients<br>with temporary crowns and<br>bridges. | Equivalent<br>The subject and primary predicate<br>device indications for use<br>statements are equivalent. The<br>subject device does not contain<br>language regarding final abutments<br>and the subject device is for<br>temporary use only.<br>The type of prosthetic restorations<br>is implied for the primary predicate<br>device, but included specifically in<br>the subject and reference device<br>indications for use statements.<br>The statement regarding the<br>maximum duration of usage has<br>been moved to the Instructions for<br>Use for the subject devices. |
| Material | Titanium alloy/TAN<br>(Ti-6Al-7Nb),<br>PMMA | Titanium alloy/TAN<br>(Ti-6Al-7Nb) | Titanium alloy/TAN<br>(Ti-6Al-7Nb),<br>PMMA | Identical<br>The materials are identical to the<br>reference device. |
| Implant to<br>Abutment<br>Connection | TorcFit® | TorcFit® | CrossFit®<br>synOcta | Identical<br>The connection is identical to the<br>primary predicate device. |
| FEATURE | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
| K Number | | K173961 | K122192 | |
| Compatibility | RB/WB and WB | RB/WB and WB | Narrow CrossFit® (NC)<br>Regular CrossFit® (RC)<br>Narrow Neck CrossFit® (NNC)<br>Regular Neck (RN)<br>Wide Neck (WN) | Identical<br>The compatibility is identical to the<br>primary predicate device. |
| Diameter or<br>Minor Oval<br>Dimension/<br>Major Oval<br>Dimension | Major oval diameter: 10 mm<br>Minor oval dimensions: 5 mm | RB/WB:<br>$Ø$ 3.8, $Ø$ 4.5, $Ø$ 6.0<br>WB:<br>$Ø$ 5.5 | Major oval diameter: 10.1 mm<br>Minor oval dimensions: 5 mm | Equivalent<br>The diameters are nearly identical<br>to the reference devices. |
| Overall<br>abutment<br>length | 15.1 and 16.1 mm | 13.5, 14.1, 15.1, 16.1, 17.1 mm | 16.3 mm | Identical<br>The overall abutment lengths are<br>identical to two of the abutment<br>lengths of the primary predicate<br>device. |
| Gingival<br>heights | 1.5 mm | 0.75, 1.5, 2.5 and 3.5 mm | 1.0 mm | Identical<br>The gingival height is identical to<br>one of the gingival heights of the<br>primary predicate device. |
| Prosthesis<br>Fixation | Screw-retained | Screw-retained | Screw-retained | Identical<br>The prosthetic fixation is identical<br>to the primary predicate and<br>reference devices. |
| Sterilization<br>Method | Non-sterile | Non-sterile | Non-sterile | Identical<br>The subject devices are provided<br>non-sterile and require the same<br>cleaning, disinfection and<br>sterilization procedure as the<br>primary predicate and reference<br>devices. |
| FEATURE | SUBJECT DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | EQUIVALENCE DISCUSSION |
| K Number | | K173961 | K122192 | |
| Duration of<br>Use | 6 months | 6 months | 6 months | Identical<br>The duration of use is identical for<br>the subject, primary predicate, and<br>reference devices. |
{7}------------------------------------------------
# Straumann® BLX Temporary Abutment, VITA CAD-Temp, PMMA
510(k) Summary
{8}------------------------------------------------
# Straumann® BLX Temporary Abutment, VITA CAD-Temp, PMMA
510(k) Summary
Table 1 – Comparison of the subject devices with the primary predicate and reference devices
{9}------------------------------------------------
## Straumann® BLX Temporary Abutment, VITA CAD-Temp, PMMA
510(k) Summary
#### Performance Testing 5.7 -
Dynamic fatigue and static strength tests were conducted according to the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and demonstrated the BLX Temporary Abutments PMMA performed as well as the reference devices.
The subject device materials, PMMA and TAN (titanium alloy), are identical to the reference device materials. The manufacturing processes, packaging, body/fluid contact are identical to the reference devices; therefore, no new issues regarding biocompatibility were raised.
The sterilization validation for the BLX Temporary Abutments PMMA, according to applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015", was leveraged from K162890.
#### Conclusion 5.8
The conclusion drawn from the nonclinical tests demonstrates that the subject devices, the BLX Temporary Abutments PMMA, are substantially equivalent to the legally marketed primary predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.